PONE-D-22-19808A portable feedback-controlled pump for monitoring eye outflow facility in conscious ratsPLOS ONE Dear Dr. Passaglia, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Oct 24 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Ted S Acott, PhD Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. As part of your revision, please complete and submit a copy of the Full ARRIVE 2.0 Guidelines checklist, a document that aims to improve experimental reporting and reproducibility of animal studies for purposes of post-publication data analysis and reproducibility: https://arriveguidelines.org/sites/arrive/files/Author%20Checklist%20-%20Full.pdf (PDF). Please include your completed checklist as a Supporting Information file. Note that if your paper is accepted for publication, this checklist will be published as part of your article. 3. Thank you for stating the following in the Competing Interests section: "I have read the journal's policy and the authors of this manuscript have the following competing interests: U.S. patents 9022968, 9314375, and 10758408." Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests). If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf. Additional Editor Comments: reviewers were generally favorable to the study, but both had several concerns that would greatly improve the manuscript - please carefully address all reviewer concerns [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: N/A Reviewer #2: I Don't Know ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: No Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Review: A portable feedback-controlled pump for monitoring eye outflow facility in conscious rats This paper describes a small, portable device for continuously measuring IOP and outflow facility in conscious, moving rats. Most systems require anesthesia and do not continuously monitor IOP and outflow facility over long time periods, so this device has great potential in expanding the type of ocular studies we can do in animals, such as circadian studies of changes in outflow facility and facility measurements without impact from anesthetic or other changes to physiology. Thus, it will garner interest from readers, and I am generally positive about its publication. Overall, this system could be really useful for researchers. The long-term measurements in awake, moving animals shown in Figures 7 and 8 are really interesting, and show great potential for work studying rhythmic changes in aqueous humor dynamics. It would be nice to see this work eventually used in an established glaucoma or ocular hypertensive rat model to further validate the IOP and facility measurement capabilities. Below are detailed comments. My main concern with this paper is that the authors have not put their data in context. Others have measured outflow facility in the rat eye, including the high-fidelity measurements of Feola et al. and previous measurements by the current authors, plus possibly others that I am not aware of. These should be compared with the data gathered here. Similarly, the flow-pressure graphs (Fig 5B) can be used to get an estimate of aqueous inflow rate by extrapolating the lines back to (estimated) EVP, at which point the (negative) pump flow rate should equal inflow rate. Obviously, there are assumptions underlying such a calculation, but it would be good to see what values the authors get. Measuring inflow rate in small eyes like the rat is very technically challenging and having more data points, with suitable caveats, is useful. Finally, what is the success rate? The authors show 3 animals that seem to have worked, but how many were tried? I know there was some information on this in previous papers, but now that the authors have more experience, a quantitative update should be given. Other comments: The first two figures provide a good overview of the device size, setup, and use. The compact nature and basic components of the device are clear from figure 1A, and it’s great that this is shown at the start of the paper. I may have liked to see the device with the cannula attached to a conscious rat as well, but this is partially addressed in figure 8. Request: please include a photo in Fig 1 of the device in use with a rat. One thing the authors do not comment on is how bubbles are avoided/dealt with. Bubbles are the bane of all who perfuse eyes, and some comments on this important topic are needed. The model shown in Figure 2A is useful to demonstrate Goldmann’s equation and visualize the device setup in more detail. It also helps with interpretation of the figures throughout the paper, as readers can reference how each variable in the future graphs relates to the rest of the system. However, I believe there is an error in Fig 2A: the R_G element should not be there. I understand this is to represent viscoelastic globe stretching, but that is not a fluid resistance element per se. Globe stretching can be represented by the compliance C_G. In any case, R_G does not seem to be used anywhere in the paper, so deleting it should not change anything so far as I can tell. While the assessment of the pump/system parameters is important (Figs 2B-D), I have some concerns. It would help if the dashed line were labeled as nominal load in Fig. 2B. More importantly, the slope of the load line in Fig 2B (and the operating range in Figs 2C and D) are not in the right range. I understand that this refers to the situation before the internal flow limiter is used, but Fig 2C somehow implies that the load that the pump sees in operation is 0.1 mmHg min/ul, which is far from the resistance of the eye (by 1-2 orders of magnitude). Fig 2C is similarly misleading. For Fig 2D, we only care about the very bottom part of the graph, right? The other lines are useless in practice, so a zoomed view of the bottom of the graph (with more data) would be useful. This whole section was confusing and should be reworked/better explained/changed. Overall, the system validation seems sound (Fig 3), showing the repeatability of the device in a benchtop setup before introducing the noise that results from measuring a live, conscious animal. A minor critique is that in Figure 3B, the measurement points for different days are all on top of each other and mostly indistinguishable, I wonder if colors would be better or if there is another way to visualize this data more clearly. Minor comment: when comparing the new system to the independently measured IOP values in anesthetized animals (Fig. 4) and measuring IOP with the new system in conscious animal (Fig. 5), the IOP as measured by the system is referred to as PE (to match the model in Fig. 2, which is helpful), but in later figures looking at diurnal patterns of IOP in rats (Fig. 6-8), it is labeled as IOP. This should be made consistent. Additionally, it’s not clear what Figure 5C is adding. It provides information for some of the other animals measured, but it is not easy to distinguish individual lines by the different shapes. It is not explicitly stated that the shapes are different animals, and it’s not easy to see how many are shown. This panel should be improved. Further, Fig. 5D would be better as a box and whisker plot rather than a bar plot (although keeping the individual data points and connecting lines). Figure 6 demonstrates the capabilities of the device, especially by showing how well the feedback algorithm functions in a moving, awake animal in 6B. The text claims that the amount of variation shown here in outflow facility across a few hours (23, 27, 28 nl/min*mmHg) is “fairly reproducible” (line 340), but does not cite other studies that justify this claim – see also major comment above. The data in figure 7 showing the differences in IOP and facility between night and day is really interesting, and I think it’s a highlight of this paper. It shows how necessary this device is because there’s so much more we can learn about diurnal changes in outflow using this device. However, it is confusing that the setup of Fig. 7A is different from B and C (light/dark intervals are inconsistent). Please fix. Same for the diurnal curves in Fig 8. While I appreciate that Figure 8 shows the cannula located in different locations on the rat eye, and that the diurnal patterns shown in 7 are still maintained in these other animals with the cannulas located at different points, it is almost impossible to actually see the cannulas in the pictures of rat eyes in figure 8A. In the middle panel it is completely impossible – presumably some of the cannula was in the AC. Zoomed in views/better labeling are needed. The discussion mentions measuring ocular viscoelasticity (line 441), and although this is included in the system model, data isn’t shown validating this measurement. Suggest just deleting this part. Reviewer #2: This is a description of a new system for measuring aqueous humor dynamics in awake rats. The basic set up consists of a small silicone tube implanted in the anterior chamber that is connected externally to a “cage-side” apparatus capable of monitoring IOP and pumping balanced salt solution into the eye at microfluidic rates equivalent to aqueous humor formation or flow. These can be controlled through feedback from the measured IOP. For validation, the authors have used anesthetized animals to directly compare IOP and flow readings obtained by their system with independent measurements made through a secondary cannula placed in the eye. In implanted animals, they show the response of pressure and flow to paradigms of, respectively, constant flow and constant pressure--approaches that are commonly used in aqueous humor dynamics studies. Responses appear to be in line with what would be expected. The system is then used to obtain outflow facility readings in a small number of awake animals over time. They find evidence of a diurnal fluctuation in IOP equivalent to reports in the literature (with night-time IOP at about 25 and day-time of 15). Interestingly, they also report that conventional outflow facility also has a diurnal variation, being higher during the day and lower at night. Up to now such measurements have not been possible as existing systems require the use of either general anesthesia or the need for studying enucleated eyes. Generally, this is a difficult manuscript to read. Much of this is due the use of a large number of acronyms, many of which do not appear intuitive, or are inconsistent. For example, CG stands for “globe wall compliance”, whereas C is generally used to indicate facility in some parts of this manuscript and generally in the aqueous humor dynamics literature. Also, line 97, they refer to “PF” mode. It’s not clear what this is, although it appears to be what they earlier define as “CP” (line 64). They persist in using this “PF” designation throughout the manuscript, without ever applying a definition. The authors need to carefully evaluate their acronyms, make sure they are consistent and use them as little as possible in order to minimize the amount of work they are asking of the reader. While a table providing definitions would be somewhat helpful, it would still be extremely cumbersome. A strong suggestion would be, in addition to a table, to spell out the terms in full in the body of the manuscript, without using these acronyms, and reserve them for the equations and figure legends. While this would increase manuscript length to some degree, the improved readability would be worth the space. Line 108, they refer to figure 2A as a “dynamic version of the Goldmann equation”. As presented, it assumes that the reader is thoroughly familiar with this equation. Restating the meaning of the Goldmann equation would make it easier to understand the parallels between the different components of their eye model and the components of the equation. In their description of the apparatus, they refer multiple times to the “pump duty cycle”. This should be defined to improve clarity. The description of the tubing implantation technique is very cryptic, referring to previously published reports. It is not entirely clear just how the silicone tubing is held in place in order to prevent movement and extrusion from the eye. The statement that they use “half-thickness scleral sutures” is difficult to believe, given that the rat sclera is exceedingly thin and the size suture and needle are not stated. Important features of the outcome of this technique that should be presented include numbers of animals implanted and success rate. Additional factors that are not presented include complications from the implantation itself, including incidence of bleeding and inflammation or infection, and how long these tubes might remain functional. The authors present results from a few animals, but the future utility of this potentially promising method will depend heavily on the reliability of these implantations. Line 225 refers to a method they have developed to reduce noise in IOP and flow measurements in awake animals “by recursively regressing raw data”. A short explanation of what this does, rather than simply presenting the mathematical equations, would improve understanding of what they are doing here. Why they performed this in every 2 hour or every 6 hour intervals is not clear. Additionally, there is no indication of numbers of animals used for each approach. In line 233, they state that “resting IOP was checked by tonometry”. They do not state if these are done with the animals awake or not. If done awake, how do they account for the possibility that restraining animals for pressure measurement (particularly with tether attached) might result in artificially high pressure readings? They do not mention what the results of these tonometer readings were. The legend for Figure 4 appears to have an error, in that it lists 4 panels: A,B,C,D, but the text and figure itself have only 3 panels. It appears that what they call panel B (in the legend) should be the right hand portion of panel A. This needs to be corrected. For this figure, and for figure 5 the black and gray traces need to contrast more. They’re very hard to tell apart. Lines 313, 332 and 544, refer to reported values for facility and various factors that may lead to variability in IOP, respectively. In all cases, the authors just reference their own work. They should include the work of others, such as that of Crawford Downs, who has published telemetry IOP data in monkeys. References to work independent of that of the authors would strengthen the validity of the relevance of values obtained by their current system. Figure 7 shows interesting data that suggests there may be a diurnal rhythm to conventional outflow facility, with flow and facility being lower at night (dark phase) and higher during the light phase. If true, this would be a new and important finding. As the authors mention, aqueous humor production is known to have a circadian rhythm, being higher during the dark. While the ability to determine flow and facility in awake animals would be unique, the variable rate of aqueous production is a factor that standard aqueous humor dynamics studies do not have to contend with. That said, it is not clear how their algorithm accounts for this factor and that this might not be influencing this result. The presentation of the dark and light phase is not consistent both within figure 7 and between Figure 7A and Figure 8. In the former the dark phase is presented in the middle; elsewhere it is the light phase that is in the middle. The need for this switch is not clear and it makes it more difficult to grasp these results, particularly when looking at these in terms of IOP, flow and facility, some of which are inverse of each other. Regarding Figure 8, the quality of anterior segment photos should be better. Even with this, the top photo (Y45) shows definite blood subconjunctivally, as well as a clot and likely fibrin on the tube tip, leading to adhesions and pupil distortion. This raises concerns that tube insertion was not without complications and that this and accompanying inflammation could possibly be affecting their readings. The middle photo (Y72), in which the tube is stated to be under the iris, one can see the tube before it goes into the pupil, but there is also several areas of what look like fibrin as well as possible inflammatory debris and some adhesion of the pupil to either the lens surface or the tube. In the bottom image (Y77) it is difficult to actually see the tube but it appears to be visible in the pupil. In fact it may actually be behind the iris here too, and there appears to be evidence of pupil distortion and likely adhesions. Again, one wonders what the effects of these changes might be on their system readings and the longevity of the system as a whole. Line 366 states that “it can be seen that prolonged perfusion had minimal impact on implanted eyes”. Given that in these animals the longest duration of observation was 7 days, this is not particularly prolonged and this statement should be modified. Finally, the statistics applied to these 3 eyes’ data to show a significant difference between dark and light phase is not clear, including just how the n’s equaling the number of days were used and what they consist of. The Discussion very reasonably points out some of the unique aspects of this system and its potential advantages. While the observation of a diurnal change in facility is potentially interesting, the number of animals it’s demonstrated in here is small, and there are no data presented to suggest that this going to be a consistent finding. Line 433 states that this report shows that “ocular hypertension can be induced in animal models without experimentally damaging outflow pathways.” While this is a potential, the concerns raised above regarding images of Figure 8 would suggest that outflow damage from the surgical implantation of the tube or its presence over a long period of time is a very real possibility. More animals, with documentation of tube failures (success rate) and complications will still be needed in order for this observation to be more convincing. The authors also mention the potential of using this implanted system as a way of delivering therapeutics or performing experimental pharmacologic manipulations. While this is true, it is not clear just how dosage would be standardized or monitored, as the drug would have to first go through the tubing before it gets to the eye, and the rate at which this would occur is unknown. This could certainly be considered as an important area for future development. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 1 |
|
PONE-D-22-19808R1A portable feedback-controlled pump for monitoring eye outflow facility in conscious ratsPLOS ONE Dear Dr. Passaglia, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Jan 26 2023 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Ted S Acott, PhD Academic Editor PLOS ONE Journal Requirements: Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. Additional Editor Comments (if provided): One reviewer had minor suggestions that would greatly strengthen this important paper - please address them [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: (No Response) Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors have been generally responsive and the manuscript is improved. However, there are still a few concerns as follows. 1. R_G element in Figure 2A. We asked that this be removed but the authors wanted to retain it, stating that it must be present because "the globe does not instantaneously gain or lose volume following step changes in pressure". While it is true that the globe does not instantaneously change in volume, the inference that R_G is required to explain this behavior is categorically false. More specifically, the globe compliance interacting with the outflow resistance and/or needle resistance creates a RC time constant that can explain globe filling dynamics. The only reason I can imagine to include R_G is to represent the viscoelastic response of the corneoscleral shell, in which case I guess the shell is being represented as a Maxwell body. I doubt the system has the sensitivity to measure this. Rather than having this extremely confusing element in the diagram which is never used, why not just delete it??? 2. Suggest moving Fig. 2B-C to supplemental. Characterization of the pump with arbitrary load is not essential for readers to understand the paper and is confusing given that the load shown is not relevant to aqueous humor drainage in rats. It may be useful for a small subset of people interested in the specific pump dynamics, but appearing in the supplemental material is fine for that purpose. Reviewer #2: the authors have adequately addressed my initial comments ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
A portable feedback-controlled pump for monitoring eye outflow facility in conscious rats PONE-D-22-19808R2 Dear Dr. Passaglia, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Ted S Acott, PhD Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: |
| Formally Accepted |
|
PONE-D-22-19808R2 A portable feedback-controlled pump for monitoring eye outflow facility in conscious rats Dear Dr. Passaglia: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Ted S Acott Academic Editor PLOS ONE |