Peer Review History

Original SubmissionOctober 12, 2021
Decision Letter - Alexander V Panfilov, Editor

PONE-D-21-32419Open-Source Low-Cost Cardiac Optical Mapping SystemPLOS ONE

Dear Dr. Syunyaev,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Dec 24 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Alexander V Panfilov, PhD

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following in your Competing Interests section:  

"NO authors have competing interests."

Please complete your Competing Interests on the online submission form to state any Competing Interests. If you have no competing interests, please state "The authors have declared that no competing interests exist.", as detailed online in our guide for authors at http://journals.plos.org/plosone/s/submit-now 

 This information should be included in your cover letter; we will change the online submission form on your behalf.

3. We note that you have stated that you will provide repository information for your data at acceptance. Should your manuscript be accepted for publication, we will hold it until you provide the relevant accession numbers or DOIs necessary to access your data. If you wish to make changes to your Data Availability statement, please describe these changes in your cover letter and we will update your Data Availability statement to reflect the information you provide.

Additional Editor Comments:

Please respond in full to the comments of the referee and please pay special attention to the comments on clarification of the method, such as re separate figure with all the steps of signal conditioning, for both cameras, including raw signal from one pixel, signals after gaussian filtering and after time averaging. Please also comment if you consider a two-second time averaging as a significant limitation which does not allow mapping of cardiac arrhythmias or unstable spontaneous rhythm where cycle length changes in a beat-to-beat manner. If so, please discuss what should be done in such cases and how it can be addressed/improved.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The study by Rybashlykov et al. presents a novel, complete low-cost solution for an optical mapping system. The authors tested an iDS UI-3130CP-M-GL camera-based system and showed that for the most of electrophysiological parameters, including AP duration and restitution, CV and CV anisotropy, as well as pharmacological testing, the system provides reasonable outcomes that are comparable with a 100-times more expensive MiCAM ULTIMA-L system.

The study is timely and innovative; it presents an important step into the development of reliable low-cost optical mapping systems which are specifically important for pharmacological testing. The manuscript is well-written and the results are well-presented. I just have some minor concerns and suggestions on how to improve data presentation and discussion.

First, it would be highly beneficial for the readers if the authors show in a separate figure all the steps of signal conditioning, for both cameras, including raw signal from one pixel, signals after gaussian filtering and after time averaging.

Two-second time averaging sounds like a significant limitation which does not allow mapping of cardiac arrhythmias or unstable spontaneous rhythm where cycle length changes in a beat-to-beat manner.

Figure S2: It appears that there is a high variability of SNR between 7 mouse hearts shown. Two hearts have a very high SNR but in other three hearts SNR is significantly lower. I was wondering if SNR depends on the quality of staining? Can the authors show the SNR images overlapped with the corresponding fluorescence images (similar to those on Fig. 3A, B and Fig. 4A, B)?

Hearts shown in Figure 3B and Figure 4A seem to be out of focus.

Figure 6: APD difference of 9 ms between the systems sounds too high (about 17% of baseline APD). Was this difference statistically significant? Where were these APDs calculated? Were they averaged throughout the entire mapped area of both ventricles? Would it be more accurate to measure APD separately in the right and the left ventricles? This may decrease the observed difference in APD obtained by different systems.

I assume that APD heterogeneity (dispersion) within the area of interest was relatively high that may also affect the average APD values shown and thus contribute to the observed difference. It would be beneficial to show APD heterogeneity measured within the area of interest for each system.

Figure 7: For both systems, APD looks very heterogeneous within the ventricles. Moreover, there are some regions with outliers: APD of 10 ms (dark blue areas) and more than 100 ms (dark red areas) seems to be not physiological. It would be helpful if the authors exclude these outliers and re-analyzed their data for Figure 5 and 6. It may significantly improve the results.

Finally, the authors may consider including a special section in the discussion that would summarize major pros and contras for the presented system. This section should include the mention that this system might be not applicable for specific studies focused on cardiac arrhythmias (see above).

