Peer Review History
| Original SubmissionMay 24, 2020 |
|---|
|
PONE-D-20-15595 Performance of the ImmuView and BinaxNOW assays for the detection of urinary and cerebrospinal fluid Streptococcus pneumoniae and Legionella pneumophila serogroup 1 antigen in Legionnaires’ disease or pneumococcal pneumonia and meningitis PLOS ONE Dear Dr. Edelstein, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised below during the review process. Please submit your revised manuscript by Aug 08 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Ray Borrow, Ph.D., FRCPath Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. Please provide additional details regarding participant consent. In the ethics statement in the Methods and online submission information, please ensure that you have specified (a) whether consent was informed; (b) what type you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed). If the need for consent was waived by the ethics committee, please include this information; (c) whether you obtained consent from parents or guardians of the children included in your study. 3. Thank you for stating the following in the Financial Disclosure section: 'PHE received funding from SSI Diagnostica to conduct this study CSJ received funding from SSI Diagnostica to conduct this study LAW received funding from SSI Diagnostica to conduct this study SSI Diagnostica: https://www.ssidiagnostica.com/ The study sponsor, SSI Diagnostica, wrote the test protocols, paid for the study reagents and supplies as well as for the performing technologist’s time. This report was written independently of SSI Diagnostica using data spreadsheets supplied by the sponsor as well as separately kept spreadsheets for the UPenn site. SSI Diagnostica reviewed the manuscript for accuracy of the reported data but otherwise had no role in its writing.' We note that you received funding from a commercial source: SSI Diagnostica a. Please provide an amended Competing Interests Statement that explicitly states this commercial funder, along with any other relevant declarations relating to employment, consultancy, patents, products in development, marketed products, etc. Within this Competing Interests Statement, please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests). If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. b. Please include your amended Competing Interests Statement within your cover letter. We will change the online submission form on your behalf. Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests 4. We note that you have included the phrase “data not shown” in your manuscript. Unfortunately, this does not meet our data sharing requirements. PLOS does not permit references to inaccessible data. We require that authors provide all relevant data within the paper, Supporting Information files, or in an acceptable, public repository. Please add a citation to support this phrase or upload the data that corresponds with these findings to a stable repository (such as Figshare or Dryad) and provide and URLs, DOIs, or accession numbers that may be used to access these data. Or, if the data are not a core part of the research being presented in your study, we ask that you remove the phrase that refers to these data. 5. Please amend either the title on the online submission form (via Edit Submission) or the title in the manuscript so that they are identical. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: I Don't Know Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Overall a good paper describing the results of a multisite study comparing the ImmunView and BinaxNOW urine antigen and CSF tests for Streptococcus pneumoniae and Legionella pneumophila. I would like to have seen more samples from patients with bacteraemic pneumococcal pneumonia tested at the UPenn site as I feel that the lack of these samples makes site comparisons for this particular test group difficult. I also have some minor points for you to consider. 1. Line 47- I don’t think it is necessary to describe the disease as “pneumonia caused by either pneumococcal pneumonia”. I would suggest something like “pneumococci” instead of “pneumococcal pneumonia” or alternatively remove the "caused by " and just list pneumococcal pneumonia. 2. Line 59- Please could you list the criteria that was used to define the children with lower respiratory infection not due to pneumococcal pneumonia. As cases of co-infections with pneumococci and other bacteria or viruses have been reported, was this considered when selecting these samples? 3. Line 172-174- The footnote for Table 1, specifically references e and f, are used to describe the total "urines" tested. However, the data in the first column of the table are supposed to represent CSF samples. Please can you clarify. 4. Line 265 - Please can you clarify as to what is meant by the ImmunView L. pneumophila test. I am assuming that it means that some of the samples reported a result for L. pneumophila when tested with the ImmunView test. 5. Line 283-286 -Please can you explain why priority was given to the ImmunView test over the BinaxNow when conducting repeat testing with insufficient sample volume. 6. Line 381-383 - Have you considered the sensitivity of the assays in regard to the different pneumococcal serotypes? 7. Table 4 - Could this table also include the figures for the number of samples reported positive and negative out of the total tested that was used to calculate the sensitivity and specificity? Reviewer #2: In the present study, Edelstein et al compared the performances of ImmuView S. pneumoniae and L. pneumophila urinary antigen tests with that of the BinaxNOW S. pneumoniae and L. pneumophila assays using frozen samples and performing the testing at three laboratories. The study sponsor SSI Diagnostica wrote the test protocols and financed the study, but the report was written independently of them. The report adds information regarding the ImmuView test performance in comparison with the BinaxNOW assays on clinical samples. My major concern is that samples are referred to as collected from pneumococcal pneumonia and Legionnaires’ disease patients but data are lacking on collection procedure and patient characteristics. When and where did the authors collect samples, and when were they analyzed? If the 55 Legionella serogroup 1 samples tested at SSI are the same 55 samples analyzed by Jorgensen et al (2015) previously this should be presented. Patient characteristics from those with suspected Legionnaires’ disease would help to interpret discordant results, especially since more than half of Legionnaires’ disease cases from the UPenn site are defined only by previous positive antigen tests (see below). It would be suitable to refrain from using the term “clinical” for defining test performance to avoid misunderstandings concerning the trial design. Another concern is the lack of criteria for pneumonia diagnosis. Usually, a majority of patients with pneumococcal bacteremia have airway infections but, sometimes, may have meningitis, otitis or other foci only. If not described, maybe the authors should refer to patients with bacteremia and not bacteremic pneumonia. Furthermore, are these urine samples collected consecutively during certain time periods, or are they selected for some reason due to special features, e.g. if a positive pneumococcal antigen test was positive in the clinical routine along with positive blood culture? Could the selection of samples explain the high sensitivity yields for pneumococcal antigen testing for both tests in the study, and for site specific differences? Line 115: How was serogroup 1 identified? Line 118: How was pneumonia defined? Line 155: Bacteremic pneumococcal pneumonia. Since the definition of Legionnaires’ disease and pneumococcal pneumonia is not clear, and based on positive microbiology results only, it should be referred to possible diagnoses. Line 157: The sentence refers to Table 1 only. Line 178: Supportive laboratory findings for patients with suspected Legionnaires disease? Line 199: Not clinical performance, since the definitions of diagnoses have not yet been clarified. Line 210: Not clinical specificity, since the definitions of diagnoses have not yet been clarified. Line 261: Not clinical sensitivity, half the samples are spiked. Line 269: Patient characteristics? Age? Previous urinary antigen test performed? Was urine sample and nasopharynx sample collected simultaneously, or colonization “at any time”? Line 340: Is it 54% (60/111)? Please refer to Table 2 If appropriate. Line 352: Due to absent test concordance…? Words missing. Line 404: Little or no information of samples tested for specificity is presented in the study. Thus, if the samples are from healthy individuals, the results only reflects specificity on such a healthy population without respiratory symptoms. It would be of interest to study the test performances on patients with non-Legionella non-pneumococcal lower airway infections or bacteremia. Line 408: Here, it is stated that the use of the term clinical sensitivity is not appropriate in the study, I agree. Lines 388-391: If the selection of urine samples are based on positive Legionella urinary antigen tests only, with no clinical data on pneumonia, samples may be true-negative from the beginning, but false-positive before boiling. This is a major issue of the study, see above. Besides the discrepant specimens, could there be more positive tests that would have turned negative after boiling? ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 1 |
|
PONE-D-20-15595R1 Performance of the ImmuView and BinaxNOW assays for the detection of urine and cerebrospinal fluid Streptococcus pneumoniae and Legionella pneumophila serogroup 1 antigen in Legionnaires’ disease or pneumococcal pneumonia and meningitis PLOS ONE Dear Dr. Edelstein, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the final minor points raised during the review process. Please submit your revised manuscript by Sep 27 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Ray Borrow, Ph.D., FRCPath Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: I Don't Know ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Thank you for submitting the revised manuscript. I am satisfied with the authors response to the reviewers’ comments and believe that the manuscript should be accepted. I would however like to suggest a couple of minor changes for the authors to consider. 1. In lines 304-305 you state “Testing of urines from children with lower respiratory tract infection, none of whom had pneumococcal pneumonia, was performed to determine the specificity of urine antigen testing...” A number of publications have highlighted that the diagnosis of pneumonia caused by pneumococci can be insensitive when based solely on culture. In fact, this is a reason in favour of the use of urine antigen diagnostic kits such as the BinaxNOW and ImmuView. For this reason, I believe that it is incorrect to state that these children did not have pneumococcal pneumonia and instead suggest that you re-word this sentence to note that the children were deemed not to have pneumococcal pneumonia based on the absence of any culture. 2. Lines 327-328 – “For the three BinaxNOW-ImmuView discordant result UPenn L pneumophila results for urines from patients with Legionnaires' disease…” – perhaps consider removing the first "result" from this sentence for clarification. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
Performance of the ImmuView and BinaxNOW assays for the detection of urine and cerebrospinal fluid Streptococcus pneumoniae and Legionella pneumophila serogroup 1 antigen in Legionnaires’ disease or pneumococcal pneumonia and meningitis PONE-D-20-15595R2 Dear Dr. Edelstein, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Ray Borrow, Ph.D., FRCPath Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: |
| Formally Accepted |
|
PONE-D-20-15595R2 Performance of the ImmuView and BinaxNOW assays for the detection of urine and cerebrospinal fluid Streptococcus pneumoniae and Legionella pneumophila serogroup 1 antigen in patients with Legionnaires’ disease or pneumococcal pneumonia and meningitis Dear Dr. Edelstein: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Prof. Ray Borrow Academic Editor PLOS ONE |