Peer Review History
| Original SubmissionSeptember 24, 2020 |
|---|
|
PONE-D-20-30175 Development of a novel three-dimensional organoid sarcoidosis granuloma model; created by adding a microfluidics system and developed granulomas to a dual-chambered lung on membrane model PLOS ONE Dear Dr. Mirsaeidi, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses all the points raised during the review process. ============================== Two experts have evaluated the manuscript. While the data are promising, the manuscript need to be substantially reorganized and rewritten. Both reviewers found that the references do not reflect the field. Ethical approvals/permissions for the use of human cells need to be added to the Methods section. ============================== Please submit your revised manuscript by Dec 05 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Mária A. Deli, M.D., Ph.D. Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. You state "to demonstrate the functionality of the device to run an experimental study, we developed ten LOMM and ten 3D OSGM as previously described." Please clarify whether these were exact replicates, or whether the ten LOMMs and OSGMs differed. If so, how did they differ from one another? 3. Thank you for stating the following in the Competing Interests section: "The authors have declared that no competing interests exist". We note that one or more of the authors are employed by a commercial company: 'Genix-Engineering, Irvine, California, USA'. 1. Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form. Please also include the following statement within your amended Funding Statement. “The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.” If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement. 2. Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc. Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. * Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf. Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: N/A ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Cacagno et al. developed an in vitro sarcoidosis granuloma model derived from the blood of patients. These 3D cellular structures were then added to a novel microfluidics system using regular Transwell system to model air liquid interface (ALI), and cell-cell interactions, cell aggregate-cell interactions and cytokine release was characterized. Generally the paper contains very interesting material related to the field, since no good lung granuloma-lung tissue (or other barrier tissue) in vitro model exists. Although non-referenced, but important papers in this research area should be cited such as (PMIDs): 28174307; 32974300; 32716613. This model is new by combining the ALI with the granuloma formation and studying cell-cell interactions. The major problems with the paper are not the findings, but with the presentation of the data. Introduction is too general, contains repetitions and the references are not always matching the text in a required fashion. Authors claim that neither human, animal or in vitro models are good to model the disease, which is important when one wants to state the limitations of different experimental systems. Here biochip models are claimed as too low throughput and cell culture models as redundant. Still authors try to present a new microfluidic-ALI-cell culture granuloma-model. Although one understands the model, the reasoning behind is not very logical. Methods section does not contain any ethical statements, although cells were isolated from patients. This has to be added. Discussion contains some methodology about the device, which should be inserted here. 10 paralells mean less than 4 repeats with the system / group? Results section is extremely redundant. Results have to be presented more in detail, not just as one would describe a figure legend. Please also double-check all references, for which I also indicated some ideas: - Ref 2, 3: the text is matching the logic of the paper, but the references are not very well chosen, they are too general. To discuss the limitations of animal models, special sarcoidosis animal models should be mentioned and discussed. Ref. 3. Is a German language paper matching the content of the text, but is also too general and is not available to read to wider audience. - Ref 4-7. - About biochip modeling and its versatility. These references are coming from 2 groups, where 2 papers is citing heart tissue monitoring. Plenty of papers are to be found about lung models, and in other tissue barrier models, please use a more broad range of citations when referring to this very wide field. - Sentence before Ref. 9. is almost the same as sentence before ref.2., so it is a repetition, please remove. - Refs 12 and 13 are the same, please correct. In general the paper contains good data, and the biochip system looks valuable, but the manuscript has to be re-formatted, carefully checked and some parts re-written. Also the title (and the text) should be re-formatted: the word "organoid" suggests a stem-cell based approach, which is not the case here. Reviewer #2: The authors present the development of an in vitro model of 3-D lung-on-chip organoid designed to mimic granuloma formation, which is characteristic of a heterogeneous multisystem disorder known as sarcoidosis. The 3-D organoid sarcoidosis model was created by adding a newly designed fluidic device, with three channels for cell culture insertion, to a previously developed dual-chambered lung-on-membrane model. Granulomas were cultured from blood samples of patients with sarcoidosis and then inserted in the air-lung interface of a microchip to create a 3-D organoid sarcoidosis model, and it was tested for cell viability with fibroblasts. They further challenged the air-lung interface of the organoid sarcoidosis model with microparticles to demonstrate the functionality of the device for experimental studies. Overall, the manuscript seems to be prepared in haste without sufficient detail for the readers to follow the narrative resulting in major deficiencies and, therefore, is not recommended for publication in its current form. The field of lung-on-a-chip has rapidly been developing in the last decade and is reaching maturity to a certain extent. Introducing granuloma formation to mimic in vitro sarcoidosis in the lungs is certainly a nice contribution. However, the text is rather confusing as to the origin of granuloma formation; is it from PBMC taken from blood of patients with sarcoidosis or is it induced by in-house developed microparticles? The images in Figure 1 seem to be taken from a culture dish; how these cultures are utilized for the development of 3-D OSGM? No protocol has been detailed. The 3 chambers in the system design shown in Figure 2 is an artefact as all chambers are identical; it is rather easy to design a similar device with 6 chambers or assemble two devices with the same number of devices. The authors should completely mute this aspect of the device. How the cross-section in Figure 3 was taken; presumably this is taken from the previously developed dual-chambered lung-on-membrane model which is not new? The endothelial layer seems to be rather sparse not quite fully confluent; is this an important issue, Can this be a reliable model? What is the role of the fibroblast cells shown in Figure 4; the lung model has already been developed and includes only epithelial and endothelial cells. The fibroblast discussion is completely superficial and should be removed although the image is nice. A similar image for the endothelial cells would be much more desired. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 1 |
|
PONE-D-20-30175R1 Novel three-dimensional biochip pulmonary sarcoidosis model PLOS ONE Dear Dr. Mirsaeidi, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses all the points raised during the review process. ============================== While the manuscript has been improved, many of the original requests remained unanswered. A thorough revision is needed addressing all the original comments, together with a marked-up copy of the manuscript showing all changes in yellow. It should be clearly described, explained and justified if figures were left out or changed. ============================== Please submit your revised manuscript by Dec 31 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Mária A. Deli, M.D., Ph.D. Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Calcagno et al. provided R1 in a timely fashion. Although several major issues have been addressed, I still do not recommend the publication in this form. This group has high quality and well written papers, such as PMID 32350353, which shows an example for this paper's aim as for quality. Although PlosOne is a lower impact factor journal than the PMID 323250353, PlosOne is Q1 in its field, therefore high quality of writing and data presentation is expected. Introduction and Discussion were modified and amended as requested. Still some references are missing, which for example work with Transwell granuloma models, such as PMID 32716613. References 28 and 29 (previously 12 and 13) are still the same, were not corrected, please correct. Ethical statement was added. Clarification about the repeat numbers was added, although in the text it is still confusing how many repeats were performed. The Figure legend makes it more understandable. The general term "organoid" was removed and corrected. My two most major concerns are the results section and presentation of the data. Some figures are missing compared to the previous version and the it is not clear why or what was corrected. Still the description of the data is very redundant, although a bit improved compared to the previous version. Also the answers for the reviewers' questions were extremely redundant and minimalistic, which is very unprecedented and unacceptable. Several questions of the previous revisons were not answered or are not found well in the text. Marked-up copy should have been provided to help the work of the reviewers. In general the paper was improved majorly. Still major revision is needed, but if authors correct the mistakes outlined, it would be recommended for publication. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
PONE-D-20-30175R2 Novel three-dimensional biochip pulmonary sarcoidosis model PLOS ONE Dear Dr. Mirsaeidi, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. ============================== There are two minor issues left to be addressed as requested by the Reviewer. ============================== Please submit your revised manuscript by Feb 19 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Mária A. Deli, M.D., Ph.D. Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Thank you for addressing my comments. Two minor comments are left, and then I suggest the paper for publication: For Fig 1. Please insert back the sentence detailing the cell types either in the figure legends or to the text. Previously text contained "The presence of multi-nucleated giant cells, lymphocytes, and macrophages which are aggregated together formed a large structured granuloma", which is now missing. This information is important, therefore I suggest its addition to the paper. Please put a short decription of the model to the Figure 2 legend - explaining the setup in a compact form. After these small modifications the manuscript is ready for acceptance. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 3 |
|
Novel three-dimensional biochip pulmonary sarcoidosis model PONE-D-20-30175R3 Dear Dr. Mirsaeidi, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Mária A. Deli, M.D., Ph.D. Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors have addressed all my comments. Now the paper is ready for acceptance. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No |
| Formally Accepted |
|
PONE-D-20-30175R3 Novel three-dimensional biochip pulmonary sarcoidosis model Dear Dr. Mirsaeidi: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Mária A. Deli Academic Editor PLOS ONE |