Reliability and performance of the IRRAflow® system for intracranial lavage and evacuation of hematomas—A technical note

  • Mette Haldrup ,

    Contributed equally to this work with: Mette Haldrup, Mojtaba Nazari, Naveed Ur Rehman, Anders Rosendal Korshoej

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Validation, Visualization, Writing – original draft, Writing – review & editing

    methalje@rm.dk

    Affiliation Department of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark

  • Mojtaba Nazari ,

    Contributed equally to this work with: Mette Haldrup, Mojtaba Nazari, Naveed Ur Rehman, Anders Rosendal Korshoej

    Roles Conceptualization, Formal analysis, Investigation, Methodology, Software

    Affiliation Department of Engineering, Electrical and Computer Engineering, Aarhus University, Aarhus N, Denmark

  • Chenghao Gu ,

    Roles Data curation, Investigation, Writing – review & editing

    ‡ CG, MR, SD, CZS, MG and FRP also contributed equally to this work.

    Affiliation Department of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark

  • Mads Rasmussen ,

    Roles Conceptualization, Investigation, Supervision, Writing – review & editing

    ‡ CG, MR, SD, CZS, MG and FRP also contributed equally to this work.

    Affiliations Department of Clinical Medicine, Aarhus University, Aarhus N, Denmark, Department of Anesthesiology, Section of Neuroanesthesia, Aarhus University Hospital, Aarhus, Denmark

  • Stig Dyrskog ,

    Roles Supervision, Writing – review & editing

    ‡ CG, MR, SD, CZS, MG and FRP also contributed equally to this work.

    Affiliation Department of Intensive Care, Aarhus University Hospital, Aarhus N, Denmark

  • Claus Ziegler Simonsen ,

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    ‡ CG, MR, SD, CZS, MG and FRP also contributed equally to this work.

    Affiliations Department of Intensive Care, Aarhus University Hospital, Aarhus N, Denmark, Department of Neurology, Aarhus University Hospital, Aarhus N, Denmark

  • Mads Grønhøj ,

    Roles Writing – review & editing

    ‡ CG, MR, SD, CZS, MG and FRP also contributed equally to this work.

    Affiliation Department of Neurosurgery, Odense University Hospital, Odense, Denmark

  • Frantz Rom Poulsen ,

    Roles Supervision, Writing – review & editing

    ‡ CG, MR, SD, CZS, MG and FRP also contributed equally to this work.

    Affiliation Department of Neurosurgery, Odense University Hospital, Odense, Denmark

  • Naveed Ur Rehman ,

    Contributed equally to this work with: Mette Haldrup, Mojtaba Nazari, Naveed Ur Rehman, Anders Rosendal Korshoej

    Roles Conceptualization, Investigation, Methodology, Software, Supervision, Writing – review & editing

    Affiliation Department of Engineering, Electrical and Computer Engineering, Aarhus University, Aarhus N, Denmark

  • Anders Rosendal Korshoej

    Contributed equally to this work with: Mette Haldrup, Mojtaba Nazari, Naveed Ur Rehman, Anders Rosendal Korshoej

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Validation, Writing – original draft, Writing – review & editing

    Affiliations Department of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark, Department of Clinical Medicine, Aarhus University, Aarhus N, Denmark

Abstract

Background

Intraventricular hemorrhage (IVH) is a severe condition with poor outcomes and high mortality. IRRAflow® (IRRAS AB) is a new technology introduced to accelerate IVH clearance by minimally invasive wash-out. The IRRAflow® system performs active and controlled intracranial irrigation and aspiration with physiological saline, while simultaneously monitoring and maintaining a stable intracranial pressure (ICP). We addressed important aspects of the device implementation and intracranial lavage.

Method

To allow versatile investigation of multiple device parameters, we designed an ex vivo lab setup. We evaluated 1) compatibility between the IRRAflow® catheter and the Silverline f10 bolt (Spiegelberg), 2) the physiological and hydrodynamic effects of varying the IRRAflow® settings, 3) the accuracy of the IRRAflow® injection volumes, and 4) the reliability of the internal ICP monitor of the IRRAflow®.

Results

The IRRAflow® catheter was not compatible with Silverline bolt fixation, which was associated with leakage and obstruction. Design space exploration of IRRAflow® settings revealed that appropriate settings included irrigation rate 20 ml/h with a drainage bag height at 0 cm, irrigation rate 90 ml/h with a drainage bag height at 19 cm and irrigation rate 180 ml/h with a drainage bag height at 29 cm. We found the injection volume performed by the IRRAflow® to be stable and reliable, while the internal ICP monitor was compromised in several ways. We observed a significant mean drift difference of 3.16 mmHg (variance 0.4, p = 0.05) over a 24-hour test period with a mean 24-hour drift of 3.66 mmHg (variance 0.28) in the pressures measured by the IRRAflow® compared to 0.5 mmHg (variance 1.12) in the Raumedic measured pressures.

