Table 1.

Experimental and clinical studies on the use of CeVOX.

More »

Table 1 Expand

Table 2.

Patients characteristics.

More »

Table 2 Expand

Fig 1.

Bland Altman plot comparing ScvO2_CeVOX_jug to ScvO2_BGA derived from measurements with jugular CVC.

ScvO2_CeVOX_jug: Central venous oxygen saturation derived from the CeVOX-device. ScvO2_BGA: Central venous oxygen saturation derived from blood gas analysis. CVC: Central venous catheter.

More »

Fig 1 Expand

Fig 2.

Bland Altman plot comparing ScvO2_CeVOX_fem to ScvO2_BGA derived from measurements with femoral CVC.

ScvO2_CeVOX_fem: Central venous oxygen saturation derived from the CeVOX-device. ScvO2_BGA: Central venous oxygen saturation derived from blood gas analysis. CVC: Central venous catheter.

More »

Fig 2 Expand

Fig 3.

Bland Altman plot comparing ScvO2_CeVOX to ScvO2_BGA derived from all measurements (jugular or femoral CVC).

ScvO2_CeVOX: Central venous oxygen saturation derived from the CeVOX-device. ScvO2_BGA: Central venous oxygen saturation derived from blood gas analysis. CVC: Central venous catheter.

More »

Fig 3 Expand

Table 3.

Univariate association of different variables to ScvO2_CeVOX–ScvO2 and the amount |ScvO2_CeVOX–ScvO2|.

More »

Table 3 Expand

Fig 4.

Correlation of the amount of the bias |ScvO2_CeVOX–ScvO2_BGA| with a prediction formula derived from the total datset (n = 120).

ScvO2_CeVOX: Central venous oxygen saturation derived from the CeVOX-device. ScvO2_BGA: Central venous oxygen saturation derived from blood gas analysis.

More »

Fig 4 Expand

Fig 5.

ROC curve comparing different predictors of |ScvO2_CeVOX–ScvO2_BGA| ≥5% in the evaluation group (n = 80).

“Prediction of |bias|”: model predicting |ScvO2_CeVOX–ScvO2_BGA| which was derived from multiple regression analysis within the evaluation group. ROC: receiver operating characteristic. AUC: area under the curve. ScvO2_CeVOX: Central venous oxygen saturation derived from the CeVOX-device. ScvO2_BGA: Central venous oxygen saturation derived from blood gas analysis.

More »

Fig 5 Expand

Fig 6.

ROC curve comparing different predictors of |ScvO2_CeVOX–ScvO2_BGA| ≥5% in the independent validation group (n = 40).

Prediction of |bias|: model predicting |ScvO2_CeVOX–ScvO2_BGA| which was derived from multiple regression analysis within the evaluation group. ROC: receiver operating characteristic. AUC: area under the curve. ScvO2_CeVOX: Central venous oxygen saturation derived from the CeVOX-device. ScvO2_BGA: Central venous oxygen saturation derived from blood gas analysis.

More »

Fig 6 Expand