Peer Review History
| Original SubmissionSeptember 17, 2021 |
|---|
|
PONE-D-21-30139Validity and repeatability of the XSENSOR in-shoe pressure measurement systemPLOS ONE Dear Dr. Parker, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please attend carefully to both reviewers' comments and address particularly the concerns of reviewer 2. Please submit your revised manuscript by Mar 07 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Peter Andreas Federolf Academic Editor PLOS ONE Journal requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. Please note that in order to use the direct billing option the corresponding author must be affiliated with the chosen institute. Please either amend your manuscript to change the affiliation or corresponding author, or email us at plosone@plos.org with a request to remove this option. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: No ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: No Reviewer #2: I Don't Know ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Thank you for the opportunity to review the article titled Validity and repeatability of the XSENSOR in-shoe pressure measurement system. This manuscript describes a technical study of an in-shoe pressure measurement system to validate its measurement capabilities. The results of the study indicate that the XSENSOR system provides reasonable estimates of pressure and provide a characterization of bias and reliability. General Comments Overall, this is a well-designed and executed study. The authors’ approach to quantifying accuracy is well founded and executed. However, I think the approach to day-to-day reliability is not as rigorous. I am generally skeptical of ICCs and their use in quantifying reliability as they have many issues that have been identified in literature (Muller and Buttner, 1994). My suggestion would be to reframe the problem of reliability as a hypothesis test with respect to a minimally important difference (MID). The authors have provided a reference that suggests clinically important changes in pressure would be on the order of 30%. Using this (or other prior work as appropriate), we can define an MID that we would expect to see in a clinical study. If we wish to be conservative, we could use the lower end of the typical range of pressure values, 100 kPa, and the reference of 30% gives us an MID of 30 kPa. We can then take the measurements for day 1 and day 2, across all pressure conditions, and treat these as repeated measures observations. In doing so, we can calculate the distribution of the differences between days and conduct a single-sided hypothesis test with respect to an MID of 30 kPa (the alternate hypothesis being that our estimated mean of differences exceeds 30 kPa). The advantage of this approach is that it provides a concrete benchmark for evaluating whether the insoles will measure pressure reliably. In my opinion, this is more informative than ICCs, and a more statistically rigorous test of insole reliability. I also suspect that the findings would favour the insoles as being highly reliable. A table that provides comparative values for the XSENSOR system, similar to Table 1 from Price et al., 2016, might be of interest to the reader. Specific Comments The introduction is a bit meandering and some of the information is related, but tangential to the main purpose of the article. There are somewhat contradictory statements as well. It is stated that comparing between different studies/systems is very difficult, while the aim of the current study is to provide such a comparison between XSENSOR and other systems. Yet, this study does not address this difficulty at all; it only provides similar quantifications for the XSENSOR as has already been done for other systems in earlier work. I’d suggest a revision of the introduction to make it more focused on the purpose of the article: evaluating the XSENSOR system for clinical use. Throughout the manuscript there are no spaces between numbers and units and these should be added throughout (e.g., 400kPa should be 400 kPa). Line 53: capitalization of diabetic not necessary Line 60: single quotes around creep not required; this is a standard physics term Line 106: single quotes around calibration file not required Line 108: single quotes around active sensels not required Line 126: this sentence is a little unclear as it may read as though there were two separate cylinders with different bases; perhaps consider rewording Line 130: “For [each of] the six conditions…” Line 138 and 144: As mentioned in my general comments, ICCs are not the ideal representation of reliability Line 144: suggest “performed” instead of “undertaken” Line numbers stopped after page 10 Page 11, First paragraph, the peak pressures for 100-400 kPa measurements seem to be within 1-8 % of the target measurements, unless there is some averaging done between day 1 and 2. Page 11, First paragraph, “… interact with evaluations of efficacy…” might be clearer as “… bias evaluations of efficacy…” Page 11, First paragraph, while some of the mean measurements did indeed show less error than peaks, this was not universally true; I suggest rephrasing this statement Page 13, Second paragraph, given that the testing in this study was done under very controlled laboratory conditions, the evaluation of reliability should reflect the caveat that calibration and measurement under in situ conditions might be less reliable than what has been shown Figure 1, the scales for the grids in this figure is not obvious, and if I am correct, the scales are not identical between S4 and S10, even though they appear to be equal in this figure. I would suggest either adding scales, or at least clarifying that the graphical representations of contact area are not to scale. Figure 2, superscript required for ‘cm2’ in the y-axis label. Figure 