Fig 1.

Schematic of the steerable needle assembly.

(A) Inner needle. (B) Outer catheter. The assembly of (A) and (B) forms the active steerable needle (A+B). The boxes show the transversal cross-section of each part.

More »

Fig 1 Expand

Fig 2.

Schematic of needle steering.

An applied force on the proximal end of the needle (Fproximal) results in deflection of the distal end of the needle (δdi). The roller supports in the middle function as flexure bearings for needle deflection.

More »

Fig 2 Expand

Fig 3.

Needle steering with the manufactured prototype.

An applied force on the proximal end of the needle results in deflection of the distal end of the needle. The needle guide in the middle functions as flexure bearings for needle deflection.

More »

Fig 3 Expand

Fig 4.

Pseudo-rigid body model of the inner needle.

The inner needle is modelled with six pseudo-rigid-body links and six torsional springs. The horizontal segments and roller supports indicate the needle guiding. Downward movement of the proximal end results in axial pushing of the bottom segment and axial pulling of the upper segment resulting in upward movement of the distal tip.

More »

Fig 4 Expand

Table 1.

Properties of the finite element model.

More »

Table 1 Expand

Fig 5.

The finite element model of the active steerable needle in normal and bent condition.

The boxes show the configuration of the inner needle during bending. The arrows indicate the direction of the shift of the segments. Stress and bending apply for the computational model with Xpr = 120 mm and θpr = 70°.

More »

Fig 5 Expand

Fig 6.

The relationship between proximal input angle and relative translation between segments for different proximal needle lengths.

The red line indicates the distal output. Markers on the distal output line relate to the vertically aligned markers on the input lines. Maximum stress and deflection is described for all proximal inputs.

More »

Fig 6 Expand

Fig 7.

Set-up and pipeline for shape reconstruction of the fixed-bent needle steering experiment.

The Robotic Ultrasound System (RobUSt) is used for shape reconstruction of the steerable instrument. For each insertion, RobUSt performs a volumetric scan of the tissue simulant, recording US images along with corresponding transducer pose data, expressed in the global reference frame at the base of the robotic manipulator. The silhouette of the instrument is segmented from each frame. All segmented data are assembled in a point cloud. The point cloud is processed using an iterative shape reconstruction algorithm, described in detail in Suligoj et al. [47]. The iterative shape algorithm first generates a series of points along the instrument (red dots). These points are used to fit a third order polynomial (green) describing the continuous shape of the instrument.

More »

Fig 7 Expand

Table 2.

Steering capability of the steerable needle in the fixed-bent needle steering experiment.

More »

Table 2 Expand

Fig 8.

Lateral deflection of the steerable needle over insertion length in the fixed-bent needle steering experiment.

The mean error ± σ in lateral direction for two different proximal bending angles and two initial straight insertion depths. The curves characterized by steering after 40 mm are superimposed on the curves of steering from <0,0,0> to determine what the influence of the initial depth is on needle steering.

More »

Fig 8 Expand

Fig 9.

Set-up of active needle steering experiment in prostate tissue simulant.

Proximal needle manipulations in Z-direction allow for steering at the distal tip. Insertion of the needle in the prostate tissue simulant and movement of the US transducer is performed bimanually. The camera visualizes the back surface of the tissue simulant in transversal plane. The prostate, pubic arch, urethra and rectum are illustrated in the figure in sagittal plane to demonstrate the clinical applicability of the steerable needle for prostate brachytherapy. The pubic arch and urethra are circumvented to reach occluded prostate tissue.

More »

Fig 9 Expand

Table 3.

The endpoint errors of the steerable needle and the reference needle per tissue simulant.

More »

Table 3 Expand

Fig 10.

Set-up of active needle steering experiment in ex-vivo tissue.

The trajectory of the steerable needle (red) over a length of 90 mm towards a target (black) located laterally from the neutral axis. Control of the steerable needle was performed in 3D. Segmentation of the needle and target from the US images are used for analysis.

More »

Fig 10 Expand