Table 1.

Panel of clinical samples used to assess the test characteristics of the RDT products.

More »

Table 1 Expand

Table 2.

Overview of RDT products and their lot numbers.

More »

Table 2 Expand

Figure 1.

Received product of Unitest (Ciriano global S.L., Zaragoza, Spain).

The delivered Unitest malaria cassette P.f-P.v kit mentions detection of antibodies while the almost identical kit on the website mentions detection of the Pf and Pv antigen. (http://www.clinica.co.za/index.php?page=shop.product_details&flypage=flypage.tpl&product_id=11&category_id=3&option=com_virtuemart&Itemid=90).

More »

Figure 1 Expand

Table 3.

Test characteristics of the different RDT products.

More »

Table 3 Expand

Table 4.

Presence of important items that need to be addressed in the instructions for use.

More »

Table 4 Expand

Figure 2.

Lancets and transfer devices delivered with the different RDT products.

CareStart and OptiMAL included a simple lancet, SDFK63 a safety seal lancet and the other products safety lancets with a retractable needle. The systems of TODA and Immunoquick (SMI) do not require a transfer device (direct contact of the test strip with the drop of blood). Sanitoets contained a calibrated pipette and OptiMAL a straw pipette. The other products included a balloon pipette. The transfer devices of OneStep and Labstix did not display a volume mark.

More »

Figure 2 Expand