Table 1.

Rapid diagnostic test products used by the participants (137 RDT products from 8 brands, used by 133 participants).

The underlined names represent the shortened names used in the text to refer to the different RDT products.

More »

Table 1 Expand

Fig 1.

Example of photographs for Alere Binax Now® Malaria Test, (n = 59).

Abbreviations: Pf = P. falciparum, Pm = P. = malariae, Po = P. ovale, non-Pf = Plasmodium non-falciparum, IFU: Instructions for use.

More »

Fig 1 Expand

Fig 2.

Instructions for Use (IFU) for Bio-Rad OptiMAL-IT, (881056, version 01/2010, included in each box containing 24 individual test packs).

The IFU did not clearly mention the (rare) possibility of a P. falciparum infection with only a Pf-pLDH band without a pan-pLDH band. According to the IFU, the test result should be reported as invalid.

More »

Fig 2 Expand

Fig 3.

Instruction for Use (IFU) for All Diag PALUTOP+4 OPTIMA® (version 10, 03/07/2012).

The possibility of a mixed infection of P. falciparum and P. vivax with P. ovale and/or P. malariae was not clearly mentioned in the IFU (not included in “8.1 Test results”, but further under “Test limits”).

More »