Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Table 1.

Interobserver reproducibility (κ score) among three observers for qualitative assessment*.

More »

Table 1 Expand

Table 2.

Results of qualitative scores* for image quality at different tube currents with AIDR3D and FBP (without AIDR3D).

More »

Table 2 Expand

Figure 1.

Axial plain chest CT images at the upper lung zone (60-year-old male weighing 76 kg).

These images were created from scan data at 240 mA (A, D), 120 mA (B, E) and 60 mA (C, F). The three upper images (A–C) were reconstructed using AIDR3D and the three lower images (D–F) were reconstructed using a conventional reconstruction mode (Boost3D). Each image pair at the same tube current was created from single row data. Image noise was obviously reduced on images with AIDR3D, particularly at lower tube currents.

More »

Figure 1 Expand

Figure 2.

Reconstructed coronal plain chest CT images (56-year-old male weighing 62 kg).

Images are arranged as in Figure 1 (A–C, with AIDR3D; D–F, without AIDR3D; A and D, at 240 mA; B and E, at 120 mA; C and F, at 60 mA). Severe image noise was observed at the upper lung zones and bottoms at 120 and 60 mA without AIDR3D (E and F), which was obviously improved using AIDR3D (B and C).

More »

Figure 2 Expand

Figure 3.

Axial plain chest CT images with a solid lung mass in the right middle lobe (75-year-old female weighing 56 kg).

Images are arranged as in Figures 1 and 2. Spiculae were found on all images, while density heterogeneity inside the mass was severe on images at 60 mA without AIDR3D (F).

More »

Figure 3 Expand

Figure 4.

Axial plain chest CT images showing a ground-glass opacity (GGO) nodule in the left apex (74-year-old female weighing 49 kg).

Images are arranged as in Figures 1 and 2. Nearly homogeneous density of the nodule was accurately depicted on images with AIDR3D at any of three tube currents (A–C). However, on images without AIDR3D (D–F), artificial density heterogeneity due to image noise increased as the tube current decreased from 240 (D) to 60 mA (F).

More »

Figure 4 Expand

Figure 5.

Axial plain chest CT images with a mediastinal setting to assess streak artifacts (55-year-old male weighing 64 kg). A

, at 240 mA with AIDR3D; B, at 120 mA with AIDR3D; C, at 60 mA with AIDR3D; D, at 240 mA without AIDR3D; E, at 120 mA without AIDR3D; F, at 60 mA without AIDR3D. Many radial streaks from the spine were apparent in the heart, particularly on the image without AIDR3D at 60 mA (F). These streaks were greatly reduced using AIDR3D (C, at 60 mA).

More »

Figure 5 Expand

Table 3.

Results of quantitative image noise (SD measurement*) at different tube currents with AIDR3D and FBP (without AIDR3D).

More »

Table 3 Expand

Table 4.

Correlations between quantitative image noise and body weight*.

More »

Table 4 Expand

Table 5.

Results of phantom study at different tube currents with AIDR3D and FBP (without AIDR3D).

More »

Table 5 Expand