Peer Review History

Original SubmissionNovember 24, 2021
Decision Letter - Purvi Purohit, Editor

PONE-D-21-37268Anti-TSH receptor antibodies (TRAb): comparison of two third generation automated immunoassays broadly used in clinical laboratories and results interpretationPLOS ONE

Dear Dr. Jose Maria Lopez Ortega,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by 31 march. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Purvi Purohit

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please provide additional details regarding participant consent. In the ethics statement in the Methods and online submission information, please ensure that you have specified what type you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed). If your study included minors, state whether you obtained consent from parents or guardians. If the need for consent was waived by the ethics committee, please include this information.

Additional Editor Comments (if provided):

The manuscript needs major revisions and is not acceptable in its current format. If the authors wish to continue they should address the queries raised by the reviewer.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: José María López Ortega et al. manuscript "Anti-TSH receptor antibodies (TRAb): comparison of two third generation automated immunoassays broadly used in clinical laboratories and results interpretation" studied the comparison of accuracy in terms of sensitivity and specificity of two automated IIIrd generation immunoassays in use for TRAb testing of a clinical laboratory, the recent EliATM method (ThermoFisher Scientific, Germany) and the Elecsys® /Cobas ECLIA (Roche Diagnostics Ltd, Switzerland). This paper is easy to understand and read. Some of the aspects to be improved to favor its understanding is the methodology and result sections. I would like to add some comments and suggestions to this paper.

1. Please write a full form of abbreviation (TRAb) {Page No.4; Line No. 67}

2. I think the cost per sample matters a lot here, it is very important to compare it, today most of the third world countries are facing the lowest level of health services, so the lowest cost of any diagnostics test is to reach every poor patient.

3. Apart from this, by doing a comparative analysis on the following points, their explanation can be helpful in improving this Manuscript.

(A) Assay Time

(B) sample volume

(C) Measuring Range

(D) Functional Sensitivity

(E) Cutoff

(F) Precision

4. Please correct typing error. {Page No.15; Line No. 277}

"For the EliATM test the reported specificity of 99.6% is very similar to the 99.4% in our study, but in our series the optimal cut-off of 3,2 IU/L is closer the upper limit of normality provided by the manufacturer, and slightly lower than the optimal cut-off of 3.8 IU/L reported in their paper."

5. Please explain "disease control" in the manuscript.

6. In Figure 3, denote each specific sample in relation to disease.

7. Many figure captions do not explain what colouring and boxing mean in many figures. The figure legend needs a lot of improvement.

8. I would like to see a critical discussion on limitations in the current study.

9. The authors perform very similar analyses to https://doi.org/10.1186/s12902-019-0363-6, except I would argue the linked paper is more thorough since they use more rigorous statistical analysis. The authors should justify what their paper adds to the field that this paper doesn’t.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Dear reviewers,

Thank you very much for the time you've dedicated to evaluate our work. Please find below the response to all the comments:

2. Reviewer 1

Comment from the author: all changes applied to the manuscript due to reviewer suggestions are highlighted in the manuscript in light blue.

2.1. Comment 1

Please write a full form of abbreviation (TRAb) {Page No.4; Line No. 67}

Response

The full form abbreviation has been written. See below:

“functional effect on TSH-R, three types of TRAb or Thyrotropin Receptor Autoantibodies (TRAb) can be considered:”

2.2. Comment 2

I think the cost per sample matters a lot here, it is very important to compare it, today most of the third world countries are facing the lowest level of health services, so the lowest cost of any diagnostics test is to reach every poor patient.

Response

We completely agree with the reviewer comment, but we were not able to conduct a health economic evaluation. It must be taken into account that, in our country, health assistance is free for all citizens, so we are focusing on offering solutions that provide better aid in the diagnosis of patients. Moreover, the price of a test is subjected to many variables (public tenders, test volume, etc.).

A reader from third world countries must evaluate the results of this study in the context of the country-specific prices for the methods, and select the best option for them.

2.3. Comment 3

3. Apart from this, by doing a comparative analysis on the following points, their explanation can be helpful in improving this Manuscript.

(A) Assay Time

(B) sample volume

(C) Measuring Range

(D) Functional Sensitivity

(E) Cut-off

(F) Precision

Response

Again we agree with reviewer comments. Some of the aforementioned assay characteristics were already described in table 2 (measuring range, functional sensitivity, and cut-off) but we have added the ones missing (assay time, sample volume, and precision). Also, we have added the following paragraph to the discussion (page 16, line 310)

“In this study, we have focused on test performance, however, other test characteristics like assay time, sample volume, and precision are also worthy of evaluation when selecting the appropriate test for a specific laboratory. Nevertheless, it is difficult to draw general conclusions, as their impact mainly depends on the type of laboratory system used, and the laboratory workflow. Sample volume is not critical in daily routine, however, lower volume sample needs are always better for patients.”

2.4. Comment 4

Please correct typing error. {Page No.15; Line No. 277}.

Response

The typing error has been corrected. See below:

“99.4% in our study, but in our series the optimal cut-off of 3.2 IU/L is closer the”

2.5. Comment 5

Please explain "disease control" in the manuscript.

