Peer Review History

Original SubmissionDecember 17, 2020
Decision Letter - Alfred S Lewin, Editor

PONE-D-20-39598

Macular pigment changes after cataract surgery with yellow-tinted intraocular lens implantation

PLOS ONE

Dear Dr. Obana,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please add a schematic of the absorbance spectra the ocular media (cornea, lens, retina [e.g. MP and other chromophores]) to help readers understand the region of high-energy, short-wavelength light. Please provide spectral transmission of the SN60 WF, XY1, YP2.2 and PCB00V IOLs.

Please ensure that your decision is justified on PLOS ONE’s publication criteria and not, for example, on novelty or perceived impact.

For Lab, Study and Registered Report Protocols: These article types are not expected to include results but may include pilot data. 

Please submit your revised manuscript by Mar 22 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Alfred S Lewin, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please provide a table of patient demographics.

3. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. Please see http://www.bmj.com/content/340/bmj.c181.long for guidelines on how to de-identify and prepare clinical data for publication. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: 1) Adding a schematic of the absorbance spectra the ocular media would be beneficial (cornea, lens, retina [e.g. MP and other chromophores]) to help readers understand the region of high-energy, short-wavelength light

2)The authors have correctly pointed it out but direct comparisons of RRS and 2-wavelength AF and HFP are inappropriate. Differences in measured MPOD among the 3 are difficult to assign meaning to.

3) Please provide spectral transmission of the SN60 WF, XY1, YP2.2 and PCB00V IOLs. This might help the reader understand the absorption contribution relative to the other ocular components

4) Although the author states that "No L/Z supplementation was utilized, a dietary questionnaire control might be useful

5) Tbl 2 and 3 show that virtually all of the MPOD and MPOV values increased at every position. As MP cannot be synthesized de novo and must be obtained through diet, I'm interested to learn more about the etiology. I ask out of clinical interest because the theory posited in Lines 74-79 states that increased transmission of SW light leads to decreased MP but the theory described in Lines 233-239 state that increased retinal irradiation following cataract surgery leads to enhanced retinal capture of circulating L/Z due to increased isomerization. Should all ECCE patients be using L/Z supplementation?

6) Lines 127-129 discussed analysis of RRS and MPOV @ 1.5 degrees measured through FAF. Comparison of these values may be inappropriate due to technique differences.

7) A table displaying change rate for time pints and eccentricities would be valuable

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Thank you for organizing a quick review. We appreciate the reviewers’ comments and have revised our manuscript accordingly.

The following details are our point-by-point responses (black font) to the reviewers’ comments (blue font, Italic). The main revised parts are indicated in the proofreading file in red.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming.

We checked our manuscript according to PLOS ONE’s style and we believe this manuscript fits it. If not, please let us know. We will revise it quickly.

2. Please provide a table of patient demographics.

Thank you for your suggestion. We added patient’s demographic data in Table 1.

3. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly.

We uploaded our data set.

Dear reviewer #1

We appreciate your immensely thorough review with positive comments and useful suggestions. We have analyzed our data and revised our manuscript according to your suggestions. The following details are our point-by-point responses to your comments.

1) Adding a schematic of the absorbance spectra the ocular media would be beneficial (cornea, lens, retina [e.g. MP and other chromophores]) to help readers understand the region of high-energy, short-wavelength light

Thank you for your comment. We added figures of spectral transmission of human crystalline lenses of different age and clear and yellow-tinted IOLs. The maximum absorption wavelength of macular pigments is 460 nm and this information was added in Figure 1B.

2)The authors have correctly pointed it out but direct comparisons of RRS and 2-wavelength AF and HFP are inappropriate. Differences in measured MPOD among the 3 are difficult to assign meaning to.

We consider that this comment is in response to the discussion from line 243 to 256.

Both two wavelength AF and HFP measure absorbance of MP, i.e. both methods measure the same thing. Therefore, the values measured by two methods should be identical, but actually there is some difference in the absolute values, although the values by two methods show high correlation. In this discussion, we compared not the absolute values but the increase rates between our study and Nolan’s study and we think this comparison was reasonable.

However, we attributed the different increase rates in two studies to the difference in the initial MP absolute values. This discussion might have some bias. Therefore, we added the following sentence in this discussion, “However, the measurement techniques used to calculate the initial MP values differed in the two studies, therefore, we should take into consideration of the difference in MP values, although the differences were thought to be minor.”

3) Please provide spectral transmission of the SN60 WF, XY1, YP2.2 and PCB00V IOLs. This might help the reader understand the absorption contribution relative to the other ocular components

Thank you for your suggestion. We added the spectral transmission of clear and yellow-tinted IOLs in Figure 1B.

4) Although the author states that "No L/Z supplementation was utilized, a dietary questionnaire control might be useful

Thank you for your comment. We asked the patients at the time of every examination and confirmed that they didn’t start taking supplements including lutein and zeaxanthin. We added the following sentence, “It was confirmed verbally that no patients started to take supplements containing lutein and/or zeaxanthin after surgery.”

5) Tbl 2 and 3 show that virtually all of the MPOD and MPOV values increased at every position. As MP cannot be synthesized de novo and must be obtained through diet, I'm interested to learn more about the etiology. I ask out of clinical interest because the theory posited in Lines 74-79 states that increased transmission of SW light leads to decreased MP but the theory described in Lines 233-239 state that increased retinal irradiation following cataract surgery leads to enhanced retinal capture of circulating L/Z due to increased isomerization. Should all ECCE patients be using L/Z supplementation?

Thank you for your comment. The mechanisms of increase in MP after surgery is unknown. But considering the autoregulatory system of humans, this result seems reasonable. Nolan et al speculated that “the increased visible light irradiation of the retina after cataract surgery could stimulate enhanced retinal capture of circulating lutein and zeaxanthin, perhaps due to a mechanism involving increased isomerization of the MP constituents under light exposure” as described in Discussion. We have no evidences to support their speculation. We think that whether MP increase or decrease depends on the extent of increased irradiation. When the irradiation is excessive (for example, in case with clear IOL implantation), consumption of MP may be higher than MP accumulation, and MP decreases.

6) Lines 127-129 discussed analysis of RRS and MPOV @ 1.5 degrees measured through FAF. Comparison of these values may be inappropriate due to technique differences.

RRS uses a different principle from AF. It measures the Raman signal of carotenoids in MP in the area of a diameter of 1 mm (= 1.6̊ ). Therefore, we compared the increase rate of MP in the similar area, i.e. MPOV within 1.5̊ eccentricity in AF and total Raman signals within 1.6̊ eccentricity. Since we compared not the absolute values but the rate of values, we consider this comparison is reasonable.

Attachments
Attachment
Submitted filename: Response to Reviewers(PLOS).docx
Decision Letter - Alfred S Lewin, Editor

Macular pigment changes after cataract surgery with yellow-tinted intraocular lens implantation

PONE-D-20-39598R1

Dear Dr. Obana,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Alfred S Lewin, Ph.D.

Section Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Formally Accepted
Acceptance Letter - Alfred S Lewin, Editor

PONE-D-20-39598R1

Macular pigment changes after cataract surgery with yellow-tinted intraocular lens implantation.

Dear Dr. Obana:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Alfred S Lewin

Section Editor

PLOS ONE