Peer Review History

Original SubmissionSeptember 22, 2021
Decision Letter - Arum Han, Editor

PONE-D-21-30672Sensitizing Drug-Resistant Cancer Cells from Blood using Microfluidic ElectroporatorPLOS ONE

Dear Dr. Hur,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Feb 21 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Arum Han, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following in the Competing Interests section: 

S.C.H. benefits financially from royalty payments from the Vortex Biosciences, Inc.

Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. 

Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf.

3. PLOS requires an ORCID iD for the corresponding author in Editorial Manager on papers submitted after December 6th, 2016. Please ensure that you have an ORCID iD and that it is validated in Editorial Manager. To do this, go to ‘Update my Information’ (in the upper left-hand corner of the main menu), and click on the Fetch/Validate link next to the ORCID field. This will take you to the ORCID site and allow you to create a new iD or authenticate a pre-existing iD in Editorial Manager. Please see the following video for instructions on linking an ORCID iD to your Editorial Manager account: https://www.youtube.com/watch?v=_xcclfuvtxQ

4. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Please see the attached comments for details. Briefly, the manuscript described a microfluidic systems integrated with electrodes for trapping and electroplating cells. The experiments were designed and executed well.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachments
Attachment
Submitted filename: 2021_PONE-D-21-30672.docx
Revision 1

We thank the reviewer for their constructive comments and suggestions and have made the necessary edits to address their concerns. Please see the following point-by-point address.

Comment #1: I have the impression that this study aims to develop a workflow that could allow direct assessment of patient samples. However, the study works with blood samples that are spiked with HCC827 wt and HCC827 GR6 cells. For proof-of-concept, spiking cells with the blood sample is reasonable. It may be better to make it clear that the workflow has not been able to work with patient sample directly yet and point out this would be the next step work.

• This is a great point raised by the reviewer. The current study focuses on spiked cell experiments to develop the electroporation workflow for future transition into utilizing patient blood samples. The authors would like to respectfully point out that vortex technology has been demonstrated to capture CTCs, and that the electroporation workflow utilizing spiked cells has been stated in the abstract (Sentence 7 “HCC827 GR6 cells spiked…”), introduction (Paragraph 4 Sentence 8 “… we detail a procedure…”), results and discussion (Section: CTC purification and subsequent drug screening, Paragraph 2 Sentence 1), and conclusion (Paragraph 2 Sentence 5 “Gefitinib-resistant cancer cells emulating drug-refractory CTCs…”) to avoid confusion. The authors agree with the comment, and both the limitation of this current study and future goals to transition into patient samples have been more clearly stated in the updated manuscript (both introduction and conclusion).

Comment #2: It is not clear whether or not flushing the device with wash buffer prior to cell collection affects the processed cells, for example, are the cells flushed away?

• The authors appreciate the reviewer’s comment. When transitioning between any molecule/drug/wash buffer solution, our solution exchange system utilizes a simultaneous co-flow transition (i.e., there is no abrupt on-off switching from one vial to another) of solution vials at 40 psi to ensure that the flow rate through the device, at any given point when switching solutions, is sufficient to retain cell trapping. Similar solution exchange protocols have been demonstrated numerous times for cell trapping in the past (Ouyang et al. Sci Rep 2017, Vickers et al. Anal Chem 2014, Yun et al. Lab Chip 2013). This information has been revised on page 7 for improved reading clarity in the updated manuscript.

Comment #3: When collecting the cells to 96-well plates, was the cell density controlled in each well? If yes, how? If not, would the cell density variation affects the study?

• We thank the reviewer for this important comment. While the exact processed cell density cannot be ascertained until the endpoint, we have identified that the number of collected cells is largely dependent on the flow condition of the device as it directly affects robust vortex formation for cell trapping. The effects of these conditions on the cell density per well were minimized by closely monitoring the condition of the device and randomizing the input voltage sequence for electroporation. Additionally, the trapping efficiency and input cell concentrations were closely monitored, and the solution exchange procedure purposely designed to retain trapped cell populations. For viability assays that measure the reducing potential of all viable cells present, impacts from cell density variations were mitigated by recording chemiluminescence readings from two collective runs. Fold-change in cell chemiluminescence from each well was calculated in comparison over relative change to lessen the effects of cellular density variations. In the supplementary information provided with the manuscript, we observed that cell density variation effects on fold-change in cell chemiluminescence were negligible. The manuscript has been revised to include these details.

Comment #4: It is observed that 25V cause bubbles in addition to other problems. I believe 20V would cause bubble as well although it might not affect the study significantly. Please comment on that.

• This is an important observation pointed out by the reviewer. In the video file “S2_Movie.mp4”, the bubble formation at the electrodes, due to contributions from both electrolysis of the ionic solution and joule heating effects, can be observed in brightfield. Without sacrificing too much electroporation performance, the magnitude of the bubbles and rate at which cellular debris from burst cells accumulate on the electrodes were lower when operating at 20V compared to 25V when electroporating HCC827 wt cells. Additionally, bubble formation at 20V does not significantly impede the trapping capacity of the vortices and are also rapidly dissipated away from continual flow into the device (i.e., the large majority of cells in the view remain uninterrupted and remain robustly trapped in vortex after electroporation sequence). These explanations and details will be included in the revised manuscript.

Comment #5: It seems the liquid sample with spiked cells is directly in contact with the electrodes. If this is correct, would the cells adhere to the electrodes quickly leading to unstable performance? Please comment whether or not a passivation layer designed on top of the electrodes to separate the liquid sample from the electrodes is possible to make the system work?

• We thank the reviewer for this insightful suggestion. As stated before, cellular debris naturally accumulates onto the charged electrodes over the course of electroporation. In the past, we have tested passivating our microfluidic devices with 5% w/v Pluronics F-127 in DPBS to minimize biofouling, but this yielded insignificant impact in deterring cell debris adhesion (data not shown). There are several works with microscale interdigitated electrode circuits utilizing thin passivation layers to physically eliminate bubble generation and cell contact between electrodes (Jungreuthmayer et al. Lab Chip 2011, Pandian et al. Biomicrofluidics 2020). These passivation layers are integral for these systems where flow is restricted, cells are in physical contact with the underlying electrodes, or bubbles directly interfere with sample collection procedures. Additionally, the relative permittivity of a passivation layer compared to liquid ionic solution suggest significantly higher voltage inputs in order to generate the required electric fields for reversible membrane permeabilization, and thus may additionally require a sophisticated microelectrode fabrication technique to follow. Because the cells are in vortex orbit and the observed bubbles in our system are dispersed rapidly, similar working performance with these mentioned systems can be achieved with a simpler fabrication process.

Comment #6: I suggest “a vacuum environment of 6e-6 torr” change to “a vacuum environment of 6e-6 torr”

• This suggestion has been included in the revised manuscript.

Thank you for your time, consideration, and comments to our initial submission. We hope that the following response and manuscript amendments meet the publishing criteria set forth by the editors, reviewers, and PLOS One.

Attachments
Attachment
Submitted filename: Response To Reviewers.docx
Decision Letter - Arum Han, Editor

Sensitizing drug-resistant cancer cells from blood using microfluidic electroporator

PONE-D-21-30672R1

Dear Dr. Hur,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Arum Han, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Formally Accepted
Acceptance Letter - Arum Han, Editor

PONE-D-21-30672R1

Sensitizing drug-resistant cancer cells from blood using microfluidic electroporator

Dear Dr. Hur:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Arum Han

Academic Editor

PLOS ONE