Peer Review History

Original SubmissionApril 25, 2023
Decision Letter - Mattias Günther, Editor
Transfer Alert

This paper was transferred from another journal. As a result, its full editorial history (including decision letters, peer reviews and author responses) may not be present.

PONE-D-23-12462To Test the Performance of an Open-Source, Low-Cost Ventilator in a Porcine Model.PLOS ONE

Dear Dr. Cho,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Sep 02 2023 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Mattias Günther

Academic Editor

PLOS ONE

Journal requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. As part of your revision, please complete and submit a copy of the Full ARRIVE 2.0 Guidelines checklist, a document that aims to improve experimental reporting and reproducibility of animal studies for purposes of post-publication data analysis and reproducibility: https://arriveguidelines.org/sites/arrive/files/documents/Author%20Checklist%20-%20Full.pdf Please include your completed checklist as a Supporting Information file. Note that if your paper is accepted for publication, this checklist will be published as part of your article.

3. To comply with PLOS ONE submissions requirements, in your Methods section, please provide additional information regarding the experiments involving animals and ensure you have included details on (1) methods of sacrifice, (2) methods of anesthesia and/or analgesia, and (3) efforts to alleviate suffering.

4. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

5. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For more information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially sensitive information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

6. Please include your full ethics statement in the ‘Methods’ section of your manuscript file. In your statement, please include the full name of the IRB or ethics committee who approved or waived your study, as well as whether or not you obtained informed written or verbal consent. If consent was waived for your study, please include this information in your statement as well.

7. Please include your tables as part of your main manuscript and remove the individual files. Please note that supplementary tables (should remain/ be uploaded) as separate "supporting information" files

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: No

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: No

Reviewer #2: I Don't Know

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: I am a little unclear what the author's expectations were for this article. They claim an open source design, one that is not thoroughly given or expanded upon. Additionally, the timelines of this topic is misaligned with the reality of current health states. The federal government has removed the state of emergency over COVID-19 and there have been several papers on this or a similar topic.

Additionally, proposing an "open-source" medical instrument like a ventilator is a very precarious position to take as the amount of safety concerns was also something brought up when teams were in the midst of the pandemic working on rapid response ventilators. There are several other papers exactly like this paper, including the ones cited and published in this journal that did porcine modeling as well. The novelty of this paper is therefore in question and the value that it adds to the journal and to the field I think is questionable. The reference list is woefully short for this kind of important paper and the depth of analysis, risk analysis, part sourcing is lacking. The animal study that was claimed was not detailed with any thoroughness or explanation. This paper is far too short and under developed for the important concepts introduced. There are many other papers, including ones the authors cite that cover this work in better detail, in PLOS ONE no less. Therefore I suggest the editor reject this submission.

Reviewer #2: The overall conclusion of this paper is well stated - this limited study describes the use of a low-cost ventilator that could be used in a crises scenario for short periods (<12hrs) of time. This paper serves as a basis for further study into the low-cost ventilator area of research.

Recommendations for Revision for Author:

There is no description of the anesthetic protocol or management for the porcine model. It is beneficial for the reader to understand what type of anesthetics were used for this study and what effect the anesthetics could have on the parameters evaluated.

Please make sure all acronyms are spelled out during their initial use.

For the lung simulator testing, the duration of testing for each condition is unclear.

Table 3 - Ti and Te for control ventilator were not provided in the chart. Are those values available? The legend of table 3 also appears cut off (no description of what PO2 is)

Recommended citation: King, W. P., Amos, J., Azer, M., Baker, D., Bashir, R., Best, C., ... & Wooldridge, A. R. (2020). Emergency ventilator for COVID-19. PLoS One, 15(12), e0244963.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Response to Referee and Editor Comments on the Paper:

“To Test the Performance of an Open-Source, Low-Cost Ventilator in a Porcine Model”

PLOS One Journal

PONE-D-23-12462

Response to Editor’s Comments:

For this revision, the Editor recommends that we consider the suggestions by the reviewer. This response note forms the main part of the requested accompanying letter. We also note here that we have been able to accommodate all the suggestions proposed by both reviewers. Below is the summary of the suggestions and our responses to them.

Reviewer #1

1. I am a little unclear what the author's expectations were for this article. They claim an open-source design, one that is not thoroughly given or expanded upon. Additionally, the timeline of this topic is misaligned with the reality of current health states. The federal government has removed the state of emergency over COVID-19 and there have been several papers on this or a similar topic. Additionally, proposing an "open source" medical instrument like a ventilator is a very precarious position to take as the amount of safety concerns was also something brought up when teams were in the midst of the pandemic working on rapid response ventilators. There are several other papers exactly like this paper, including the ones cited and published in this journal that did porcine modeling as well. The novelty of this paper is therefore in question and the value that it adds to the journal and to the field I think is questionable. The reference list is woefully short for this kind of important paper and the depth of analysis, risk analysis, part sourcing is lacking. The animal study that was claimed was not detailed with any thoroughness or explanation. This paper is far too short and underdeveloped for the important concepts introduced. There are many other papers, including ones the authors cite that cover this work in better detail, in PLOS ONE no less. Therefore, I suggest the editor reject this submission.

