Peer Review History

Original SubmissionJuly 13, 2020
Decision Letter - Jianguo Wang, Editor

PONE-D-20-21415

Evaluating lubricant performance to reduce COVID-19 PPE-related skin injury

PLOS ONE

Dear Dr. Masen,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

==============================

ACADEMIC EDITOR:

  • I received comments and recommendations from three reviewers. Two reviewers are medicial experts and one is the expert in mechanics. Their recommendations are mixed but the comments may be helpful for further improvement.
  • Your research is interesting and we hope the result of this research may be helpful to the current community. Therefore, please also pay some attentions to the potential by-effects. 

==============================

Please submit your revised manuscript by Sep 27 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Jianguo Wang, PhD

Academic Editor

PLOS ONE

Journal requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that you have stated that you will provide repository information for your data at acceptance. Should your manuscript be accepted for publication, we will hold it until you provide the relevant accession numbers or DOIs necessary to access your data. If you wish to make changes to your Data Availability statement, please describe these changes in your cover letter and we will update your Data Availability statement to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: N/A

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Healthcare workers around the world are experiencing skin injury due to the extended use of personal protective equipment (PPE) during the COVID-19 pandemic. This study aims to provide a practical lubricating solution for frontline medical staff working a 4+ hours shift wearing PPE.

A literature review into skin friction and skin lubrication was conducted to identify products and substances that can reduce friction. We evaluated the lubricating performance of commercially available products in vivo using a custom-built tribometer. The results can improve our understanding of evaluating lubricant performance to reduce COVID-19 PPE-related skin injury. The detailed comments are presented below:

1. The study was approved by the Imperial College London Science, Engineering and Technology Research Ethics Committee (SETREC), reference 20IC5999. Please provide evidence.

2. In this study, the experimental programme was performed over a one month period, from mid-May to mid-June 2020. Please list your work for each period in detail

3. In this study, there are some formulas. But the formula is not numbered. Besides, each variable needs to be explained.

4. There are so many authors for this article. A statement for your authors' contribution and conflicts of interest are necessary.

Reviewer #2: This is a very interesting experiment.In this study, a 44 year old caucasian man was recruited to test the effectiveness of a series of lubricants to reduce friction and shear force in the left forearm (instead of facial skin), and hope find which lubricants has a protective effect for more than 4 hours.

However, the experimental design did not consider the following factors:

1.Participants: The subjects' health status was unknown, such as height, weight, etc.Only one caucasian male was selected and does reflect the friction and lubricant effect of women and other skin color health care workers on their skin ?

2.Setting: The volar forearm was selected as the experimental site. Does affect the friction coefficient measured after using the lubricant due to the different thickness of between the stratum corneum on the volar forearm and the facial skin ?Does affect the friction coefficient measured after lubricant use due to different perspiration of facial skin and forearm skin?In this study, authors measured the static friction coefficient , while the skin of medical staff wearing PPE against covid-19 produced dynamic friction coefficient. How to explain the correlation between the two?In the fight against covid-19, the skin damage caused by wearing PPE was related to the increased friction between skin and PPE caused by pressure, rapid exercise and sweating.Does consider the increase of friction force under multi-factors when lubricant is used to reduce friction?

3.Experimental products: The author selected four kinds of 19 different products(Table 1 on page7-8) to be used in the same part of the same person in a month. The friction coefficient of each product was measured three times at 1, 2 and 4 hours after use. Will the frequent use of different products change the microclimate of the skin and thus affect the friction coefficient?

Analysis of the results and discussion of this study has the following doubts:

1.Figures 4A to 4F showed schematic rather than measurement results. What does the author want to tell the readers?

2.In the discussion, the authors inferred the results based on these diagrams. Can they reflect the real situation?

