We thank the editor and the referees for their constructive comments. We have addressed
the substantive concerns that appear to relate to placing our protocol in the wider
context of other related methods and ensuring that aspects of validation, clarity,
and the reproducibility of the protocol are made more clear.
Below we respond first to the PLOS editorial comments and then below, point by point,
to the referees in the PLOS template. All major changes have also been copied into
the rebuttal.
PLOS
Main points:
Discrepancy with the publication criteria for Lab Protocol type of article. In addition
to the general PlosOne criteria for publication, the manuscripts should provide clear
evidence that the protocol works and that it is not presenting describing routine
methods, or extensions or modifications of routine methods. In this context, a clear
outline of the novelty of this system along with a more detailed idea of what was
done with this system are required for publication.
RESPONSE: We thank the editor for this comment. We have enhanced citations of similar
work that supports the motivation and context of the work and places our work more
effectively in the wider literature, as well as highlight that this protocol is the
first description of this type of system adapted for experimental evolution of algae
in addition to being used for modelling herbicide resistance evolution.
Specifically the novelty of the system is addressed in the Introduction and is expanded
on in detail in the Discussion and have been pasted below in response to Reviewer
#1.
Furthermore, to make clear what the system was used for, we have modified our presentation
of data and enhanced the evidence it provides that the protocol can track trends and
variability in the abundance of organisms among replicates.
Finally, we have also added pilot data demonstrating the applicability of the system
to the intended use of experimental herbicide resistance evolution, showing how evolution
in action can be followed through population density data and that this effect is
clearly distinguishable from among population variation, day-to-day variation or the
variation introduced by equipment failures. Details of these changes have been pasted
below in response to the relevant comments by Reviewer #2.
Please, ignore the request of providing a detailed description of methods by Reviewer
1
RESPONSE: Thank you.
Take into carful consideration all points raised by Reviewers, particularly attention
should be paid to request related to validation and reproducibility.
RESPONSE: Thank you for this advice. We have improved our description of the motivation
and context of the work, extended the description and discussion of biological variation
of experimental interest vs. variation caused by equipment failure as well as addressed
further possible issues and improvements in more detail which are the key points raised
by the reviewers. The changes have been pasted below in response to the relevant comments.
Align manuscript presentation with journal requirements for Lab Protocol as presented
in the Template available on the dedicated link (https://journals.plos.org/plosone/s/submission-guidelines#loc-lab-protocols).
RESPONSE: Thank you for this advice. We have done this, see responses below:
1. Please ensure that your manuscript meets PLOS ONE's style requirements, including
those for file naming. The PLOS ONE style templates can be found at
https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and
https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf
RESPONSE: Thank you for pointing this out. We have done the following to address this:
Changed the title of Table 1 to conform to style guidelines.
Changed file names of the supporting information protocol file as well as figures
to “S1_File.pdf” and “S1_Fig1.tiff” etc.
Added a Supporting Information section (see pasted below in response to point 6).
Removed the Funding Statement from the manuscript (see below).
Uploaded the protocol to Protocols.io.
2. To comply with PLOS ONE submissions requirements, please provide the Protocols.io
DOI in the Methods section of the manuscript using this format: “The protocol described
in this peer-reviewed article is published on protocols.io, https://dx.doi.org/10.17504/protocols.io[........] and is included for printing as supporting information file 1 with this
article.” Please also provide the Protocols.io DOI in the “Protocol DOI” field of
the submission form (via “Edit Submission”). For more information, please see our
submission guidelines: https://journals.plos.org/plosone/s/submission-guidelines#loc-guidelines-for-specific-study-types
RESPONSE: Thank you for pointing this out, the Protocols.io DOI has now been included
in the manuscript as well as in the relevant field in the submission form.
3. We note you have not provided a Protocol.io PDF version of your protocol. As noted
in our submission requirements, please upload a Protocol.io PDF version of your protocol
as a Supporting Information file and name the file ‘S1 file’. Please update your Supporting
Information Captions if necessary. If you have not yet uploaded your protocol to Protocols.io
you are welcome to use the Protocols.io customer service code ‘PLOS2021.’ When using
this customer code while submitting to Protocols.io, please make reference to your
PLOS ONE submission, including your PLOS ONE manuscript number. With this customer
code, Protocols.io editorial staff will import and format your protocol at no charge.
For more information, please see our submission guidelines: https://journals.plos.org/plosone/s/submission-guidelines#loc-guidelines-for-specific-study-types
RESPONSE: Thank you for this information, we have now provided a Protocols.io pdf
version of the protocol as instructed.
4. Thank you for stating the following in the Acknowledgments Section of your manuscript:
"EMH was supported by a scholarship from The University of Sheffield. Special thanks
to Allison Blake and Lynsey Gregory for technical support in the lab. "
We note that you have provided funding information that is not currently declared
in your Funding Statement. However, funding information should not appear in the Acknowledgments
section or other areas of your manuscript. We will only publish funding information
present in the Funding Statement section of the online submission form.
Please remove any funding-related text from the manuscript and let us know how you
would like to update your Funding Statement. Currently, your Funding Statement reads
as follows:
"EMH was funded by a University of Sheffield scholarship. The funders had and will
not have a role in study design, data collection and analysis, decision to publish,
or preparation of the manuscript."
Please include your amended statements within your cover letter; we will change the
online submission form on your behalf.
RESPONSE: Thank you for pointing this out. The funding information has been removed
from the manuscript. The current Funding Statement is correct and can be kept as is.
5. We note you have included a table to which you do not refer in the text of your
manuscript. Please ensure that you refer to Table 1 in your text; if accepted, production
will need this reference to link the reader to the Table.
RESPONSE: Thank you for noticing this, the table in question had been mislabelled
as a figure which has been amended.
6. Please include captions for your Supporting Information files at the end of your
manuscript, and update any in-text citations to match accordingly. Please see our
Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.
RESPONSE: Thank you for pointing this out. We have now included the following Supporting
information section (lines 385–407):
Supporting information
S1 File
Step-by-step protocol, also available on protocols.io.
S1 Fig1
Simplified schematic of the mesostat system. The mesostat system includes medium container(s),
a pump, culture chamber(s) with sampling needle, overflow bottle(s), a gas washing
bottle along with the medium and air influx lines and the culture efflux line.
S1 Fig2
Photographs of the mesostat system. A) The complete setup just after inoculation with
algae, running an experiment with six levels of treatments applied through the media
lines. B) Close-up of medium siphon through medium container lid. C) Close-up of connection
between pump tubing and silicone tubing used throughout array. D) Close-up of culture
chambers rubber bung with the four hypodermic needles, capped sampling needle to the
left in foreground, steel efflux needle to the right in foreground, steel aeration
needle in the middle, and pink plastic medium influx needle in the background. E)The
overflow chamber (left) and the culture chamber at steady state (right) with the efflux
line running between them.
S1 Fig3
Assembly schematic; the media lines.
S1 Fig4
Assembly schematic; the chambers.
S1 Fig5
Assembly schematic; the aeration line.
REFEREE COMMENTS
2. Has the protocol been described in sufficient detail?
Descriptions of methods and reagents contained in the step-by-step protocol should
be reported in sufficient detail for another researcher to reproduce all experiments
and analyses. The protocol should describe the appropriate controls, sample sizes
and replication needed to ensure that the data are robust and reproducible.
Reviewer #1: No
Reviewer #2: Yes
3. Does the protocol describe a validated method?
The manuscript must demonstrate that the protocol achieves its intended purpose: either
by containing appropriate validation data, or referencing at least one original research
article in which the protocol was used to generate data.
Reviewer #1: No
Reviewer #2: Yes
4. If the manuscript contains new data, have the authors made this data fully available?
The PLOS Data policy requires authors to make all data underlying the findings described
in their manuscript fully available without restriction, with rare exception (please
refer to the Data Availability Statement in the manuscript PDF file). The data should
be provided as part of the manuscript or its supporting information, or deposited
to a public repository. For example, in addition to summary statistics, the data points
behind means, medians and variance measures should be available. If there are restrictions
on publicly sharing data—e.g. participant privacy or use of data from a third party—those
must be specified.
Reviewer #1: N/A
Reviewer #2: Yes
5. Is the article presented in an intelligible fashion and written in standard English?
PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles
must be clear, correct, and unambiguous. Any typographical or grammatical errors should
be corrected at revision, so please highlight any specific errors that need correcting
in the box below.
Reviewer #1: Yes
Reviewer #2: Yes
6. Review Comments to the Author
Please use the space provided to explain your answers to the questions above. You
may also include additional comments for the author, including concerns about dual
publication, research ethics, or publication ethics. (Please upload your review as
an attachment if it exceeds 20,000 characters)
Reviewer #1: Major comments:
Which cultures/associations have been used? What were their initial parameters (cell
densities, DW, etc.)??
RESPONSE: Thank you for this comment. We have included the starting concentrations
for each experiment in Table 1 as well as added the cultures used to Methods section
(see pasted below, from lines 65–68):
In all of the presented experiments, Chlamydomonas reinhardtii strain Sager’s CC-1690
wild-type 21 gr was used, obtained from the Chlamydomonas Resource Centre (University
of Minnesota, St Paul, MN, USA) core collection.
Mere including another paper as a supplementary file with the method description is
not enough—at least, a brief recapitulation of the key methods used should be provided.
RESPONSE: We have ignored this on advice from the Editor.
The section ”Representative results and observations” is actually a mix of Methods
and Results.
RESPONSE: We have moved the parts that constituted methods to a subsection in the
Methods (pasted below), conforming to the format used by other papers published by
PLOS ONE (e.g. https://doi.org/10.1371/journal.pone.0263071 or https://doi.org/10.1371/journal.pone.0259202), see pasted below (copied from lines 60–110):
Experimental design for validation data
Replicability
Presented below are control data from four separate experiments using the linked protocol
to show replicability. The conditions and relevant differences for these experiments
are summarised in Table 1, unless otherwise stated the experimental conditions correspond
to those outlined in the protocol. In all of the presented experiments, Chlamydomonas
reinhardtii strain Sager’s CC-1690 wild-type 21 gr was used, obtained from the Chlamydomonas
Resource Centre (University of Minnesota, St Paul, MN, USA) core collection. Two different
dilution rates were used in the experiments: 0.3/day and 0.15/day. The former was
based on the dilution rates used in previous experiments using chemostat populations
of similar species that this system was designed for (e.g. [17, 18]), the latter was
used as an alternative lower rate to decrease the consumption of growth medium as
well as wear and tear on the pump tubing.
Applicability to experimental herbicide resistance evolution
Six mesostat chambers in experiment C were allowed a week to reach steady state before
the glyphosate treatment was introduced. Shock injections of 38 ml were performed
as described in the protocol bringing two chambers each to concentrations of 0 mg/L
(controls), 100 mg/L and 150 mg/L glyphosate (analytical standard, PESTANAL®). Both
of the chosen glyphosate concentrations are above the minimum inhibitory concentration
for C. reinhardtii of 97.5 mg/L [30].
Common problems
We have provided data from three common problems that present with this type of system:
a leak, contamination and algal clumping, all from experiment D. These were spontaneous
events and the data presented here aims to show how to identify their signal in the
population density data and distinguish it from normal variation among populations.
The leak in this example resulted in elevated dilution of a single chamber for roughly
four hours due to a clamp securing the pump tubing cassette coming undone. In the
case of the contamination event, all of the presented six chambers had been disconnected
from the array six days before bacterial contamination was observed under the microscope
in four chambers, with the remaining two unaffected by the contamination event. The
clumping phenotype was not receiving control medium but presented in a population
undergoing treatment with a sublethal dose of glyphosate.
Sample processing
Population density was in all cases determined through flow cytometry (Beckman Coulter
CytoFLEX), using CytExpert (Beckman Coulter) to gate and count events detected in
the PerCP-A channel (Excitation: 488nm, Emission: 690/50 BP). This channel is used
to detect chlorophyll a and represents a robust method for estimating algal density
[34] which was further validated against manual haemocytometer counts for this system.
Data handling
All statistical analyses were carried out in R (version 4.0.5, [35]), using the lme4
package [36] to fit a linear mixed effects model with log-transformed population density
as the response, dilution rate and experiment as fixed effects, and day and chamber
as random effects with varying intercepts. The significance of the fixed effects was
tested using the Anova() function from the car package [37] and confirmed through
parametric bootstrapping using the pbkrtest package [38].
The slope of population density decline was estimated between days 6–16 with the package
emmeans [39] after fitting a linear mixed effects model with the log-transformed population
density as the response, treatment and day as fixed effects as well as day and chamber
as random effects with varying intercepts.
The authors underline the economical aspects of their system but never present cost
estimation/breakdown.
RESPONSE: We thank the reviewer for this comment — however the point of the paper
and described method is to provide an example of how easily this type of system can
be built from parts readily found from all major lab material suppliers and parts
that are likely to already be present in most wet labs, rather than buying a premade
set.
We have provided a detailed list of components as part of the protocol and, the prices
and availability of them is subject to change and to vary internationally. Our emphasis
is on that the vast majority of those parts could be substituted for something similar
according to need.
To add clarity around this issue, the following sentence has been added to the Materials
list in the protocol:
Other than the pump and pump tubing, all of the pieces are fairly standard pieces
found in many wet labs and similar products can be obtained easily from all major
scientific suppliers.
Furthermore, the How to Improve or Modify subsection in the Discussion section addressing
substitution of parts has been expanded to the following (lines 342–347):
While the pump and pump-tubing are integral to the design and also the most expensive
parts, all other parts could be easily substituted depending on availability or cost
constraints. The materials list provided in the protocol can be used as a guide for
the dimensions and properties of the part, but primarily aims to illustrate how this
type of system can be built from parts already found in most wet labs rather than
buying a pre-made set.
As a result of overall lack of structure, it is difficult to understand which scientific
goal was set and achieved in this study, and the whole manuscript is more like a technical
note with a limited novelty than a scientific paper.
RESPONSE: We thank the referee for this comment. We have made every effort to follow
the guidelines for a protocol submission which is quite different to a study with
a scientific goal. To make the objective and application of the protocol more clear,
we have increased the context of the work and the data/examples.
First, with respect to context, our protocol thus contributes to an emerging landscape
of protocols and design criteria for experimental evolution, and highlights specifically
how to implement this for algae, larger volumes and higher replication for experimental
evolution within the eco-evolutionary context with the specific view to apply it to
herbicide resistance evolution.
We are aware of and reference and discuss two additional manuscripts that have added
evidence for how to specifically expand the Miller et al (2013) platform, initially
developed for yeast work. One (Tonoyan et al. 2020. Construction and Validation of
A Low-cost, Small-scale, Multiplex Continuous Culturing System for Microorganisms)
focuses on developing a cost-effective system for working with bacteria and the other
(Ekkers et al. 2020. The Omnistat: A Flexible Continuous-Culture System for Prolonged
Experimental Evolution) focuses on a bespoke, highly refined system for microbial
communities. The unique and complementary aspects of our system are now expressed
more clearly in the introduction and more fully developed in the discussion (see changes
pasted below):
Introduction Content:
Lines 39–55: Here we describe a multiplexed small-scale DIY chemostat array system
(dubbed “mesostats”) adapted from the ministat array developed by Miller et al. [22]
to suit experimental evolution of algae, in contrast to the so far described designs
specifically intended for yeast [21, 22, 26] and bacterial cultures [23, 27]. Our
system uses common algal model species Chlamydomonas reinhardtii, with the specific
goal to use it as a herbicide resistance evolution model. C. reinhardtii is an established
model species for herbicide resistance evolution [8, 28–30] and molecular analysis
of herbicide resistance mutations [31–33], but all studies to date have used batch
cultures. We present the full protocol for assembling and maintaining a 16-chamber
mesostat array by a single person as well as control data illustrating the ability
of the system to track trends and variability in the abundance of organisms among
replicates. We also present pilot data illustrating the ability to use the mesostats
to evolve resistance in C. reinhardtii to growth inhibiting herbicide glyphosate.
Furthermore, we have included data from this
system illustrating the signal of common problems like leaks, contamination and cell
clumping, showing how to distinguish it from biological variation as well as how to
prevent and address these problems if they occur. We also outline the ways in which
this system could be further modified and avenues of future research.
Discussion Content:
Lines 179–208: Chemostats offer a number of advantages over batch cultures for long-term
experimental evolution research. Precise control of selective pressures in a chemically
constant environment without evolutionary bottlenecks along with a link between growth
rate and dilution rate constitute a useful conceptual framework for modelling evolutionary
adaptation and population dynamics. This system adds to the small, but growing, number
of efforts to produce simple but scalable, multiplexed DIY chemostats from cheaper
materials that are possible to build and maintain by a single person [21–27], and
is the first of its kind for experimental evolution of algae, specifically the evolution
of herbicide resistance in model species C. reinhardtii. There are three substantive
changes from the Miller et al. ministats [22], one system specific and two generic
changes to suit experimental evolution with continuous sample extraction. Firstly
the system was adapted to suit the study species C. reinhardtii, including light and
a lower dilution rate, which distinguishes the system from previous DIY chemostat
arrays developed for
maintenance of yeast [21, 22, 26] and bacterial cultures [23, 27]. Secondly, a needle
and syringe system was added to facilitate easy, sterile access to the culture for
the removal of samples. This allows sampling from the middle of the active culture
rather than relying on the overflow. The efflux only samples from the top and the
overflow chamber constitutes a wholly different environment without continuous dilution,
build-up of waste products and increased evaporation, making them unrepresentative
samples of the chamber populations. Furthermore this simplifies addition of cells
or treatment compounds directly to the chambers, eliminating the risk of contamination
that comes with disconnecting the medium influx or efflux channels. While sampling
ports have been described before (e.g. [23, 25]) our simplification and combination
with syringe extraction allows manual sampling with minimal contamination. The third
change is an increase in the chamber volume to allow larger population sizes and possible
future introduction of several trophic levels. Furthermore, this increases increases
the amount of sample that can be extracted on a regular basis, extending the possibilities
for the types of assays that can be performed to characterise evolution in action,
as most of the
previous DIY chemostat arrays have been limited by their small working sizes [22,
23, 25]. Lastly, there were several changes to specific materials to lower the overall
costs.
Furthermore, we have included additional pilot data demonstrating how the system may
be applied to herbicide resistance evolution specifically, using growth inhibiting
herbicide glyphosate. The added results have been pasted below (from lines 130–145):
Applicability to herbicide resistance evolution
Fig 2 shows the population densities of the four glyphosate treated populations and
two control populations for 24 days following glyphosate treatment introduction. The
glyphosate treated chambers exhibit population decline at a rate approximate to (150
mg/L, slope = -0.14, SE = 0.006) or below (100 mg/L, slope = -0.098, SE = 0.006) the
dilution rate of 0.15/day. In the same timespan, the control populations exhibit an
overall slight increase in population density (slope = 0.022, SE = 0.006), possibly
reflecting adaptation to the mesostat environment. The onset of the population decline
appears to be immediate for the 150 mg/L glyphosate treatment, whereas it occurs roughly
5 days after the glyphosate injection for the 100 mg/L glyphosate treatment. This
is likely due to the 100 mg/L glyphosate treatment being just on the cusp of the minimum
inhibitory concentration, enabling the populations to maintain growth for a short
while before the herbicidal action is apparent. After 15 and 18 days respectively
of population density decline, the 100 mg/L populations increase in cell density again,
suggesting the populations have evolved resistance to the glyphosate, whereas the
150 mg/L populations never show evidence of resistance.
The authors can be advised to set forth a scientifically relevant problem and solve
it with the proposed mesostat(s) to demonstrate its applicability.
RESPONSE: Thank you for this suggestion. We have added additional data that provides
insight into the nature of the relevant problems and questions that can be addressed
via this protocol.
We note that this work is a protocol via which three additional papers will be submitted.
Minor comments:
Introduction: Here I describe a… — please rephrase.
The authors mention “…this thesis..” in the section called “Limitations”, but this
is an article, not a thesis, right?
RESPONSE: Thank you for noticing these mistakes, they have been corrected.
Reviewer #2: The manuscript describes a development of a low cost mesoscale continuous
culturing system, which could be applied to investigate adaptive evolution of microalgae.
The method has potential utility for research and describes some new developments,
and the methods and materials are described in sufficient details for reproducibility.
However some improvements are necessary for method validation and other points as
indicated below.
line 77-78, replicability: it is not clear why only these dilution rates (0.15 and
0.3 /day) were applied. Is there a range of dilution rates, based on which the ideal
dilution rate (i.e. to achieve steady state) could be established? Did the authors
test some 'extreme' range of dilution rates to test the plasticity of the setup?
RESPONSE: Thank you for this comment. The chosen dilution rates were based on previous
studies using similar organisms that this system was designed for, along with testing
a lower dilution rate did not lead to major changes in the steady state population
density, with the intention to decrease consumption of growth medium as well as wear
and tear on the pump tubing. It was not intended to be a stress test. We have now
addressed this in the Methods section (see pasted above in response to comment 3 by
Reviewer #1 and on line 68–73 of the revised document).
Figure 2: the results of this experiment are not entirely clear, because there seems
to be quite a large variations in treatments A-E as well, which did not undergo leak
test. Conditions A-E need to be better explained.
Figure 3: There is some variability of population density in the control, non-contaminated
cultures as well, so these results cannot be judged with certainty. Authors refer
to 'likely' source of contamination, could not this be better verified with deliberately
contaminating some of the cultures, e.g. by inoculating some bacteria?
RESPONSE: Thank you for these two comments. There are many examples that the variability
we see in the unaffected populations here is normal for this type of system (e.g.
Fussmann et al. 2000 Science. Crossing the Hopf Bifurcation in a Live Predator-Prey
System; Yoshida et al. 2003 Nature. Rapid Evolution Drives Ecological Dynamics in
a Predator-Prey System; Becks et al. 2012 Ecology Letters. The Functional Genomics
of an Eco-Evolutionary Feedback Loop: Linking Gene Expression, Trait Evolution, and
Community Dynamics. Note all are cited in the manuscript).
While we cannot do more experiments to cause deliberate leaks or contamination, we
believe the data presented in this manuscript from spontaneous faults to the system
serves to effectively illustrate that the data signal of these errors, compared to
normal variability within and among the unaffected populations, is very readily identified.
To make this clearer we have highlighted the distinction between normal variability
among replicates by highlighting variation in the replicability section as well as
improving the description of the features of variation caused by faults in the common
problems section (see changes in paragraphs pasted below).
We have also added in clumping as a common problem based on your suggestion below,
and pilot data for the population density response to the growth inhibiting herbicide
glyphosate to demonstrate that the effect of a herbicide treatment is readily distinguishable
from normal variation and equipment faults.
Changes copied from the revised document:
Among population variation, lines 123–129: The level of variation observed in this
data set is normal for this type of system [16–18] and can be divided into among population
variation and day-to-day variation. Among population variation is primarily caused
by the biology of the system as these are separate, genetically heterogeneous populations
on separate evolutionary trajectories. Day-to-day variation is however at least partly
caused by limitations in sample processing. Both are discussed in more detail in the
Discussion section, as well as how to reduce or circumvent the latter in particular.
Leak, lines 149–159: Compared to the expected among population and within-population
day-to-day variation observed in the chambers that did not experience a leak, three
crucial differences together make this the characteristic signal of over-dilution:
1) While similarly large day-to-day fluctuations in the measured density occur in
the presented data set, day effects present across chambers. The rapid reduction in
population density for chamber F between days 34 and 35 is only apparent in that chamber,
whereas a similar reduction between days 37 and 38 is seen in all of chambers A–E.
2) The reduction in population density in chamber F results in a lower population
density than otherwise observed in the data set (by roughly 3 x 105 cells/ml). 3)
The reduced population density is observed in chamber F for several days after day
35, rather than recovering by the next day like seen for chambers A–E after day 38.
Contamination, lines 166–170: Furthermore, while there is considerable variation among
all populations, the signal of contamination in the data is clearly distinguished
from the expected among population variation and day-to-day variation by the fact
that it is a consistent, long-term population-density decline without recovery 12
days after the contamination event.
Clumping, lines 174–177: The data signal here is an artefact of the limitations of
the instrument being unable to accurately distinguish individual cells within aggregates,
resulting in huge fluctuations in estimated cell density considerably larger than
and out of step with the day-to-day variation observed in the other populations.
Finally, as sources of variation are of pivotal importance in ecology and evolution
studies, a subsection has also been added to the discussion section on sources of
variation, highlighting those where it cannot be minimised as well as suggesting methods
to reduce (pasted below from lines 210–292):
Sources of variation and how to minimise it
The data presented here illustrates the expected variation between cultures and how
to identify the signal of equipment failure, such as a leak, or contamination. We
also demonstrate that the system can be used to evolve resistance to growth inhibiting
herbicide glyphosate, and that the signal of herbicidal action is apparent as a population
density decline, followed by an increase after the population has evolved resistance.
The herbicidal effect is clearly distinguishable from the expected variation under
control conditions, and given enough time, the resistant population is expected to
settle at a new steady state.
The variation among replicate populations observed here is normal [16–18] and expected
as they constitute separate, genetically heterogeneous populations on separate evolutionary
trajectories. Even when using a single founder population, the genetic bottleneck
caused by splitting it between populations as well as the dynamics during the establishment
phase of batch-like growth dynamics [6] will result in similar but distinct populations
by the time they reach steady state. Effort should be made to ensure that all chambers
receive the same levels of light and aeration as well as consistent dilution with
the same medium, and starting variation could be eliminated through starting with
clone populations at a high enough concentration to effectively avoid the establishment
phase. However, the among population variation is generally of scientific interest
to experimental evolution studies and should be investigated rather than eliminated.
Conversely, while day-to-day variation within a population is also normal for this
type of system, it is also partly caused by limitations to the sampling protocol.
The data presented here was obtained from measurements performed on living cells that
had the opportunity to grow and divide between sample extraction from the mesostat
chambers and sampling processing. While this is an unavoidable source of variation,
it can however be reduced by minimising the time that passes and working in a controlled
temperature environment. If the experimental design allows, the cells can be immobilised
by using e.g. Lugol’s solution before counting with flow- or haemocytometry. It is
also possible to control for this variation by including sampling day as a source
of error in statistical models applied to the data
The among population and day-to-day within population variation are however both clearly
distinguishable from the data signal of common faults like leaks, contamination and
clumping. While these faults are likely to be detected before they become apparent
in the population density data, leaks causing significant over-dilution are apparent
within a few hours while clumping and contamination can be observed under a microscope,
it is important to understand how they affect the data so that an informed decision
can be made on how to handle it. While the population density is always expected to
quickly return to steady state after over-dilution, the increased flushing out of
cells constitutes an evolutionary bottleneck and the changed growth conditions may
affect other traits of the population not visible in the population density data and
data collected subsequent to a major leak should thus be treated with caution. The
leak presented here was caused by equipment failure resulting in over-dilution, but
smaller leaks often occur as the pump tubing wears out with long term use, which can
lead to under-dilution of the connected chambers. Both are best prevented by regular
inspection of the pump parts for irregularities
Bacterial contamination is another common risk in long-running continuous cultures
[15], and is best prevented by working in a sterile environment and minimising the
points at which contamination can enter the system. The main contamination risk presents
when disconnecting any part of the array, such as when switching medium containers,
or when extracting samples, and particular care should be taken to keep the connecting
parts sterilised during. The example presented here is the only instance of contamination
observed across eight separate experiments each lasting more than a month and happened
when the chambers were disconnected from the array for a longer period of time and
removed from the sterile environment. Even so, only four out of six chambers showed
evidence of contamination under the microscope 12 days after the contamination event,
despite all of the chambers in question sharing a medium source. This suggests that
the system is robust in terms of contamination not spreading between the chambers.
While regular microscopy inspection of cell samples for contamination is recommended,
this can be laborious with a large number of replicates and the characteristic population
density decline provides another opportunity to detect and isolate the problem.
Lastly, chemostat populations being under a selective pressure to evolve phenotypes
reducing their risk of being flushed out is an often cited issue with the method [7,
15], presenting as adhesion to the chamber walls and cell flocculations. While this
phenomenon has as of yet never been observed under control conditions with this protocol,
there was one instance of cells exhibiting a clumping phenotype while under treatment
with a sublethal dose of glyphosate, making it possible it was a response to the treatment
rather than the mesostat environment. In C. reinhardtii there are two distinct types
of clumping: cell aggregations of separate cells and palmelloid colonies that share
a cell wall [40–42]. Both have been found to be an induced response as well as heritable
[40, 41, 43–45], meaning that once they become common in a population they may be
hard to get rid of [40]. Palmelloids are small enough that they will not cause blockages,
but due to the shared cell wall they cannot be disassociated through bubbling or by
vortexing a sample. Cell aggregations can be considerably larger, however they are
also possible to break apart through vortexing, and vigorous bubbling of the cultures
often prevents their formation [7]. How much of a problem clumping is depends on the
experiment, i.e. it becomes a problem if it hinders sample processing and when it
is thought to be an artefact of the chemostat environment rather than in response
to the applied treatment. For population density measurements by flow cytometry as
presented here, clumping considerably reduces the accuracy of the measurements as
each clump is counted as a single particle, increasing day-to-day variation. In this
case, manual haemocytometry could give a better estimate but this is considerably
more laborious.
Besides the common problems investigated, are there any potential problems with the
introduced mesoscale chemostat system? For example changes in system pressure, aeration,
problems with cell clumping, clogging of the tubing system and needles, etc. The impact
of these potential problems should be mentioned and possible solutions offered to
circumvent these issues.
RESPONSE: Thank you for this idea. The prevention of problems relating to aeration
and blockages is addressed by the Daily Maintenance section of the protocol, and we
have expanded the detail around this (pasted below). We have also included this set
of issues and solutions in the Discussion (see pasted below). Furthermore, we have
extended the Common Problems section to include representative data for the data signal
of clumping. This is also addressed in the Discussion and linked to detecting variation
due to protocol issues vs. biological variation (see above).
Protocol, Step 18.2 Check culture chamber airflow is satisfactory and equal. If a
culture chamber has low airflow, check if the filters are blocked, first the filter
on the adjoining collection chamber, then the filter connecting to the manifold, and
replace if necessary. The most common cause of filter blockage is them becoming wet,
either by an efflux blockage or low pressure resulting in the culture entering the
aeration needling and tubing, or by high ambient humidity. Make a note of airflow
problems data is being collected if it is possible the problems have been present
for more than an hour. If filters are often blocked, consider changing the ambient
humidity.
Protocol, Step 18.4 Check culture levels are even and do not deviate from the 380
ml line. If the medium level is too low, check the media influx tubing and needle
for blockages. The most likely points for blockages are inside the pump tubing and
any needles due to their narrow gauges. If medium influx is normal, adjust the efflux
needle, and check if the level is back to normal in a couple of hours (time needed
to wait dependent on flow rate and total volume deviation). If the level is too high,
examine the efflux tubing and needle for blockages, along with the collection chamber
filter. When unblocked, the culture chamber level should return to normal volume relatively
quickly. Always make a note of culture level changes if data is being collected.
Discussion, lines 316–327: Another potential problem involves insufficient aeration
or efflux blockages causing over- or under-pressure in the chambers. Provided the
air supply is sufficient, the most common reason for low or uneven bubbling is blocked
air filters, usually because they have become damp. If the air filters frequently
become damp, it may be due to too high ambient humidity. Not enough bubbling may cause
sedimentation and stratification of the culture, as well as selection for phenotypes
that sink so they avoid being flushed out, or it may instead cause the culture to
rise through the aeration needle instead of through the efflux needle. Clogging of
the media line is uncommon, but can occur if not properly sterilised and contamination
is allowed to grow. This is often apparent as a reduction in flow into and out of
the affected chambers. The daily maintenance part of the protocol outlines how to
spot and address these problems.
line 39: perhaps plural pronoun should be used throughout, as there are multiple authors
of the work.
RESPONSE: We have made this change. Thank you for pointing this out.
line 118-119 - if the work was the basis of a thesis, add the relevant reference of
the thesis.
RESPONSE: See response to Reviewer #1 above — this work is the first paper submitted
for publication from a thesis, the three papers that have utilised this system are
currently in preparation for publication before the end of the year.
line 128-129: 'the system must therefore be carefully balanced against the resulting
biology' -this is unclear, please clarify/define this more precisely.
RESPONSE: Thank you for this comment. This statement was referring specifically to
cost saving measures like lowering the dilution rate also changing the selective pressure
experienced by the population. We have edited the relevant section to make this clearer
(pasted below from lines 302–307):
One way to conserve medium and treatment components is to lower the dilution rate,
which in the experiments presented here had no effect on population density in the
chambers. However, this changes the selection pressures experienced by the populations
as well as their doubling rate [15]. The logistics of the system and any cost saving
measures must therefore be carefully balanced against the resulting biology, taking
into account the desired selective pressure, cell cycle stage and generation time.
line 133-135: are there pilot experiment performed on the potential effects of sampling
frequency on population density and potential disruption of the system?
RESPONSE: Thank you for this idea. Unfortunately we have no data available on this,
it is an assumption made from theory – the constancy of the environment and thus the
selective pressure is based upon continuous dilution. However, see the referenced
Fischer et al. 2014 (“The Exponentially Fed Batch Culture as a Reliable Alternative
to Conventional Chemostats”) as this is something they try to address with their design.
line 148-150: In a high performance chemostat system, these sensors are indeed important,
therefore the authors need to demonstrate the feasibility of introducing these sensors
in their system.
RESPONSE: There is a notable example, the “omnistat”, fitting many sensors into a
DIY system (Ekkers et al. 2020. The Omnistat: A Flexible Continuous-Culture System
for Prolonged Experimental Evolution). Our mesostats are however an example of a simpler
DIY system where sensors are an optional extra, afforded by the fact that it is overall
larger and the lids thus have the space for them if needed for the experiment in question.
The text has been amended to reference this paper and clarify this detail (pasted
below from lines 340–341):
As the chamber lids are relatively large, sensors to monitor e.g. pH or CO2 levels
could also be fitted through additional ports (see [25]).
The light system lacks some details. What was the light source applied? Could the
authors provide the emission spectrum of the lamp? Is it possible to provide diurnal
light cycles? Is it possible to change the spectral composition of the light source?
So could it be tuned for different photosynthetic organisms with different photosynthetic
pigment or light harvesting antenna composition?
RESPONSE: Thank you for these ideas. The light sources are addressed in the overview
of the design in the protocol: “When growing photosynthetic organisms such as algae,
even light levels for all chambers are best maintained by a light table as well as
fitting strip lights around the chambers.” This has been extended in the Overview
of the design section of the protocol (see pasted below) and the lights have been
added to the Materials list. The current light system design is very simple, as that
is sufficient for C. reinhardtii and the questions we used the system to address,
but the majority of your suggested modifications would indeed be possible to implement,
and we have now enhanced this detail in the How to modify or improve section (see
pasted below).
Overview of the design: The light system
The light is provided by LED strip lights mounted around the chambers and between
the two rows of chambers, as well as a DIY light box consisting of LED strip lights
and a semi-transparent plastic top to diffuse the light. Equal light from all angles
is essential to ensure even algal growth in the chambers. A light box is not necessary,
but convenient and can be used for providing light to batch cultures or growth assays
of subsamples.
Discussion, lines 352–361: The light system here is rudimentary but sufficient for
C. reinhardtii growth [40], using warm white light LED strip lights mounted around
the chambers along with DIY light box also consisting of white light LED strip lights
and a semi-transparent plastic top to diffuse the light. The light box is not necessary,
but convenient for maximising light from all angles. Under control conditions 24h
light was used, but it is possible to fit a timer to the outlet connecting the lights
to instead provide a diurnal light cycle. Coloured semi-transparent plastic could
be used to provide light only from a specific part of the light spectrum, but it would
also be possible to mount specialist lighting around the mesostats if tuning for a
specific photosynthetic organism or experiment is desired.
In the Discussion, authors describe several potential future improvements and research
opportunities, but how feasible are these practically for the introduced system? For
example, temperature control seems to be unsolved for the current stage of development,
and the application of e.g. a water bath may need a significant redesign of the entire
system. The feasibility of the proposed applications and future developments should
be compared and analyzed with other existing continuous culturing systems.
RESPONSE: Thank you for this comment. We appreciate this request which we interpret
as a request for the context and potential of these types of platforms for a variety
of experimental evolution work. We have increased our discussion of the context of
our work by comparing it to other protocols (see new Discussion lines 179–208, pasted
above in response to reviewer #1), and identified a few ways to extend our protocol
with minimal modification (see pasted below).
Lines 346–351: Any water-tight, sterilisable container can be used for culture chambers
if suitable lids can be manufactured, such as falcon tubes [23] or commonly available
lab glassware [25]. The controlled temperature room can be replaced with water baths
(note however that this requires mounting the lights up on the sides of the water
baths), and portable aquarium pumps can be used instead of building infrastructure,
increasing flexibility in where the system can be housed.
Lines 356–361: Under control conditions 24h light was used, but it is possible to
fit a timer to the outlet connecting the lights to instead provide a diurnal light
cycle. Coloured semi-transparent plastic could be used to provide light only from
a specific part of the light spectrum, but it would also be possible to mount specialist
lighting around the mesostats if tuning for a specific photosynthetic organism or
experiment is desired.
Lines 363–375: We have used this system for experimental evolution of herbicide resistance
in algae by adding glyphosate as a shock injection and then continuously through the
growth medium, however, this setup is also easily adaptable depending on the research
question. The herbicide treatment could also be applied gradually through the medium
or through series of shock injections in a ratchet protocol [28] and investigate to
what level the resistance can be pushed and at what speed. The dilution rate and thus
the cell growth rate is set by the pump speed, tubing thickness and culture volume,
so running chambers with different dilution rates simultaneously would be possible
with different pump tubing thicknesses, multiple pumps or multiple culture volumes,
depending on the range required. Furthermore, the use of multiple light tables with
opaque partitions between cultures would allow testing for an interaction with light
level, or the chambers could be kept in water baths at different temperatures to determine
the effect of temperature.
For example, the multiple trophic level experiments referenced were all single chemostat
systems, and our system was specifically designed for the same types of populations
with regards to size and light levels but allowing simultaneous replication. This
is now made clear in the closing paragraph (excerpt from lines 381–387 pasted below).
Lines 342–345: The predator-prey cycles of rotifer Brachionus calyciflorus and C.
reinhardtii as well as Chlorella vulgaris have been successfully studied and modelled
using chemostat environments (e.g. [17,18]) and our setup allows simplified simultaneous
replication for this type of system that can be maintained by one person. Competition
could also be introduced to the system through using multiple algal strains and monitoring
their frequencies or through expanding the culture ecosystem to include other algal
species or bacteria [48].
With regards to temperature control, as the system is waterproof, introducing water
baths for temperature control is as simple as placing the chambers in a water bath.
Strip lights can be mounted out of the water on the sides and above, which is something
we have done in the past for batch cultures. We have now clarified in the text that
minor adjustments to light placement would be required (see above or on line 350).
7. PLOS authors have the option to publish the peer review history of their article
(what does this mean?). If published, this will include your full peer review and
any attached files.
If you choose “no”, your identity will remain anonymous but your review may still
be made public.
Do you want your identity to be public for this peer review? For information about
this choice, including consent withdrawal, please see our Privacy Policy.
Reviewer #1: No
Reviewer #2: No
[NOTE: If reviewer comments were submitted as an attachment file, they will be attached
to this email and accessible via the submission site. Please log into your account,
locate the manuscript record, and check for the action link "View Attachments". If
this link does not appear, there are no attachment files.]
While revising your submission, please upload your figure files to the Preflight Analysis
and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first
register as a user. Registration is free. Then, login and navigate to the UPLOAD tab,
where you will find detailed instructions on how to use the tool. If you encounter
any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.
- Attachments
- Attachment
Submitted filename: Response to Reviewers.pdf