Peer Review History
| Original SubmissionJuly 31, 2020 |
|---|
|
PONE-D-20-23937 Negligible-cost microfluidic device fabrication using 3D-printed interconnecting channel scaffolds PLOS ONE Dear Dr. Hughes, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Nov 02 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Zhan Li, PhD Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2.Thank you for stating the following in the Acknowledgments Section of your manuscript: [The authors would like to acknowledge the training & pump-priming funding award through BristolBridge (grant number EP/ M027546/1) under the Engineering and Physical Sciences Research Council (EPSRC) Bridging the Gaps between the Engineering and Physical Sciences and Antimicrobial Resistance cross-council AMR initiative, as well as pump-priming awarded by the Civil, Aerospace & Mechanical Engineering (CAME) School, University of Bristol. Harry Felton’s work reported in this paper has been undertaken as part of the Twinning of digital-physical models during prototyping project at the University of Bristol, which is funded by the EPSRC [grant number EP/R032696/1]. Andrea Diaz Gaxiola’s work has been funded by ONACYT, MEXICO (http://www.conacyt.gob.mx/).] We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form. Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows: [The author(s) received no specific funding for this work.] Please include your amended statements within your cover letter; we will change the online submission form on your behalf. 3. We note that you have stated that you will provide repository information for your data at acceptance. Should your manuscript be accepted for publication, we will hold it until you provide the relevant accession numbers or DOIs necessary to access your data. If you wish to make changes to your Data Availability statement, please describe these changes in your cover letter and we will update your Data Availability statement to reflect the information you provide. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly Reviewer #2: Partly ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors presented a 3D printing process for the development of interconnecting microchannel scaffolds with a resolution down to 100 micron and a nearly negligible cost. The work is interesting, the 3D printed microchannel devices showed a good promise towards real-world applications, but before I can consider recommending publication in PLOS One I would suggest the authors to address the following issues: 1. To keep the manuscript written in a concise way, the authors may consider removing some of the figures in the manuscript. For example, Fig 1 is a typical process that is not necessarily placed in the introduction section of the manuscript; fig 2 describes the results of a previous work of others and therefore, the figure shall not be used in this manuscript; fig 2-5 may be put together and some of the information can be put in the supplementary. 2. I would suggest the authors not to use words like "novel" "significantly" "drastically" repeatedly. 3. The authors demonstrated several 3D printed device prototypes and discussed their printed microchannel characterisations in a detailed manner. In this perspective, the readers can easily evaluate the 3D printing results. However, it is not possible to determine whether the printed devices perform well as there lacked control experiments of microchannel devices fabricated via other means in this work or there are no references to the studies of others to allow the readers to compare the device performance. I would suggest the authors to consider put more efforts on the discussion of the device performance. Reviewer #2: The paper presented a microfluidic device fabrication paradigm based on 3D printing technology. The modularized design and manufacturing of the channel scaffolds accelerated the realization of microfluidic systems with low-cost. Through the authors giving thorough descriptions of every detail of the printing settings, this manuscript's major issue is the lack of sound analysis on explaining the reason for choosing fabrication parameters. Some other technical comments: 1. The descriptions on the 3d printing technology, 3d printer, and the slicing software in this work is trivial since the layer manufacturing has already been employed in the microfluidic and microelectronics fields with a lot of achievements, referring to ref. [1]. The manuscript should be shortened with a concise introduction of the settings related to mold design and fabrication. 