Response

We did not include a “disease control group” if it is defined as “healthy population”. However, we included 60 samples received for TRAb measurement corresponding to diagnostics unrelated to thyroid pathologies. With the exception of a couple of patients with Diabetes Mellitus type I, samples at this group corresponded to patients under control and not affected by autoimmune diseases or under clinical processes of a minor entity. We have modified the text accordingly to make it clear (pag 7, line 131):

“Group 7: Miscellaneous (MISC), 60 cases referred to our laboratory because of diseases unrelated to thyroid pathology. This group was considered the control group in the study.”

2.6. Comment 6

In Figure 3, denote each specific sample in relation to disease.

Response

We would like to denote each specific sample in relation to disease as the reviewer comments, however a modification like the proposed requires the involvement of the statistician and, due to the short timespan we’ve had, It has not been possible. As it is a clear improvement, but we do not consider it critical, we leave it to your criteria to include the modification for the final publication if you think it is essential, and of course, the manuscript is accepted.

2.7. Comment 7

Many figure captions do not explain what coloring and boxing mean in many figures. The figure legend needs a lot of improvement.

Response

We completely agree with reviewer's comments. We have modified figure captions in order to better explain them. See below the modified figure captions:

“Fig 1. Patient distribution in different clinical groups according to the final diagnosis.

Fig 2. Passing-Bablok regression. Correlation between EliA™ and Elecsys® TRAb. Trab are displayed in UI/l. Dotted line (- - - -) corresponds to identity. Blue line corresponds to Passing Bablock regression and double dotted line (-- – – –) to 95% CI bands. Tests values are displayed as orange circles.

Fig 3. Bland-Altman plot. Inter-assay agreement between ELIA™ and Elecsys® TRAb. The dashed line (........) corresponds to bias. Lower (-8.036) and upper (6.195) 95% limits of agreement are plotted as speckled lines. Y-axis plots the difference scores of ELIA™ and Elecsys® TRAb measurements against the mean for each studied sample.

Fig 4. Box-plot of Kruskal-Wallis analysis. A. Elecsys®TRAb levels in the different groups enrolled in the study. B. EliATM TRAb levels in the different groups enrolled in the study. TRAb levels are displayed as data ranks obtained by transforming to continue data ranks using Roche’s test. For each box plot, the central line represents the median; the boxes limits represent the upper and lower quartile; black points are outliers (a value more than 1.5 times the interquartile range above/below the interquartile values); and the black lines are the whiskers, which extend from the interquartile ranges to the maximum values that are not classed as outliers.

NDGD, newly diagnosed untreated GD patients; GD-R, GD patients in remission; GD-T, GD patients under treatment; AIT, Autoimmune Thyroiditis; MNG, Multinodular Goiter non-toxic; OTD, Other Thyroid Diseases; Misc, Miscellaneous.

Fig 5. ROC (Receiver operating curve) analysis for EliATM and Elecsys®. ROC curve evaluates the accuracy of both methods at manufacturer cut-off values, showing the tend to the specificity of EliATM assay.”

2.8. Comment 8

I would like to see a critical discussion on limitations in the current study.

Response

We especially agree with the reviewer. All manuscripts must include a description of the limitations of the study in the discussion section and our manuscript does not include it. Thus, have included the following paragraph in the discussion (page 13, line 248):

“Our study has two main limitations, first, we did not include a healthy patients group as the negative control, which is always good to have. Second, we did not evaluate the performance of both tests for patients' disease follow-up and relapse prediction. This is also a key point when deciding which test to include in the daily routine, and our next objective as follow-up performance data is scarce on TRAb measurement methods.”

2.9. Comment 9

The authors perform very similar analyses to https://doi.org/10.1186/s12902-019-0363-6, except I would argue the linked paper is more thorough since they use more rigorous statistical analysis. The authors should justify what their paper adds to the field that this paper doesn’t.

Response

Struja’s manuscript is really interesting for groups working in the field. However, although Struja’s comparative study includes EliATM anti-TSH-R, it does not include Elecsys® anti-TSH-R. EliATM and Elecsys® are two of the main used methods in our environment and that's why we decided to perform this specific comparison, to have a head-to-head comparison between two methods that have never been compared before, and that can be useful for professionals working in the field.

In fact, we highlight the novelty of the comparison in the manuscript (pag 13, line 261).

“Articles comparing different TRAb assays including the EliATM or the Elecsys® tests have been published, but to our knowledge, this is the first head-to-head direct comparison (Table 4).”

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Purvi Purohit, Editor

Anti-TSH receptor antibodies (TRAb): comparison of two third generation automated immunoassays broadly used in clinical laboratories and results interpretation

PONE-D-21-37268R1

Dear Dr. José María López Ortega

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Purvi Purohit

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Congratulations to the authors. The manuscript has been revised satisfactorily and can now be accepted for publication.

Formally Accepted
Acceptance Letter - Purvi Purohit, Editor

PONE-D-21-37268R1

Anti-TSH receptor antibodies (TRAb): comparison of two third generation automated immunoassays broadly used in clinical laboratories and results interpretation

Dear Dr. López Ortega:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Purvi Purohit

Academic Editor

PLOS ONE