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

We thank the editor and the reviewer for their questions. We hope that we were able to address the concerns in the new revision. In particular, we have revised the text and the figures and did and additional experiment in order to improve Figures 4,8 quality. We have also added a separate figure with all the steps of signal conditioning for both cameras. The limitations of a two-second time averaging is also discussed in the new version of the manuscript.

The text below is also attached to the submission as a separate file.

"We thank the reviewer for the positive response to our manuscript. We have carefully considered the suggestions, ran an additional experiment and revised the text and the figures. For the convenience, the reviewer’s questions are colored with cyan font below.

Reviewer #1: The study by Rybashlykov et al. presents a novel, complete low-cost solution for an optical mapping system. The authors tested an iDS UI-3130CP-M-GL camera-based system and showed that for the most of electrophysiological parameters, including AP duration and restitution, CV and CV anisotropy, as well as pharmacological testing, the system provides reasonable outcomes that are comparable with a 100-times more expensive MiCAM ULTIMA-L system.

The study is timely and innovative; it presents an important step into the development of reliable low-cost optical mapping systems which are specifically important for pharmacological testing. The manuscript is well-written and the results are well-presented. I just have some minor concerns and suggestions on how to improve data presentation and discussion.

First, it would be highly beneficial for the readers if the authors show in a separate figure all the steps of signal conditioning, for both cameras, including raw signal from one pixel, signals after gaussian filtering and after time averaging.

We have added the S2 figure showing the signal during all steps of signal processing. We thank the reviewer for the suggestion and believe that new figure makes the signal quality of the camera much more clear for a potential reader.

Please see lines 134-140:

“The signals after all intermediate steps of signal conditioning for both cameras are shown in the S2 Figure. Raw iDS-recorded signal was, in general, noisier as compared to signals from MiCAM camera. While spatial filtering reduced the SNR to a level adequate to measure upstroke for activation sequence and conduction velocity (CV) analysis, both spatial and temporal averaging were required to measure APD. Therefore, only the former step of signal conditioning was used for the activation sequence and conduction velocity (CV) measurements.“

Two-second time averaging sounds like a significant limitation which does not allow mapping of cardiac arrhythmias or unstable spontaneous rhythm where cycle length changes in a beat-to-beat manner.

Indeed, this is a significant limitation. It should be noted though that in our publication we didn’t use the time averaging to measure the activation sequences and the conduction velocities. The reason behind that is that the upstroke of a signal is more robust and easier to measure then slower repolarization. Nevertheless the system might not allow the user to accurately measure the activation sequences in case of arrhythmia, because in general the signal tend to be more noisy in this case. We have added the text discussing both the limitations of the optical mapping system and differences between the measurements of APD and activation sequence.

Lines 244-246:

“RMSD between corresponding recordings by two cameras was equal to 4 cm/s even though ensemble averaging over 2 second recording was not used for these measurements ”

Please also see new lines 134-140 quoted above.

Lines 299-315 discussing the limitations are also added and quoted at the very end of this Response document.

Figure S2: It appears that there is a high variability of SNR between 7 mouse hearts shown. Two hearts have a very high SNR but in other three hearts SNR is significantly lower. I was wondering if SNR depends on the quality of staining? Can the authors show the SNR images overlapped with the corresponding fluorescence images (similar to those on Fig. 3A, B and Fig. 4A, B)?

We have added the amplitude of the fluorescence to the figure S2 (our original submitted version) but now the figure S3 (the new version of the manuscript). We believe that although there is some correlation between SNR and signal intensity, some other factors are involved, namely the uneven illumination of heart surface by a LED and overall viability of the tissue.

Hearts shown in Figure 3B and Figure 4A seem to be out of focus.

Thank you for pointing out this issue, we have changed the figure 3B (mouse heart) and redone the experiment with the rat heart (Fig. 4, and Fig. 8B,D). We believe that the quality of both the heart images and the recordings themselves have improved.

Figure 6: APD difference of 9 ms between the systems sounds too high (about 17% of baseline APD). Was this difference statistically significant? Where were these APDs calculated? Were they averaged throughout the entire mapped area of both ventricles? Would it be more accurate to measure APD separately in the right and the left ventricles? This may decrease the observed difference in APD obtained by different systems.

I assume that APD heterogeneity (dispersion) within the area of interest was relatively high that may also affect the average APD values shown and thus contribute to the observed difference. It would be beneficial to show APD heterogeneity measured within the area of interest for each system.

Figure 7: For both systems, APD looks very heterogeneous within the ventricles. Moreover, there are some regions with outliers: APD of 10 ms (dark blue areas) and more than 100 ms (dark red areas) seems to be not physiological. It would be helpful if the authors exclude these outliers and re-analyzed their data for Figure 5 and 6. It may significantly improve the results.

Indeed, the difference between the APDs as measured by two cameras is the issue. In a previous version of the manuscript the APD was averaged over the whole mapped area of both ventricles, excluding the regions with high SNR and over-saturating areas. The difference was not statistically significant, but p-value was relatively low (p=0.1). We have redone the analysis separating the ventricles and excluding the APD outliers. While the results improved, but the mean difference is still relatively high: 7+/- 12 ms for LV (p=0.30, paired t-test), 5+/-9 ms for RV (p=0.31). We believe that there are two reasons behind these differences: 1) some changes due to imperfect perfusion happened to the hearts between the measurements by two optical systems and 2) field of view was not exactly the same.

We thank the reviewer for the suggestion and have modified Figs 5 and 6. We have also outlined the region included in statistical analysis in the new version of Fig. 7 to make it clear for a reader. The numbers throughout the text were changed in accordance with new analysis.

Please also see new lines 149-152:

“APD outlier boundaries were calculated as Q1 - 1.5*IQR and Q3 + 1.5*IQR, where Q1 and Q3 denote first and third quartile respectively and IQR = Q3 - Q1 (interquartile range). Any value exceeding those boundaries was considered an outlier and was excluded from statistical analysis.”

Lines 177-178:

“APD80 measurements in left ventricle by two cameras differed by 7 ± 12 ms in control and 5 ± 5 ms in 4-AP; in right ventricle: 5 ± 9 ms in control, 6 ± 5 ms in 4-AP.”

Lines 256-258:

“This is also consistent with the fact that unpaired comparison between two cameras resulted in high p-values (unpaired t-test, p=0.42 in control, p=0.52 in 4-AP, compare to paired t-test in Fig. 5).“

Finally, the authors may consider including a special section in the discussion that would summarize major pros and contras for the presented system. This section should include the mention that this system might be not applicable for specific studies focused on cardiac arrhythmias (see above).

Thank you we have added the section. Please, see lines 299-315:

“We summarize the advantages and limitations of the proposed optical system as follows. (1) The price is more than 20 times cheaper than a specialized camera, making the system suitable for educational purposes or for the applications requiring the use of several cameras such as panoramic and multi-parametric mapping. (2) We have shown that despite the relatively low SNR, AP waveform, activation sequence and CV could be accurately measured after the signal processing via compatible open-source Rhythm software. It should be noted, that while only spatial binning was used for the activation sequence and CV, precise measurement of the AP waveform required temporal averaging over 2 second recordings. This requires that the AP waveform is limited to regular periodic heart rhythm while, for example, the recording of alternans is not always feasible. Moreover, arrythmia usually results in complex propagation patterns reducing overall SNR of the recording, which is likely to obstruct activation sequence recording as well. (3) As an example of practical application we have demonstrated in this study that drug effects on AP waveform can be measured by the optical-mapping system. However, as exemplified by atrial recordings in Fig. 4 C, thin myocardium tissue reduces the quality of the signal, which might hinder the use of the cheaper system in experiments involving the mapping of the monolayer cell cultures.“"

Attachments
Attachment
Submitted filename: Response.docx
Decision Letter - Alexander V Panfilov, Editor

Open-Source Low-Cost Cardiac Optical Mapping System

PONE-D-21-32419R1

Dear Dr. Syunyaev,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Alexander V Panfilov, PhD

Section Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The authors successfully addressed all my questions and concerns. I do not have additional comments.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Formally Accepted
Acceptance Letter - Alexander V Panfilov, Editor

PONE-D-21-32419R1

Open-source low-cost cardiac optical mapping system

Dear Dr. Syunyaev:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Prof. Alexander V Panfilov

Section Editor

PLOS ONE