3.The authors has come to the following conclusion:Results indicate that the use of emollients and moisturizing creams is to be discouraged when wearing PPE for long duration as they may result in excessive shear forces acting on the skin. Talcum powder, a lanolin containing petrolatum, and a coconut oil-cocoa butter-beeswax mixture provide excellent long-lasting lubrication.But there seems to be little evidence to support.

Minors

1.There are some spelling mistakes in the text.

2.The author described it in the abstract:This research was funded by the Imperial College COVID-19 Response Fund. But on page 10 it was written:The research funder did not play a role in the research.What does that mean?

Reviewer #3: Very interesting and relevant study looking at lubricants to protect skin health in those using PPE during COVID-19. I have some minor points which require clarification.

1. Friction is only one element of the boundary conditions that causes skin damage. Pressure, shear and an altered microclimate will also effect skin health. This is introduced in the manuscript but perhaps a little miss-leading in figure 1. If the same pressure and shear is applied to the skin, its likely to cause similar tissue deformation regardless of lubricant.

2. In addition, Figure 1 should include ‘pressure and shear induced skin injury’ – shear will not be acting in isolation, unless you are referring to tissue shear strain.

3. Did you recreate the material interface of PPE devices when assessing the friction properties? You discuss the implications of skin epidermal absorption, occlusion and water retention, but does this also apply for the PPE interface material?

4. In the introduction it would be worth highlighting PPE devices which are typically implicated in damage, for example respiratory protective equipment (FFP3 masks)

5. NHS England and NHS Improvement [4] do not recommend petrolatum to the skin. This has been reported in one manuscript with limited evidence for its efficacy.

6. Many PPE devices that are implicated in skin damage e.g. the aforementioned FFP3 masks do not employ PDMS at the interface.

https://www.3m.co.uk/3M/en_GB/company-uk/3m-products/~/3M-Aura-Disposable-Healthcare-Respirator-FFP3-Valved-1873V-/?N=5002385+3292799385&rt=rud

7. Brill et al. [17] evaluated the pressures required to secure non-invasive ventilation masks, this may differ to the pressures observed in PPE equipment.

8. I appreciate the forearm is convenient for testing, however does this replicate the skin properties of the face, for example the highly porous sites at the nose? As you’ve highlighted there are large differences in friction for unlubricated skin, which are related to variations in sebum, hydration and mechanical properties

9. Strip waxing the forearm would remove layers of the stratum corneum and result in increased transepidermal water loss. Could this have effected your results? Why not use a commercial hair removal cream?

10. Did you measure the biophysical properties of the skin/check for blanching erythema? Seems like the single participant was put through a large number of mechanical tests.

11. In-between tests the PDMS was applied, so the interactions between skin and PDMS under mechanical loading conditions was not tested? - you have addressed this to some extent in the discussion

12. Results are very descriptive, some inclusion of data, or for example percentage change values from baseline would add context for the reader.

13. No mention of the discussion regarding the functionality of PPE. Indeed, FFP3 masks must retain a seal with the skin in order to filter >95% of particles. Any additional sliding at the interface could be to the detriment of the devices functionality. There must be a careful interplay between maintaining skin health and PPE function.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: Yes: Peter Worsley

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

We would like to thank the reviewers for their positive comments and helpful suggestions. Our detailed response to the reviewers comments was uploaded in a separate document.

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Jianguo Wang, Editor

Evaluating lubricant performance to reduce COVID-19 PPE-related skin injury

PONE-D-20-21415R1

Dear Dr. Masen,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Jianguo Wang, PhD

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The revisions are satisfactory to this reviewer. The manuscript is recommended for publication in its current form.

Reviewer #3: Thank you for addressing the comments, I have no further issues. The manuscript has its limitations, although this is in part due to the restrictions in lab based studies during COVID19.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #3: No

Formally Accepted
Acceptance Letter - Jianguo Wang, Editor

PONE-D-20-21415R1

Evaluating lubricant performance to reduce COVID-19 PPE-related skin injury

Dear Dr. Masen:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Jianguo Wang

Academic Editor

PLOS ONE

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .