Table 1.

Radionuclide Properties.

More »

Table 1 Expand

Figure 1.

Photon energy spectra as measured by the Canberra HPGe detector for A) background, B) F-18, C) Zr-89 and D) I-124.

Insets show close-up of 511 keV annihilation photon peak upon which calibrations were based.

More »

Figure 1 Expand

Figure 2.

Activity versus calibration setting number as measured on a Capintec CRC-15R dose calibrator for A) Zr-89 and B) I-124, measured in different geometries.

Filled symbols identify the curves corresponding to the measurements with the copper filter; open symbols without the filter. Triangles correspond to the small volume in the 5 mL syringe, circles to the 3 mL volume in the 5 mL syringe and squares to the 10 mL volume in the liquid scintillation vial. Note that for I-124 the filled squares almost completely obscure the filled circles.

More »

Figure 2 Expand

Table 2.

Recommended Calibration setting Numbers.

More »

Table 2 Expand

Table 3.

Error in Radioactivity when Standard Calibration Setting Numbers are Used.

More »

Table 3 Expand

Figure 3.

Plot of photon energy versus sensitivity for a Capintec CRC-15R dose calibrator.

(Reprinted with permission of Capintec Inc. from the CRC-15R Owner's Manual).

More »

Figure 3 Expand