Fig 1.

An electrospraying setup for stent-coating.

(A) Schematic of an electrospraying setup for stent-coating using a circular shielding electrode placed right underneath the needle and above the aluminum plate. Continuous coating on Express coronary stent was achieved by exposed to the electrospraying mist for 30 min: (B) bare metal stent and (C) coated stent. Scale bar: 500 μm.

More »

Fig 1 Expand

Fig 2.

Deposited droplet size histogram obtained from different eletrospraying modes.

Four eletrospraying modes were analyzed: (A) micro-dripping mode at electric field of 1.3 kV/cm, (B) spindle mode at electric field of 1.4 kV/cm, (C) cone-jet mode at electric field of 1.5 kV/cm, and (D) precession mode at electric field of 1.6 kV/cm. The optical microscopy images of the deposited droplets are shown in the inset Figs with scale bars of 100 μm.

More »

Fig 2 Expand

Fig 3.

Formation of primary electrosprayed droplets.

(A) Schematic diagram of the droplets formed from a jet through Rayleigh’s capillary breakup under a cone-jet mode during electrospraying process. (B) The diameter of deposited droplets vs. the square root of flow rate using the improved electrospraying setup with circular shielding electrode as shown in Fig 1A.

More »

Fig 3 Expand

Fig 4.

Breakup of electrosprayed droplets due to Coulombic fission.

Schematic (left) and optical microscopy (right) images of deposited droplets collected from electrspraying mist at different tip-to-collector distance of (A) 3 cm, (B) 6 cm and (C) 9 cm, using a same scale bar of 200 μm. Fragmented droplets were formed in (B) and (C) due to Coulombic fission. The electrospraying process was performed using a constant tip-to-plate distance of 9 cm.

More »

Fig 4 Expand

Fig 5.

Coating morphology obtained at different electrospraying modes.

SEM images of Express coronary stents coated by electrosprayed droplets obtained at different electrospraying modes: (A) cone-jet mode, (B) precession mode, and (C) multi-jet mode. These electrospraying modes were achieved at an increasing electric field of (A) 1.5 kV/cm, (B1) 1.6 kV/cm, (B2) 1.7 kV/cm, and (C) 1.8 kV/cm, respectively. Round and stretched deposited droplets were highlighted in empty yellow triangles (A) and solid yellow triangles (B), respectively.

More »

Fig 5 Expand

Fig 6.

SEM image of a stent coated with large electrosprayed droplets (dia. ~120 μm).

A non-uniform coating with large pores was observed on the stent due to dewetting of the liquid coating caused by ineffective evaporation and solidification of the deposited droplets during coating process.

More »

Fig 6 Expand

Fig 7.

Coating morphology controlled by Coulombic fission.

SEM images of Express coronary stents coated by electrosprayed droplets obtained at different flow rates with varying degree of Coulombic fission: (A) 0.5 mL/h, (B) 0.4 mL/h, (C) 0.3 mL/h, and (D) 0.2 mL/h.