Peer Review History

Original SubmissionMay 14, 2020
Decision Letter - Selvakumar Subbian, Editor

PONE-D-20-14345

SARS-CoV-2 detection in setting of viral swabs scarcity: are MRSA swabs and viral swabs equivalent?

PLOS ONE

Dear Dr. Federman,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

As mentioned by Reviewer#1, this study was done using limited number of samples from one location.  This should be mentioned in the text as a limitation of this study. All other comments of all reviewers should be addressed on the revised manuscript in addition to the rebuttal letter.

Please ensure that your decision is justified on PLOS ONE’s publication criteria and not, for example, on novelty or perceived impact.

Please submit your revised manuscript by Jul 25 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Selvakumar Subbian, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating in your ethics statement

"The protocol was deemed a quality improvement project and granted exemption by the VA Connecticut Research Office and deemed that review by the VA Connecticut Investigational Review Board and R and D Committee was not needed."

Please clarify whether the  VA Connecticut Research Office is your ethics committee.

Please also clarify whether informed consent was specifically waived by your ethics committee.

If not, please discuss whether all data were fully anonymized before you accessed them or whether written informed consent was obtained from patients.

3. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter.

4. We noted in your submission details that a portion of your manuscript may have been presented or published elsewhere:

'The entire manuscript was submitted simultaneous with MedRxiv and published there. There was supposed to be simultaneous submission to your journal but there was a problem with submission (linking of my account with the account of a Dr. David Feillin- your office knows about this)'

Please clarify whether this publication was peer-reviewed and formally published.

If this work was previously peer-reviewed and published, in the cover letter please provide the reason that this work does not constitute dual publication and should be included in the current manuscript.

5. PLOS requires an ORCID iD for the corresponding author in Editorial Manager on papers submitted after December 6th, 2016. Please ensure that you have an ORCID iD and that it is validated in Editorial Manager. To do this, go to ‘Update my Information’ (in the upper left-hand corner of the main menu), and click on the Fetch/Validate link next to the ORCID field. This will take you to the ORCID site and allow you to create a new iD or authenticate a pre-existing iD in Editorial Manager. Please see the following video for instructions on linking an ORCID iD to your Editorial Manager account: https://www.youtube.com/watch?v=_xcclfuvtxQ

6. Please ensure that you refer to Figure 1 in your text as, if accepted, production will need this reference to link the reader to the figure.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: No

Reviewer #2: N/A

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The manuscript entitled SARS-CoV-2 detection in setting of viral swab scarcity: are MRSA swabs and viral swabs equivalent? describes the use of MRSA swabs with Amies media and Viral swab with universal transport media for sample collection from SARS-CoV-2 infected patients. Authors indicated that the MRSA swabs yields the comparable results as that of viral swabs. However, this manuscript failed to address the following;

1. Sample size is less

2. Sampling was done at only one center; sampling in different screening center by different technicians will require to exclude personal variation in sample collection.

3. As the Amies media is not known as medium of choice for sample collection for viral diagnosis and it is expected that the sample will be stored for varying time period at room temperature prior to transport to lab to store at -20C; the duration of stability of RNA in the Amies media must be tested prior to recommending them for SARS-CoV-2 sample collection.

4. References not included for several sentences in the introduction; For example- "The spectrum of disease severity ranges from asymptomatic disease to severe pulmonary disease including acute respiratory distress syndrome, multisystem organ failure, and death. Genetic sequencing performed early in the course of the outbreak led to the

development of diagnostic testing".....include references.

5. The following sentence is not clear and need to be re-written : We sampled 81 unique patients, of which the viral swabs were positive on the DiaSorin test in 19.

6. It is needless to mention "Additionally, while our study was conducted only in Veterans, we do not believe there

should be a difference in test results between Veterans and non-Veterans for the positive or

negative predictive value of testing for SARS-CoV-2 infection"

7. Under methods, the sentence "virals wabs (BD H192(07) flexible mini-tip flocked " need to be written as "viral swabs (BD H192(07) flexible mini-tip flocked"

8. The very short and incomplete sentences [ for example "The viral swab specimens were nasopharyngeal" needs to be merged with neighboring sentence or can be re-written appropriately.

Reviewer #2: The tremendous need for swab test justifies the work by Daniel et.al. and this would for sure be helpful in terms of increasing the capacity to test. The manuscript is technically sound and the data support the conclusions. Although the sample number is small, the importance of work makes it acceptable.

Percentage correlation of the positive and negative test of two different swabs together with results from Cepheid Xpert Xpress ® SARS-CoV-2 assay justify the conclusions with appropriate control.

Minor revisions: The authors should describe the primers used for the amplification of CoV2 RNA with details of the amplification conditions in the method section.

The manuscript is well written and acceptable with minor reversions.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: Yes: Alok Choudhary

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

see accompanying document

Attachments
Attachment
Submitted filename: Response to Reviewers PLOS ONE.docx
Decision Letter - Selvakumar Subbian, Editor

SARS-CoV-2 detection in setting of viral swabs scarcity: are MRSA swabs and viral swabs equivalent?

PONE-D-20-14345R1

Dear Dr. Federman,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Selvakumar Subbian, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: (No Response)

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: (No Response)

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: (No Response)

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: (No Response)

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: (No Response)

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Authors have addressed all the comments.

However, due to the small sample size tested for the evaluation of MRSA swabs, it will be more appropriate if the title of manuscript changed as "Pilot study on the applicability of MRSA swabs for SARS-CoV-2 sampling in setting of viral swabs scarcity".

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Santhamani Ramasamy

Formally Accepted
Acceptance Letter - Selvakumar Subbian, Editor

PONE-D-20-14345R1

SARS-CoV-2 detection in setting of viral swabs scarcity: are MRSA swabs and viral swabs equivalent?

Dear Dr. Federman:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Selvakumar Subbian

Academic Editor

PLOS ONE