Peer Review History

Original SubmissionSeptember 2, 2019
Decision Letter - David Fyhrie, Editor

PONE-D-19-24675

Novel multi-energy X-ray imaging methods: Experimental results of new image processing techniques to improve material separation in computed tomography and direct radiography

PLOS ONE

Dear Mr Heckert,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

We would appreciate receiving your revised manuscript by Mar 12 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

David Fyhrie

Academic Editor

PLOS ONE

Journal requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. It is unclear why you have selected 'No - some restrictions will apply', 'For the multi-energy computed tomography data it may not be possible to share the whole dataset because of limited space (its around 500-800 Gbs). Some exemplaric slices and a scaled down dataset will also be made available via figshare.' Please clarify the nature of these restrictions.

PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For more information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially sensitive information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The authors introduce an empirical approach to perform a pixelwise fitting of the grey level of both radiographic images and a series of volumetric images corresponding to a CT scan measured with different X-ray tube voltages. The radiographs are fitted using a cubic interpolation while the CT scans are fitted using an exponential function. I believe the results are very interesting and depict a new method for empirical material differentiation on CT scans using conventional hardware and multiple X-ray energies. Therefore, I suggest its publication after the changes specified below. These are mostly required to justify the effectiveness of the approach, as well as to improve the scientific and technical rigor of the paper. Major points: - I do not think the relationship between the proposed fitting and the X-ray imaging model is clear. This mathematical relationship is, in my opinion, critical for replication of the results in the case of other materials. Since the authors claim the method to be qualitative rather than quantitative, my suggestion is for the authors to use a phantom with other materials to show the empirical method can do material differentiation in other cases too (Not as supplemental material with video 2 but as part of the discussion). - Related to the previous comment, further discussion is needed on the selection of the cubic interpolation for the radiography case. Furthermore, how did the authors get to the conclusion on how to use \\lambda, \\gamma and \\delta to generate the material images? The relation with the physical quantities is not clear from the theory developed in the paper. If the authors have explained this in a previous publication, please use a reference (line 125). - The notation is confusing. The authors do a nice work getting to equation (3) but then they go from V(t) to V(U) in equation (4) without an explanation and say it’s a derivation. This needs to be explained. Furthermore, V(U) is supposed to be the measured intensity at each tube voltage U, however, they claim to do a pixel/voxel grey level fitting for the CT reconstruction using the expression V(U) too. The latter is very confusing since this gray level corresponds to the attenuation coefficient quantities reconstructed for each voltage, these are not intensities. Please change the notation or clarify this accordingly. - Related to the previous comment, please explain why the second fit is on the linear attenuation coefficients rather than on the intensities as derived in equation (4). - Seems to me from the manuscript that the user needs prior information of the materials present in the sample to be able to identify which one corresponds to each parameter fitted. Is this correct? If not, please clarify How can one know which material is in the sample if the \\rho \\Z images are not quantitatively correct? - I believe there is not enough detail on the fitting to replicate the results of the paper. The authors do not give information on the type of software written, size of images, number of voltages, size of the step in the tube voltage, selection of n, size of the sample. Among others. Please elaborate, this information is often key for the reconstructions and cannot be derived from the manuscript. Minor points: - In the abstract the authors claim there are not stablished multi-energy methods. What about photon counting detectors? The statement needs to be revised. Maybe clarify what they mean with multi-energy method. On the other hand, to my knowledge there is one paper that relies on photon counting detectors to attain \\rho Z images [**] that is close a multi-energy method. Please elaborate. - In equation (1), e (which I assume is the charge of the electron) is not defined.

- In equation (4), K is not defined. Previous K in equation (2) came from the Kramer’s law but with the simplification in equation (3), K is not present anymore. Is the K in equation (4) different? Please, define. - The link for the fitting program MultiE-GUI does not work. I was able to find a software with the same name that seems to correspond to the description in the text, URL: https://figshare.com/articles/MultiE-GUI_zip/9758390/1. Please fix this. - If I understand correctly the approach in reference [1] is not used to fit /mu due to the lack of correction of the quantum efficiency at higher X-ray energies. Could you show the results that you get when you try performing the fit with your data? From what I saw on the .zip that I downloaded you have the option of performing this fit. I think this discussion will help the readers see a clear motivation of the proposed approach. - There are some typos along the paper. Please revise carefully before resubmitting. Some examples below: o Line 34: that instead of than o Line 112: change to: account for this behavior computationally o Line 182: An image instead of a image o Line 206: To an acceptable instead of to a acceptable o Line 249: A more instead of an more o Line 251: parameter instead of paramter [**] Busi, Matteo & Kehres, Jan & Khalil, Mohamad & Olsen, Ulrik. (2019). Effective atomic number and electron density determination using spectral x-ray CT. 2. 10.1117/12.2519851.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachments
Attachment
Submitted filename: ReviewSpectral.pdf
Revision 1

Dear Editor,

please find the response to your email in the uploaded cover letter and the reviewers comments in the letter called "Response to Reviewers"

Kind Regards,

Mirko Heckert

Attachments
Attachment
Submitted filename: rebuttal letter-2.docx
Decision Letter - David Fyhrie, Editor

Novel multi-energy X-ray imaging methods: Experimental results of new image processing techniques to improve material separation in computed tomography and direct radiography

PONE-D-19-24675R1

Dear Dr. Heckert,

We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements.

Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication.

Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

With kind regards,

David Fyhrie

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Formally Accepted
Acceptance Letter - David Fyhrie, Editor

PONE-D-19-24675R1

Novel multi-energy X-ray imaging methods: Experimental results of new image processing techniques to improve material separation in computed tomography and direct radiography

Dear Dr. Heckert:

I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

For any other questions or concerns, please email plosone@plos.org.

Thank you for submitting your work to PLOS ONE.

With kind regards,

PLOS ONE Editorial Office Staff