Peer Review History

Original SubmissionSeptember 28, 2022
Decision Letter - Guocan Yu, Editor

PONE-D-22-26863Diagnostic performance of the GenoType MTBDRplus VER 2.0 line probe assay for the detection of isoniazid resistant Mycobacterium tuberculosis in EthiopiaPLOS ONE

Dear Dr. Shewki Moga Siraj,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by 30 Dec 2022. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Guocan Yu

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf.

2. Please provide additional details regarding participant consent. In the ethics statement in the Methods and online submission information, please ensure that you have specified what type you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed). If your study included minors, state whether you obtained consent from parents or guardians. If the need for consent was waived by the ethics committee, please include this information.

3. We note that you have stated that you will provide repository information for your data at acceptance. Should your manuscript be accepted for publication, we will hold it until you provide the relevant accession numbers or DOIs necessary to access your data. If you wish to make changes to your Data Availability statement, please describe these changes in your cover letter and we will update your Data Availability statement to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: No

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This study followed scientific methods to analyze observations and to present results successfully, study design is more likely to be a cross-sectional study, a design fits for studying test's performance.

Reviewer #2: This manuscript assesses the Diagnostic performance of the GenoType MTBDRplus VER 2.0-line probe assay for the detection of isoniazid resistant Mycobacterium tuberculosis in Ethiopia.

The subject remains a point of interest, however looking into the current form the manuscript seems to have major needed correction, mostly clarifying and updating the mix-ups between the methods and result section. The discussion part is well written, and a few minor edits are required there too.

Reviewer #3: Dear Editorial manager

Thank you for giving chance to review this manuscript. It is important, especially to observe the detection level of INH resistance by LPA among phenotypically Hr-TB and MDR-TB isolate. However, it needs major revision. Here below are my comment for Authors.

1. In line 40 ---the katG 315 mutation was observed in 94.3% (n=33) of MDR-TB by phenotypes, but in the Table 5 it is written, 97% similar for katG 315 mutation was observed in 71% (n=44) of Hr-

40 TB phenotypes, but not in the table.

2. Line 41-and 42 – need revision. The percent is not correct

3. Why Table 1and 2 sensitivity and NPV differed while the number of cases were similar. What is there difference also? This make also confusion to me to get the message in the result and discussion part.

4. Table 4 calculation is not right please revised it.

5. Table 5 cite reference for Nucleotide change

6. In the figure 1 among 62 INH resistance 14 were not INH resistance, but in the table 1and 2- written as 12 was not INH resistance by the LPA

7. line 232 and 233 need reference

8. What about patient characteristics?

9. While 36 phenotypically INH resistant isolates were required for assessing the sensitivity of GenoType MTBDRplus LPA VER 2.0. The Author used 62 INH resistant isolates which is good but based on what proportion the Author compere with Pan-susceptible. It is also better to calculate sample size for phenotypically MDR-TB and INH resistant isolates separately.

10. line 270-275 what is it’s important to discuss here about RIF resistant mutant

11. The study doesn’t show the diagnostic accuracy of LPA. Sensitivity, Specificity, PPV and NPV are different from accuracy

12. Why the Author left to interpret the presence and absence of wild-type probes using the LPA card.

13. It is not clear Why the Author recommend additional genes to be evaluated for INH resistance? Is that because of the detection of INH resistance was lower among Hr-TB cases compared to MDR-TB isolates. If so first it will better to calculate sample size for phenotypically INH resistant proportional and MDR-TB separately then compare each other.

14. Define INH, HPV and NPV under each table

15. No acknowledgment

16. add Author contribution

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Layth Al-Salihi

Reviewer #2: No

Reviewer #3: Yes: Yeshiwork Abebaw Asaye

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachments
Attachment
Submitted filename: Comment to Diagnostic performance of the GenoType MTBDRplus VER 2.0 LPA PLOS ONE.docx
Revision 1

Reviewer #1 I have incorporated all of your suggestions into my revision. Thank you for the detailed and useful comments.

Reviewer #2 I have incorporated all of your suggestions into my revision. Thank you for the detailed and useful comments.

Dear: Reviewer #3

We have addressed your comments across the manuscript. Most of numerical related comments were very useful, so we have now applied the proposed corrections in our manuscript. Given that the paper is aimed on evaluation of diagnostic assay, we decided not to include demographic data of patients. We would like to thank you again for sparing the time to write so many detailed and useful comments. I look forward to hearing from you

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Guocan Yu, Editor

PONE-D-22-26863R1Diagnostic performance of the GenoType MTBDRplus VER 2.0 line probe assay for the detection of isoniazid resistant Mycobacterium tuberculosis in EthiopiaPLOS ONE

Dear Dr. Shewki Moga Siraj,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by 2023.2.2. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Guocan Yu

Academic Editor

PLOS ONE

Additional Editor Comments:

Dear authors,

I find that you did not respond to reviewer 3's comments, please clarify. Please indicate the specific line number after the modification. Removal of modification marks, modifications can be marked in red. Please revise and resubmit, including all reviewers' comments.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 2

Reviewer 1: I have incorporated all of your suggestions into revised manuscript. They were very helpful. Thank you

Reviewer 2: I have incorporated all of your suggestions into revised manuscript. They were very helpful. Thank you

Reviewer 3: I have addressed most points raised into my revised manuscript and put explanation for recommendation which do not require revision. They were very helpful. Thank you

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Guocan Yu, Editor

PONE-D-22-26863R2Diagnostic performance of the GenoType MTBDRplus VER 2.0 line probe assay for the detection of isoniazid resistant Mycobacterium tuberculosis in EthiopiaPLOS ONE

Dear Dr. %Shewki Moga Siraj,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by 2023.4.15. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Guocan Yu

Academic Editor

PLOS ONE

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

Reviewer #4: (No Response)

Reviewer #5: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: Yes

Reviewer #4: No

Reviewer #5: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

Reviewer #4: No

Reviewer #5: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: No

Reviewer #4: Yes

Reviewer #5: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

Reviewer #4: Yes

Reviewer #5: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: This manuscript assesses the Diagnostic performance of the GenoType MTBDRplus VER 2.0-line probe assay for the detection of isoniazid resistant Mycobacterium tuberculosis in Ethiopia.

All the major and minor comments have been adequately addressed, and I have no further comments. I recommend publishing this article.

Reviewer #4: Summary

This was a retrospective study that aimed at assessing the performace of GenoType MTBDRplus version 2.0 (Hain ver. 2) in detecting isoniazid resistance among M tuberculosis complex (MTBC) clinical isolates. Samples were taken from the third Ethiopian national drug resistance survey conducted between August 2017 and December 2019. Phenotypic drug susceptibility testing (DST) with MGIT determined bacillary susceptibility to isoniazid. Phenotypic DST results were available for 1552 MTBC isolates out of 2268 isolates. Among these 1552 isolates, 62 were isoniazid-resistant and rifampicin-susceptible, and 35 were multidrug-resistant (MDR, resistant to both isoniazid and rifampicin). These 97 isoniazid-resistant isolates together with 40 randomly selected isoniazid-susceptible isolates were included in the analysis. They found that Hain ver. 2 was 77.4% sensitive and 100% specific for detecting isoniazid resistance among rifampicin-susceptible isolates, and 94.3% sensitive and 100% specific for detecting isoniazid resistance among MDR isolates. Additionally, they also calculated positive and negative predictive values (PPV and NPV) using the included samples. They concluded that Hain ver. 2 was significantly more sensitive for detecting isoniazid resistance among rifampicin-susceptible isolates than MDR isolates. They also suggested that, using Hain ver.2 as stand-alone test for detecting isoniazid resistance among rifampicin-susceptible cases may miss isoniazid-resistant cases, who would then be given an ineffective treatment regimen. They recommended that additional genes be evaluated for isoniazid resistance to improve the sensitivity of Hain ver.2 for rifampicin-susceptible TB cases.

Major comments

1. This study did not provide any information about the respective prevalence of isoniazid resistance among rifampicin-susceptible isolates and rifampicin-resistant isolates.

2. It is uncertain whether the 40 randomly selected isoniazid-susceptible isolates were rifampicin-susceptible. It may be better to evaluate the specificity of Hain ver. 2 for isoniazid resistance using isoniazid-susceptible isolates with matched bacillary susceptibility to rifampicin among rifampicin-susceptible and rifampicin-resistant isolates, respectively.

3. PPV and NPV were erroneously estimated in this study without due consideration of the pre-test probability of isoniazid resistance. The study sample of 62 isoniazid-resistant and rifampicin-susceptible isolates plus an arbitrarily determined number of randomly selected isoniazid-susceptible isolates would not give any information about the pre-test probability of isoniazid resistance among rifampicin-susceptible isolates. Neither would the study sample of 35 MDR isolates plus the same batch of isoniazid-susceptible isolates give any information about the pre-test probability of isoniazid resistance among rifampicin-resistant isolates.

4. The performance of Hain ver.2 is heavily dependent on the pre-test probability of isoniazid resistance. This is why WHO has recommended that phenotypic DST should be done when Hain ver. 2 does not detect isoniazid resistance and the pre-test probability of isoniazid resistance is high. In general, isoniazid resistance is highly correlated with rifampicin resistance. It may be reasonable to expect a prevalence of isoniazid resistance in the range of 5-15% among rifampicin-susceptible TB cases, and about 90% among rifampicin-resistant TB cases. In a theoretical population of 1000 rifampicin-susceptible TB patients, with a pre-test probability of 15% for isoniazid resistance, if Hain ver. 2 is 77.4% sensitive and 100% specific, Hain ver. 2 would detect 116 cases and miss 34 cases. There are no false-positive cases and 850 true-negative cases. PPV is 100% and NPV is 96.2%. On the other hand, in a theoretical population of 1000 rifampicin-resistant TB patients, with a pre-test probability of 90% for isoniazid resistance, if Hain ver. 2 is 94.3% sensitive and 100% specific, Hain ver. 2 would detect 849 cases and miss 51 cases. There are no false-positive cases and 100 true-negative cases. PPV is 100% and NPV is 66.2%. Thus, with reference to sensitivity and specificity values found in this study, Hain ver. 2 has excellent PPV for isoniazid-resistant case regardless of rifampicin susceptibility, and a more reliable negative result when the patient has rifampicin-susceptible TB than when the patient has MDR-TB.

5. It has been shown that, if the TB treatment regimen is properly formulated with inclusion of ethambutol, missing initial isoniazid resistance does not affect the treatment success rate among rifampicin-susceptible TB cases treated with rifampicin, ethambutol and pyrazinamide for six months in comparison with drug susceptible TB treated with the standard six-month regimen. The situation is very different when rifampicin resistance is missed.

6. There may be a typo error regarding the P value (line 39 in abstract, line 172 in results). The P value was probably 0.04 rather than 0.4.

7. In summary, study findings did not appear to substantiate the conclusion.

Minor comments

1. The Clopper-Pearson exact method was probably used in estimating the 95% confidence interval. The modified Wald method might be more accurate.

Reviewer #5: The authors have satisfactorily addressed the previous reviewers’ comments. The revised manuscript is clearer and much improved compared to the original submission. I recommend accepting the manuscript for publication. I have added a couple of very minor stylistic comments for consideration.

Line 69: Please add “the” before “added.”

Lines 78-79: Suggest “…, can additionally be performed on smear-negative clinical specimens given its higher sensitivity than version 1.0.”

Line 116: Replace “was” with “were.”

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: No

Reviewer #4: No

Reviewer #5: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 3

Reviewer #2: We very much appreciate the time and effort for reviewing our manuscript and constructive feedback. Thank you for recommending our manuscript for publication

Reviewer #4: We very much appreciate the time and effort for reviewing our manuscript and constructive feedback. We have incorporated your suggestion into our revision and thank you for consideration of our revised manuscript

Reviewer #5: We very much appreciate the time and effort for reviewing our manuscript and constructive feedback. We have incorporated all of your suggestion into our revision. Thank you for recommending our manuscript for publication

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Guocan Yu, Editor

Diagnostic performance of the GenoType MTBDRplus VER 2.0 line probe assay for the detection of isoniazid resistant Mycobacterium tuberculosis in Ethiopia

PONE-D-22-26863R3

Dear Dr. Shewki Moga Siraj,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Guocan Yu

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Formally Accepted
Acceptance Letter - Guocan Yu, Editor

PONE-D-22-26863R3

Diagnostic performance of the GenoType MTBDRplus VER 2.0 line probe assay for the detection of isoniazid resistant Mycobacterium tuberculosis in Ethiopia.

Dear Dr. Siraj:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Guocan Yu

Academic Editor

PLOS ONE

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .