PONE-D-22-20756Single-Layered Compact Broadband Circularly Polarized Implantable Antenna for Medical ApplicationsPLOS ONE

Dear Dr. Ahmmad,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

ACADEMIC EDITOR:Authors should revise the manuscript addressing all the issues raised by reviewers. Note that you do not need to cite any suggested reference articles by the reviewer(s) if you find them not-related to your work.

Please submit your revised manuscript by Oct 07 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Muhammad Zubair

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following financial disclosure: 

"NO"

At this time, please address the following queries:

a) Please clarify the sources of funding (financial or material support) for your study. List the grants or organizations that supported your study, including funding received from your institution. 

b) State what role the funders took in the study. If the funders had no role in your study, please state: “The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.”

c) If any authors received a salary from any of your funders, please state which authors and which funders.

d) If you did not receive any funding for this study, please state: “The authors received no specific funding for this work.”

Please include your amended statements within your cover letter; we will change the online submission form on your 

3. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter.

4. Please amend either the title on the online submission form (via Edit Submission) or the title in the manuscript so that they are identical.

5. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Additional Editor Comments:

Authors should revise the manuscript addressing all the issues raised by reviewers. Note that you do not need to cite any suggested reference articles by the reviewer(s) if you find them not-related to your work.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This article presents a single-layered compact broadband circularly polarized implantable antenna for medical applications. This antenna is intended to function in the ISM Band, which operates at 2.45GHz. Promising results have been achieved and well discussed in the well-organized manuscript. So, the results have been experimentally validated, and highlighted by providing a fair comparison with state-of-the-art. Although the concept and idea of this work were found very interesting and they seem attractive for the antenna and propagation society, authors are encouraged to address the following comments prior to final recommendation.

1) How can the bandwidth of the antenna be further improved?

2) What is the advantage of having dual functionality in free space and in human tissue?

3) Some important Abbreviations should be fully defined.

4) To design the antenna structures for implantable applications, the compact size is very important, therefore, authors are requested to mention some miniaturization methods in the introduction section along with proper references. Below are some helpful suggestions.

"A Comprehensive Survey of "Metamaterial Transmission-Line Based Antennas: Design, Challenges, and Applications"", IEEE Access, vol. 8, pp. 144778-144808, 2020.

“Miniaturized Planar-Patch Antenna Based on Metamaterial L-shaped Unit-Cells for Broadband Portable Microwave Devices and Multiband Wireless Communication Systems" IET Microwaves, Antennas & Propagation, Volume 12, Issue 7, 13 June 2018, p. 1080 – 1086.

“Compact Single Layer Travelling-Wave Antenna Design Using Metamaterial Transmission-Lines" Radio Science, Volume 52, Issue 12 December 2017, Pages 1510–1521.

“Periodic Array of Complementary Artificial Magnetic Conductor Metamaterials-Based Multiband Antennas for Broadband Wireless Transceivers” IET Microwaves, Antennas & Propagation, Volume 10, Issue 15, 10 December 2016, p. 1682 – 1691.

“Miniature CRLH-based ultra wideband antenna with gain enhancement for wireless communication applications”, ICT Express, Volume 2, Issue 2, June 2016, Pages 75–79.

"New Compact Antenna Based on Simplified CRLH-TL for UWB Wireless Communication Systems", International Journal of RF and Microwave Computer-Aided Engineering, Volume 26, Issue 3, March 2016, pages: 217–225.

“Metamaterial-Based Antennas for Integration in UWB Transceivers and Portable Microwave Handsets” International Journal of RF and Microwave Computer-Aided Engineering, Volume 26, Issue 1,January 2016, pages: 88–96.

"New Compact Printed Leaky-Wave Antenna with Beam Steering", Microwave and Optical Technology Letters, Volume 58, Issue 1, January 2016, Pages: 215–217.

"UWB Antenna Based on SCRLH-TLs for Portable Wireless Devices", Microwave and Optical Technology Letters, Volume 58, Issue 1, January 2016, Pages: 69–71.

"Composite Right-Left-Handed-Based Antenna with Wide Applications in Very-High Frequency-Ultra-High Frequency Bands for Radio Transceivers" IET Microwaves, Antennas & Propagation, Volume 9, Issue 15, 10 December 2015, p. 1713 – 1726.

“Bandwidth and Radiation Specifications Enhancement of Monopole Antennas Loaded with Split Ring Resonators” IET Microwaves, Antennas & Propagation, Volume 9, Issue 14, 19 November 2015, p. 1487 – 1496.

“The Resonating MTM Based Miniaturized Antennas for Wide-band RF-Microwave Systems”Microwave and Optical Technology Letters, Volume 57, Issue 10, pages 2339–2344, October 2015.

“Novel UWB Miniaturized Integrated Antenna Based on CRLH Metamaterial Transmission Lines” AEUE Elsevier- International Journal of Electronics and Communications, Volume 69, Issue 8, August 2015, Pages 1143–1149.

“Compact Antenna based on a Composite Right/Left Handed Transmission Line” Microwave and Optical Technology Letters, Volume 57, Issue 8, pages 1785–1788, August 2015.

“Printed planar patch antennas based on metamaterial”, International Journal of Electronics Letters, Volume 2, Issue 1, Jan 2014, pp 37-42.

“Design and Modeling of New UWB Metamaterial Planar Cavity Antennas with Shrinking of the Physical Size for Modern Transceivers,” International Journal of Antennas and Propagation, vol. 2013, Article ID 562538, 12 pages, 2013. doi:10.1155/2013/562538.

4) Some grammatical mistakes should be incorporated.

5) All figures should be properly cited in the main text.

6) The quality of the figure should be further improved.

7) Authors need to highlight the novelty of the proposed work.

8) Some old references should be replaced with the latest ones.

9) Reference part can be improved as per above mentioned suggestions.

Reviewer #2: In the introduction chapter, I suggest to introduce some more text and a few more references related to the actual practical applications (existing or future) of the proposed antenna solution. In other words to further elaborate on a few specific biomedical cases where the proposed antenna can be used. Also, add some further explanation about how and where in the human body the proposed antenna is implanted. Otherwise, I recommend the paper for publication.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 

Revision 1

PONE-D-22-20756

Single-Layered Compact Broadband Circularly Polarized Implantable Antenna for Medical Applications

PLOS ONE

………………………………………………………………………………………...

Reviewer Comments to the Author

Reviewer 1

This article presents a single-layered compact broadband circularly polarized implantable antenna for medical applications. This antenna is intended to function in the ISM Band, which operates at 2.45GHz. Promising results have been achieved and well discussed in the well-organized manuscript. So, the results have been experimentally validated, and highlighted by providing a fair comparison with state-of-the-art. Although the concept and idea of this work were found very interesting and they seem attractive for the antenna and propagation society, authors are encouraged to address the following comments prior to final recommendation.

Comment_1: How can the bandwidth of the antenna be further improved?

Response_1: We are thankful to the reviewer for this question. The bandwidth of the antenna can be further improved by varying the slots and by increasing the size of the antenna. Since, for the applications of biomedical implants, it is necessary to keep the size of the antenna as compact as possible.

……………………………

Comment_2: What is the advantage of having dual functionality in free space and in human tissue?

Response_2: The advantage of having dual functionality of the antenna both in free space and in human tissue is that there is no any need to design an external antenna to communicate with the antenna inside human body. The antenna will work for both at the same time. That will reduce the cost, time and efforts.

……………………………

Comment_3: Some important Abbreviations should be fully defined.

Response_3: We are thankful to the reviewer for pointing out the concerns related to the abbreviations. It is done in the revised manuscript.

……………………………

Comment_4: To design the antenna structures for implantable applications, the compact size is very important, therefore, authors are requested to mention some miniaturization methods in the introduction section along with proper references. Below are some helpful suggestions.

"A Comprehensive Survey of "Metamaterial Transmission-Line Based Antennas: Design, Challenges, and Applications"", IEEE Access, vol. 8, pp. 144778-144808, 2020.

“Miniaturized Planar-Patch Antenna Based on Metamaterial L-shaped Unit-Cells for Broadband Portable Microwave Devices and Multiband Wireless Communication Systems" IET Microwaves, Antennas & Propagation, Volume 12, Issue 7, 13 June 2018, p. 1080 – 1086.

“Compact Single Layer Travelling-Wave Antenna Design Using Metamaterial Transmission-Lines" Radio Science, Volume 52, Issue 12 December 2017, Pages 1510–1521.

“Periodic Array of Complementary Artificial Magnetic Conductor Metamaterials-Based Multiband Antennas for Broadband Wireless Transceivers” IET Microwaves, Antennas & Propagation, Volume 10, Issue 15, 10 December 2016, p. 1682 – 1691.

“Miniature CRLH-based ultra wideband antenna with gain enhancement for wireless communication applications”, ICT Express, Volume 2, Issue 2, June 2016, Pages 75–79.

"New Compact Antenna Based on Simplified CRLH-TL for UWB Wireless Communication Systems", International Journal of RF and Microwave Computer-Aided Engineering, Volume 26, Issue 3, March 2016, pages: 217–225.

“Metamaterial-Based Antennas for Integration in UWB Transceivers and Portable Microwave Handsets” International Journal of RF and Microwave Computer-Aided Engineering, Volume 26, Issue 1,January 2016, pages: 88–96.

