Morphological changes in eyes of amblyopic patients through quantification of mean reflectivity.

PLOS ONE

Dear Dr. Estephan,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

A learned reviewer has offered several constructive criticism. Manuscript need to be revised appropriately addressing these comments. 

Please submit your revised manuscript by Jun 24 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Sanjoy Bhattacharya

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. Please see http://www.bmj.com/content/340/bmj.c181.long for guidelines on how to de-identify and prepare clinical data for publication. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: In general:

The topic of the article is about the differences of some OCT parameters between amblyopic eyes and healthy eyes which might be of interest. However, as they applied a non-standardised, not widely used reflectivity estimation method on exported OCT images, the article should include detailed description of the reflectivity measurements and calculations and the accuracy and reproducibility of the method used.

Detailed comments and questions:

The title is not clear enough as it is not containing the place of changes in eyes or the reflectivity of which structure.

Abstract

The authors showed significant differences between healthy and amblyopic eyes, however the screening abilities of the detected changes are not examined, so conclusion about it is not supported by the results.

Methods

The definition and measurement/determination method(s) of reflectivity is not described in details in the Methods. However, this would be crucial as the reflectivity is the main outcome measure in this article. Please, provide also the reproducibility and accuracy values for the reflectivity measurements used in this study.

Results

Table 1. There were possibly significant demographic differences between patients and healthy controls (like: maximum age, sex distribution). Please, include statistical comparisons between patients and controls for the parameters and address these differences in the discussion and disclose that these differences do not cause bias.

The interpretation of the results need to be very restrained as the differences were found significant at the level of p<0.05 and only few of them with p<0.01 and none of them at p<0.001.

Discussion

In line 198: “an U/S study” – please, introduce the abbreviation.

The statement in the sentence in lines 205-207 would need supporting literature reference or justification.

There are no figure legends in the manuscript.

For approving a test for screening needs validation, first. Have to know what is the sensitivity, specificity, etc. of the new diagnostic method. And not only a comparison with healthy eyes but comparisons with other eye diseases are necessary to judge before the statement about the usefulness of the new diagnostic modality.

We know that both retinal thickness and choroidal thickness depend on the axial eye length. The longer eyeballs have thinner layers and possibly with different reflectivity. What was the range of the axial eye length of the patients and controls? Were there differences in axial eye length between AE, FE and HE? Need to show if there is a correlation between axial eye length and the reflectivity of the layers.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Reviewer #1: In general:

The topic of the article is about the differences of some OCT parameters between amblyopic eyes and healthy eyes which might be of interest. However, as they applied a non-standardised, not widely used reflectivity estimation method on exported OCT images, the article should include detailed description of the reflectivity measurements and calculations and the accuracy and reproducibility of the method used.

We calculated two ICCs (intraclass correlation coefficients) the first was an intra-rater reliability measure wherein one rater repeated the measurements four times over twelve regions of interest from a randomly chosen participant, the second was an inter-rater reliability measure wherein four independent raters measured twelve regions of interest from a randomly chosen participant. We added this relevant information to the abstract, materials and methods, the results and discussion sections (lines 20-21, 28, 130-141, 158-170, 237-244). All added parts including tables are highlighted in yellow in the revised manuscript.

Detailed comments and questions:

The title is not clear enough as it is not containing the place of changes in eyes or the reflectivity of which structure.

Addressed

Abstract

The authors showed significant differences between healthy and amblyopic eyes, however the screening abilities of the detected changes are not examined, so conclusion about it is not supported by the results.

Addressed. Mention of its screening and diagnostic potential was removed from the abstract.

Methods

The definition and measurement/determination method(s) of reflectivity is not described in details in the Methods. However, this would be crucial as the reflectivity is the main outcome measure in this article. Please, provide also the reproducibility and accuracy values for the reflectivity measurements used in this study.

Addressed. More details were added as well as images in the supporting information section. All added parts are highlighted in yellow in the revised manuscript

Results

Table 1. There were possibly significant demographic differences between patients and healthy controls (like: maximum age, sex distribution). Please, include statistical comparisons between patients and controls for the parameters and address these differences in the discussion and disclose that these differences do not cause bias.

The interpretation of the results need to be very restrained as the differences were found significant at the level of p<0.05 and only few of them with p<0.01 and none of them at p<0.001.

