Evaluation of peripheral intravenous line securement devices under clinically relevant loading and perspiratory conditions
Fig 5
Box and whisker plots of peak forces during pull-out (Stage 1) testing for each of the four devices tested: micropore tape, Grip-Lok, Javelo, and Medow. The results of a Friedman test with post-hoc Wilcoxon signed-rank tests with a Benjamini-Hochberg correction procedure are displayed. n = 10. ***p-adj < 0.001; ns = no significance.