3, the vertical alignment of all plots makes reading the results for the 500 and 600 kPa tests difficult. I would suggest staggering the results slightly left and right to make reading them easier to read. Supplementary file, the Excel file could use a little formatting/clean-up to make it easier to interpret; for example, the headers for each section could be bolded/formatted as table headings; some headers are bolded already, but many aren’t Reviewer #2: General comments: This manuscript investigates the validity and day-to-day repeatability of an XSensor in-shoe pressure-sensing insoles during standardized loading protocols in comparison to known pressures and contact areas. Overall, I agree with the authors on the importance for independent research groups to confirm the validity and reliability of commercial measurement systems. I also get the impression that the measurements were carried out carefully. In an original research article, however, the generated data should be presented and discussed in a way that provides value to other researchers. In my opinion, the submitted manuscript is lacking a clear structure, the investigated outcome variables are not well justified, and there are no criteria for what constitutes sufficient validity and reliability. My main critical points are: 1) The outcome variables are not well prepared and justified in the introduction. The authors talk about mean and peak pressure over varying duration, contact areas at different pressures, creep, insole consistency, etc.. It is unclear, why each variable is of interested and in which context each variable is relevant. The aim statement should clearly state the variables that the authors plan to use to determine the insoles’ validity and repeatability. Then, the variables should be described in the methods. Particularly, the authors should avoid to report results that were not prepared in the methods (creep, insole consistency). 2) What constitutes sufficient validity and repeatability? The authors talk about various applications of pressure-sensing insoles, e.g. running biomechanics [2] or in patients with neuromotor disorders [3-5]. Comparing pressures between slightly modified running shoes during running OR comparing pressures between healthy vs. diseased patients during standing or walking will lead to very different requirements of validity and repeatability for a pressure-sensing system. Therefore, the authors cannot make a sweeping statement in the abstract to say that “the Xsensor system is appropriate for clinical assessment that require multiple assessments”. Therefore, the criteria that indicate sufficient validity and repeatability need to be developed carefully in the context of a certain research area and are only relevant in this context. Those criteria should not only be based on relative measures (e.g. relative error or ICC). Such measures certainly have their strengths but also weaknesses (e.g. Koo, Terry K., and Mae Y. Li. 2016. “A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research.” Journal of Chiropractic Medicine 15 (2): 155–63.) The authors should consider including absolute measures such as Bland-Altman limits of agreement (see Atkinson, G., and A. M. Nevill. 1998. “Statistical Methods for Assessing Measurement Error (reliability) in Variables Relevant to Sports Medicine.” Sports Medicine 26 (4): 217–38.). Limits of agreement would provide an intuitive way of determining whether the measurement error grows with increasing pressure or measurement duration. Finally, the authors should come up with clinically relevant or meaningful changes in pressure in absolute physical units (or maybe relative to body weight) that occur in a certain research context and use those values to build a framework for validity and repeatability. 3) The authors criticize previous studies for their limited external validity. In this experiment however, there is no information on the loading rate. I assume that the pressure was built up slowly and was not comparable to a pressure profile to running or walking. Is there anyway for your system to simulate a dynamic loading profile? Currently, I would argue that the results of this manuscript are relevant for standing only. Specific comments: Line 72-73: ICCs of around 0.5 are not usually considered to show “high intra class correlation”. Line 73-81: The relevance of previous findings on the Tekscan system is unclear for this study on the XSensor system. It would seem more relevant to explain if some measurement systems have a larger error than clinically relevant underfoot pressure changes in clinical populations. Line 95: It is unclear how the “suitability for clinical and biomechanical assessment” is defined. There should be a-prior standards about the validity and repeatability for their use in clinical research and biomechanical research. Line 99: When looking at the XSensor webpage (https://xsensor.com/), the available systems are “Intelligent Insoles Clinical” and “Intelligent Insoles Pro”. Please explain how the XSensor X4 system relates to these two available systems. Line 115: How long were the measurements? Line 117: Where exactly were those sensels and how many were excluded? Line 118: sensel underlined? Line 125: Please briefly describe the testing device to generate pressure during the regional protocol. How accurate is this device? Line 129: How long were the measurements in the regional protocol? Line 130: Where was the regional pressure applied? Line 138: This is the first time that a between-day study design is mentioned. This must be mentioned in the aims of this study and at the beginning of the methods. In general, while it is appreciated that the previously developed protocol is cited [reference 18], the most important aspects of this protocol must still be mentioned. Like how many days passed between the