Addressed. We are in total agreement, and we clarified in the discussion section that we were not able to adjust for these possible confounding factors due to the small sample size. Furthermore, the p-value is affected by the sample size: the larger the sample size, the smaller the p-value will be. We sought to address this by reporting the effect sizes of significant findings.(lines 149, 187, 281, 284, 305, 352) Unlike significance tests, effect size is independent of sample size.[1, 2]

Discussion

In line 198: “an U/S study” – please, introduce the abbreviation.

Addressed. Changed to ultrasound.

The statement in the sentence in lines 205-207 would need supporting literature reference or justification.

Addressed. Citation added.

There are no figure legends in the manuscript.

Addressed.

For approving a test for screening needs validation, first. Have to know what is the sensitivity, specificity, etc. of the new diagnostic method. And not only a comparison with healthy eyes but comparisons with other eye diseases are necessary to judge before the statement about the usefulness of the new diagnostic modality.

We mentioned that this tool might have potential screening and/or diagnostic value. To further clarify this claim, we qualified it with the statement that the tool needs another study dedicated for calculating sensitivity and specificity, with a larger sample size and with another research design in lines 291. Ultimately we aim make this method usable in the clinical setting, our team has started to collect a larger sample size and we are working on to improve and approve the test. If the statement is still objectionable we can omit it altogether.

We know that both retinal thickness and choroidal thickness depend on the axial eye length. The longer eyeballs have thinner layers and possibly with different reflectivity. What was the range of the axial eye length of the patients and controls? Were there differences in axial eye length between AE, FE and HE? Need to show if there is a correlation between axial eye length and the reflectivity of the layers.

To the best of our knowledge, axial length is only associated with a change in choroidal volume[3] and thickness[4] and not reflectivity. Axial length is associated with a change in RNFL reflectivity since RNFL reflectivity is directionally dependent[5] (myopes and hyperopes are very likely to have differences in cell bundle orientation).

We tried to account for axial length by performing another set of analysis excluding myopes and strabismic participants. The results were very similar and consistent to previous statistical analysis that did not account for axial length, except for ROI#4 which included the choriocapillaris and the Sattler’s layer. This might be due to the smaller sample size and/or the difficulty in delineating the borders of ROI#4. It could also be due to changes in thickness and volume which could be speculatively affecting reflectivity. We added the statistical results in the results section and we mentioned this matter in the discussion. (lines 215-233,337-347) (all added parts including tables are highlighted in yellow in the revised manuscript)

1. Thiese MS, Ronna B, Ott U. P value interpretations and considerations. Journal of thoracic disease. 2016;8(9):E928-e31.

2. Sullivan GM, Feinn R. Using Effect Size-or Why the P Value Is Not Enough. Journal of graduate medical education. 2012;4(3):279-82.

3. Barteselli G, Chhablani J, El-Emam S, Wang H, Chuang J, Kozak I, et al. Choroidal volume variations with age, axial length, and sex in healthy subjects: a three-dimensional analysis. Ophthalmology. 2012;119(12):2572-8.

4. Flores-Moreno I, Lugo F, Duker JS, Ruiz-Moreno JM. The relationship between axial length and choroidal thickness in eyes with high myopia. American journal of ophthalmology. 2013;155(2):314-9.e1.

5. Huang XR, Knighton RW, Feuer WJ, Qiao J. Retinal nerve fiber layer reflectometry must consider directional reflectance. Biomed Opt Express. 2016;7(1):22-33.

Attachments
Attachment
Submitted filename: Response to reviewers.docx
Decision Letter - Sanjoy Bhattacharya, Editor

Morphological changes in amblyopic eyes in choriocapillaris and Sattler’s layer in comparison to healthy eyes, and in retinal nerve fiber layer in comparison to fellow eyes through quantification of mean reflectivity: a pilot study.

PONE-D-21-10099R1

Dear Dr. Estephan,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Sanjoy Bhattacharya

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: (No Response)

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: All comments were addressed and included in the manuscript. The present form of the manuscript is suitable for publication.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Formally Accepted
Acceptance Letter - Sanjoy Bhattacharya, Editor

PONE-D-21-10099R1

Morphological changes in amblyopic eyes in choriocapillaris and Sattler’s layer in comparison to healthy eyes, and in retinal nerve fiber layer in comparison to fellow eyes through quantification of mean reflectivity: a pilot study.

Dear Dr. Estephan:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Sanjoy Bhattacharya

Academic Editor

PLOS ONE

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .