Peer Review History

Original SubmissionAugust 12, 2024
Decision Letter - Joshua William Giles, Editor

PONE-D-24-34534Biomechanical Comparison of Various Bone Reduction Forceps in Interfragmentary Compression and Area of Compression in an Experimental Model of Lateral Humeral Condylar FracturesPLOS ONE

Dear Dr. Baskette,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please make the revisions suggested by the reviewer to increase the clarity of your methods, results, and discussion.

Please submit your revised manuscript by Dec 05 2024 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org . When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols .

We look forward to receiving your revised manuscript.

Kind regards,

Joshua William Giles, Ph.D.

Academic Editor

PLOS ONE

Journal requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. When completing the data availability statement of the submission form, you indicated that you will make your data available on acceptance. We strongly recommend all authors decide on a data sharing plan before acceptance, as the process can be lengthy and hold up publication timelines. Please note that, though access restrictions are acceptable now, your entire data will need to be made freely accessible if your manuscript is accepted for publication. This policy applies to all data except where public deposition would breach compliance with the protocol approved by your research ethics board. If you are unable to adhere to our open data policy, please kindly revise your statement to explain your reasoning and we will seek the editor's input on an exemption. Please be assured that, once you have provided your new statement, the assessment of your exemption will not hold up the peer review process.

3. Please remove your figures from within your manuscript file, leaving only the individual TIFF/EPS image files, uploaded separately. These will be automatically included in the reviewers’ PDF.

4. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Additional Editor Comments:

The authors have provided a clear description of their work, but a few areas should be improved based on the reviewer comments.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Please state in title and in the abstract that the study is on CANINE lateral humeral condylar fractures.

Fracture Model paragraph, last sentence: The second osteotomy direction and the reasoning for this is unclear. The cut started at the epicondyle and was directed 45 deg disto-lateral to proxo-medial? A figure would be helpful.

Interfragmentary Compression paragraph: Please double-check your sensor is 4040, that sensor is not listed on Tekscan’s website. It is important to know due to resolution of the sensor.

Please describe the calibration procedure. Calibration of these sensors are key to accurate data.

Bone Reduction Forcep Groups: Somewhere you should spell out DPS to DePuy Synthes or reference Johnson & Johnson.

Data Collection:

“converted into a line graph” is not enough detail. Please describe if/how you defined the region of compression. Was the peak pressure the single “sensel” of the highest pressure, or was it the average pressure across the region and the peak of the average over the recorded time.

How was the position of the condyle repeatedly placed back into the original/best location for each compressive test? A mal-positioned (rotated) condyle would mean mating faces were not contacting and would change the pressure recorded. A non-repeatable positioning would also give large variability to “Area of Compression”. If you analyze repeatability of the 3 tests per forceps type, would give an indication of reliability of your technique and therefore forceps comparisons.

Please state whether or not the sensor extended beyond the mating faces of the fracture fragments. If not, how would you ensure the same region of the fragments contacted the sensor. Similarly, state if there is any “fragment” portion of the bone contacting parent bone that is not have sensing elements in-between. This would make a big difference, if there was load supported by mating faces but not being sensed.

How were the 3 repeat measures handled in statistical ANOVA?

In the methods, you state “The bone forceps were then engaged until the maximum interfragmentary compression prior to bone fragment fracture, forcep disengagement, or loss of reduction was achieved.”

There are no results given on whether any bone fractured, if so, how were the other forceps measured? There are no results given on if forceps stayed engaged or slipped of easily. That would be very useful information. Perhaps a high fleeting peak compression was made, but the reduction not able to be held, that is not useful clinically.

Discussion:

Sentence: “Based on the compression achieved by each group, all five types of bone reduction forceps are appropriate for holding the fracture in reduction during fixation.” There is no indication from your data that the pressures measured would hold during fixation. You would need a reference related to holding power, or perhaps to the other studies you mention later in the discussion.

Paragraph 5 (the comparison against reference 14): It is difficult to make any comparisons with this study to your study, given the lack of specifying in both studies if the measures are average pressure across the fragment, then the maximum of this data over time, or if the “peak pressure” or “interfragmentary compression” is the peak value (max pressure in any sensel at any time point). These are very different measurements.

Figure 4. Please state if the images are of the same bone. Are the images all at the same scale? They are all varying in shape. The reader cannot tell if shape is related to the bone contacting itself or it is simply a different bone.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean? ). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy .

Reviewer #1: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/ . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org . Please note that Supporting Information files do not need this step.

Revision 1

Thank you for all of your feedback. That was very helpful in realizing where the paper could be improved. The reviewer comments, as well as replies, have been attached to the submission as well as pasted below.

Journal Requirements:

1. All manuscript stylings have been adjusted to meet PLOS ONE’s style requirements. These include titling of the article and article sections, citations, and author information.

2. Data will be made available publicly.

3. Figures have been removed from the manuscript file.

4. All references are accurate and available when searching either by citation or DOI.

Reviewer Comments:

Reviewer comment: Please state in title and in the abstract that the study is on CANINE lateral humeral condylar fractures.

Reply: This has been adjusted.

Reviewer comment: Fracture Model paragraph, last sentence: The second osteotomy direction and the reasoning for this is unclear. The cut started at the epicondyle and was directed 45 deg disto-lateral to proxo-medial? A figure would be helpful.

Reply: The second osteotomy direction was performed in the usual direction of the fracture through the lateral ramus seen in patients with lateral humeral condylar fracture. This was added to the end of the sentence and a figure was added (Fig 1) following both osteotomies.

Reviewer comment: Interfragmentary Compression paragraph: Please double-check your sensor is 4040, that sensor is not listed on Tekscan’s website. It is important to know due to resolution of the sensor.

Reply: Thank you for noticing. The correct sensor is 4041. Here are the specifications: https://www.tekscan.com/products-solutions/pressure-mapping-sensors/4041

Reviewer comment: Please describe the calibration procedure. Calibration of these sensors are key to accurate data.

Reply: Calibration was achieved through placement of a 10 kg calibrated weight on the wire film sensor while resting on a level surface. The pressure sensor was then set to display pressure in kPa, as well as to show the maximum differences among the pixels measuring the pressure. This is now included in the interfragmentary compression section.

Reviewer comment: Bone Reduction Forcep Groups: Somewhere you should spell out DPS to DePuy Synthes or reference Johnson & Johnson.

Reply: Thank you, this has now been included in the bone reduction forcep groups section.

DATA COLLECTION

Reviewer comment: “converted into a line graph” is not enough detail. Please describe if/how you defined the region of compression. Was the peak pressure the single “sensel” of the highest pressure, or was it the average pressure across the region and the peak of the average over the recorded time.

Reply: The data was converted into a line graph by plotting pressure over time. This allowed us to identify where the pressure plateaued. The Tekscan software then calculates the average area of compression, compressive force, and peak pressure for this time period. This is now included in the data collection section (lines 129-132 in copy with tracked changes).

Reviewer comment: How was the position of the condyle repeatedly placed back into the original/best location for each compressive test? A mal-positioned (rotated) condyle would mean mating faces were not contacting and would change the pressure recorded. A non-repeatable positioning would also give large variability to “Area of Compression”. If you analyze repeatability of the 3 tests per forceps type, would give an indication of reliability of your technique and therefore forceps comparisons.

Reply: The fracture, having been artificially generated, allowed for uncomplicated reduction of the lateral condyle and precise interdigitation with the proximal humerus. This process was made easier by the presence of the condylar bridge and the excellent visualization of the joint anatomy, ensuring clarity and accuracy in the operation. The fracture fragment was anatomically reduced with digital manipulation prior to the placement of the bone reduction forceps to ensure that the measurements were consistent and accurate. This is now mentioned in the Interfragmentary Compression section. Additionally, after the compression was achieved, the contact area was verified to check for any decrease in size.

Reviewer comment: Please state whether or not the sensor extended beyond the mating faces of the fracture fragments. If not, how would you ensure the same region of the fragments contacted the sensor. Similarly, state if there is any “fragment” portion of the bone contacting parent bone that is not have sensing elements in-between. This would make a big difference, if there was load supported by mating faces but not being sensed.

