Peer Review History
| Original SubmissionMarch 17, 2023 |
|---|
|
PONE-D-23-08013A 3D-printed condom intrauterine balloon tamponadePLOS ONE Dear Dr. Piaggio, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. ============================== Dear Dr. Piaggio, Thank you for submitting your manuscript and thank you for your patience. We have been able to receive two reviews, however, we feel that your manuscript requires at least one review by a researcher in the biomedical engineering field, which has taken longer than we had anticipated. For now, please respond to the comments by the first reviewers and return the manuscript with responses to their comments as we work on securing a review from a biomedical engineering researcher, whose comments might come at the next round of the review process. Please submit your revised manuscript by Jun 17 2023 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Daniel Semakula, M.D. MPH, PhD Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability. ""Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized. Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access. We will update your Data Availability statement to reflect the information you provide in your cover letter Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly Reviewer #2: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: I Don't Know Reviewer #2: N/A ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: No Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: No Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The paper is addressing an important intervention component to manage PPH esp in low resource setting. However the technical aspects are not clear and comparison with available standard methods is not explicitly described. The contents of the UBT need to be described in detail what material, size, applicability and how this use is supposed to be different than others. I do not agree with the advantage of being able to sterilise this UBT as there are multiple low cost dispensable UBTs available. Hence its not clear to the reader how this UBT is better in either the contents, material used, ease of use, capacity of the device to hold the fluid, duration that could be kept in the uterus. The figure of inflated balloon also does not visually satisfy that it takes shape of the uterus such that its entire inner surface gets pressure from the inflated balloon I suggest the paper be reviewed by biomedical engineer expert to understand the nitty-gritties. Every device has to meet an international quality standard (ISO) . There is no mention how this device falls within those standards Authors must also be aware that 3 D printers may not be easily available in Low resource settings for which this device is being developed. Reviewer #2: This is very important development that should tested with real PPH cases, where the other management options like uterotonics are also exacting pressure on the uterus. Few comments for the authors to clarify 1. The abstract results are written as conclusion of the prototyping but not the results. Please state the results 2. Lines 2 and 3: Define the abbreviation LRSs and LMIC since they appear for the first time. There are several abbreviations that must be defined other than assuming that the readers know what it is. 3. You will a reference for this statement “Postpartum haemorrhage is defined as heavy bleeding after birth and it can be classed as “primary” if there is a 500-ml blood loss (>2 litres in case of severe haemorrhage) within the first 24 hours after birth, and as “secondary” if there is abnormal or heavy vaginal bleeding between 24 hours and 12 weeks after birth.” 4. Lines 18-20, the authors mention that the uterine balloon is among the first line management option, they need to check the citation well. It is for refractory management of PPH. 5. In line 82-83, The parts that underwent 3D printing were: 1) A valve 2) A flow-stopper 3) A modified bottle cap 4) A uterus model. What about the condom uterine balloon. It is not clear how this is different from ordinary condom 6. With several parts making the system, what were the different points of leakage of water. 7. Although the circular economic model is mentioned in the introduction and the discussion, the authors have not demonstrated how it applied in the methods and results. Which parts are re-usable. 8. What is the likely cost of the device ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Sam Ononge ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.
