Peer Review History

Original SubmissionAugust 25, 2024
Decision Letter - Minsoo Chun, Editor

PONE-D-24-34482Tin prefiltration in computed tomography fails to alter radiation-induced gene expression and double-strand break formationPLOS ONE

Dear Dr. Abend,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Nov 28 2024 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Minsoo Chun, Ph.D.

Academic Editor

PLOS ONE

Journal requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that you have included the phrase “data not shown” in your manuscript. Unfortunately, this does not meet our data sharing requirements. PLOS does not permit references to inaccessible data. We require that authors provide all relevant data within the paper, Supporting Information files, or in an acceptable, public repository. Please add a citation to support this phrase or upload the data that corresponds with these findings to a stable repository (such as Figshare or Dryad) and provide and URLs, DOIs, or accession numbers that may be used to access these data. Or, if the data are not a core part of the research being presented in your study, we ask that you remove the phrase that refers to these data.

3. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: General Comments

The manuscript titled "Tin prefiltration in computed tomography fails to alter radiation-induced gene expression and double-strand break formation" presents an interesting study on the molecular biological response to the tin prefilter technique in CT. The authors examine both radiation-induced double-strand break (DSB) formation and differential gene expression (DGE) in blood samples exposed to different CT protocols. The study is relevant in the context of reducing radiation dose without compromising image quality, and the results indicate that this technique does not significantly affect radiation-induced molecular changes when compared to standard CT protocols.

Specific Comments:

1. Abbreviations, Units, and Sentence Structure:

The entire manuscript requires a thorough revision to ensure consistent use of abbreviations, correct application of units in the biological endpoints measured, and the construction of complete sentences.

2. Image Quality Technique:

The description of image quality techniques should be more detailed and clearer. It is essential to specify whether the quality of the images was evaluated objectively or subjectively.

3. Title Suggestion:

I recommend revising the title to one of the following options for clarity and impact:

- Tin prefiltration in computed tomography does not significantly alter radiation-induced gene expression and DNA double-strand breaks."

- Minimal effects of tin prefiltration on radiation-induced gene expression and DNA double-strand break formation in computed tomography."

4. Abstract Revision (Page 3):

In the conclusion of the abstract, instead of "our ex vitro," it is better to write "the current ex-vivo study."

5. miRNA Notation:

The first miRNA should be correctly noted as "MicroRNA-667."

6. Dose Measurement Methodology:

The methodology section should clearly explain how the dose was measured or calculated, whether through dosimetry or other methods.

7. Control Blood Samples:

The exclusion criteria for the healthy control blood samples should be specified. It would have been better to compare the same donor blood samples before and after irradiation. If controls excluded certain criteria such as smoking or chemotherapy, this should be clearly stated.

8. Discussion on DGE Values:

Although the paper concludes that tin prefiltration does not alter radiation-induced DGE or DSB formation, it also notes a significant tendency toward lower DGE values for the Sn 150 kV low-dose protocol. This finding needs clarification. The discussion should be expanded to better explain the biological implications of this trend toward lower DGE values. Additionally, it should be explicitly stated whether these changes, though not statistically significant, could still hold clinical relevance.

Reviewer #2: The authors have presented a careful study of an assessment of the tin pre-filtration technique for CT scanning. This technique has the ability to reduce radiation exposure during CT scans by eliminating the lower energy radiation from the scanners which gets absorbed by a thin layer of tin (Sn), allowing only the higher energy radiation. This is best suited for deep and high contrast imaging. A thorough evaluation of the possible protective effect of this technology is significant and this reviewer believes it has been done carefully, and with suitable sampling (at least for the DNA DSB marker yH2AX), and with appropriate assays. The authors chose to concentrate on the DNA damage aspects of CT scanning. This reviewer believes this study was straightforward, and presents reliable data that is easily decipherable and would be of value to the field. Care was also taken to control for dose levels, and several iterations were used for a full comparison. It was also commendable that only yH2AX foci co-localized with 53BP1 were counted, and yH2AX is known to mark non-DSB lesions as well. The number of patient/samples (6) could have been higher, which might have helped reducing variability in the transcriptomics and miRNA (but that was acknowledge by the authors).

some minor edits:

line 242 Fig 2. yH2AX foci: the scoring and tabulating methodologies for this assay are sound. according to the method to score these the number of foci per cell of unexposed is subtracted from the exposed. The resulting increase in foci count is not that high, it is less than 1 foci per cell, but nevertheless significant. This reviewer would be heartened to see some example images. The increase in foci might not be perceivable, but showing example imaging (even if in the supplement) would boost confidence in the technique.

line 370-371

suggested change to

"can distinguish between whole-body and partial-body irradiation [33]."

line 374

suggested change to

"low dose irradiation has not been"

line 403 and 404

change deregulated to 'dysregulated'

line 407

change genome to 'transcriptome'

line 415

suggested change to

"This is a limitation of this study, and which was attempted to be minimized by performing the trial scans."

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Hariwan A. Mohammed

Reviewer #2: Yes: Aris A. Polyzos

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachments
Attachment
Submitted filename: General Comments.docx
Revision 1

Dear Editor and Reviewers,

thank you for your valuable comments. Please find our point-by-point response in the attached file "Response to Reviewers".

Kind regards

Attachments
Attachment
Submitted filename: Response to Reviewers_SIS_MA_V2.pdf
Decision Letter - Minsoo Chun, Editor

Tin prefiltration in computed tomography does not significantly alter radiation-induced gene expression and DNA double-strand break formation

PONE-D-24-34482R1

Dear Dr. Abend,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. If you have any questions relating to publication charges, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Minsoo Chun, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #2: The authors responded very thoroughly to the initial review comments provided for the manuscript.

Including figures of the quantified data for DNA damage (imaging of gamma-H2AX nuclear foci) allows a reader or investigator trying to recapitulate these findings to properly evaluate both the data and the quality of the signal. This was the chief issue this reviewer had with the original manuscript. Otherwise, it is very thoroughly detailed and annotated and in this reviewers opinion a worthy publication for the community.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Hariwan Abdulkareem Mohammed

Reviewer #2: Yes: Aris Polyzos

**********

Formally Accepted
Acceptance Letter - Minsoo Chun, Editor

PONE-D-24-34482R1

PLOS ONE

Dear Dr. Abend,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

If revisions are needed, the production department will contact you directly to resolve them. If no revisions are needed, you will receive an email when the publication date has been set. At this time, we do not offer pre-publication proofs to authors during production of the accepted work. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few weeks to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Minsoo Chun

Academic Editor

PLOS ONE

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .