Fig 1.

Flow Diagram.

Fig 1 shows the process of selecting recall reports as Flow Diagram in line with the PRISMA guidelines. The Flow Diagram includes two search methodologies; the link one is the process of a systematic search performed for English articles in academic databases; the right one is from data sources provided by different regulatory authorities from 193 United Nations Member States. A total of 103 recall reports have been analyzed in this study.

More »

Fig 1 Expand

Table 1.

Number of cardiac implants and total recall reports.

More »

Table 1 Expand

Fig 2.

Product categories of recalled cardiac implants.

Fig 2 indicates the recall rate of six different cardiac implant categories in the 103 recall reports analyzed in this study. Of all 103 recall reports, the ratio of categories is shown: ICDs 40.8%; pacemakers 14.5%; stents 14.5%; as well as CRTs 12.7%; leads 9.7% and implantable artificial organs 7.8%.

More »

Fig 2 Expand

Table 2.

Ratio of recall reasons among cardiac implant medical devices.

More »

Table 2 Expand

Table 3.

The death and serious injury reports from 103 identified recall reports.

More »

Table 3 Expand

Table 4.

Different recall management according to the same recall event in different regulatory authorities.

More »

Table 4 Expand