Conclusion

Bolting of the IRRAflow® catheter using the Medtronic Silverline® bolt is not recommendable. Increased irrigation rates are recommendable followed by a decrease in drainage bag level. ICP measurement using the IRRAflow® device was unreliable and should be accompanied by a control ICP monitor device in clinical settings.

Introduction

Intraventricular hemorrhage (IVH) is a severe condition with poor outcomes and high mortality rates [1, 2]. Currently, IVH treatment relies on passive drainage of cerebral spinal fluid (CSF) with external ventricular drainage (EVD) to decrease the intracranial pressure (ICP) and facilitate hematoma evacuation3,4. Recent studies have shown that clot removal could be accelerated using intraventricular fibrinolysis (IVF) leading to improved survival and functional outcomes [1, 2]. However, no significant treatment advances have been made in the past decades, and minimally invasive techniques to achieve better management are warranted.

A new technology, IRRAflow® (IRRAS AB), was introduced to accelerate IVH evacuation by minimally invasive wash-out. The IRRAflow® system performs active and controlled intracranial irrigation and aspiration with physiological saline while monitoring and maintaining a stable ICP. The IRRAflow® system is FDA-approved (2019) and CE-marked (2018) for the treatment of IVH and has currently been used to treat more than 600 patients worldwide (postmarket surveillance data; J. Unser, MBA, IRRAS AB, written communication, January 2023). Currently, there are only two published case reports evaluating IRRAflow® in the treatment of IVH (DRIFT and Hummingflow [3, 4] and four randomized controlled trials and one case-control study have presently been initiated to test IRRAflow® treatment of IVH in different settings (registered trials: ARCH (NCT05118997), ACTIVE (NCT05204849) [5]).

The IRRAflow® control unit has several adjustable settings, including irrigation rate, drain valve opening pressure, bolus volume, and drainage bag level. However, there are no current recommendations for best-practice treatment settings, and no studies have investigated the physiological effects of varying these parameters. IRRAflow® uses an internal ICP sensor connected to the catheter irrigation line to exert supervisory control over the system’s operational features, including irrigation and drainage. Documentation of sensor reliability is therefore crucial for safe device operation. To our knowledge, no former studies have evaluated these aspects of the IRRAflow® system to ensure patient safety and validate grounds for clinical decision-making using the technology.

In this paper, we addressed these gaps in the current knowledge on IRRAflow® implementation and IVH lavage using an ex vivo model and data from patients treated in a randomized, controlled safety/feasibility study of IRRAflow® performance (Active Study [5]). Specifically, we 1) conducted an extensive exploration of the physiological impact of different settings of the IRRAflow® control unit (design space exploration); 2) evaluated and described the data output from IRRAflow®; 3) conducted extensive validation and reliability tests of IRRAflow® volume injections and ICP readouts. Our experiments shed important light on the physiological effects and feasibility of different treatment implementations and propose a basis for clinical guidelines in the treatment of IVH using IRRAflow®.

Materials and methods

The IRRAflow® device

IRRAflow® is an advanced drainage system that combines gravity-driven drainage with periodic irrigation and ICP monitoring. The system operates in cycles beginning with a 1-second irrigation phase, followed by a 9-second ICP measurement phase, and ends with an aspiration phase with variable duration (10 to 170 seconds) defined by the user (Fig 1).

thumbnail
Fig 1. Lab setup and pressure signals.

1. Saline bag, 2. IRRAflow® device, 3. Bolt-connected dual lumen IRRAflow® catheter, 4. Raumedic device for ICP monitoring, 5. Raumedic ICP monitor probe, 6. Acrylic chamber, 7. Valves, 8. Drainage bag, 9. schematic illustration of the proximal pressure sensing in the internal IRRAflow® ICP monitor showing a spiked CPRS signal from IRRAflow® compared to a synchronized and stable ICP signal obtained using the Raumedic probe at the distal catheter tip. The IRRAflow® current pressure signal (CPRS, red solid line), the IRRAflow® ICP signal (gray solid line), and the Raumedic ICP signal (blue solid line). The CPRS shows a rapid pressure increase during the first second irrigation phase followed by a stabilizing decrease during the 9s ICP measurement phase, and finally a further decrease during the aspiration phase. The CPRS has a lower sampling rate (red dots) and a lower signal to noise ratio (SNR) compared to the Raumedic ICP signal. The IRRAflow® ICP curve represents a single point value in time, sampled from the final CPRS value at the end of the measurement phase, when the CPRS signal has settled. This value is then maintained at a constant until the end of the next measurement phase. The RLC circuit is shown above the irrigation line (R, resistance; L, inductance; C, compliance). Furthermore, the cycle operations of the IRRAflow® are illustrated.

https://doi.org/10.1371/journal.pone.0297131.g001

The system uses a peristaltic pump to deliver 0.5 or 1 ml of saline infusion over 1 second during the irrigation phase. Users can set the cumulative irrigation rates between 20 ml/hour and 180 ml/hour depending on the duration of drainage and hence the frequency of bolus infusions.

IRRAflow® monitors the ICP using pressure transducers located in the cassette of the device, coupled to the tip of the catheter via the irrigation tubing of the tube set. The sensor measures ICP as a hydrostatic pressure, so it must be placed in a liquid environment to function properly. The pressure measured at the pressure transducer is equivalent to the pressure at the catheter tip when the irrigation line is fluid-filled, unobstructed, the pressure transducer is positioned at the same height as the tip of the catheter, and the liquid is still and not in motion. The decoupling of the pressure transducer from the tip of the catheter allows for the pressure transducer signal to be re-zeroed in situ. An ICP value is output to the user at the end of the 9s Measurement Phase, and this displayed average pressure is displayed throughout the Drain and Irrigation Phase until updated after the next Measurement Phase (Fig 1). The intrinsic ICP readout during the Monitoring Phase serves as supervisory control of drainage and irrigation. The measured pressure is then compared to the Drain Above Setting. If the measured pressure is lower than the Drain Above Setting, the IRRAflow System remains in Measurement Mode until the measured pressure exceeds the Drain Above Setting. If the measured pressure is higher than the Drain Above Setting, the control unit enters Drainage Mode. The IRRAflow System monitors the ICP signal throughout the entire IRRAflow cycle and is identified as the current pressure signal (CPRS), which can be displayed upon user prompt.

The Drainage Phase can vary between 10 and 170 seconds, depending on the user-set irrigation bolus volume (0.5 or 1 ml of saline infusion over approximately 1 second) and the user-set irrigation flow rate, which range from 20 ml/hour to 180 ml/hour. IRRAflow controls drainage via a solenoid pinch-type valve within the cassette of the device. Drainage is controlled by the user via the “Drain above” setting. When the measured pressure during the Measurement Phase exceeds the Drain Above Setting, the system enters the Drainage Phase, the solenoid pinch valve opens and liquid is allowed to drain. If the measured pressure is below the Drain Above Setting, the system enters and remains in the Measurement Phase (no drainage and no irrigation occur) until the measured pressure is greater than the Drain Above Setting.

Like a standard EVD, IRRAflow® drainage is gravity-driven. However, with a standard EVD, the user adjusts the level of the drainage line relative to the catheter tip to define the pressure gradient driving the aspiration. With the IRRAflow® the drainage is driven by the gradient between the ICP and the level of the drainage bag, when the ICP is within the operational range, which can be set from -15 cmH2O to -103 cmH2O below the catheter tip. This results in suction of CSF during drainage, with gradients ranging from 0 to -75 mmHg.

The lab setup

To allow versatile investigation of multiple device parameters, we designed an ex vivo lab setup replicating the hydrodynamic conditions of the intracranial space (Fig 1). We used a stiff 10x10x7 cm acrylic box with a 10 mm wall thickness. We filled it with 200 ml of room-temperature physiological saline (viscosity: 0.7 mPa) and 500 ml of atmospheric air at 23 degrees Celsius to attain physiological system compliance (ΔV/ΔP) of 0.60–0.70 ml/mmHg [6]. The temperature of the fluid was controlled by the Raumedic PtO2 monitor and a thermostat in the lab. Two airtight manual valves with 10 mm circular openings provided interior chamber access. To establish access points for ventricular drains, pressure sensors, and syringes, we fixed two catheter bolts (Silverline®, Spiegelberg GmbH & Co), a Raumedic Neurovent-P bolt, and a female luer-lock at 1 cm height from the chamber bottom.

We inserted a dual-lumen IRRAflow® catheter (version 2.0 ICGS020) 50 mm deep through the Silverline bolt and the catheter tip was placed precisely in the hight of the “0” reference on the IRRAflow device. The IRRAflow® cassette was connected to the IRRAflow® control unit (IRRAS, version 3.0). Tubing and saline bag attachment were prepared following the product guidelines. We inserted a Raumedic Neurovent-P pressure sensor at the height of the IRRAflow® catheter tip to continuously monitor the chamber pressure equivalent to the ICP (Fig 1). The Neurovent-P sensor and the IRRAflow® device were calibrated and zeroed before all experiments. We measured the saline input and output on a scale before each experiment to accurately monitor the amount of irrigated saline. We set the IRRAflow® alarm limits to the maximum range and the Drain Above setting to -99 mmHg to maintain device operation under extreme conditions (S2 File).

