Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Fig 1.

Schematic of the experimental setup used to measure the performance of the FFR.

The elements labeled in orange were used for the generation of NaCl submicron particles. These particles were then released into the test chamber, and filtration performance was calculated using the measuring devices labeled in purple.

More »

Fig 1 Expand

Fig 2.

Schematic representation of the experimental design used with volunteers.

The cycle used for large-scale implementation in the hospital is identified by the red dotted line.

More »

Fig 2 Expand

Fig 3.

A 40-m3 decontamination room with 8 metal racks equipped with hooks, used for large-scale implementation of FFR treatment at the IUCPQ-UL.

More »

Fig 3 Expand

Table 1.

Minimum filtration performances measured at 85 L/min a.

More »

Table 1 Expand

Table 2.

Percentage of FFRs that passed fit tests after treatment.

More »

Table 2 Expand

Table 3.

Aeration time required to achieve FFRs hydrogen peroxide emissions below 0.1 ppm a,b.

More »

Table 3 Expand