2. In an FDM, process planning, which includes slicing and print path planning, is vital in determining print quality. Instead of listing all the printer settings, to improve the printing quality (section 2.2.2), the authors are suggested to further investigating the key parameters (like slicing height, line width, infill pattern, et al.) that affect the material deposition. Many works related to the print quality improvement of FDM has been overlooked, such as ref. [2,3]. Furthermore, the qualitative evaluation results listed in Table 2 seems meaningless for other practitioners or other print parts. The quantified analysis of the reason for unsuccessful printing should be made. 3. What are the advantages of this work over other 3d printed microfluidic devices, such as ref? [4]. To the best of my understanding, the device fabricated by the method of this work is simply two dimensional (the height is constant). Can the method be extended to a more complex device with variable height? Moreover, the authors may show some comparisons with others’ works. The 3d printed microfluidic device is an exciting topic, and the low-cost one is desperately needed in the academic sector. The author presented a complete process in realizing the fast fabrication of the device. However, the current manuscript is more like an introduction to instructing practitioners to design and manufacture the microfluidic device step-by-step. The manuscript has too much trivial information related to the print settings, which could be shortened, while lacks sufficient explanations on the key factors that affect the deposition accuracy or print surface quality. More quantified analysis related to the relationship between print parameters and the accuracy of the fabricated part should be added. The effectiveness of more demonstrations of the device can also be validated. Based on the above concerns, I suggest a major revision of this paper. Reference: [1] R. D. Sochol et al., “3D printed microfluidics and microelectronics,” Microelectronic Engineering, vol. 189. pp. 52–68, 2018. [2] A. Dey and N. Yodo, “A systematic survey of FDM process parameter optimization and their influence on part characteristics,” J. Manuf. Mater. Process., vol. 3, no. 3, p. 64, 2019. [3] G. Q. Jin, W. D. Li, and L. Gao, “An adaptive process planning approach of rapid prototyping and manufacturing,” Robot. Comput. Integr. Manuf., vol. 29, no. 1, pp. 23–38, 2013. [4] S. Tsuda et al., “Customizable 3D printed ‘Plug and Play’ millifluidic devices for programmable fluidics,” PLoS One, vol. 10, no. 11, pp. 1–13, 2015. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 1 |
|
PONE-D-20-23937R1 Negligible-cost microfluidic device fabrication using 3D-printed interconnecting channel scaffolds PLOS ONE Dear Dr. Hughes, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Feb 01 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Zhan Li, PhD Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: I Don't Know ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Thank the authors for the revision of this manuscript. The authors have addressed my previous querries and concerns. I therefore recommand publication in Plos One. Reviewer #2: The authors have made significant improvements in the revised version of the manuscript. Most of my concerns have alleviated. The 3d-printing-based microfluidic device fabrication method introduced in the paper substantially lowers the barrier for undergraduate students and other industry practitioners with low cost and easy-to-implement processes. The authors elaborate on all the 3d modeling, printer setting, and fabrication parameters. Before considering for publication, some minor issues are expected to be resolved: 1. Page 2, Figure 1, I cannot find figure 1(g), which is introduced in the caption of Figure 1. 2. Table 1, the unit of the build plate temperature is wrong. 3. Make sure all the literature is correctly referenced (e.g., Line 213). ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
PONE-D-20-23937R2 Negligible-cost microfluidic device fabrication using 3D-printed interconnecting channel scaffolds PLOS ONE Dear Dr. Hughes, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Feb 06 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Zhan Li, PhD Academic Editor PLOS ONE Additional Editor Comments (if provided): There are still some errors and typos as pointed out by the reviewer, which are not yet corrected, e.g., "...Images e-g) show characteristic ridges of 101 channels manufactured using SLA moulds printed at an angle to the substrate plane to improve print resolution..." in the caption of Fig. 1. [Note: HTML markup is below. Please do not edit.] [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 3 |
|
Negligible-cost microfluidic device fabrication using 3D-printed interconnecting channel scaffolds PONE-D-20-23937R3 Dear Dr. Hughes, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Zhan Li, PhD Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: |
| Formally Accepted |
|
PONE-D-20-23937R3 Negligible-cost microfluidic device fabrication using 3D-printed interconnecting channel scaffolds Dear Dr. Hughes: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Zhan Li Academic Editor PLOS ONE |
Open letter on the publication of peer review reports
PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.
We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.
Learn more at ASAPbio .