"New Compact Printed Leaky-Wave Antenna with Beam Steering", Microwave and Optical Technology Letters, Volume 58, Issue 1, January 2016, Pages: 215–217.

"UWB Antenna Based on SCRLH-TLs for Portable Wireless Devices", Microwave and Optical Technology Letters, Volume 58, Issue 1, January 2016, Pages: 69–71.

"Composite Right-Left-Handed-Based Antenna with Wide Applications in Very-High Frequency-Ultra-High Frequency Bands for Radio Transceivers" IET Microwaves, Antennas & Propagation, Volume 9, Issue 15, 10 December 2015, p. 1713 – 1726.

“Bandwidth and Radiation Specifications Enhancement of Monopole Antennas Loaded with Split Ring Resonators” IET Microwaves, Antennas & Propagation, Volume 9, Issue 14, 19 November 2015, p. 1487 – 1496.

“The Resonating MTM Based Miniaturized Antennas for Wide-band RF-Microwave Systems «Microwave and Optical Technology Letters, Volume 57, Issue 10, pages 2339–2344, October 2015.

“Novel UWB Miniaturized Integrated Antenna Based on CRLH Metamaterial Transmission Lines” AEUE Elsevier- International Journal of Electronics and Communications, Volume 69, Issue 8, August 2015, Pages 1143–1149.

“Compact Antenna based on a Composite Right/Left Handed Transmission Line” Microwave and Optical Technology Letters, Volume 57, Issue 8, pages 1785–1788, August 2015.

“Printed planar patch antennas based on metamaterial”, International Journal of Electronics Letters, Volume 2, Issue 1, Jan 2014, pp 37-42.

“Design and Modeling of New UWB Metamaterial Planar Cavity Antennas with Shrinking of the Physical Size for Modern Transceivers,” International Journal of Antennas and Propagation, vol. 2013, Article ID 562538, 12 pages, 2013. doi:10.1155/2013/562538.

Response_4: We are thankful to the reviewer for these useful suggestions. The above references are cited in the revised manuscript.

……………………………

Comment_5: Some grammatical mistakes should be incorporated.

Response_5: The grammatical errors are incorporated in the revised manuscript.

……………………………

Comment_6: All figures should be properly cited in the main text.

Response_6: Authors are thankful to the reviewer for this suggestion. The comment has been addressed in the revised manuscript.

……………………………

Comment_7: The quality of the figure should be further improved.

Response_7: We are thankful to the reviewer for this suggestion. The comment has been addressed in the revised manuscript.

……………………………

Comment_8: Authors need to highlight the novelty of the proposed work.

Response_8: The comment has been addressed in the revised manuscript. We have highlighted the novelty of the proposed work.

……………………………

Comment_9: Some old references should be replaced with the latest ones.

Response_9: The old references are replaced with the latest ones.

……………………………

Comment_10: Reference part can be improved as per above mentioned suggestions.

Response_10: Authors are thankful for this worthy suggestion.

………………………………………………………………………………………………

Reviewer 2

In the introduction chapter, I suggest to introduce some more text and a few more references related to the actual practical applications (existing or future) of the proposed antenna solution. In other words, to further elaborate on a few specific biomedical cases where the proposed antenna can be used. Also, add some further explanation about how and where in the human body the proposed antenna is implanted. Otherwise, I recommend the paper for publication.

Response: Authors are very thankful to the reviewer for this useful comment. We have incorporated new references in the revised manuscript. Since, from the title as well as the related applications, it is clear that the antenna is being utilized for biomedical implants for in-body communication. The proposed antenna can be utilized in several tissues of the human body such as the skin, fat or muscle. In our case, the antenna is being implemented in Skin tissue. Also, we have highlighted the section where we have given the description of the antenna use inside skin tissue. We have added and highlighted the explanation in the revised manuscript.

………………………………………………………………………………………………

Attachments
Attachment
Submitted filename: Authors Response.docx
Decision Letter - Muhammad Zubair, Editor

Single-Layered Compact Broadband Circularly Polarized Implantable Antenna for Medical Applications

PONE-D-22-20756R1

Dear Dr. Hussein,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Muhammad Zubair

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

The authors have addressed the concerns raised by the reviewers. The manuscript can be editorially accepted for publication.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: (No Response)

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Authors have successfully addressed the reviewer's concerns. So, looking at the quality of the revised manuscript which shows a significant improvement than its initial version, there are no more technical comments.

Reviewer #2: The authors have, in my opinion, addressed the issues raised by the reviewers, such that I can recommend the paper for publication in PLOS ONE.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

**********

Formally Accepted
Acceptance Letter - Muhammad Zubair, Editor

PONE-D-22-20756R1

Single-Fed Broadband CPW-Fed Circularly Polarized Implantable Antenna for Sensing Medical Applications

Dear Dr. Hussein:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff