Peer Review History
| Original SubmissionJune 25, 2019 |
|---|
|
PONE-D-19-17736 Head to head comparison of two commercial fecal calprotectin kits as predictor of Mayo endoscopic subscore and mucosal TNF expression in ulcerative colitis PLOS ONE Dear Dr Goll, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Standardization is a frequent problem with assays such as fecal calprotectin. Diagnostic cut-off determined with one assay do not necessary apply for another assay. It is useful to compare different producer’s tests and I like the attempt to use an independent parameter as TNF transcripts. Unfortunate the authors only compare two of the many available assays for fecal calprotectin, extending to more would have increased the usefulness of the paper. I wonder if the authors have thought about directly comparing the assays with spiked samples? Using a given amount of WBC and stool from non-human species could have provided exact limit of detection/quantification and could have normalized the assays. It seem rather unclear how many samples/patients that have been used in the present paper, I would suggest providing a flowchart. Have the authors got permission to use data from the authors and journal of the previous paper (Scand J Gastroenterol. 260 2018 Jul 3;53(7):825–30). We would appreciate receiving your revised manuscript by Sep 07 2019 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript:
Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Pal Bela Szecsi, M.D. D.M.Sci. Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: No Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Regarding the manuscript by Goll et al. Today F-calprotectin assays are widely used and is an important part of the workup of patients with suspected IBD. Evaluation assays and the correlation with TNF expression are thus important. I think the data is clearly presented and I have only minor comments. Abstract line 25: I am not completely enthusiastic about the use of calibration in this context. Calibration is more often used for assay calibrations and here the authors are relating the values to clinical states which is much more complex and also more valuable. I believe that the word calibration here may be misinterpreted by some readers. I thus suggest that the authors try and change calibration to another word. Page 3, line 45: To some extent, part of… Please remove either as this is a duplication Page 3 line 56. Add a space before (9-11) Page 4: The extraction process: How similar were the sample extractions. Same sample? Were the samples mixed before sampling? Page 5 and 6 method comparison: Most likely the authors had results that were above and below the measuring ranges of the assays. How were such samples handled? Values set at the limits? Analyzed rediluted? Omitted? Page 7 line 164: Add a space between andhead-to Table 1: Please add a table legend that also should include the units for the F-calprotectin assays. Figure 3: please add the units Please a sentence about availability of data. The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository Reviewer #2: Material and methods When the authors refer to the two Calprptectin kits, they have to specify which methodology they use, and they put it in results, at the end of Inter-assay variation. Nor do they indicate the cut-off from which they consider the TNF activity of the biopsies to be positive. The total number of samples used for the comparative study is not understood, as they indicate that data from another study have been used, but everything is very unclear. In addition, the controls for which they were used are not well known, as no results are given either in Table 1. Results The first fragment of results should actually be included in Material and methods (lines 102 to 112), because there are not results on it. I do not understand how lowering the cut-off increases specificity and lowers sensitivity (lines 144 and 145). Discussion The results have been compared with few previous results, hence the scarce bibliography. Table 1 The legends of the abbreviations are missing. Errata Some errors should be corrected, as in Line 60: Calpro (Calpro Norway), and some words together, without spaces. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 1 |
|
PONE-D-19-17736R1 Head to head comparison of two commercial fecal calprotectin kits as predictor of Mayo endoscopic subscore and mucosal TNF expression in ulcerative colitis PLOS ONE Dear Dr Goll, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. The manuscript have greatly improved, but some minor issues still remain. Please address the point reviewer #2 has raised. We would appreciate receiving your revised manuscript by Nov 01 2019 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript:
Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Pal Bela Szecsi, M.D. D.M.Sci. Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: (No Response) Reviewer #2: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: (No Response) Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: (No Response) Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: (No Response) Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: (No Response) Reviewer #2: Abstract - Line 35: In the sentence "and at higher values Calprest measured higher values than Calpro", the authors should indicate the average increase increase of Calprest over Calpro in % at values >1000 mg/Kg. - Lines 37 and 38: the authors use the terms "reasonable precision" and "transcript values reasonably well" subjectively, rather than objective identifiers in quantitative terms. - Line 58: the authors continue to use the term 'calibrating'. Instead, it would be more appropriate to use the terms "adjustment" or "setting", better than tuning. Material and methods - Line 66: delete "Patients with UC were recruited first" because it is repeated later; add "as part of a prospective project at the Department of Gastroenterology, University hospital of North Norway" behind (12), on line 70. - Line 97: add manufacturer and locality of TaqMan assays for TNF. - Line 11: GraphPad Software, La Jolla, CA, USA, best put in parentheses. Results - Lines 145-146: I disagree that when the cut-off goes down, the sensitivity drops and the specificity increases. When lowering the cut-off, more cases of UC are diagnosed (sensitivity increases), but also other non-inflammatory diseases, with a greater number of false positives (specificity decreases). These data should be reviewed. I do not understand how you have obtained these results; it is not understandable or reasonable. In fact, to increase specificity, the cut-off increase above 100 later (lines 150-151). - Line 146: correct "maksimal". - Line 153: separate "totalRNA". - Lines 160 and 161: it is noteworthy that the Youden indices obtained with the MES + Calprotectin in both trials are very low (0.13 and 0.24), and the authors do not refer to it in the Discussion. Table I: The result of the 16 controls has not been included in the trial comparison. Figure 2: The authors should add the line of the line and refer in the text that the correlation between the calprotectin results of two samples from the same patient is weak by both methods (Pearson's r of 0.53 and 0.55). Figure 5: the authors must add in the ROC curves the kit that corresponds to each color (blue and red), and also in the legend, in brackets TNF <7500 after “Normalized mucosal TNF gene expression”. Conclusion Add some mention to the prediction of inflammation or its cure (MES), as you indicated in the summary. References - Reference 11 (line 260), change 1936.e1. for 36.e1. - Reference 20: the year is incorrect; volume and pages are missing and the authors are incompletely cited. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
PONE-D-19-17736R2 Head to head comparison of two commercial fecal calprotectin kits as predictor of Mayo endoscopic sub-score and mucosal TNF expression in ulcerative colitis PLOS ONE Dear Dr Goll, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Some minor issues still needs your attention, please correct. We would appreciate receiving your revised manuscript by Nov 29 2019 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript:
Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Pal Bela Szecsi, M.D. D.M.Sci. Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: (No Response) Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: (No Response) Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: (No Response) Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: (No Response) Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: (No Response) Reviewer #2: The authors have greatly improved the manuscript quality, there are only minor details to modify before publication. These are the following: Authors' contributions: delete Summary (line 17). Keywords: delete "calibration". Lines 145-145. Now I understand the objective of the authors: different cut-off points according to the MES searched (positive/negative). That is correct, but clinically this is difficult to establish. Laboratory tests usually have a single cut-off point, and results that are above/below are considered positive/negative (or vice versa). Otherwise, a grey area of results is established, difficult to classify. I accept your explanation. Results Add a full stop, new paragrah behind of"...for Calprest." (line 160). Conclusions Add to the end of the Conclusions the statement that the authors include in the Abstract, but that does not appear in the manuscript text (already indicated in the previous review): "Both kits were more accurate in the prediction of active inflammation (MES 2-3), but less for the prediction of mucosal healing (MES 0) and normalization of the expression of the mucosal TNF gene ". Figure 2: I was referring to the straight line of identity, which you have already added by the authors. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. |
| Revision 3 |
|
Head to head comparison of two commercial fecal calprotectin kits as predictor of Mayo endoscopic sub-score and mucosal TNF expression in ulcerative colitis PONE-D-19-17736R3 Dear Dr. Goll, We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements. Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication. Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. With kind regards, Pal Bela Szecsi, M.D. D.M.Sci. Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: |
| Formally Accepted |
|
PONE-D-19-17736R3 Head to head comparison of two commercial fecal calprotectin kits as predictor of Mayo endoscopic sub-score and mucosal TNF expression in ulcerative colitis Dear Dr. Goll: I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. For any other questions or concerns, please email plosone@plos.org. Thank you for submitting your work to PLOS ONE. With kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Pal Bela Szecsi Academic Editor PLOS ONE |
Open letter on the publication of peer review reports
PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.
We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.
Learn more at ASAPbio .