Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Figure 1.

Demonstration of the VSOP uptake by the MSC.

Micrographs of MSC after Prussian Blue staining after cultivation with (A) or without (B) iron oxide nanoparticles VSOP incubation (3 mM). The multiple blue dots in (B) represent VSOP particles in the cells.

More »

Figure 1 Expand

Figure 2.

Layers of the agarose phantoms.

Layer structure of group A): 2 phantoms contained layers bearing 1,000, 10,000, and 100,000 labeled MSC, and 2 phantoms comprising layers with 0, 100, and 500 labeled MSC. Layer structure of Group B): One phantom contained a layer configuration bearing 1,000, 10,000, and 100,000 labeled MSC as well as a 4th layer with 100,000 non labeled MSC. The other phantom contained layers with 100, 500, and 5,000 labeled MSC as well as one more layer without MSC. Slices containing no MSC or unlabeled MSC are indicated by white numbers.

More »

Figure 2 Expand

Table 1.

Sequence parameters.

More »

Table 1 Expand

Figure 3.

Demonstration of the fitting of the signal intensity curve to eliminate artificial effects due to sequence, gradient, and field inhomogeneities.

Each value on the x-axis represents one slice of an aquired sequence. The datapoints on the x-axis therefore may vary from sequence to sequence, depending on the amount of aquired slices. The mean signal intensity per slice (as by ROI analysis) is shown for each individual slice on the y-axis, resulting in a curve with a varying resolution, depending on the amount of slices of each sequence. The original data is shown in (A). Negative peaks of signal loss indicate cell-containing layers. A signal inhomogeneity along the B0 axis corresponding to the long axis of the agarose phantom is superimposed, resulting in a drop of signal intensity in the first and last slices of the sequence. To remove this signal inhomogeneity without compromising cell derived peaks for further analysis a curve was fitted to the data after removal of the cell bearing slices (B) and subsequently subtracted from the original data which resulted in a straightened curve, showing the peaks on a linear baseline (C).

More »

Figure 3 Expand

Figure 4.

Representative phantom MR images for evaluation of the detection limit, images of Group A.

Cross sectional images of results from group A at different cell concentrations and with different pulse sequences. The cell concentrations are indicated in the top row, the pulse sequences are indicated on the left according to Table 1. In sequence A1, the sequence with the highest sensitivity, a clear signal loss can bee seen at cell concentrations of 500, whereas the other sequences do not show any signal loss at this concentration. There is also some signal loss at 100 cells. Between the T2* sequence (A5) and a SWI sequence at the same voxel size (A2) no obvious difference can be seen, although the T2* weighted sequence was rated slightly more sensitive than the SWI sequence.

More »

Figure 4 Expand

Figure 5.

Representative phantom MR images for evaluation of the detection limit, images of Group B.

Cross sectional images of results from group B at different cell concentrations and with different pulse sequences. The cell concentrations are indicated in the top row, the pulse sequences are indicated on the left according to Table 1. A diffuse signal loss at a concentration of 500 cells can be seen in B9. In B11 the signal loss at 500 cells is not obvious at inspection. Both sequences were rated sensitive for a concentration of 500 cells by ROI analysis. The higher resolution of B9 can be easily identified. The main differences between these sequences are in-plane resolution and the averages, and therefore the scanning time. Comparing B9 and B1 (8 channel knee coil and 12 channel head coil) B1 appears to be noisier although the imaging parameters remain the same. In some slices an apparently irregular shape of the container can be noted. This is due to small air inclusions at the side of the gel bloc that result in signal voids melting with the black background. The actual form, size and material of the container was the same in all phantoms.

More »

Figure 5 Expand

Table 2.

Results of group A and group B.

More »

Table 2 Expand

Figure 6.

Further analysis of the rater based evaluation.

Images were rated and the Blinded Rater Value (BRV) was established as described in the methods section. Significant differences (p<0.05) are indicated by brackets. The grey area between a BRV of 1.25 and 3.75 indicates BRVs regarded as “probably not detectable” or “probably detectable”. (A) Differences in BRV in relation to slice thickness of the sequence are demonstrated, regardless of other sequence parameters, sequence type or amount of labeled cells. A lower slice thickness resulted in significantly lower BRVs indicating improved detectability of cells. (B) BRVs in relation to the number of labeled cells per layer are shown. Layers with higher cell counts result in significantly lower BRVs. (C) Relationship between BRV and the imaging sequence type (SWI or T2* weighted) is shown. Differences were not statistically significant.

More »

Figure 6 Expand

Figure 7.

Results of the ROI based evaluation of examinations with the 12 channel coil.

Results obtained by using SWI (phantoms B1 to B4) and T2* (phantoms B5 to) sequences are listed separately. According to Figure 3, on the x-axis the individual slices of the phantoms are represented consecutively, with the mean signal intensity for each slice on the y-axis (the actual axes have been removed to make the figure more clear). The dashed lines represent the 0.95 confidence interval, the straight lines represent the mean signal intensity. The amount of labeled MSC per layer is indicated for each graph in numbers below the graph. The graphs represent the changes of the mean signal intensity with each data point representing the mean signal intensity measured in a particular slice. Sequences with a lower slice thickness (i.e. better resolution of data points) depict the signal loss due to labeled MSC better than sequences with a higher slice thickness (i.e. B1 vs. B8). A similar slice thickness leads to a relatively great similarity of the performance of T2* and SWI sequences (i.e. B1 vs. B5). These graphical results are also noted quantitatively in Table 2. There are positive and negative peaks present in regions below the 100 cells layer (i.e. B2 or B5). These are artificial peaks due to inhomogeneities of the phantoms in the 0 cell layer and in the 100,000 non labeled MSC layers.

More »

Figure 7 Expand

Figure 8.

Results of the ROI based evaluation of examinations with the 8 channel coil.

As in Figure 7 results obtained by using SWI (phantoms B9 to B12) and T2* (phantoms B13 to B16) sequences are listed separately. The form of the graphs is the same as in figure 7. Again sequences with a lower slice thickness (i.e. better resolution of data points) depict the signal loss due to labeled MSC better than sequences with a greater slice thickness (i.e. B11 vs. B16). With the knee coil, significant peaks at layers containing 500 labeled MSC could be identified (i.e. B9, see also table 2).

More »

Figure 8 Expand

Figure 9.

Signal intensity versus artifact-index (SivA).

Statistically significant differences (p<0.05) are indicated by brackets above the data points. Larger amounts of cells per layer generate a higher SIvA indicating a better differentiation between signal drops induced by artifacts and signal drops induced by labeled cells.

More »

Figure 9 Expand