Peer Review History

Original SubmissionNovember 15, 2021
Decision Letter - Rafael da Costa Monsanto, Editor

PONE-D-21-36271Cochlear implant positioning and fixation using 3D-printed patient specific surgical guides; a cadaveric studyPLOS ONE

Dear Dr. Markodimitraki,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Specifically:Although the manuscript is interesting and reads well, both reviewers made comments on the feasibility of such system on daily operations. Furthermore, some of the results could be presented in a more informative manner to reach a broader community of surgeons.

Please submit your revised manuscript by Mar 21 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Rafael da Costa Monsanto, M.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf  and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please include the name and URL of the donation center in your ethics statement in the manuscript Methods and online submission form.

3. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter.

4. Please include your full ethics statement in the ‘Methods’ section of your manuscript file. In your statement, please include the full name of the IRB or ethics committee who approved or waived your study, as well as whether or not you obtained informed written or verbal consent. If consent was waived for your study, please include this information in your statement as well.

Additional Editor Comments (if provided):

Please address the comments made by the reviewers. Also, please make sure the article suits the journal requirements on competing interest and disclosures

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: N/A

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Thank you for the opportunity to review the study titled "Cochlear implant positioning and fixation using 3D-printed patient specific surgical guides; a cadaveric study". Authors developed a 3-D printed surgical guide for cochlear implantation.

Despite the main idea of patient-specific surgical guides and treatment, which I personaly believe is the future of medicine, there are few points I'd like the authors to explain.

1. Why not include Advanced Bionics' devices in the study?

2. In line 113, what is the meaning of STL?

3. In line 136 is said that soft tissue was meticulously removed. During CI surgery, it is preferable not to remove any soft tissue in order to avoid dehiscence of surgical wound.

4. It was not clear if authors found a significant difference between brands.

5. Do authors believe that the 3-D printed surgical guides are financialy suitable for daily activities?

Reviewer #2: Dear authors

The study “Cochlear implant positioning and fixation using 3D-printed patient specific surgical guides; the cadaveric study” aims to create of a surgical guide for a more accurate bone well drilling for the receiver/stimulator of the cochlear implant. I congratulate the authors for the work done. The topic is relevant, having a considerable importance for cochlear implant surgeons.

General Recommendation: Regarding the main theme, I see two questions that could be answered throughout the text. First: what would be the feasibility of carrying out the entire pre-operative planning and 3D printing in other services, using specific software, materials and specific 3d printers for the planning and production of the guide? Second: I believe it would be interesting to discuss in more detail the influence of using those surgical guide on the surgical and planning time.

Intro: The intro writing is good, we can understand the importance of the study, and the challenges that the authors want to solve. I have a specific question on the sentence “The internal part of the cochlear implant, also known as the receiver/stimulator (R/S) device, is designed to reside in close proximity to the pinna, without any interference with the external processor”. What would be the influence of the external processor in the receiver/stimulator? I understand that the external processor cannot be related to the pinna, but the receiver/stimulator will be related with the external processor.

Materials and methods: the authors say that a 45º plane relative to the Frankfurt horizontal plane is originated in the external auditory canal. I suggest the authors to further expand on whether the Frankfurt horizontal plane crosses the external auditory canal, or if it is above or below this. At the same part where the authors say: “the position for the Cochlear and MedEl models was determined so that the anterior part of the implant (receiver/stimulator) would be embedded whilst allowing the posterior part (magnet with coil) to rest on the skull”, something about the OTICON model could be added as well. In the next sentence is important to explain what is the ‘’STL of the cochlear implant’’, there is no mention in the rest of the text about the meaning of STL.

Surgical workflow: the last sentence, ‘’Drilling out of the bony wells of each specimen was performed simultaneously bilaterally, though each side of a specimen was implanted and scanned sequentially, in order to assess the depth of the bony bed without scattering created by the implant.’’ You need to explain how could be possible to drill and implant simultaneously in the same specimen head, maybe you did it bilaterally but not at the same time.

Analysis: it is difficult to understand which comparisons will be made. in my opinion, it would be more important to focus on the idea of comparison between the planned images and the images of the implanted specimen using the surgical guide. This topic point is not explored in depth. The technical details are interesting, but I think there is a need for a more didactic explanation. The help of images is important, but some terms that are not well worked by the author make reading difficult, such as row, pitch and yaw, as well as Euler angles, technical terms that end up making reading difficult and making the passage less interesting

Results: the results are expressed in the text, and in the table with the help of graphics. In the same way as in the topic of analysis, a lot of numerical information is placed, with not sufficient explanation, making the reading and understanding challenging. Even tables and graphs contain a lot of information, but interpretation is extremely difficult. It is difficult to understand which data is important. A more didactic approach would be interesting, better organizing the data, perhaps separating rotational from translational deviations, or some other type of division.

Discussion: The discussion is very well written; it helps a lot in understanding the idea of the study. A few points I would just like to comment:

In the excerpt: ''A challenge we faced during this study was finding the balance between optimizing the surgical guides' accuracy, while also maintaining the low level of invasiveness that is exercised in clinical practice.'' The radiation that patients are exposed in CBCT is one of the invasive points of the IC protocol, especially in young children. In services that only use MRI prior to surgery, would it be possible to carry out planning for the creation of the surgical guide? Question that refers to the initial theme reported by me in the general recommendations on the feasibility of the technique.

When analyzing the contact area of the surgical guide with the temporal bone ''The first designs of our surgical guide had a small contact area and also did not include the mastoid bone. We found that using both the external meatus rim and the mastoid bone as contact areas for the surgical guide gave the best results. These conditions were met by seven cases. '' It suggests that the authors used guides of different sizes in some cases, this should be said in the materials and methods too, I think it would be interesting to report this change in design that occurred throughout the study.

Conclusion: the conclusion of the text is satisfactory.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Maybe there has been a confusion, in the last resubmission I revised the data availability statement as requested. However, I received the same feedback as last time. I hope that you can evaluate the current statement.

Decision Letter - Rafael da Costa Monsanto, Editor

Cochlear implant positioning and fixation using 3D-printed patient specific surgical guides; a cadaveric study

PONE-D-21-36271R1

Dear Dr. Markodimitraki,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Rafael da Costa Monsanto, M.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Thank you for addressing the comments made by the reviewers. Congratulations for the excellent work.

Reviewers' comments:

Formally Accepted
Acceptance Letter - Rafael da Costa Monsanto, Editor

PONE-D-21-36271R1

Cochlear implant positioning and fixation using 3D-printed patient specific surgical guides; a cadaveric study

Dear Dr. Markodimitraki:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Rafael da Costa Monsanto

Academic Editor

PLOS ONE