Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Fig 1.

Schematic presentation of portable Raman spectrometer.

The output from the diode laser (300mW continuous power) is first collimated, unwanted spectral side lobes and fluorescence are filtered, then redirected by the dichroic mirror and finally focused (by the f/0.55 lens) just below the skin surface. Whilst both the intense reflected/scattered light, fluorescence, and generated Raman photons are collected by the f/0.55 lens, the dichroic mirror and the long-pass filter ensure that only the latter two contributions to the spectrum are focused (by the f/1.3 lens) on the entrance slit of the spectrometer.

More »

Fig 1 Expand

Fig 2.

Simulation of signal acquisition density distribution.

Panels (A)-(D) show the simulated density distribution of the generated Raman signal in the skin for collection depths of 150, 200, 250, and 300μm, respectively. The skin is modelled as a turbid medium featuring both bulk and surface scattering.

More »

Fig 2 Expand

Fig 3.

Box plot of performance versus collection depth.

35 patients were grouped into three depth ranges as indicated: 140–200 um, n = 14; 240–270 um, n = 12; 280–320 um, n = 9. The box shows upper and lower quartile and median. The threshold of 0.5 was supported by the CV results and the grading of glucose similarity for the regression vector, above this value the performance was satisfactory.

More »

Fig 3 Expand

Fig 4.

Independent validation of the last five days of analysis.

The measured glucose concentrations are plotted as a function of reference values in a consensus error grid. The measurement critical depth was 240–320 um and the ISUP was greater than 0.5.

More »

Fig 4 Expand

Table 1.

Performance of cross-validated (CV) and independent validations (IV).

More »

Table 1 Expand

Fig 5.

Raman spectrum and independent predictions for an average-performing patient.

(A) Average Raman spectrum from the thenar region (blue), Raman spectrum of glucose in an aqueous solution (red), and regression vector of PLS calibration model (black). (B) Plot of the independently predicted glucose concentration and the reference value as a function of time from the last day of calibration. The dashed lines are for guidance only.

More »

Fig 5 Expand