|
| Revision 1 |
|
PONE-D-23-08013R1A 3D-printed condom intrauterine balloon tamponadePLOS ONE Dear Dr. Piaggio, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Jan 07 2024 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Hugh Cowley Staff Editor PLOS ONE Additional Editor Comments: Your manuscript has been evaluated by two reviewers: one from the previous round (Reviewer 1) and one new reviewer with expertise in biomedical engineering (Reviewer 3). Their comments are appended below. Reviewer 1 is satisfied with the changes you have made. Reviewer 3 recognizes the changes, but has stated that clinical experiments and a comparison with standard methods are necessary to support the conclusions of this study. Recognizing that this work was intended as a pre-clinical investigation, clinical validation will be a necessary step, but reasonable to consider outside the scope of the current study. However, demonstration of performance against existing comparable balloon tamponades would help validate the claim that this newly proposed concept has comparable (or at least practical given the other stated advantages) functionality. Please also ensure that any claims of potential efficacy are appropriately qualified with statements clarifying that no clinical experiments have been carried out. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #3: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #3: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: N/A Reviewer #3: No ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #3: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #3: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: (No Response) Reviewer #3: I really appreciate all your hard work to write this article. however, it still needs more efforts to finalize it. A set of clinical experiments is needed to validate the proposed catheter. Also, a comparison with one of the standard methods is highly required to support the concluded results. best regards, ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #3: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
PONE-D-23-08013R2A 3D-printed condom intrauterine balloon tamponade: design, prototyping, and technical validationPLOS ONE Dear Dr. Piaggio, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Dear authors, please correct the manuscript by separating the single paragraph of the discussion into 3 or 4 and leave it in accordance with the following reviewers' recommendations: "this version of the manuscript has more typos than the previous one (e.g. periods in titles of some subsections of the " Results" section and the "Discussions and Conclusions" section consists of only one long paragraph). Also, the reference section needs to be revised carefully (e.g. ref 23 and 24 are repeated in 37 and 38).", and "the quality of Figure 1 must be improved and the Figure 2 showing the complete prototype, including the used pump." Please submit your revised manuscript by Apr 25 2024 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Ricardo Ney Oliveira Cobucci, Ph.D Academic Editor PLOS ONE Journal Requirements: Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. Additional Editor Comments: Dear authors, please correct the manuscript by separating the single paragraph of the discussion into 3 or 4 and leave it in accordance with the following reviewers' recommendations: "this version of the manuscript has more typos than the previous one (e.g. periods in titles of some subsections of the " Results" section and the "Discussions and Conclusions" section consists of only one long paragraph). Also, the reference section needs to be revised carefully (e.g. ref 23 and 24 are repeated in 37 and 38).", and "the quality of Figure 1 must be improved and the Figure 2 showing the complete prototype, including the used pump." [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #3: (No Response) Reviewer #4: All comments have been addressed Reviewer #5: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #3: No Reviewer #4: Yes Reviewer #5: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #3: No Reviewer #4: I Don't Know Reviewer #5: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: (No Response) Reviewer #3: Yes Reviewer #4: Yes Reviewer #5: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #3: No Reviewer #4: Yes Reviewer #5: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: (No Response) Reviewer #3: I reject this paper mainly for the lack of solid statistical comparisons between the proposed system and other systems which are already in the market. Although, the authors showed some comparisons in separate tables, I am still not satisfied by the point of comparisons or the qualitative results used. Also, the comments from the previous round were not completely addressed. The manuscript is still not acceptable for publication and this version of the manuscript has more typos than the previous one (e.g. periods in titles of some subsections of the "Results" section and the "Discussions and Conclusions" section consists of only one long paragraph). Also, the reference section needs to be revised carefully (e.g. ref 23 and 24 are repeated in 37 and 38). Reviewer #4: (No Response) Reviewer #5: The paper presents a proposal for a hemorrhages intrauterine tamponade prototype that may be occur in the postpartum. Although this technique is not new, the paper has a target audience in low-resource settings, propose the use of recycled material and presents innovation of 3D printing use for some parts of the prototype, which facilitates the dissemination to the proposed device. I noted that the requested revisions were met. The paper target all the requirements to the publication and I believe it has good potential to encourage more research in this study area. In my point of view, the paper should be accepted for publication. My suggestion is the quality of Figure 1 to be improve and that the Figure 2 showing the complete prototype, including the used pump. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #3: No Reviewer #4: Yes: Itamir de Morais Barroca Filho Reviewer #5: Yes: Jan Erik Mont Gomery Pinto ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 3 |
|
A 3D-printed condom intrauterine balloon tamponade: design, prototyping, and technical validation PONE-D-23-08013R3 Dear Dr. Piaggio, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. If you have any questions relating to publication charges, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Ricardo Ney Oliveira Cobucci, Ph.D Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: |
| Formally Accepted |
|
PONE-D-23-08013R3 PLOS ONE Dear Dr. Piaggio, I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now being handed over to our production team. At this stage, our production department will prepare your paper for publication. This includes ensuring the following: * All references, tables, and figures are properly cited * All relevant supporting information is included in the manuscript submission, * There are no issues that prevent the paper from being properly typeset If revisions are needed, the production department will contact you directly to resolve them. If no revisions are needed, you will receive an email when the publication date has been set. At this time, we do not offer pre-publication proofs to authors during production of the accepted work. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few weeks to review your paper and let you know the next and final steps. Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. If we can help with anything else, please email us at customercare@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of PROFESSOR Ricardo Ney Oliveira Cobucci Academic Editor PLOS ONE |
Open letter on the publication of peer review reports
PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.
We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.
Learn more at ASAPbio .