Fig 1.

Bland–Altman plots of agreement between swept-source optical coherence tomography and partial coherence interferometry.

(A) axial length (AL) for 175 eyes, (B) anterior chamber depth (ACD) for 104 eyes. The solid line shows the mean difference (bias). Upper and lower dashed lines imply 95% limits of agreement.

More »

Fig 1 Expand

Table 1.

Mean value, mean difference and 95% limit of agreement of axial length in 175 eyes and anterior chamber depth in 104 eyes readings of the two biometry devices.

More »

Table 1 Expand

Fig 2.

Bland–Altman plots of agreement in the evaluation of the corneal curvature.

(A) flat K (K1), (B) steep K (K2), (C) J0, (D) J45. The solid line shows the mean difference (bias). Upper and lower dashed lines imply 95% limits of agreement.

More »

Fig 2 Expand

Fig 3.

Preoperative astigmatism using double-angle plot of two biometry.

More »

Fig 3 Expand

Table 2.

Mean value, mean difference, and 95% limit of agreement (LoA) of keratometric values (K1, K2, J0 and J45) between two biometry.

More »

Table 2 Expand

Table 3.

Pearson correlation of ocular parameters reading between two biometry.

More »

Table 3 Expand