Peer Review History

Original SubmissionJune 23, 2021
Decision Letter - Etsuro Ito, Editor

PONE-D-21-20548

Evaluation of the clinical performance of a magnetic force-assisted electrochemical immunoassay for the detection of SARS-CoV-2 antigens

PLOS ONE

Dear Dr. Lee,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please describe the limitation of your present study, which was commented by one of the reviewers.

Please submit your revised manuscript by Sep 17 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Etsuro Ito

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please upload a new copy of Figure 1 as the detail is not clear. Please follow the link for more information: " ext-link-type="uri" xlink:type="simple">https://blogs.plos.org/plos/2019/06/looking-good-tips-for-creating-your-plos-figures-graphics/" https://blogs.plos.org/plos/2019/06/looking-good-tips-for-creating-your-plos-figures-graphics/

3. Please expand the acronym “BBB” (as indicated in your financial disclosure) so that it states the name of your funders in full.

This information should be included in your cover letter; we will change the online submission form on your behalf.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

Reviewer #3: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The Authors evaluated the MARK-B COVID-19 Ag test (MARK-B, BBB, Sungnam, Republic of Korea) that is a portable point-of-care device based on the use of a magnetic force-assisted electrochemical immunoassay (MESIA) designed to detect SARS-CoV-2 antigens. Notably, this newly developed immunoassay, MESIA, reportedly suitable for measuring low concentrations of proteins for cancer detection and with the potential to detect any other biomarkers.

The evaluation was done against RT PCR as the gold standard for SARS COV-2 detection and another immunochromatographic manual antigen assay for SARS CoV-2, showing a significant higher correlation with RT PCR and a significant higher sensitivity over the standard antigen assay. Overall, this data confirms that MESIA may provide enhanced sensitivity for point-of-care devices that involve the detection of SARS-CoV-2.

The study is well conducted, each section is well written and developed, statistical measures correctly applied. There is one general limitation in this study:

The MARK-B test has not been CE/FDA-cleared or approved so far for SARS CoV-2, but these validations are pending. Therefore, this is a clear limitation of the study conducted by the Authors, in terms of extension and commercialization of this device. The Authors should discuss this issue and clarify if there are any on-going news on this important issue which is essential for the scientific community. Moreover, the test is not cited by FIND as the date of today July 15 among the 653 antigen immunoassays for SARS CoV-2 enlisted in the web site https://www.finddx.org/covid-19/pipeline/?section=immunoassays#diag_tab

Is this just for Asia? The Authors must clarify it.

Minor issues

Material and Methods:

The semi-quantitative/quantitation system used by MESIA for their electric signal should be specified by the Authors as they use it in the Figures, as well as the nature of the cut-off value and the type of antigen detected by this system.

Reviewer #2: This paper describes the evaluation of a magnetic force-assisted electrochemical immunoassay -based test compared to other commercial rapid Ag tests by application to clinical samples. Authors have successfully compared the results provided by this device with those from other methodologies applying them to a large number of samples. In my opinion, the information offered in the article is very interesting, the electrochemical sensor has demonstrated a good behavior, but the experimental work and the new data in the article are not relevant from the point of view of scientific research. I consider that the paper should not be published in its current state in PLOs One.

Reviewer #3: This is a clear well-written paper describing the performance of a new Ag test for diagnosis of COVID-19 infection. As an independent of assessment of the test, the publication provides important data to users who may be considering use of the test. I recommend publication after revision to address some minor weaknesses in the paper:

1. The main weakness is associated with the comparison to an existing commercial Ag test and the interpretation of the comparison

a. In the abstract and main text their are locations where the authors describe their paper as comparing their test to commercial antigen tests. These statements should be clarified to indicate that they compared to a (as in one) antigen test, and isn't necessarily reflective of how the test would perform to the range of available tests.

b. Some discussion is warranted regarding the criteria for the selection of the comparator test, and the relative performance of this test. Relative to the many Ag tests that are now available commercially, does the published data for the SDQ test place it at the top, middle or bottom of the range for clinical sensitivity and accuracy?

c. It would be helpful to provide some testing with a reference sample (for example, one of the available viral culture reference samples or a sample of recombinant nucleocapsid protein) that could allow direct comparison of the analytical performance of the antigen assays to each other (and potentially other commercial tests) and determination of whether the clinical performance is a direct reflection of the analytical performance).

2. The volume of UTM used to extract the sample should be provided and, if applicable, a discussion of how the volume of UTM may differ from the volume of extraction buffer used in the standard swab protocols for the MESIA and SDQ tests, and how that may affect performance.

3. Ct should be proportional to Log2(RNA concentration) since it represents the number of RNA doubling steps required to achieve a detectable PCR signal. It is therefore probably most appropriate to plot Log2(MESIA current) vs. Ct, which should provide a slope approaching 1. If the log(current) vs. Ct is not linear, that should be discussed.

4. In the methods section, it isn't very clear how the MESIA technique works. I think readers would provide a little more detail or, alternatively, a reference to where they could go for more details.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Valeria Ghisetti, MD, Director of the laboratory of Microbiology and Virology, Amedeo di Savoia Hospital, Turin, Italy

Reviewer #2: No

Reviewer #3: No

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

.

Attachments
Attachment
Submitted filename: Response_to_Reviewers.docx
Decision Letter - Etsuro Ito, Editor

Evaluation of the clinical performance of a magnetic force-assisted electrochemical immunoassay for the detection of SARS-CoV-2 antigens

PONE-D-21-20548R1

Dear Dr. Lee,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Etsuro Ito

Academic Editor

PLOS ONE

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #3: Some minor comments/suggestions based on the response to my original review:

1. The description of the assay technology in your response to the reviewers comments is much clearer than in the edits to the manuscript. I suggest just using the text from the response in the manuscript.

2. The line in the log2 antigen vs. Ct count plot should be calculated using the log2 transformed values and should look like a straight line in the plot. Interestingly, the slope of that line will be about -0.3 indicating that an 8-fold change in nucleic acid levels (3 Ct counts) corresponds to roughly a 2-fold change in the antigen assay signal (roughly a cube root dependence). The table of assay signals for the virus titration also shows roughly a cube root dependence.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #3: No

Formally Accepted
Acceptance Letter - Etsuro Ito, Editor

PONE-D-21-20548R1

Evaluation of the clinical performance of a magnetic force-assisted electrochemical immunoassay for the detection of SARS-CoV-2 antigens

Dear Dr. Lee:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Prof. Etsuro Ito

Academic Editor

PLOS ONE