Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Table 1.

The tumor volumes, respiratory-tumor motion, and tumor margins for six cases.

More »

Table 1 Expand

Fig 1.

3D printed lung phantom.

(a) The GTV volume and location of the six patient cases, (b) Design of a 3D printed lung phantom, (c) Rendering image (left) and fabricated 3D-printed lung phantom (right).

More »

Fig 1 Expand

Fig 2.

4D phantom with 3D printed lung phantom.

The inserts are either a 3D printed lung phantom (inhomogeneous) or a homogeneous phantom.

More »

Fig 2 Expand

Fig 3.

The lung tissue volume (LTV) of the phantom.

(a) The LTV is filled with 0.3 mm strips with a 2 mm air gap to match the lung density. (b) The green area is the LTV in the phantom.

More »

Fig 3 Expand

Fig 4.

Patient-specific respiratory data.

Six patient-specific breathing patterns were acquired from the Varian RPM system during planning CT scan for each patient.

More »

Fig 4 Expand

Fig 5.

4D-CT scan of the 4D-lung phantom for case P4.

(a) A respiratory motion device and the 3D-printed lung phantom insert. The EBT3 film could be placed inside phantoms. (b) 4D-CT image set of the lung phantom was sorted in phases. The images listed from left to right correspond to the breathing phases of end-inhale (0%), mid-exhale (30%), end-exhale (50%), and mid-inhale (70%), respectively. The 17.9 mm full motion was reduced to 3.6 mm residual motion in the 30–70% gating window. (c) VMAT plan dose distribution for the lung phantom at the end-exhale (50%) phase.

More »

Fig 5 Expand

Fig 6.

Image registration between the patient CT and lung phantom CT.

Both CT image were used in the coronal plane.

More »

Fig 6 Expand

Fig 7.

Dosimetric evaluation of the phantoms with lung density reassignment.

(a) Axial and sagittal images of the lung phantom are shown. The phantom lung regions were implemented with a mesh structure with 2 mm spacing. The plan dose of 2 arc VMAT is shown in the coronal plane. (b) The CT number of the phantom LTV was replaced by the average lung CT number in the patient’s image (-813 HU). The dose distributions based on the original and reassigned images were compared in the gamma index map. (c) For the HU reassignment with air (-1000 HU), the GPR was 62.5%.

More »

Fig 7 Expand

Table 2.

HU value of GTV and LTV for six cases and the GPRs (3%/0 mm) compared under various material assignments of LTV in the lung phantoms.

More »

Table 2 Expand

Fig 8.

The analysis of the gamma passing rates (2%/1 mm) between the calculated dose and the measured dose comparing VMAT techniques under the static and gating conditions using both homogeneous and inhomogeneous phantom (3D-printed patient-specific moving phantom) in the P2 case.

The treatment planning dose was applied using the AAA or the AXB dose calculation algorithm.

More »

Fig 8 Expand

Table 3.

The gamma analysis (2%/1 mm criteria) results of the homogeneous and inhomogeneous phantoms (patient-specific lung phantom) for the six cases.

More »

Table 3 Expand