Response: Appreciate the comments and agree with the reviewed on the numerous reports for low-cost ventilators that have been produced during the COVID pandemic. However, we would like to provide reasoning and support on why our LCV study is unique compared to others that have been previously published. In a PubMed search for “low-cost ventilator”, there were 14 relevant published papers in the first 150 searches. Only two papers have a similar design using a solenoid valve to control flow rate. Furthermore, our study is the first in providing physiological gas exchange (e.g. arterial blood gas or ABG) data from an LCV ; in addition, compares the gas exchange data to that of a control ventilator in our animal model. Out of all the 14 LCV studies, physiological gas exchange has not been assessed to our knowledge. Of the two other solenoid valve designs, our LCV provides are more compact, portable design in comparison. Given the scarcity of data using solenoid valves in LCV and our gas exchange data, we feel that our study is novel and will add to the current literature. We also appreciate that more discussion could be added, and we have included that in our revision as well as including more references to support the discussions and our findings as well. The title, introduction and discussion paragraphs have been revised to highlight what stands out in our design in addition to emphasizing that we have physiological gas exchange data that has not been reported previously for LCV during the COVID pandemic.

Reviewer #2

1. There is no description of the anesthetic protocol or management for the porcine model. It is beneficial for the reader to understand what type of anesthetics were used for this study and what effect the anesthetics could have on the parameters evaluated.

Response: Agree. Added in the methods section.

2. Please make sure all acronyms are spelled out during their initial use.

Response: Agree. Our manuscript has been revised.

3. For the lung simulator testing, the duration of testing for each condition is unclear.

Response: Agree. Our manuscript has been revised.

4. Table 3 - Ti and Te for control ventilator were not provided in the chart. Are those values available? The legend of table 3 also appears cut off (no description of what PO2 is)

Response: The Ti and Te chosen for this study resulted from calibration using our lung simulator. This provided us with the tidal volumes demonstrated in the porcine model. Legend for Table 3 has been corrected.

5. Recommended citation: King, W. P., Amos, J., Azer, M., Baker, D., Bashir, R., Best, C., ... & Wooldridge, A. R. (2020). Emergency ventilator for COVID-19. PLoS One, 15(12), e0244963.

Response: Agree. Citation has been added.

Attachments
Attachment
Submitted filename: LCV Response Letter.docx
Decision Letter - Mattias Günther, Editor

PONE-D-23-12462R1Development and Performance Evaluation of a Solenoid Valve Assisted Low-Cost Ventilator on Gas Exchange and Respiratory Mechanics in a Porcine Model.PLOS ONE

Dear Dr. Cho,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by May 16 2024 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Mattias Günther

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #3: (No Response)

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #3: The paper, “Development and Performance Evaluation of a Solenoid Valve Assisted Low-Cost Ventilator on Gas Exchange and Respiratory Mechanics in a Porcine Model.” has been presented. Technically, the article presents good scientific information and may be accepted for publication after clarifying some unclear and/or missing points:

1. Write the applications in the introduction section.

2. Add quantitative results in the abstract.

3. Provide references to support your results.

4. The results section needed to be improved as it seems to be a laboratory report rather than a scientific investigation.

5. Reconstructed the conclusion part.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #3: Yes: Dr. Rajkumar Sivanraju

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 2

Figure 2 was revised and the "Dell" Logo is now covered and blacked out. Thank you

Attachments
Attachment
Submitted filename: LCV Response Letter2.docx
Decision Letter - Mattias Günther, Editor

Development and Performance Evaluation of a Solenoid Valve Assisted Low-Cost Ventilator on Gas Exchange and Respiratory Mechanics in a Porcine Model.

PONE-D-23-12462R2

Dear Dr. Cho,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. If you have any questions relating to publication charges, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Mattias Günther

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #3: The paper, “Development and Performance Evaluation of a Solenoid Valve Assisted Low-Cost Ventilator on Gas Exchange and Respiratory Mechanics in a Porcine Model” has been presented. The revised manuscript is improved a lot and the authors are responded well for the queries raised by me in my previous review. Now I accept this manuscript for publication.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #3: Yes: Dr. Rajkumar Sivanraju

**********

Formally Accepted
Acceptance Letter - Mattias Günther, Editor

PONE-D-23-12462R2

PLOS ONE

Dear Dr. Cho,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

If revisions are needed, the production department will contact you directly to resolve them. If no revisions are needed, you will receive an email when the publication date has been set. At this time, we do not offer pre-publication proofs to authors during production of the accepted work. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few weeks to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff