Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Table 1.

Inventory of specimens used to evaluate the Ov-16 lateral flow test.

More »

Table 1 Expand

Figure 1.

Signal strength of test and control line as a result of location of sample application.

Diluted Ov-16-positive plasma samples were applied to either the conjugate pad upstream of the dried conjugate (BC), directly onto the conjugate (OC) or on the nitrocellulose upstream of the test and control lines (NC). Line scans show average pixel intensity across the width of the nitrocellulose strip.

More »

Figure 1 Expand

Figure 2.

Schematic and pictures of Ov-16 soft cassette test.

A) top view schematic of the basic Ov-16 lateral flow strip components, with compressed cellulose at the wick end noted (although Ahlstrom 243 absorbent material was also used when specified as standard wick material). B) Side view schematic of the test as assembled in the soft cassette housing, before fluid actuation. C) Side view schematic of a running test shown, with expansion of cellulose resulting in lift of overlay material and blood sample filter. D) Top and side views of the soft cassette housing, with test components labeled on top view.

More »

Figure 2 Expand

Figure 3.

Thickness over time of three compressed hydrophilic materials.

Sample thicknesses were measured after exposure to 37°C and 80% relative humidity over time.

More »

Figure 3 Expand

Figure 4.

Fraction of tests that failed due to blood on membrane after dry.

Tests that had blood on the membranes after the test was dry were counted as failed tests. This failure rate is shown as a function of the volume, blood filter, and wick material. A minimum of 10 tests were run for each condition shown.

More »

Figure 4 Expand

Figure 5.

Example soft cassette tests run with standard wick (Ahlstrom 243) or compressed cellulose wick.

All tests were run with positive-spiked blood samples. Vivid samples shown below were run with 5 µl of blood and Cytosep samples shown were run with 30 µl of blood.

More »

Figure 5 Expand

Table 2.

Performance of the Ov-16 lateral flow test and ELISA on a panel of sera samples against Onchocerca volvulus (Ov) status and Ov-16 enzyme immunoassay (ELISA).

More »

Table 2 Expand

Table 3.

Performance of assay-actuated lateral flow test on spiked whole-blood samples compared to ELISA and the lateral flow test with sera.

More »

Table 3 Expand

Table 4.

Sensitivity and specificity of the lateral flow test at different read times.

More »

Table 4 Expand