Peer Review History
| Original SubmissionAugust 23, 2021 |
|---|
|
PONE-D-21-27242Evaluation of GENECUBE® HQ SARS-CoV-2 for anterior nasal samples and saliva samples with a new rapid examination protocolPLOS ONE Dear Dr. Suzuki, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. ACADEMIC EDITOR: As appended below, the reviewers have raised major concern/critique and suggested further justification/work to consolidate the findings. Do go through the comments and amend the MS accordingly. Please submit your revised manuscript by Nov 05 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, A. M. Abd El-Aty Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. Please amend your current ethics statement to address the following concerns: a) Did participants provide their written or verbal informed consent to participate in this study? b) If consent was verbal, please explain i) why written consent was not obtained, ii) how you documented participant consent, and iii) whether the ethics committees/IRB approved this consent procedure. 3. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match. When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section. 4. Thank you for stating the following financial disclosure: "This study was supported financially by TOYOBO Co., Ltd. PSS provided the MagDEA Dx SV and magLEAD Consumable Kit for the evaluation of the rapid protocol with magLEAD. TOYOBO Co., Ltd., provided support in the form of salaries to author Akio Sugiyama and Masashi Michibuchi, lecture fees to author Hiromichi Suzuki, and advisory fees to author Hiromichi Suzuki. Hiromichi Suzuki also received advisory fees from PSS. The funder did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript." We note that one or more of the authors is affiliated with the funding organization, indicating the funder may have had some role in the design, data collection, analysis or preparation of your manuscript for publication; in other words, the funder played an indirect role through the participation of the co-authors. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please do the following: a. Review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. These amendments should be made in the online form. b. Confirm in your cover letter that you agree with the following statement, and we will change the online submission form on your behalf: “The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section. 5. Thank you for stating the following in the Acknowledgments Section of your manuscript: "PSS provided the MagDEA Dx SV and magLEAD Consumable Kit for the evaluation of the rapid protocol with magLEAD. TOYOBO Co., Ltd., provided support in the form of salaries to author Akio Sugiyama and Masashi Michibuchi, lecture fees to author Hiromichi Suzuki, and advisory fees to author Hiromichi Suzuki. Hiromichi Suzuki also received advisory fees from PSS. The funder did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript." We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form. Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows: "This study was supported financially by TOYOBO Co., Ltd. PSS provided the MagDEA Dx SV and magLEAD Consumable Kit for the evaluation of the rapid protocol with magLEAD. TOYOBO Co., Ltd., provided support in the form of salaries to author Akio Sugiyama and Masashi Michibuchi, lecture fees to author Hiromichi Suzuki, and advisory fees to author Hiromichi Suzuki. Hiromichi Suzuki also received advisory fees from PSS. The funder did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript." Please include your amended statements within your cover letter; we will change the online submission form on your behalf. 6. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability. Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized. Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access. We will update your Data Availability statement to reflect the information you provide in your cover letter. 7. We note that Figure 1 in your submission contain [map/satellite] images which may be copyrighted. All PLOS content is published under the Creative Commons Attribution License (CC BY 4.0), which means that the manuscript, images, and Supporting Information files will be freely available online, and any third party is permitted to access, download, copy, distribute, and use these materials in any way, even commercially, with proper attribution. For these reasons, we cannot publish previously copyrighted maps or satellite images created using proprietary data, such as Google software (Google Maps, Street View, and Earth). For more information, see our copyright guidelines: http://journals.plos.org/plosone/s/licenses-and-copyright. We require you to either (1) present written permission from the copyright holder to publish these figures specifically under the CC BY 4.0 license, or (2) remove the figures from your submission: a. You may seek permission from the original copyright holder of Figure 1 to publish the content specifically under the CC BY 4.0 license. We recommend that you contact the original copyright holder with the Content Permission Form (http://journals.plos.org/plosone/s/file?id=7c09/content-permission-form.pdf) and the