A sound coding strategy based on a temporal masking model for cochlear implants
Fig 9
Comparison between the patient performance in the clinical setting (measured with the HSM sentence test at 10 dB SNR during the yearly control visit, Table 1) and the benefit from using TPACE with a T½ of 0.5 ms.
The results of the first and second study are presented in green and blue, respectively. Regression between the clinical performance of the subject and the benefit from temporal masking was statistically significant for both studies (R² and p value are shown with the respective color in the plot).