Peer Review History

Original SubmissionMay 10, 2021
Decision Letter - Peter Dziegielewski, Editor

PONE-D-21-14821

An automatic water-occluding device to enable laryngectomee participation in water activities

PLOS ONE

Dear Dr. Erath,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

ACADEMIC EDITOR:  I agree with the reviewers. Please address their concerns and revise the paper to be a proof of concept with a warning that it does not prove safety. A warning is needed that the device is not FDA (or similar) approved and further safety data is needed before it can be recommended for patients.

Please submit your revised manuscript by Sep 05 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Peter Dziegielewski, MD, FRCSC

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please provide additional details regarding participant consent. In the ethics statement in the Methods and online submission information, please ensure that you have specified whether consent was informed.

3. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match. 

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

4. Please include your full ethics statement in the ‘Methods’ section of your manuscript file. In your statement, please include the full name of the IRB or ethics committee who approved or waived your study, as well as whether or not you obtained informed written or verbal consent. If consent was waived for your study, please include this information in your statement as well. 

5. We note that Figures 1, 3, 4 and 5 in your submission contain copyrighted images. All PLOS content is published under the Creative Commons Attribution License (CC BY 4.0), which means that the manuscript, images, and Supporting Information files will be freely available online, and any third party is permitted to access, download, copy, distribute, and use these materials in any way, even commercially, with proper attribution. For more information, see our copyright guidelines: http://journals.plos.org/plosone/s/licenses-and-copyright.

We require you to either (1) present written permission from the copyright holder to publish these figures specifically under the CC BY 4.0 license, or (2) remove the figures from your submission:

a. You may seek permission from the original copyright holder of igures 1, 3, 4 and 5 to publish the content specifically under the CC BY 4.0 license. 

We recommend that you contact the original copyright holder with the Content Permission Form (http://journals.plos.org/plosone/s/file?id=7c09/content-permission-form.pdf) and the following text:

“I request permission for the open-access journal PLOS ONE to publish XXX under the Creative Commons Attribution License (CCAL) CC BY 4.0 (http://creativecommons.org/licenses/by/4.0/). Please be aware that this license allows unrestricted use and distribution, even commercially, by third parties. Please reply and provide explicit written permission to publish XXX under a CC BY license and complete the attached form.”

Please upload the completed Content Permission Form or other proof of granted permissions as an "Other" file with your submission. 

In the figure caption of the copyrighted figure, please include the following text: “Reprinted from [ref] under a CC BY license, with permission from [name of publisher], original copyright [original copyright year].”

b. If you are unable to obtain permission from the original copyright holder to publish these figures under the CC BY 4.0 license or if the copyright holder’s requirements are incompatible with the CC BY 4.0 license, please either i) remove the figure or ii) supply a replacement figure that complies with the CC BY 4.0 license. Please check copyright information on all replacement figures and update the figure caption with source information. If applicable, please specify in the figure caption text when a figure is similar but not identical to the original image and is therefore for illustrative purposes only.

6. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information. 

Additional Editor Comments:

Thank you for your submission. This is an interesting paper about a potentially helpful device in laryngectomee patients. However, I agree with the reviewers that this study doe not prove safety. Publishing it as such may put patient's in harm's way and could be dangerous. This needs to be revised to be a proof of concept paper and there needs to be a warning that this study does not prove safety or recommend the use of the device until further investigation is carried out and FDA approval (or similar) is passed.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

Reviewer #2: Yes

Reviewer #3: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: N/A

Reviewer #2: N/A

Reviewer #3: I Don't Know

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: In their manuscript, the authors describe their novel device designed to allow laryngectomy patients the ability to engage in water activities.

There are some unneccessary commercial items listed around line 47.

the largest issue is the ethics of this study. the authors do not differentiate the risk of the various water sports (ie fishing from a dock vs paddleboarding or kayaking).

the itegrity of the "waterproof" adhesive is not addressed.

this may be better suited as a proof of concept manuscript with explicit discussion that is not meant for actual patient use at this time

Reviewer #2: A well written manuscript about a new device for larygectomees' patients. As device was only tested in normal people, it remains untested in real patients and there may be unforeseen issues. I suggest to add these into the discussion. What are the unforeseen circumstances that may arise and what measures can be instituted to overcome them?

Reviewer #3: The aim of this paper was to establish proof of concept for a tracheostoma occluding device for use in the event of unanticipated immersion in water. With such a safety device, individuals who have undergone total laryngectomy may be more willing to participate in various recreational water activities, thus improving their quality of life, which is significantly compromised after such surgery. For those who already participate in water activities, this type of device may increase survival rates of those who experience accidental submersion. The design was based on survey responses of individuals who have undergone total laryngectomy from a local support group. Based on that feedback, design objectives included automatic water occlusion, comfortable respiration, compatibility with alaryngeal speech including use of a tracheoesophageal voice prosthesis, and comfort for long wear time.

In reference to Line [8] of the Introduction, bilateral vocal fold paralysis and intractable aspiration are not causes of laryngeal dysfunction, but results. Cancer, trauma, or neurological insult are common etiologies of laryngeal impairment. In line [15], the authors go on to state that the primary function of the larynx is speech production. Many would argue that the primary function is airway protection, which would support the premise for the device.

The design and thus the name of the device is a bit confusing. SToma-snORKEL (STORKEL) implies a breathing device, not an arresting device. When the airway is submerged in water, the airway is closed on multiple levels, e.g., lips close, vocal folds adduct, diaphragm and lungs stop expanding, until the airway (typically the nose or mouth) has surfaced. For laryngectomees, the problem is twofold. First, the airway of a laryngectomee is redirected to the tracheostoma, and the vocal folds are no longer present to provide protection. Second, the stoma is positioned distally in the neck, which will often remain submerged when an individual is in deep water, unless they can float on their back and allow the stoma to rise above the surface.

Based on that premise, one can see why a "snorkel" type device would be advantageous, to enable respiration, should the stoma remain submerged for an extended period of time. This concept could be highlighted, especially in the event that a life preserver may not provide enough buoyancy to keep the stoma above the waterline.

While the occluding feature of the float valve was intensively assessed, the occlusion at the level of the stoma was not evaluated. If the main purpose of the device is to protect the stoma, then this should be assessed. Individuals who have undergone total laryngectomy often experience suboptimal baseplate seals due to sensitive skin and stoma landscape. Individuals who use a TEP experience additional challenges of excess pressure exerted on the baseplate when they occlude their stomas to generate TEP speech, thus often "break the seal." A broken seal while using this device would be life threatening. This could be easily tested on several laryngectomees (with IRB approval) by just having them speak and cough with the device on, and assessing if the seal stays intact.

The additional limitations and additional questions for future testing and design modification are suggested. In addition, did the authors present their design to the original laryngectomees who completed the design survey?

It is commendable that the needs of individuals who have undergone total laryngectomy have caught the attention of mechanical engineers. Collaborating with clinicians in the field of head and neck surgery, and speech-language pathology may all the more expedite the effort.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: Yes: Professor Dr Baharudin Abdullah

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Please see attached files.

Attachments
Attachment
Submitted filename: Response_to_Reviewers_FINALdocx.pdf
Decision Letter - Peter Dziegielewski, Editor

An automatic water-occluding device to enable laryngectomee participation in water activities

PONE-D-21-14821R1

Dear Dr. Erath,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Peter Dziegielewski, MD, FRCSC

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: N/A

Reviewer #3: I Don't Know

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #3: The authors have adequately addressed my comments raised in a previous round of review. Specifically, that this is a proof of concept study, testing was not completed on individuals who have undergone laryngectomy, and the initial source of drowning is submersion of the stoma.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #3: No

Formally Accepted
Acceptance Letter - Peter Dziegielewski, Editor

PONE-D-21-14821R1

An automatic water-occluding device to enable laryngectomee participation in water activities 

Dear Dr. Erath:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Peter Dziegielewski

Academic Editor

PLOS ONE