Reply: The sensor was always oriented perpendicular to the interfragmentary interface. This allowed us to ensure there was at least a small amount of the sensor on each side that was in excess, ensuring the entire area of compression was included within the sensor. The complete surface of the condyle was fully covered by the sensor, ensuring that every inch of the surface was accurately mapped during the test. Additionally, utilizing a larger sensor than the tested area provided us with an extra “spare" sensor for greater reliability.

On the other hand, despite it being almost impossible to position the sensor perfectly during every testing cycle, it didn’t matter which part of the sensor was mapping the contact, as long as the intercondylar area was fully covering the sensor. This is because the full size of the sensor maintained the same quality in terms of density, resolution, and sensitivity. This means the captured data was not limited by the sensor position, but instead, represented the actual contacted force mapping

Reviewer comment: How were the 3 repeat measures handled in statistical ANOVA?

Reply: The three repeat measurements were included in the total data set, meaning that there were 21 sets of data for each forcep. The average peak pressure and contact area for each forcep were then compared using statistical ANOVA to evaluate for significant differences between the averages.

Reviewer comment: In the methods, you state, “The bone forceps were then engaged until the maximum interfragmentary compression prior to bone fragment fracture, forceps disengagement, or loss of reduction was achieved.”

There are no results given on whether any bone fractured, if so, how were the other forceps measured? There are no results given on if forceps stayed engaged or slipped of easily. That would be very useful information. Perhaps a high fleeting peak compression was made, but the reduction not able to be held, that is not useful clinically.

Reply: This is a good point. No bones were fractured during placement, which allowed us to include all of our data sets. Loss of reduction and/or forcep disengagement did occur; however, this was infrequent and could have resulted from user error, so this data was not included. This is now discussed in the results section (145 to 149 in the tracked changes copy).

DISCUSSION

Reviewer comment: Sentence: “Based on the compression achieved by each group, all five types of bone reduction forceps are appropriate for holding the fracture in reduction during fixation.” There is no indication from your data that the pressures measured would hold during fixation. You would need a reference related to holding power, or perhaps to the other studies you mention later in the discussion.

Reply: This sentence was poorly written and was intended to discuss the ability to hold the fragments in reduction, which was not clear. However, reference to one of the papers mentioned later in the discussion was added as recommended to tie in the discussion of compression, as well (lines 166-168 in marked copy).

Reviewer comment: Paragraph 5 (the comparison against reference 14): It is difficult to make any comparisons with this study to your study, given the lack of specifying in both studies if the measures are average pressure across the fragment, then the maximum of this data over time, or if the “peak pressure” or “interfragmentary compression” is the peak value (max pressure in any sensel at any time point). These are very different measurements.

Reply: This is valid. The Tekscan software, however, is used in both studies, meaning that peak pressure is the highest pressure generated in a single cell, allowing us to compare both studies. The idea behind using the peak pressure is that the pressure should be the greatest in the center of the fragment at the level of the ideal placement for the transcondylar screw.

Reviewer comment: Figure 4. Please state if the images are of the same bone. Are the images all at the same scale? They are all varying in shape. The reader cannot tell if shape is related to the bone contacting itself or it is simply a different bone.

Reply: The images are not all of the same bone, but the variance also occurs due to the use of different forceps and their placement. The scale is, however, the same in all images. This has been added to the description of the figure.

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Joshua William Giles, Editor

Biomechanical comparison of various bone reduction forceps in interfragmentary compression and area of compression in an experimental model of canine lateral humeral condylar fractures

PONE-D-24-34534R1

Dear Dr. Baskette,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager®  and clicking the ‘Update My Information' link at the top of the page. If you have any questions relating to publication charges, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Joshua William Giles, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Thank you for thoroughly addressing the reviewer comments.

Reviewers' comments:

Formally Accepted
Acceptance Letter - Joshua William Giles, Editor

PONE-D-24-34534R1

PLOS ONE

Dear Dr. Baskette,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

If revisions are needed, the production department will contact you directly to resolve them. If no revisions are needed, you will receive an email when the publication date has been set. At this time, we do not offer pre-publication proofs to authors during production of the accepted work. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few weeks to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Professor Joshua William Giles

Academic Editor

PLOS ONE