Patient data

The in vivo data set for this study was collected prospectively from patients who were a part of the Active study, a randomized controlled phase 2 study testing the safety and feasibility of IRRAflow®. Patient recruitment occurred from January 2022 to November 2022. The data was obtained from patients with severe spontaneous primary or secondary IVH treated with IRRAflow®. Patients were followed for 3 month and the authors had full written permission to access patients information. All patients had a Raumedic Neurovent-P monitor placed in the same hemisphere as the IRRAflow® catheter to obtain correlated datasets. Ethical approval was obtained from the Central Denmark Committee on Health Research Ethics (journal number 1-10-72-34-21) and participant consent was collected from legal authorized representatives for all patients in accordance with relevant Danish legislation and the Helsinki Declaration.

Study design and hypotheses

Experiment 1—Bolt compression cap tightness test.

In this experiment, we aimed to evaluate the compatibility between the IRRAflow® catheter and the Silverline f10 bolt from Spiegelberg GmbH & Co as the IRRAflow is placed using a tunneling technique (S1 File). We tested the following hypotheses:

  1. Leakage would occur when the bolt was only loosely tightened around the IRRAflow® catheter.
  2. The IRRAflow® catheter would be increasingly constricted with an increasing number of bolt revolutions.

This was important to assess the clinical feasibility of implementing bolt mounting of IRRAflow® and to determine the appropriate settings of the ex vivo setup. To test these hypotheses, we performed five IRRAflow® irrigation tests of 1-hour duration for at different bolt revolutions with and without the catheter insertion stylet in situ. We evaluated the IRRAflow® irrigation performance and the integrity of the catheter fixation to determine the optimal settings, ideally preventing obstruction and leakage.

Experiment 2—IRRAflow® design space exploration.

In this experiment, we aimed to evaluate the physiological and hydrodynamic effects of varying the IRRAflow® settings. We conducted an extensive design space exploration with comprehensive variations of the following device parameters in different combinations which were the settings used in the Active study:

  1. Irrigation rates: 20 ml/h, 90 ml/h, and 180 ml/h.
  2. Drainage bag levels: +39 cm, 0 cm, -19 cm, -29 cm, -39 cm and -49 cm.

For each test, we performed five repetitions of 10-minute duration. In addition, the Drain Above parameter was set to -99 and alarm limits to -99 to 99 to ensure continuous device operation throughout the experiments.

Experiment 3—Accuracy of the IRRAflow® injection volume.

This experiment aimed to assess the accuracy and reliability of saline volumes injected by the IRRAflow® system. We hypothesized that increasing pressure within the chamber would compromise the internal saline pump, resulting in a reduced injection volume. To test this, we conducted repeated consecutive 1 ml bolus infusion cycles in a closed system, gradually increasing chamber pressure. Our observations utilized a scale measuring the injected volume in grams. IRRAflow® infusions were stopped when chamber pressure exceeded the upper critical limit of 99 mmHg. We repeated the experiment with different baseline chamber volumes (100–500 ml) to ensure stable results and repeated the tests 5 times at each baseline volume.

Experiment 4—Accuracy and reliability of the IRRAflow® pressure sensor.

In this experiment, we aimed to evaluate the accuracy and reliability of the internal IRRAflow® ICP monitor by evaluating the current pressure signal (CPRS) output of IRRAflow® in two different ways:

  1. Assessment of the CPRS during the irrigation phase of the IRRAflow® operation cycle compared to the CPRS during manual injections using a syringe. A Raumedic device was used as a control unit.
  2. Assessment of the CPRS during the aspiration cycle and comparison of the pressure curves with the Raumedic-derived pressure signals.

Furthermore, we measured deviations between the IRRAflow® ICP signals compared to the Raumedic ICP signals in both the ex vivo setting and in real patient data. The real patient data were extracted in a time-synchronized manner for two patients and in an unsynchronized manner for two additional patients. The null hypothesis was that there was no difference in the pressure values between the two devices.

To evaluate sensor stability, we conducted five drift tests of 24-hour duration for both IRRAflow® and the Raumedic ICP sensors. These tests were performed with different baseline chamber pressures (6, 8, and 10 mmHg).

Statistical methods

Statistical analyses were carried out using MATLAB, MathWorks. Drainage, irrigation, and chamber pressures were presented as means (+/- standard error (SE)) in all experiments. For experiments including in vivo data the distributions of the unsynchronized deviations were quantified for each patient using a two-sample t-distribution test. The drift data was analyzed using a Gardner-Altman plot and further, a histogram made from which mean drift, variance, and a corresponding correlating p-value between the data from the two devices was calculated.

Results

IRRAflow® data output

The IRRAflow® data output was extracted using the data export function. The data structure comprised pressure curves and a variety of values (i.e., CPRS, System and treatment status, ICP, upper and lower alarm limits, drain above setting, infusion volume (ml/hour), cycle time (sec), etc.) (S3 File). The pressure readouts consisted of two variables: 1) the ICP signal, which is a constant value sampled at the end of the ICP measurement phase and maintained throughout the remaining cycle, and 2) the current pressure signal (CPRS), which gives the real-time ICP signal throughout the entire cycle (Figs 1 and 2).

thumbnail
Fig 2.

(A) Illustrates the pressure in the chamber under different bolt revolutions. ICP curves for 6, 7, 8 and stylet guided (Guide Wire, GW) revolutions on the bolt are shown. (B) Schematic figure showing the difference between the metal enforced Silverline® catheter (left) and the IRRAflow® dual lumen catheter (right). No metal reinforcement resulted in gradual compression and obstruction of the IRRAflow® catheter.

https://doi.org/10.1371/journal.pone.0297131.g002

Our data showed that the IRRAflow® CPRS was obtained at varying sampling rates between 0.2 and 10 Hz. For example, during 1 second of irrigation, the sampling rate varied between 5 and 10 Hz. During 9 seconds of ICP measurement in the aspiration phase, the sampling rate was 0.2 Hz (Fig 1). The IRRAflow® pressure signal was obtained at a much lower sampling rate compared to the Raumedic device, which was sampled at 100 Hz.

Experiment 1 –Bolt compression cap tightness test

This experiment was conducted to evaluate the performance of the catheter system at different bolt revolution settings. Results showed that at 6 revolutions or below, the catheter system did not exhibit constriction, and the integrity of the irrigation and drainage lines was preserved. However, we observed that the catheter was not properly fixed, retractable, and leaked during irrigation (Table 1, Fig 2).

thumbnail
Table 1. Experiment 1 –Bolt compression cap tightness test.

Mean irrigation and drainage for the bolt constriction test.

https://doi.org/10.1371/journal.pone.0297131.t001

At 7 and 8 revolutions, we did not observe any leakage. However, at 7 revolutions, we observed constriction of the drainage line resulting in a progressive increase in chamber pressure, which reached 20 mmHg after 5 min and 30 sec and 99 mmHg after 28 minutes, indicating preserved irrigation performance but compromised drainage (Table 1, Fig 2). At 8 revolutions, both irrigation and drainage lines were fully constricted. The chamber pressure was reliably 0 mmHg according to the Raumedic sensor. However, the IRRAflow® sensor simultaneously measured 300 mmHg, indicating significantly compromised and unreliable pressure sensing (Table 1, Fig 2).

To alleviate the constriction, we attempted to secure the bolt around the catheter with the stylet (guidewire) in place and retracting the stylet before initiating irrigation. However, this option was not clinically feasible as the ICP reached 99 mmHg after 20 minutes. Therefore, the remainder of experiments was reliably performed at 6.5 revolutions, at which setting the chamber was tight, with no leakage, while irrigation/drainage was preserved, although the catheter was still easily retractable.

Experiment 2—IRRAflow® design space exploration

The results of the design space exploration are shown in Fig 3.

thumbnail
Fig 3. IRRAflow® settings and physiological implications.

Pressure signals obtained at different IRRAflow® combinations of irrigation rate and drainage bag levels (A. 20/hour and 19 cm, 29cm and 39 cm, B. 90 ml/h and 19cm, 29cm and 39 cm and C. 180 ml/h and 19cm, 29cm and 39 cm (blue: Raumedic pressure signal, Red: IRRAflow® pressure signal). Error signals showing the pressure difference between the IRRAflow® and Raumedic pressure readouts are shown on the right side of each panel along with tables of the corresponding means and variances error in mmHg. The time required to achieve a stable chamber pressure varied with different settings and balanced fluid exchange required adjustment of drain bag to accommodate the irrigation rate.

https://doi.org/10.1371/journal.pone.0297131.g003

At a low irrigation (20 ml/hour), the optimal drainage bag level was 0 cm, resulting in nearly balanced fluid exchange with a mean net drainage of 5.7 ml (SE ± 1.77) and a mean chamber pressure of -1.78 mmHg (SE ± 0.23) over the course of 10 minutes. Conversely, over-drainage was observed at bag heights of -19 cm, -29 cm, and -39 cm.

At 90 ml/hour irrigation positioning of the drainage bag at 0 cm or -19 cm resulted in mean net over drainage of 3.85 ml (SE ±1.7) and 3.04 ml (SE ±2.1), respectively, corresponding to end chamber pressures of -0.7 (SE ± 0.19) and 4.8 mmHg (0.23), respectively (Fig 3).

At 180 ml/hour irrigation, the optimal drainage bag level was -29 cm, with a mean net over drainage of 2.5 ml (SE± 1.1) resulting in a slightly negative steady state end mean pressure of -6 mmHg (SE ± 0.39). At a drain bag height of -19 cm, slight underdrainage was observed (0.8 ml (SE ± 0.11) mean saline accumulation), resulting in a mean pressure increase of 6 mmHg (SE ± 1.98) (Fig 3). The mean difference and variance between the ICP signals measured by IRRAflow® and the Raumedic devices is shown in Fig 3 and in S1 Fig. Overall, we found that the mean difference between the two signals varied between 0.89–17.07 mmHg (mean variance error 0.59–9.73 mmHg). This observation was mainly a result of under-sampling of the IRRAflow® ICP, which represented a single sample ICP obtained at the end of the ICP measurement phase which was then maintained throughout the drainage phase. The device therefore failed to capture and report the subsequent negative pressure that arose during the drainage phase.

Experiment 3—Accuracy of the IRRAflow® injection volume

The results from the IRRAflow® injection volume tests are shown in S1 Fig. Bolus injections performed by the IRRAflow® device were stable and reliable in terms of volume at various chamber pressures. This reliability was validated by a stable pressure increase for each bolus and further a stable average mass of saline injected with each bolus (1.0061g/bolus, SE ± 0.059) compared with the anticipated mass of 1 g/bolus (equivalent to 1 ml/bolus as reported by the device 1000) (S1 Fig and S4 File).

Experiment 4—Accuracy and reliability of the IRRAflow® pressure sensor

During the irrigation phase, we observed spikes in the CPRS signal ex vivo with high amplitudes of up to 200 mmHg (Fig 4).

thumbnail
Fig 4. Reliability of the IRRAflow® pressure readout.

Panel A shows spikes observed in the current pressure signal (CPRS) during the irrigation phase when the saline injections are performed by IRRAflow® as bolus injections (red). In comparison no spikes were observed following manual bolus injections using a syringe (blue). Panel B shows the negative pressure values during the aspiration phase. The blue line represents the Raumedic obtained signal and the red line represents the IRRAflow® obtained signal. Panel C shows a representation of the IRRAflow® (CPRS and ICP) and Raumedic pressure signals, respectively, obtained from the ex vivo model Infusion results in a clear pressure increase of 200 mmHg, which is consistent across all signals. In comparison, panel D shows similar data obtained from a patient. Contrary to the ex vivo data in panel C, we observed no increase in pressure in the ICP upon bolus infusion. This was thought to be partly due to the low signal to noise ratio (SNR), given the ICP variations due to the heart beat, but also likely related to a lack of pressure transmission from the catheter tip to the Raumedic device. Panel D shows that no negative pressures are observed in the Raumedic data (blue) during the aspiration phase as is observed in the IRRAflow® data (red line during aspiration phase).

https://doi.org/10.1371/journal.pone.0297131.g004

These values were approximately 15 times higher than the corresponding ICP value measured using the Raumedic device (Fig 4). The CPRS then decayed exponentially throughout the subsequent 9 seconds ICP measurement phase to reach a nearly stable plateau at the end. However, during the decay phase, the signal was only sampled twice. Contrary to the IRRAflow® CPRS, the Raumedic ICP signal did not exhibit spike features following saline injection with IRRAflow®. Rather, we observed an expected pressure increase (8 mmHg, SE ± 4.2) for all three pressure signals reaching a stable plateau, which was maintained throughout the rest of the ICP measurement phase (Fig 4C). We observed no spikes in the CPRS or IRRAflow® signals during manual 1 ml bolus injection using a syringe indicating that the spikes were an irrigation artifact inherent to the IRRAflow® system (Fig 4).

During the aspiration phase of the IRRAflow® cycle, we observed a decay of the CPRS and Raumedic signals ex vivo to reach negative values between -7 and -13 mmHg (Fig 4), reflecting the active aspiration of liquid out of the chamber during this phase (Fig 4B). A similar pressure decay in the CPRS signal was observed in the patient data, although it was not transmitted to the Raumedic ICP sensor (Fig 4D). This was thought to be partly due to the low signal to noise ratio (SNR), given the ICP variations secondary to the heartbeat, but also likely related to a lack of pressure transmission from the catheter tip to the Raumedic device.

In all experiments, we observed discrepancies between the IRRAflow® CPRS signal and the corresponding Raumedic ICP signal. In the idealized ex vivo setting, the mean deviation was 3.8 mmHg (SE ±2.6) (Fig 5), while the synchronized data set retrieved from two real patients treated with IRRAflow® had periodical deviations with IRRAflow® ICP values up to 10–15 mmHg higher than values measured using Raumedic (mean 12.0 mmHg, SE ± 3.6 mmHg).

thumbnail
Fig 5. Discrepancies in the IRRAflow® vs Raumedic obtained pressure signals.

A. Synchronized patient data. The upper curve shows the signal recorded during 6 hours from patient 3 (red: IRRAflow® and blue: Raumedic) and below the corresponding histogram of the synchronized ICP signals from patient 3 and 4. B shows the difference between the two signals (mean difference 4.5 mmHg and variance 16.4 mmHg) and the corresponding histogram of the difference in ICP signals (mean 3.8 mmHg, SE 2.6, p<0.001). C. shows histograms of unsynchronised ICP signals from patient 1 and 2 (mean difference 6.31 mmHg, variance 28.6, p<0.0001).

https://doi.org/10.1371/journal.pone.0297131.g005

In addition, the unsynchronized dataset revealed a significant difference in the ICP values comparing the IRRAflow® ICP values to the Raumedic ICP values (patient 1: mean difference 6.31 mmHg, variance 28.6, p<0.0001 and patient 2: mean difference 1.41 mmHg, variance 8.29, p<0.0001, normally distributed) (Fig 5) with the IRRAflow® values higher than Raumedic values. Importantly, the error was periodically very high, with significant deviations up to 19 mmHg between the two devices. These periodical errors in ICP measurement were often related to catheter occlusions if the catheter tip was placed in blood or had displaced into the brain parenchyma and not surrounded by liquid.

Furthermore, we observed a significant mean drift difference of 3.16 mmHg (variance 0.4 mmHg, p = 0.05) over a 24-hour test period with a mean 24-hour drift of 3.66 mmHg (Var 0.28 mmHg) in the IRRAflow® measured pressures compared to 0.5 mmHg (variance 1.12 mmHg) in the Raumedic measured pressures (Fig 6).

thumbnail
Fig 6. 24 hour drift tests.

Panel A shows a histogram of the recorded drift data from the Raumedic (blue) and the IRRAflow® (red). Panel B shows the Gardner-Altman plot and the mean difference of the recorded signals. Panel C shows the 24 hours drift test recording and below the difference is shown.

https://doi.org/10.1371/journal.pone.0297131.g006

Lastly, we observed a low signal to noise ratio (SNR). During 1 second of injection in the irrigation phase the frequency of the signal was found to be overlapping with the heartbeat in the in vivo data. In the ex vivo model, with the absence of heartbeat, changes in pressure during 1 second of irrigation with 1 ml of saline were easily observed (Fig 4C). The same change in pressure was not observed in the patient data (Fig 4B).

Discussion

IRRAflow® is a new and potentially promising technology for automated and controlled irrigation/aspiration treatment of IVH using tunneled dual-lumen silicone catheters. Currently, there are no clinical guidelines describing how to implement and set up the IRRAflow® device for approved indications, including IVH. In this study, we conducted a comprehensive assessment of the implications and effects of different IRRAflow® settings as well as an evaluation of the reliability of IRRAflow® volume injections and pressure measurements. Our study was based on an idealized ex vivo setup as well as patient data and led to four overall recommendations:

Firstly, we do not recommend the IRRAflow catheters to be placed on a Silverline bolt. Previous studies have established that bolt fixation of external ventricular catheters is associated with reduced risk of catheter displacement, CSF leakage and CNS infection. For that reason, it is potentially compelling to use conventional and approved bolt EVD technologies to fixate the IRRAflow® catheter during treatment. However, our study showed that this approach is indeed not safe or feasible, when combining the IRRAflow® catheter (version 2.0 ICGS020) with the widely used Spiegelberg Silverline® bolt, as this was associated with improper fixation, leakage, or catheter constriction. This combination represents a patient hazard and should be avoided.

Secondly, caution should be taken when deciding on device settings to ensure no over-drainage when using the IRRAflow system. We found that higher irrigation rates should be followed by lower drain bag heights to create a higher negative aspiration gradient during the shortened drainage period, to achieve a balanced saline influx and efflux.

For clinical implementation, the drain bag height could be adjusted to achieve a stable ICP in the desired range for any given irrigation rate. Elevated drainage bag heights may result in net intracranial CSF accumulation and corresponding undesired ICP elevation. Similarly, very low drainage bag levels may result in severe over-drainage. Low drainage bag levels potentially result in very negative pressure gradients during the aspiration phase. This may be associated with an increased risk of undesirable aspiration of ependyma, or blood clots followed by catheter occlusions, and we recommend to avoid very negative drainage bag levels.

Thirdly, we recommend that an additional parenchymal ICP sensor is used in conjunction with IRRAflow® to ensure stable and reliable pressure sensing during irrigation and for future systems, we recommend that the pressure sensor be integrated into the catheter tip to avoid artifacts and more accurately represent ICP continuously. These recommendations are based on the findings that the IRRAflow® pressure sensor is located proximally to the catheter tip and found to be prone to significant spike artifacts resulting from saline infusion and subsequent pressure stabilization. The sensor requires steady state conditions with no liquid movements in the catheter to be accurate. This represents a significant problem as the IRRAflow® operational cycle is governed and supervised by this internal pressure monitor. False pressure readouts may therefore result in unsafe device operation or frequent alarms.