measurements, etc.? Why were the durations 0,2, 10, 30 seconds selected? What is a duration of 0 seconds anyway? How exactly were relative and absolute errors calculated (e.g. the ones in Table 2)? Lines 134-144: This section is difficult to follow. For example: What happened to the different pressures in the regional protocol (50, 110, 200 kPa)? Why were peak pressures investigated? Is “load” equal to pressure? Why did the authors only investigate ICCs as a relative measure of reliability? In my opinion, absolute measures of repeatability and validity such as Bland-Altman limits of agreement would be much more intuitive in their interpretation and it would be easier to observe changes in agreement with the known measures as a function of held duration or applied pressure. This section needs to be structured better and the analysis approach should be justified. Line 143: Please explain exactly what type of ICC model was used and how the individual values were combined into the respective results. Different ICC models can lead to different results given the same data. Figure 1: Please add a color scale so the reader can assess changes in pressure. Please indicate where in Figure 1a the regional pressures were applied. Line 154: Why is there no analysis of the measured pressures in the regional protocol? This type of regional loading may affect the insole’s validity in measuring the correct mean and peak pressure. Table 1: There is no description of this analysis in the methods. Lines 160-167: It is unclear where those values come from? (e.g. S10 (23.09±15.92kPa; 9%)). Are those averages? Line 164-165: There has been no mention of a time-dependent analysis of peak pressure in the methods section. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Maurice Mohr [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.
|
| Revision 1 |
|
PONE-D-21-30139R1Validity and reliability of the XSENSOR in-shoe pressure measurement systemPLOS ONE Dear Dr. Parker, Thank you for submitting your manuscript to PLOS ONE. We invite you to submit a revised version of the manuscript that addresses the remainig points raised during the review process. Both reviewers point out a number of small issues. Please adress them carefully. Please submit your revised manuscript by Oct 29 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Peter Andreas Federolf Academic Editor PLOS ONE Journal Requirements: Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: (No Response) Reviewer #2: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Thank you to the authors for their comprehensive response to the comments. I agree that the manuscript is much improved and I appreciate them taking the time to revise the content and analysis. I have only a few minor remaining comments based on the revised version with changes highlighted Line 76: “discrete” misspelled Line 97: “that” seems superfluous Line 98: wording is a bit unclear; is it supposed to be “highly correlated relationships”? Line 108: capitalization of international not needed Line 120: missing a period. Lines 142-143: inclusion of pressure gauge company redundant with information in previous paragraph. Line 145: change to “were calculated” Lines 224-225: I am confused by the statement that no measurements exceeded the analytical goal, when one of the 50 kPa measurements was off by 20%. Line 340: change to “these data” Reference 11: misspelled “pressure” Reviewer #2: General comments: I would like to congratulate the authors on the extensive revisions to the manuscript. The inclusion of Bland-Altman limits of agreement and a clinically meaningful reliability target has significantly increased the value of this manuscript to the readership. Please consider a few remaining but minor comments/questions below. ^ Specific comments: Line 69: discrete Line 74: average pressure across sensors within a discrete region Line 96-102: I am still under the impression that this part of the introduction does not add value to the manuscript. The importance of evaluating the contact area has been demonstrated further up already. I do not understand the meaning of “Additionally, that when considering repeated measures most systems demonstrate high relationships between the two days or measures”. Line 173: intra-class correlation coefficient Line 220-221: This sentence sounds like a contradiction. 20% > 15% Figure 1: Is there any reason for the measurement at 100kPA showing the opposite trend of all other loading conditions? On average Figure 3 does not seem to show this opposite behavior at 100kPA. Figure 4: (c) and (d) labels are flipped in the figure Table 1&2: Does the correlation column represent the correlations before or after log-transformation? After transformation, the correlations should be much lower or rather non-existent. Also, there are strong negative and positive correlations (e.g. Table 1, s10, 100kPA vs. 500kPA). Do the authors have an explanation for this? I would expect a consistent trend where d2 shows higher pressures than d1 across all loading conditions? ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Maurice Mohr ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.
|
| Revision 2 |
|
Validity and reliability of the XSENSOR in-shoe pressure measurement system PONE-D-21-30139R2 Dear Dr. Parker, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Peter Andreas Federolf Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: |
| Formally Accepted |
|
PONE-D-21-30139R2 Validity and reliability of the XSENSOR in-shoe pressure measurement system Dear Dr. Parker: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Peter Andreas Federolf Academic Editor PLOS ONE |
Open letter on the publication of peer review reports
PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.
We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.
Learn more at ASAPbio .