following text: “I request permission for the open-access journal PLOS ONE to publish XXX under the Creative Commons Attribution License (CCAL) CC BY 4.0 (http://creativecommons.org/licenses/by/4.0/). Please be aware that this license allows unrestricted use and distribution, even commercially, by third parties. Please reply and provide explicit written permission to publish XXX under a CC BY license and complete the attached form.” Please upload the completed Content Permission Form or other proof of granted permissions as an "Other" file with your submission. In the figure caption of the copyrighted figure, please include the following text: “Reprinted from [ref] under a CC BY license, with permission from [name of publisher], original copyright [original copyright year].” b. If you are unable to obtain permission from the original copyright holder to publish these figures under the CC BY 4.0 license or if the copyright holder’s requirements are incompatible with the CC BY 4.0 license, please either i) remove the figure or ii) supply a replacement figure that complies with the CC BY 4.0 license. Please check copyright information on all replacement figures and update the figure caption with source information. If applicable, please specify in the figure caption text when a figure is similar but not identical to the original image and is therefore for illustrative purposes only. The following resources for replacing copyrighted map figures may be helpful: USGS National Map Viewer (public domain): http://viewer.nationalmap.gov/viewer/ The Gateway to Astronaut Photography of Earth (public domain): http://eol.jsc.nasa.gov/sseop/clickmap/ Maps at the CIA (public domain): https://www.cia.gov/library/publications/the-world-factbook/index.html and https://www.cia.gov/library/publications/cia-maps-publications/index.html NASA Earth Observatory (public domain): http://earthobservatory.nasa.gov/ Landsat: http://landsat.visibleearth.nasa.gov/ USGS EROS (Earth Resources Observatory and Science (EROS) Center) (public domain): http://eros.usgs.gov/# Natural Earth (public domain): http://www.naturalearthdata.com/ [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly Reviewer #2: Partly Reviewer #3: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: No Reviewer #2: Yes Reviewer #3: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: No Reviewer #2: Yes Reviewer #3: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes Reviewer #3: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: As a whole this is a good and convincing paper, but I have some concerns – both minor and major. Journal policy concern: Authors state that there will be some restrictions on data sharing, without specifying what those are. They later state that all relevant data are in the manuscript. This seems internally contradictory. Minor concern: Line 69: Authors state that “This reagent was approved in October, 2021.” Who was the approving body? Major concern: Authors note that if discrepancies were found between GeneCube analysis and the RTPCR reference method, a third method was employed to resolve the discrepancy. This procedure, known as “discrepant analysis” is biased. [1–7]. Bias results from the fact that testing of non-discrepant samples (based on two assays) may, in principle, result in a discrepancy when the “resolver” test is applied; however, there is no chance to detect this discrepancy because the resolver is never applied to the non-discrepant tests. This may give the appearance that both of the initial test methods are better than they actually are. What the authors did was not as bad as it could have been since they apparently did not use the discrepant analysis to create or adjust sensitivity numbers. However, use of a composite reference standard for assessing the performance of methods does not suffer from this bias [8,9], and gives a more accurate portrayal of the relative sensitivity of both methods. While the impact of this inappropriate technique on the conclusions is negligible, its appearance in the literature promotes its further use. I think the paper would be significantly improved (and shortened somewhat) if none of the discrepant analysis is included. Minor concern: Sampling from a single nares is to be avoided, as there is sufficient data on respiratory infections to show that sampling from both nares improves sensitivity. I include one reference, but there are many others I haven’t taken the time to look up [10]. I think the authors should note in their discussion the fact that sampling both nares is preferable. Minor concern: The number of samples assessed in the LOD study is too small to draw strong conclusions about the LOD (see CLSI document EP7 and https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2556583/). Pooling of sample types (UTM, nasal swab, saliva) to achieve N is not appropriate, since these samples may have differing sources of assay interference. I think it is fair to say the authors have approximated the LOD, however. 1. Miller WC. Bias in discrepant analysis: When two wrongs don’t make a right. Journal of Clinical Epidemiology. 1998;51: 219–231. doi:10.1016/S0895-4356(97)00264-3 2. Hadgu A. Discrepant analysis: A biased and an unscientific method for estimating test sensitivity and specificity. Journal of Clinical Epidemiology. 1999;52: 1231–1237. doi:10.1016/S0895-4356(99)00101-8 3. Hadgu A. The discrepancy in discrepant analysis. Lancet. 1996;348: 592–593. doi:10.1016/S0140-6736(96)05122-7 4. McAdam AJ. Discrepant analysis and bias: A micro-comic strip. Journal of Clinical Microbiology. American Society for Microbiology; 2017. pp. 2878–2879. doi:10.1128/JCM.00969-17 5. Hadgu A, McAdam AJ. Discrepant analysis is an inappropriate and unscientific method (multiple letters) [4]. Journal of Clinical Microbiology. 2000. pp. 4301–4302. 6. Miller WC. Can we do better than discrepant analysis for new diagnostic test evaluation? Clinical infectious diseases : an official publication of the Infectious Diseases Society of America. 1998. pp. 1186–1193. doi:10.1086/514996 7. Green TA, Black CM, Johnson RE. Evaluation of bias in diagnostic-test sensitivity and specificity estimates computed by discrepant analysis. Journal of Clinical Microbiology. 1998;36: 375–381. doi:10.1128/jcm.36.2.375-381.1998 8. Baughman AL, Bisgard KM, Cortese MM, Thompson WW, Sanden GN, Strebel PM. Utility of composite reference standards and latent class analysis in evaluating the clinical accuracy of diagnostic tests for pertussis. Clinical and Vaccine Immunology. 2008;15: 106–114. doi:10.1128/CVI.00223-07 9. Tang S, Hemyari P, Canchola JA, Duncan J. Dual composite reference standards (dCRS) in molecular diagnostic research: A new approach to reduce bias in the presence of imperfect reference. Journal of Biopharmaceutical Statistics. 2018;28: 951–965. doi:10.1080/10543406.2018.1428613 10. van Wesenbeeck L, Meeuws H, D’Haese D, Ispas G, Houspie L, van Ranst M, et al. Sampling variability between two mid-turbinate swabs of the same patient has implications for influenza viral load monitoring. Virology Journal. 2014;11. doi:10.1186/s12985-014-0233-9 Reviewer #2: The manuscript submitted by Naito et al entitled “Evaluation of GENECUBE HQ SARS-CoV-2 for anterior nasal samples and saliva samples with a new rapid examination protocol” details a study that evaluates a molecular assay for detection of SARS-CoV-2. The assay was previously evaluated for NP specimens, but this assay evaluated the use of both NS, saliva, and saliva with a rapid processing protocol. In total 359 NS were evaluated and 240 saliva samples. Overall, the results demonstrated that the assay was accurate for all specimen types tested; However, there are some discrepancies in numbers that need to be clarified to ensure the proper comparisons were made. There are also a few additional clarifications needed prior to publication listed below: Major Comments For the enriched positive patients that had confirmed COVID-19 results, how was this detected. Did patients come back in after the results came through or was it a rapid test. Also, by cherry picking these patients you mess with the pre-test probability and would likely improve the assay performance, especially as the LoD appears to be better for GENECUBE vs RT-PCR assay. I would suggest breaking out the results as Total, prospective, call back subjects to be clear and see effects on true prospective testing. The definition of the rapid method is lacking in specifics. What steps were modified in the maglead extraction process that reduced the method by 20 minutes. As this is written it would not be possible to replicate the study or adopt the new method for a clinical lab to validate and use for faster TAT. Ln223-227: This is a bit unclear. There are 3 FP results based on rapid. 1 was the FP found in 4a, but the other 2 are new. Why were these 2 specimens tested 8 times on the lightmix test? One was picked up 50% of the time and the other was 12.5%. Ln245-250: Why is table 4a and 4b not using the NP PCR result as the reference method. This analysis was done is ST2, which should be the results used for the evaluation and not comparing saliva tested on the two platforms (Is the reference PCR validated for saliva?). If you use the presentation of data in table 4a and 4b this would suggest that saliva with GENECUBE is more sensitive, but if you compare it to the PCR from NP (gold standard) then it seems like it would be slightly less sensitive of a sample type. I would delete 4a and 4b and replace with data comparing to NP. This is the same for NS comparison data as ln 257 indicates that PPA was actually 88.1% compared to the 98.1% you present in table 3. Minor Comments Ln 73: Was the pos and negative concordance both 99.7% in the NP study as you only give 2 values for the overall, pos-, neg concordance. Were discordant specimens tested on the Xpert run as package insert (i.e. testing directly from VTM). Table 2: The current layout is a bit confusing at first with the spacing for “Standard method with magLead…” LoD study: The standard method is to perform 20 replicates at each concentration in pooled negatives and the LoD is then defined as the lowest concentration that was 19/20. What was the rationale for using multiple matrices and at the end having 24 tests for GENECUBE and 12 for the reference method. Ln197: Was the copies/mL determined based on the CT value and the LoD of the specimen? If so what was the CT value for the reference test and the Xpert assay. Why were the saliva samples frozen initially? Could this have skewed results as there is some data that a freeze thaw can help sensitivity of saliva samples. Reviewer #3: Thank you for inviting me to peer review this manuscript. The authors have studied the GENECUBE ® HQ SARS-CoV-2 using anterior nasal samples and saliva samples by using a new protocol. This study aims to evaluate the detection of SARS-CoV-2 with saliva samples for the first time and using a rapid protocol. Here are some comments which could be considers to improve the manuscript: 1. Line 69 “This reagent was approved in October, 2021.”