Modified full-face snorkel masks as reusable personal protective equipment for hospital personnel

Fig 5

Exhalation valve performance testing.

a) Exhalation valve on mask. b) Setup for measuring the leak flow of the chin valve of a Decathlon FreeBreath mask, showing the connection to an 8mm tube. c) Connection to an open water tank—system at equilibrium (chin valve opened). The mask is lifted up (chin valve closed) to create a water column of 30mm, which means that a negative pressure of 30mmH2O is applied to the chin valve. d) an example time series curve of water column height in experiments described in c). e) exhalation valve leakage test results of 2 valves from Dolfino Frontier and 5 valves from Decathlon, compared with the NIOSH standard as designated in red square. All valves passed the NIOSH standard. f) Flow rate versus suction pressure before use (dry valve, as per NIOSH standards) and after use. This test was specifically to test if valve performance suffers in the humid environments of user breathing. To the contrary, evidence suggests that moisture tends to improve valve performance, as shown in this data. g) Schematic of a circular chin valve. The force balance determines the time scale of chin valve closure. h) Time-lapse images of the chin valve opening and closing.

Fig 5

doi: https://doi.org/10.1371/journal.pone.0244422.g005