Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Table 1.

Comparison of PURE system with RTS system (E. coli crude cell extract based cell-free system).

More »

Table 1 Expand

Figure 1.

Optimization of PURE system as measured by active Fluc produced by supplementing different concentrations of EF-Tu, Ts, G; EF4; RF1, 2, 3 and RRF.

(a). Active Fluc produced at different EF-Ts, Tu and G concentrations. The table below shows the actual concentration increase of EF-Ts, Tu and G in the PURE system. (b). Active Fluc produced at different EF4 concentrations. (c). Active Fluc produced at different RF1, 2, 3 and RRF concentrations. The table below shows the actual concentration increase of RF1, 2, 3 and RRF in the PURE system. Fluc activities were measured in relative luminescence unit by luciferase assay and PURE system reaction without supplement was set as control. Error bars are ± standard deviations, with n = 3.

More »

Figure 1 Expand

Figure 2.

Optimization of PURE system as measured by functional Fluc produced by adding macromolecular crowding agents.

(a). Active Fluc produced in the system at different BSA concentrations. (b). The time course (kinetics) of transcription measured by Fluc mRNA yields with the Quant-iT RiboGreen RNA reagent with and without the presence of 15.5 µM BSA. (c). Active Fluc produced in the system at different PEG-6000 concentrations. In (a) and (c) Fluc activities were measured in relative luminescence units by luciferase assay and PURE system reaction without supplement was set as control. Error bars are ± standard deviations, with n = 3.

More »

Figure 2 Expand

Figure 3.

Optimization of PURE system as measured by functional Fluc produced by adding chaperone systems GroEL/ES and DnaK/DnaJ/GrpE.

(a). Active Fluc produced at different GroEL/GroES concentrations. (b). Active Fluc produced at different DnaK/DnaJ/GrpE concentrations. Fluc activities were measured in relative luminescence unit by luciferase assay and PURE system reaction without supplement was set as control. Error bars are ± standard deviations, with n = 3.

More »

Figure 3 Expand

Figure 4.

Optimization of PURE system as measured by functional Fluc produced by adjusting tRNA, ATP and GTP concentrations.

(a). Increasing tRNA concentration by 56 A260 units/ml boosts functional Fluc yield by 25%. The table below shows the actual concentration increase of EF-Ts, Tu, G and tRNA in each reaction. Reaction 1 is taken as control. In reaction 2, 3, 4 and 5, EF4; RF 1, 2, 3, RRF; GroEL/GroES and BSA were added at their optimized concentrations. (b). Increasing Mg2+ concentration decreases functional Fluc yield in the original PURE system. (c). Increasing Mg2+, ATP and GTP concentrations has little effect on final yield of functional Fluc in our optimized PURE system with optimized concentrations of EF-Ts, Tu, G; EF4; RF 1, 2, 3, RRF; GroEL/GroES and BSA. Fluc activities were measured in relative luminescence unit by luciferase assay and PURE system reaction without supplement was set as control. Error bars are ± standard deviations, with n = 3.

More »

Figure 4 Expand

Figure 5.

Optimization of PURE system with the best combination of EF-Ts, Tu, G; EF 4; RF 1,2,3, RRF; GroEL/GroES; BSA and tRNA concentrations as measured by functional Fluc produced.

Fluc activities were measured in relative luminescence unit by luciferase assay and PURE system reaction without supplement was set as control. Error bars are ± standard deviations, with n = 3.

More »

Figure 5 Expand