Furthermore, the extent of sensor drift was significantly higher for IRRAflow compared to Raumedic, which may indicate a need for regular and frequent calibration. The American National Standard Institute (ANSI)/the Association for the Advancement of Medical Instruments (AAMI) has recommended that ICP monitors should be reliable within the ICP range of 0–100 mmHg. Especially in the ICP range of 0–20 mmHg, the drift should be within ± 2 mmHg [7].

In the ex vivo tests the Raumedic device was chosen as the control ICP monitor as this monitor was used in the clinical trial evaluating the IRRAflow where placement of an additional catheter for ICP measurement would be a more invasively approach compared to the placement of only the Raumedic probe even though the ICP measurement technologies would have been more comparable if for instance a Silverline catheter was used for control ICP measurement.

Conclusion

Near-balanced irrigation can be achieved by adjusting the IRRAflow® drainage bag to accommodate different irrigation rates. It is not possible to provide a clear guideline to define the appropriate settings. internal IRRAflow® ICP monitor is based on an unfavorable proximal pressure-sensor technology and is unreliable compared to the Raumedic. We recommend that use of the IRRAflow® device is accompanied by a separate ICP monitor for improved patient safety. The IRRAflow® volume injections are reliable at all physiological ICP values.

Supporting information

S1 File. Figure a and b showing the reverse tunneling and fastening of the IRRAflow® catheter.

Figure c showing the bolted standard passive external ventricular drain (EVD), Speigelberg, Silverline®.

https://doi.org/10.1371/journal.pone.0297131.s001

(DOCX)

S2 File. Room temperature, water amount, device settings, and experiments repetitions in the compliance test and Experiments 1–4.

https://doi.org/10.1371/journal.pone.0297131.s002

(DOCX)

S3 File. Data output from the IRRAflow control unit.

(SysS: System status, TrtS: treatment status, ICP: intracranial pressure, Cprs: current pressure, HLim: Upper alarm limit, LLim: Lower alarm limit, PM: Preset Mode (1 or 0), (system operating on default settings or not, PPrs: Preset pressure (drain above), Ivol: infusion volume (ml/hour), CycT: cycle time (sec), stat1 HEX printout of various system status, Stat2: HEX printout of various system status, ErrCode: Errorcode).

https://doi.org/10.1371/journal.pone.0297131.s003

(DOCX)

S1 Fig. Figure showing change in baseline pressure following IRRAflow® bolus injections and table showing the mean actual injected volume at each bolus injection in grams.

https://doi.org/10.1371/journal.pone.0297131.s005

(TIFF)

References

  1. 1. Hanley DF, Lane K, McBee N, Ziai W, Tuhrim S, Lees KR, et al. Thrombolytic removal of intraventricular haemorrhage in treatment of severe stroke: results of the randomised, multicentre, multiregion, placebo-controlled CLEAR III trial. Lancet. 2017;389(10069):603–11. pmid:28081952
  2. 2. Haldrup M, Miscov R, Mohamad N, Rasmussen M, Dyrskog S, Simonsen CZ, et al. Treatment of intraventricular hemorrhage with external ventricular drainage and fibrinolysis. A comprehensive systematic review and meta-analysis of complications and outcome. World Neurosurg. 2023.
  3. 3. Rajjoub K, Hess RM, O’Connor TE, Khan A, Siddiqui AH, Levy EI. Drainage, Irrigation, and Fibrinolytic Therapy (DRIFT) for Adult Intraventricular Hemorrhage Using IRRAflow® Self-Irrigating Catheter. Cureus. 2021;13(5):e15167.
  4. 4. Carrera DA, Marsh LM, Roach JJ, Sarangarm P, Cole CD, Torbey MT, et al. HummingFlow: novel single twist-drill access for ventricular drainage, irrigation, monitoring, and automated local drug delivery in subarachnoid hemorrhage. Journal of Neurosurgery. 2023:1–6. pmid:37856891
  5. 5. Haldrup M, Rasmussen M, Mohamad N, Dyrskog S, Thorup L, Mikic N, et al. Intraventricular Lavage vs External Ventricular Drainage for Intraventricular Hemorrhage: A Randomized Clinical Trial. JAMA Network Open. 2023;6(10):e2335247–e. pmid:37815832
  6. 6. Kiening KL, Schoening WN, Lanksch WR, Unterberg AW, editors. Intracranial Compliance as a Bed-Side Monitoring Technique in Severely Head-Injured Patients. Intracranial Pressure and Brain Biochemical Monitoring; 2002 2002//; Vienna: Springer Vienna.
  7. 7. Kawoos U, McCarron RM, Auker CR, Chavko M. Advances in Intracranial Pressure Monitoring and Its Significance in Managing Traumatic Brain Injury. Int J Mol Sci. 2015;16(12):28979–97. pmid:26690122