Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Table 1.

SEC-HPLC system conditions.

More »

Table 1 Expand

Table 2.

Summary of FFF-MALS system conditions.

More »

Table 2 Expand

Table 3.

SEC-HPLC test results of AMP-glucagon and ELI-glucagon.

More »

Table 3 Expand

Fig 1.

A representative SEC-HPLC chromatogram for expired AMP-glucagon: A dimer peak was observed at approximately 5.8 minutes.

More »

Fig 1 Expand

Fig 2.

A representative SEC-HPLC chromatogram for recently released AMP- glucagon: No dimer peaks were detected.

More »

Fig 2 Expand

Fig 3.

A representative SEC-HPLC chromatogram for expired ELI-glucagon: A dimer peak was observed at approximately 5.8 minutes.

More »

Fig 3 Expand

Fig 4.

A representative SEC-HPLC chromatogram for non-expired ELI-glucagon: A dimer peak was observed at approximately 5.8 minutes.

More »

Fig 4 Expand

Table 4.

Equivalence evaluation of aggregation of AMP-glucagon vs. ELI-glucagon By SEC-HPLC.

More »

Table 4 Expand

Fig 5.

Typical chromatograms of LS2 signal for glucagon samples: No peak differences were observed between new (a, b) and old (c, d) lots in the two glucagon products.

More »

Fig 5 Expand

Fig 6.

A Representative FFF chromatogram for AMP-glucagon: The molecular weight of glucagon was approximately 3,504 Da.

More »

Fig 6 Expand

Fig 7.

A representative FFF chromatogram for ELI-glucagon. The molecular weight of glucagon was approximately 3,720 Da.

More »

Fig 7 Expand

Table 5.

Glucagon samples’ weighted-average molar masses by FFF/UV-MALS.

More »

Table 5 Expand

Table 6.

BSA standard sample MW summary.

More »

Table 6 Expand

Fig 8.

Detection limit of glucagon samples: A reconstituted glucagon sample was diluted to 0.1% and 0.5% of its original concentration.

For the 0.1%, the UV, LS2, and RID, no glucagon signal was observed at 7 minutes. For the 0.5%, a very low signal was found at 7 minutes, indicating that the FFF-MALS test is unable to detect glucagon at contents lower than 0.5%.

More »

Fig 8 Expand

Fig 9.

Detection limit of BSA (positive control): The 2 mg/mL BSA sample was diluted to 0.4 and 0.2 mg/mL as positive control samples which is equivalent to 10% and 5% of glucagon sample loaded respectively.

The BSA peak should appear at 13–16 minutes based on the UV signals.

More »

Fig 9 Expand