. The date needs to be corrected. 2. Line 86, the situation of the infected cases is not explained. How did the authors select them and in which phase of disease they were? Did they have symptoms or they were asymptomatic cases? 3. It seems that the authors have collected the anterior nasal samples and saliva samples from different cases. It was better to take both sample types from each studied case to be able to compare them and investigate the accuracy of the samples to detect the infection. (It seems that such work is mentioned very briefly in the Discussion) 4. There is no explanation for positive and negative controls. 5. New Method and Standard Method are not explained in the Materials and Methods. In Line 133, “For the rapid protocol, in the preparation of saliva samples, purification and extraction processes were adjusted, and the total process time was shortened to approximately 10 minutes.” How this happened and what is the main difference making the novel method this short? 6. The obtained resulted are not discussed deeply. As an example, Line 181 “The detection rate was 100% down to 1000 copies/mL for pooled …” by considering these results, is this method useful for early detection or it is just useful for the chronic cases? How could a Dr. decide this strategy is a good choice for a specific case (like asymptomatic cases, accurate or chronic infections)? 7. Line 186. Compared with the standard test, how could the Rapid Strategy decrease the LoD of the nasopharyngeal samples from 1000 copies/mL to 500 copies/mL? (not discussed again). 8. Line 245 “In our current study, we used 60 saliva samples with positive nasopharyngeal sample results for SARS-CoV-2 (Supplementary Table 2), and the rapid protocol detected SARS-CoV-2 in most samples (98.3%; 59/60)”. It is the first time that the authors are refer to performing some tests on the nasopharyngeal and saliva samples taken from one case. It is better to be explained (materials and methods) and reported (results) in the previous sections before the Discussion. 9. The authors could also add graph(s) to report the results in a more useful way. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No Reviewer #3: Yes: Tina Shaffaf [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 1 |
|
PONE-D-21-27242R1The evaluation of the utility of the GENECUBE HQ SARS-CoV-2 for anterior nasal samples and saliva samples with a new rapid examination protocolPLOS ONE Dear Dr. Suzuki, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Jan 13 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, A. M. Abd El-Aty Academic Editor PLOS ONE Journal Requirements: Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed Reviewer #3: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: (No Response) Reviewer #2: Yes Reviewer #3: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: (No Response) Reviewer #2: Yes Reviewer #3: I Don't Know ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: (No Response) Reviewer #2: Yes Reviewer #3: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: (No Response) Reviewer #2: Yes Reviewer #3: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: (No Response) Reviewer #2: Thank you for the changes to the manuscript, which improves the clarity of study. One minor suggestion I would have is to add a limitation about the lack of low viral load samples tested in the evaluation. In ln 202 only 3 samples were CT >30. This is a small proportion of your positives and that could have improved concordance as too few were near the limit of detection. Reviewer #3: (No Response) ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No Reviewer #3: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
The evaluation of the utility of the GENECUBE HQ SARS-CoV-2 for anterior nasal samples and saliva samples with a new rapid examination protocol PONE-D-21-27242R2 Dear Dr. Suzuki, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, A. M. Abd El-Aty Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #2: (No Response) ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #2: No |
| Formally Accepted |
|
PONE-D-21-27242R2 The evaluation of the utility of the GENECUBE HQ SARS-CoV-2 for anterior nasal samples and saliva samples with a new rapid examination protocol Dear Dr. Suzuki: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Prof. A. M. Abd El-Aty Academic Editor PLOS ONE |
Open letter on the publication of peer review reports
PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.
We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.
Learn more at ASAPbio .