Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Fig 1.

LARI mechanical design. (A) CAD cross section of LARI: (1) A Lid heating block, (2) Lid housing, which is hinged to the base housing, (3) Raspberry Pi single-board computer with LCD touchscreen interface, (4) Base housing, (5) Main control PCB, (6) Sensor PCB, (7) Tube heating block, shown in cross-section through a tube pocket and cartridge heater, (8) Cartridge heater, (9) LED PCB, (10) Through-holes in heating block allow light transmission from the LEDs, through the sample tubes, and on to the sensors, (11) Sample tubes are loaded into machined pockets in the block. (B) LARI shown with the housing open. All components are visible including the lid and tube heating blocks, PCBs, Raspberry Pi/touchscreen, and cabling. (C) Fully assembled LARI with lid closed. (D) Fully assembled LARI with lid open, showing sample tubes loaded.

More »

Fig 1 Expand

Table 1.

Bill of materials for critical components.

More »

Table 1 Expand

Fig 2.

LARI block diagram, including all principal components and touchscreen UI layout.

More »

Fig 2 Expand

Fig 3.

LARI user workflow.

Typical steps for performing a LARI experiment.

More »

Fig 3 Expand

Fig 4.

Thermal performance of LARI.

(A) Transient temperature response over two temperature set points as measured at three different tube locations: T1 (leftmost tube), T4 (directly over left cartridge heater), and T7 (center of block). Also shown are the software-reported temperature readings as well as temperature monitored with the calibration RTD within the block. Tube temperature lags block temperature by about 10–20 s, with outermost tube lagging the worst. Inset shows full 40-minute run. (B) Steady-state thermal camera image of LARI with tube block at 69° C and lid block at 80° C. (C) Steady-state sample tube temperature across 5 LARIs as measured with an external thermocouple inserted in water in each tube, showing excellent temperature uniformity.

More »

Fig 4 Expand

Fig 5.

Characterizing LARI signal performance using black PN dye.

(A) Signal plots across all 12 tubes in a single LARI run where a serial dilution series of dye samples were swapped in at 5 min intervals. (B) Average signal readings vs. dye concentration across 5 LARIs (12 tubes per LARI). These readings were fit to a 4-parameter logistic (4PL) regression model as shown with the dotted line. (C) 4PL parameters (midpoint, slope, and signal amplitude) do not show significant differences among different tube positions or LARIs, indicating variability is primarily due to LED and sensor variability.

More »

Fig 5 Expand

Fig 6.

Colorimetric LAMP assay performance.

(A) Example colorimetric LAMP assay signals across all 12 tube locations from two different assay runs: one with all positive samples and another with all negative (NTC) samples. Markers indicate TTR for each curve. Inset image shows buffer color at the end of an assay run for both positive and negative samples. (B) Colorimetric LAMP assay performance across six independent runs across five LARI instruments, all using the same manufacturing lot of assay reagents. No discernable trends are visible either from run-to-run or instrument-to-instrument.

More »

Fig 6 Expand

Fig 7.

Tube-to-tube variability.

(A) Signal amplitudes for positive assays across all tube positions at t = 5 min (before amplification) and t = 15 min (after amplification). (B) Signal amplitudes for a negative assay at the same timepoints. (C) Ratios of signals for positive assays at t = 15 min to signals at t = 5 min all tube positions. (D) Signal slope at TTR for positive assays across all tube positions. (E) TTR for positive assays across all tube positions.

More »

Fig 7 Expand