Peer Review History

Original SubmissionApril 29, 2020
Decision Letter - Peter Schwenkreis, Editor

PONE-D-20-12562

Usability, occupational performance and satisfaction evaluation of a smart environment controlled by infrared oculography by people with severe motor disabilities

PLOS ONE

Dear Dr. Sime,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

For acceptance, it is crucial that you fully consider the comments and criticism of the reviewer.

Please submit your revised manuscript by Apr 17 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Peter Schwenkreis

Academic Editor

PLOS ONE

Additional Editor Comments :

-the paper needs to be substantially shortened and condensed

-details and photos that might allow to identify the participants should be omitted

-"discussion" mus be separated from "results"

-Please explain in more detail why "data cannot be shared publicly because of Ethics Committee terms"

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that your paper includes detailed descriptions of individual patients/participants. As per the PLOS ONE policy (http://journals.plos.org/plosone/s/submission-guidelines#loc-human-subjects-research) on papers that include identifying, or potentially identifying, information, the individual(s) or parent(s)/guardian(s) must be informed of the terms of the PLOS open-access (CC-BY) license and provide specific permission for publication of these details under the terms of this license. Please download the Consent Form for Publication in a PLOS Journal (http://journals.plos.org/plosone/s/file?id=8ce6/plos-consent-form-english.pdf). The signed consent form should not be submitted with the manuscript, but should be securely filed in the individual's case notes. Please amend the methods section and ethics statement of the manuscript to explicitly state that the patient/participant has provided consent for publication: “The individual in this manuscript has given written informed consent (as outlined in PLOS consent form) to publish these case details”.

3. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

4. Thank you for stating the following in the Financial Disclosure section:

"AB and TBF received a award in Google Research Awards for Latin America, of Google Inc.

The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

We note that you received funding from a commercial source: Google Inc.

Please provide an amended Competing Interests Statement that explicitly states this commercial funder, along with any other relevant declarations relating to employment, consultancy, patents, products in development, marketed products, etc.

Within this Competing Interests Statement, please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include your amended Competing Interests Statement within your cover letter. We will change the online submission form on your behalf.

Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests

5. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This is an interesting article, however it needs major modification prior to being eligible for publication. Each section needs to be condensed. While there is no word limit, PLOS ONE asks that papers be presented concisely - and this paper does not meet that criteria. For example, there is no need for a separate introduction and literature review - these should be combined and condensed. Similarly the description of the materials used could be reduced. The reference to "volunteers" should be changed to "participants" throughout the manuscript. I am also uncomfortable with the inclusion of the case study vignettes for each participant, I believe these have the potential to identify participants, especially in conjunction with their demographic data. There is no need for this level of detail. Also, the results and discussion sections need to be separated out. It is very unusual to see them presented together and does not meet the manuscript organisation guidelines specified for PLOS One (available here: https://journals.plos.org/plosone/s/submission-guidelines#loc-length). I am also not comfortable of the inclusion of photos of participants in their home environments. These are extremely identifying, even though the faces have been obscured.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachments
Attachment
Submitted filename: PONE-D-20-12562_reviewer.pdf
Revision 1

Response to Reviewers

Title: Usability, occupational performance and satisfaction evaluation of a smart environment controlled by infrared oculography by people with severe motor disabilities

In response to the reported pending issues, the following information is required:

- Point 1: The paper needs to be substantially shortened and condensed.

Answer to point 1: As recommended by the reviewer, the introduction and review have been combined and condensed. The description of the materials and instruments has been reduced. The results and discussion were separated and reorganized.

- Point 2: Details and photos that might allow to identify the participants should be omitted.

Answer to point 2: The vignettes with the details and photos of the participants have been removed and only the sociodemographic information that contributed to the discussion was maintained.

- Point 3: About the funding received of the Google Inc.

Answer to point 3: Google Inc. only supported the research through Google’s Latin America Research Awards to two authors, Teodiano Freire Bastos-Filho and Alexandre Luís Cardoso Bissoli, and this does not alter our adherence to Plos ONE policies on sharing data and materials. This statement can also be found in the cover letter, in the Competing Interests Statement section, as recommended

- Point 4: Please explain in more detail why "data cannot be shared publicly because of Ethics Committee terms".

Answer to point 4: Our data contained information that identified the participants. This information has been omitted and the data are being shared in the Supporting Information files.

- Point 5: Please amend the methods section and ethics statement of the manuscript to explicitly state that the patient/participant has provided consent for publication: “The individual in this manuscript has given written

informed consent (as outlined in PLOS consent form) to publish these case details”

Ansewer to point 5: About the use of personal data, participants or their guardians signed the Free and Informed Consent Form (FICF), which guarantees that all personal data are confidential and private, even after the publication of the research. The data protection has been guaranteed by identifying participants by codes, instead of their names or initials, and by not including information such as date of birth and address. Detailed information about the participants health condition had already been excluded from the manuscript after the original decision letter. The terms with the highlighted subsection in both the original version in Portuguese and the translated version in English were included in the list of documents for submission. The current situation in Brazil, due to the COVID-19 pandemic is very serious, and it has been very difficult to meet the participants to obtain their signature on the PLOS specific consent form. Since they are people with disabilities and comorbidities, they have been in quarantine, as a measure to prevent spread of the Sars-CoV-2 virus. The methods section and ethics statement of the manuscript were amended to explicitly state that the participants has provided consent for publication, through the FICF.

Sincerely,

Mariana Midori Sime (e-mail: mariana.sime@ufes.br)

Vitória-ES (Brazil), May 01, 2021.

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Peter Schwenkreis, Editor

PONE-D-20-12562R1

Usability, occupational performance and satisfaction evaluation of a smart environment controlled by infrared oculography by people with severe motor disabilities

PLOS ONE

Dear Dr. Sime,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

In the revised version, you should further shorten the "conclusions" section. Besides, regarding the small sample size of the statistical analysis (n=4), it is necessary to add a paragrah where you discuss the potential limitations of the study with respect to generalizability and reproducibility.

Please submit your revised manuscript by Jul 29 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Peter Schwenkreis

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Thank you for addressing the comments provided to you through the first review. I believe this paper is worth publishing. I feel some minor amendments are now required. I have noted these in the tracked changes version of your manuscript. I also want to confirm how the FIM was administered, as it appears this was via self report from the way this section is written (the FIM must be administered by a credentialed assessor). I also feel the conclusion could be condensed further, and there could be further consideration of the limitations of this paper (case studies, difficult to generalise etc.)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachments
Attachment
Submitted filename: PONE-D-20-12562_R1.pdf
Revision 2

In response to the reported pending issues, the following information is required:

- Point 1: The conclusion could be condensed further.

Answer to point 1: The Conclusions section has been shortened, as solicited.

- Point 2: It is necessary to add a paragraph where the potential limitations of the study are discussed.

Answer to point 2: A paragraph describing the limitations of the study resulting from its small sample size has been included in the Conclusions section.

- Point 3: Explain how the FIMTM was administered.

Answer to point 3: Regarding the questioning of the reviewer about the application of the FIMTM, it has been included in the text that the instrument was applied by a certified researcher through interviews, as validated in Brazil. Validation of the FIMTM in Brazil was performed by Riberto et al. (2004), which is reference no. 63 in the paper.

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Peter Schwenkreis, Editor

Usability, occupational performance and satisfaction evaluation of a smart environment controlled by infrared oculography by people with severe motor disabilities

PONE-D-20-12562R2

Dear Dr. Sime,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Peter Schwenkreis

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Formally Accepted
Acceptance Letter - Peter Schwenkreis, Editor

PONE-D-20-12562R2

Usability, occupational performance and satisfaction evaluation of a smart environment controlled by infrared oculography by people with severe motor disabilities

Dear Dr. Sime:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Peter Schwenkreis

Academic Editor

PLOS ONE