Peer Review History

Original SubmissionFebruary 24, 2022
Decision Letter - Gheyath K. Nasrallah, Editor

PONE-D-22-05585Performance evaluation of the Roche Elecsys Anti-SARS-CoV-2 immunoassays by comparison with neutralizing antibodies and clinical assessmentPLOS ONE

Dear Dr. Tabe

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. The table fonts are very small. The figures are hazy and blurry. The abstract should include detailed results with number. Please correct and resubmit

Please submit your revised manuscript by 20/3/2022. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Gheyath K. Nasrallah

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match. 

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

3. Thank you for stating the following in the Competing Interests section: 

"The reagents used in this study were provided by Roche, but the study was performed by scientifically proper methods without any bias.  "

Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. 

Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf.

4. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

"Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

March 23, 2022

Ms. Agatha Macaraig

Straive Editorial Assistant

RE: Resubmission of the revised manuscript titled “Performance evaluation of the Roche Elecsys Anti-SARS-CoV-2 immunoassays by comparison with neutralizing antibodies and clinical assessment” by Takei S., et al.

Ms. Macaraig,

We answer the requirements of the journal requirements as follows;

Reply (Answer; A) to the comments and questions (Comment; C):

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

C1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

A1. We corrected the manuscript following PLOS ONE's style requirements, including those for file naming.

C2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

A2. We stated the financial disclosure in the cover letter.

C3. Thank you for stating the following in the Competing Interests section:

"The reagents used in this study were provided by Roche, but the study was performed by scientifically proper methods without any bias. "

Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests). If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf.

A3. We included our updated Competing Interests statement in the cover letter as follows;

"The reagents used in this study were provided by Roche, but the study was performed by scientifically proper methods without any bias. This does not alter our adherence to PLOS ONE policies on sharing data and materials.”

C4. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

"Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter.

A4. We corrected the Data Availability statement as follows; “All relevant data are within the paper.“

Sincerely,

Yoko Tabe, M.D., Ph.D.

Professor

Department of Clinical Laboratory Medicine,

Juntendo University, School of Medicine

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Gheyath K. Nasrallah, Editor

PONE-D-22-05585R1Performance evaluation of the Roche Elecsys Anti-SARS-CoV-2 immunoassays by comparison with neutralizing antibodies and clinical assessmentPLOS ONE

Dear Dr. Tabe,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Bothe reviewers has serious concerns about the methodology. Further Rev2 believes that the manuscript is poorly written and organized. Please make sure that you address all of comments of the reviewers before you submit  a revised version.   

Please submit your revised manuscript by Jun 23 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Gheyath K. Nasrallah

Academic Editor

PLOS ONE

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: (No Response)

Reviewer #2: (No Response)

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: No

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: No

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: No

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Takei et al. investigated the correlation between the neutralizing antibody titre measured by two commercially available immunoassays and a standard virus neutralization assay. The measurement values of Elecsys Anti-SARS-CoV-2 S and neutralization assays were highly correlated and significantly higher values has been measured in the group of severe and critically ill cases. The study is of interest as serological tests are essential tools to evaluate the neutralizing antibodies targeting the receptor binding domain of the SARS-CoV-2 spike protein.

The authors might consider some minor comments:

Section Materials and methods, Neutralization assay:

It would be interesting to read which strain of SARS-CoV-2 was used in the assay, as the neutralizing activity of the antibodies may be different against the original SARS-CoV-2 strain and the emerging variants of concern. There may be differences between the neutralizing antibody titer measured by the commercially available immunoassays and the protective immunity, depending on the viral variant.

Section Results, Table 3:

Is there any information available regarding the SARS-CoV-2 vaccination status of these patients? Or they were all immunologically naïve prior to the acute infection?

In addition to the diagnostic PCR assays, has a variant determination been performed in the study group?

Reviewer #2: In this study, the authors investigated whether measuring anti-SARS-CoV-2 Spike antibodies can predict neutralizing activities in patients’ sera. A commercial assay (Elecsys® Anti-SARS-CoV-2) was utilized for measuring anti-S and anti-N SARS-CoV-2 antibodies. This was then compared to neutralizing titers utilizing an authentic virus neutralization assay.

Major comments:

1- The article is poorly written. All sections need to be improved. There are missing info in the abstract and methods. The authors did not discuss similar studies nor compared their results to literature in their discussion.

2- A major concern in the study in the correlation analysis. The authors indicated that 146 serum samples were tested, with a positivity rate of >70% for both binding and neutralization assays. However, the correlation figures (Fig 1A and 1B) included only few data points (n=6). All data points (n=146) need to be plotted along with a properly drawn fitted line, in addition, r, and r2 values need to be both mentioned in order to assess the strength of the correlation.

3- The abstract needs to be improved. The numbers and percentages need to be added accordingly with each finding.

4- Patients sample details needs to be better described in the Methods section. There is no information on how many samples were collected per patient. What is the number of inpatients and outpatients? Also, what is the duration between different time-points?

5- Figures and legends are poorly presented. Please fix the following:

a.Figure 1 needs re-plotting after including all the results.

b.In figure 2, better to label the lower panels with other labels (E,F and G), better than referring to them as “the lower panel of panel A, B or C”

c.Figure 3 is not clear at all. It is better to use colors to differentiate between patients.

d.In figure 3, why did they choose patients who were tested more than 3 time points only? Patients with 3 time-points can be included in the analysis especially if there is a good time gap between each time-point.

e.All legends need to be expanded to clearly indicate the type of test, analysis, and sample numbers.

6- In line 260, it is mentioned: “no significant reduction in antibody levels tested was observed until 69 days after the onset of symptoms”. I believe this was observed in 2 patients only whose samples were available at that time-point. This is not enough to draw conclusions. The limitation of sample size need to be clearly mentioned in the discussion.

Minor comments:

1- In the abstract, change "levels of neutral antibodies in COVID-19" to "neutralization antibodies"

2- The catalog numbers of the used commercial assays need to be mentioned in the methods.

3- Kindly indicate the type/source of the real virus used in the neutralization assay.

4- In the methods section, it is better to change “Clinical Backgrounds” to “Clinical characteristics”.

5- Table 1. If age is represented as range, please add this to the label of the row “Age range (average)”

6- Whenever “Elecsys® Anti-SARS-CoV-2” is mentioned, it is better to add “anti-N”, so it does not confuse the reader. This applies to figures and tables too.

7- Table 1 and 3 have the same title. Kindly change the title of table 3 to “inpatients”.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: Yes: Maria K. Smatti

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 2

Reply (Answer; A) to Reviewer’s comments and questions (Comment; C):

Reviewer 1

COMMENTS by Reviewer #1:

Takei et al. investigated the correlation between the neutralizing antibody titer measured by two commercially available immunoassays and a standard virus neutralization assay. The measurement values of Elecsys Anti-SARS-CoV-2 S and neutralization assays were highly correlated and significantly higher values has been measured in the group of severe and critically ill cases. The study is of interest as serological tests are essential tools to evaluate the neutralizing antibodies targeting the receptor binding domain of the SARS-CoV-2 spike protein.

COMMENTS (continued):

The authors might consider some minor comments:

C1: Section Materials and methods, Neutralization assay:

It would be interesting to read which strain of SARS-CoV-2 was used in the assay, as the neutralizing activity of the antibodies may be different against the original SARS-CoV-2 strain and the emerging variants of concern. There may be differences between the neutralizing antibody titers measured by the commercially available immunoassays and the protective immunity, depending on the viral variant.

A1: Thank you for the valuable suggestions. We added the sentence in the Materials and Methods section and discussed about the differences between the neutralizing antibody titer and the protective immunity, which might depend on the viral variant (L120-123, L275-282).

C2: Is there any information available regarding the SARS-CoV-2 vaccination status of these patients? Or they were all immunologically naïve prior to the acute infection?

A2: This study targeted only pre-vaccinated patients, who were immunologically naive prior to infection. We added this information to the Materials and Methods section (L98).

C3: In addition to the diagnostic PCR assays, has a variant determination been performed in the study group?

A3: Unfortunately, no variant determination has been performed in the study group. We mentioned this as the study limitation and discussed the SARS-CoV-2 strain used in the neutralization assay and the SARS-CoV-2 variant prevalent in Japan during the study period (L275-282).

Reviewer #2:

In this study, the authors investigated whether measuring anti-SARS-CoV-2 Spike antibodies can predict neutralizing activities in patients’ sera. A commercial assay (Elecsys® Anti-SARS-CoV-2) was utilized for measuring anti-S and anti-N SARS-CoV-2 antibodies. This was then compared to neutralizing titers utilizing an authentic virus neutralization assay.

C1: The article is poorly written. All sections need to be improved. There are missing info in the abstract and methods. The authors did not discuss similar studies nor compared their results to literature in their discussion.

A1: Following the reviewer’s suggestion, we added more details in the Abstract and Materials and Method section (L44-55, L91-98). We also discussed and compared our results to previous literatures (L248-253, L263-269).

C2: A major concern in the study in the correlation analysis. The authors indicated that 146 serum samples were tested, with a positivity rate of >70% for both binding and neutralization assays. However, the correlation figures (Fig 1A and 1B) included only few data points (n=6). All data points (n=146) need to be plotted along with a properly drawn fitted line, in addition, r, and r2 values need to be both mentioned in order to assess the strength of the correlation.

A2: One hundred and forty-six samples were included in the correlation analysis. Although some of the plotted sample dots appear to overlap in the graphs in Fig 1A and 1B, all of 146 samples are included in these figures.

We appreciate the reviewer’s comments about the r2 values. We added the r2 along with the r values (L51-52, L155-156).

C3: The abstract needs to be improved. The numbers and percentages need to be added accordingly with each finding.

A3: Following the reviewer’s suggestion, we added the numbers and percentages in the Abstract section (L44-55).

C4: Patients sample details needs to be better described in the Methods section. There is no information on how many samples were collected per patient. What is the number of inpatients and outpatients? Also, what is the duration between different time-points?

A4: We added the detailed information of sample numbers and duration in the Materials and Methods section (L91-97).

C5: Figures and legends are poorly presented. Please fix the following:

a. Figure 1 needs re-plotting after including all the results.

b. In figure 2, better to label the lower panels with other labels (E,F and G), better than referring to them as “the lower panel of panel A, B or C”

c. Figure 3 is not clear at all. It is better to use colors to differentiate between patients.

d. In figure 3, why did they choose patients who were tested more than 3 time points only? Patients with 3 time-points can be included in the analysis especially if there is a good time gap between each time-point.

e. All legends need to be expanded to clearly indicate the type of test, analysis, and sample numbers.

A5:

a. One hundred and forty-six samples were included in the correlation analysis. Although some of the plotted sample dots appear to overlap in the graphs in Fig 1A and 1B, all of 146 samples are included in these figures.

b. We labeled the lower panels with D, E, and F in Figure 2.

c. We used colors to differentiate between patients in Figure 3.

d. We included patients with 3 time-points in the analysis and corrected the sentence in Table 3 abbreviation (L218) and the Result section as follows: “Chronological changes in the results of the neutralization assay, Elecsys® anti-S assay, and Elecsys® anti-N assay were examined simultaneously in 23 inpatients who were tested in three time points or more.” (L203-205).

e. Following the reviewer’s suggestion, we added the type of test, analysis, and sample numbers in legends of Fig 1 (L164-170) and Fig 2 (L192-200).

C6: In line 260, it is mentioned: “no significant reduction in antibody levels tested was observed until 69 days after the onset of symptoms”. I believe this was observed in 2 patients only whose samples were available at that time-point. This is not enough to draw conclusions. The limitation of sample size need to be clearly mentioned in the discussion.

A6 Following the reviewer’s suggestion, we deleted the sentence “no significant reduction in antibody levels tested was observed until 69 days after the onset of symptoms” and mentioned the limitation of this study regarding sample size (L274-275).

C7: In the abstract, change "levels of neutral antibodies in COVID-19" to "neutralization antibodies"

A7: Corrected (L58).

C8: The catalog numbers of the used commercial assays need to be mentioned in the Materials and Methods.

A8: We added the catalog numbers of the used commercial assays in the methods section (L102-106).

C9: Kindly indicate the type/source of the real virus used in the neutralization assay.

A9: Information of SARS-CoV-2 strain used in the neutralization assay was provided in the Materials and Methods section (L120-123).

C10: In the methods section, it is better to change “Clinical Backgrounds” to “Clinical characteristics”.

A10: Following the reviewer’s suggestion, we changed “Clinical Backgrounds” to “Clinical characteristics” (L86).

C11: Table 1. If age is represented as range, please add this to the label of the row “Age range (average)”

A11: Corrected “Age, y (average)” to “Age range (average)” (Table 1).

C12: Whenever “Elecsys® Anti-SARS-CoV-2” is mentioned, it is better to add “anti-N”, so it does not confuse the reader. This applies to figures and tables too.

A12: Following the reviewer’s suggestion, we corrected “Elecsys® Anti-SARS-CoV-2-S” to “Elecsys ® Anti-SARS-CoV-2-S (Elecsys® anti-S)” and “Elecsys® Anti-SARS-CoV-2” to “Elecsys ® Anti-SARS-CoV-2 (Elecsys® anti-N)” in manuscripts, figures, and tables.

C13: Table 1 and 3 have the same title. Kindly change the title of table 3 to “inpatients”

A13: Thank you for your kind and detailed review. We corrected “patients” to “inpatients” in the title of Table 3.

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Gheyath K. Nasrallah, Editor

Performance evaluation of the Roche Elecsys ® Anti-SARS-CoV-2 immunoassays by comparison with neutralizing antibodies and clinical assessment

PONE-D-22-05585R2

Dear Dr. Tabe,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Gheyath K. Nasrallah

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: (No Response)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: No

**********

Formally Accepted
Acceptance Letter - Gheyath K. Nasrallah, Editor

PONE-D-22-05585R2

Performance evaluation of the Roche Elecsys® Anti-SARS-CoV-2 immunoassays by comparison with neutralizing antibodies and clinical assessment

Dear Dr. Tabe:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Gheyath K. Nasrallah

Academic Editor

PLOS ONE

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .