Optimization and Validation of High-Performance Chromatographic Condition for Simultaneous Determination of Adapalene and Benzoyl Peroxide by Response Surface Methodology

  • Yi-Cheng Chen ,

    Contributed equally to this work with: Yi-Cheng Chen, Pi-Ju Tsai

    Affiliation School of Pharmacy, College of Pharmacy, Kaohsiung Medical University, Kaohsiung, Taiwan, Republic of China

  • Pi-Ju Tsai ,

    Contributed equally to this work with: Yi-Cheng Chen, Pi-Ju Tsai

    Affiliation Department of Business Administration, I-Shou University, Kaohsiung, Taiwan, Republic of China

  • Yaw-Bin Huang,

    Affiliation School of Pharmacy, College of Pharmacy, Kaohsiung Medical University, Kaohsiung, Taiwan, Republic of China

  • Pao-Chu Wu

    pachwu@kmu.edu.tw

    Affiliation School of Pharmacy, College of Pharmacy, Kaohsiung Medical University, Kaohsiung, Taiwan, Republic of China

Abstract

The aim of this study was to develop a simple and reliable high-performance chromatographic (HPLC) method for simultaneous analysis of adapalene and benzoyl peroxide in pharmaceutical formulation by response surface methodology (RSM). An optimized mobile phase composed of acetonitrile, tetrahydrofuran and water containing 0.1% acetic acid at a ratio of 25:50:25 by volume was successfully predicted by using RSM. An isocratic separation was achieved by using the condition. Furthermore, the analytical method was validated in terms of specificity, linearity, accuracy and precision in a range of 80% to 120% of the expected concentration. Finally, the method was successfully applied to the analysis of a commercial product.

Introduction

Acne vulgaris is a common chronic skin disease characterized by comedones, papules, and pustules and is associated with negative sequelae including scarring and pigmentation, which causes a high degree of psychosocial suffering [13]. It affects about 85% of individuals between the ages of 12 to 24 years [4,5]. Topical antibiotics, keratolytics, and retinoids are the main first-line treatment options for comedogenic and mild inflammatory acne. Combination therapy is often used in many clinical settings in order to target different pathogenic factors [4,68].

Adapalene, (6-[3-(1-adamantyl)-4-methoxyphenyl]-2- naphthoic acid) (Fig. 1) is a synthetic naphthoic acid derivative, third-generation retinoid with slightly better efficacy and light stability [9,10]. Retinoids can reverse the abnormal follicular desquamation and have anti-inflammatory effects. Topical retinoids are used in the treatment of mild-to-moderate acne and are also used (off-label) to treat keratosis pilaris as well as other skin conditions. In comparison with other topical retinoids, adapalene shows lower incidence of side effects and exhibits better tolerability profile, hence is clinically above the rest in the treatment of acne vulgaris [1012]. Benzoyl peroxide (Fig. 1) can increase skin turnover, clearing pores and reducing the bacterial count. It also has anticomedogenic and mild keratolytic effect [8,9,13,14]. Hence, benzoyl peroxide is one of the most widely prescribed drugs in acne therapy. A combination of adapalene and benzoyl peroxide shows good stability and a good clinical tolerance [15], and has also been found to have better tolerability than BPO monotherapy in terms of cutaneous tolerability [9]. Thus, the first topical acne preparations combining adapalene 0.1% and benzoyl peroxide 2.5% were approved by the US Food and Drug Administration in 2008 [9,16].

thumbnail
Fig 1. Chemical Structure of adapalene and benzoyl peroxide.

https://doi.org/10.1371/journal.pone.0120171.g001

There are some reported analytical methods for the determination of adapalene [15,17,18] or benzoyl peroxide alone [1921] by high-performance liquid chromatography (HPLC) and capillary electrophoresis in bulk or in pharmaceutical dosage form. The purpose of this study was to develop and validate a simple HPLC method for simultaneous analysis of adapalene and benzoyl peroxide in pharmaceutical product. In order to obtain an optimal analytical condition in a short time period and minimum trails, a statistical optimization tool based on response surface methodology (RSM) with a constrained mixture design [22,23] was used. The optimization procedure involved systematic design to minimize the number of trails, and analysis the response surfaces to realize the influence of independent factors and to obtain the optimal condition. In this present study, RSM was used to search for the appropriate mobile phase for the analysis method. Then, the analytical method was validated for specificity, linearity, accuracy, precision and robustness according to the guidelines for assay methods of the International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) [24] and applied to the analysis of a commercial product.

Materials and Methods

2.1. Materials

Adapalene with purity of 98% was purchased from Glenmark Generics Limited (Mumbai, India). Benzoyl peroxide with purity of 75% was purchased from Yuh Tzong Enterprise Ltd (Kaohsiung, Taiwan). Acetonitrile and methanol were obtained from Echo Chemical Co. (Miaoli, Taiwan). Tetrahydrofuran was obtained from J.T.Baker (Pennsylvania, U.S.A.). Purified water was prepared using Milli-Q (Millipore, MA, U.S.A.) All other chemicals and reagents used in the experiment were analytical grade. The commercial gel product (Epiduo) containing 0.1% adapalene and 2.5% benzoyl peroxide was purchased from Galderma (Levallois-Perret, France).

2.2. Standard solution preparation

The stock solutions of adapalene and benzoyl peroxide were prepared by simultaneously dissolving 4 mg adapalene and 100 mg benzoyl peroxide in 50 ml methanol. The standard solution was prepared by diluting the stock solution with methanol to achieve a final nominal concentration of 40 μg/ml adapalene and 1 mg/ml benzoyl peroxide.

2.3. Instrumentation and chromatographic conditions

The HPLC system used consisted of a chromatographic pump (L-7100), autosampler (L-2200), and an UV detector (L-4000H) (Hitachi, Japan). The LiChrosorb C18 (250×4.6mm, particle size 5 μm) column and guard column 4–4 packed with LiChrosper 100 RP-18 (Merck, Darmstadt, Germany) were used. The analytical column and the guard column were operated at ambient temperature. The injection volume was 20 μL. Detection was performed at 270 nm. The flow rate was set at 1.0 ml/min.

To obtain an optimal mobile phase for the analytical method, the statistical optimization tool based on a RSM with a constrained mixture design [22,23] was used. According to the USP 36 [18], the mobile phase contains acetonitrile, tetrahydrofuran, trifluoroacetic acid, and water at ratios of 21: 16: 0.01: 13 for determination of adapalene. USP 34 [21] reported a gradient mobile phase consists of acetonitrile and 0.1% glacial acetic aqueous solution ranged from 82/12 to 60/40 in 30 mins for benzoyl peroxide analysis. Hence, tetrahydrofuran (X1), acetonitrile (X2) and water containing 0.1% acetic acid (X3) was chosen as the variable factors of mobile phase in this study. The levels of X1, X2 and X3 were set from 10–30, 50–70 and 15–25%, respectively. Based on the constrained mixture model provided by Design-Expert software, ten model mobile phase compositions were arranged randomly. Two replicated batches of R02 vs R10 and R03 vs R09 were used to estimate the reproducibility by lack of fit. The compositions of mobile phase are listed in Table 1. The tailing factor (TF), retention time (RT), theoretical plates (TP) and resolution of adapalene and benzyl peroxide (AB/R) were used to evaluate the availability of mobile phases.

thumbnail
Table 1. The composition and analysis parameters of the HPLC method.

https://doi.org/10.1371/journal.pone.0120171.t001

2.4. Method validation

The HPLC method was validated for specificity, linearity, accuracy, precision and robustness according to the guidelines for assay methods of the International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) [24].

As Specificity.

The standard solution was used to assess the retention times. Blank chromatogram was recorded to check for interfering matrix peaks. Finally, chromatogram of real sample solution was reviewed for critical peak separations. Whenever appropriate, chromatographic resolution was calculated.

As Linearity.

To evaluate the linearity of the method, five standard solutions (80%~120% of nominal concentration) were prepared. Each concentration level was analyzed in triplicate. The linearity study was performed over two days. The peak areas were plotted versus the respective concentrations, and a linear regression analysis was used to obtain the equation and correlation coefficients. The limit of detection was the lowest amount that could be detected S/N >3.

As Precision/Accuracy/Recovery.

The intra-day precision and accuracy were evaluated by measuring five replicates of the samples at three concentrations levels (80, 100 and 120% of nominal concentration) on the same day, whereas the inter-day precision and accuracy were estimated using three validation batches on three consecutive days. The precision was calculated as the relative standard deviation (RSD%), and the accuracy was defined as the relative error (E%). Recovery was calculated as the ratio of the amount of drug from spiked gel to the amount of analyte added.

As Robustness.

The robustness was performed by modifying the experimental conditions. The chromatographic parameters including peak area, retention time, theoretical plates, and tailing factor were evaluated for different detected wavelengths (268 and 272 nm) and tetrahydrofuran concentrations (24 and 26%) and flow rate (0.9 and 1.1 mL/min).

2.5. Application of the analytical method in pharmaceutical gel

Accurately weighed amounts of pharmaceutical gel 2 g corresponding to 2 mg of adapalene and 50 mg of benzoyl peroxide was added into a 50 ml-volumetric flask. About 40 ml of methanol was added to this volumetric flask and sonicated in an ultrasonic bath for 15 min with intermittent shaking, diluted to volume with methanol, and mixed well. The sample solution of concentration (nominal concentration) was 40 μg/ml of adapalene and 1 mg/ml of benzoyl peroxide. Before being analyzed by HPLC, a portion of sample solution was centrifuged in tube caps at 12000 rpm for 10 min and filtered through a 0.45 μm Durapore membrane filter. A solution of blank gel containing the same excipients with commercial gel treated by the same procedure was used as a blank sample in the spectrophotometric method.

Results and Discussion

3.1. Optimization of mobile phase

The parameters of analysis capability of HPLC method in different mobile phase compositions are listed in Table 1. The TF, RT and TP were 1.27–1.49, 5.9–25.3 min and 7643–9217 for adapalene, and 1.12–1.20, 2.8–4.6 min and 7816–9732 for benzoyl peroxide respectively. The AB/R ranged from 13 to 22. These results indicated that the composition of the mobile phase had significant effect on the analysis capability of this analytical method.

To determine the effect of composition of mobile phase on the analysis capability, the values of TF and RT of adapalene and benzoyl peroxide were set as responses. The TP and AB/R in all conditions were far greater than the requirement of 4000–6000 and 1.5, so does not assess in this study. These responses and three independent variables (X1, X2 and X3) were statistically analyzed based on the response-surface methodology using a computer program (Design-Expert software) [22,25]. The results of multiple regression analysis for all parameters studied are summarized in Table 2. The probability (p-value) of the model was less than 0.05, indicating that the fitted model could describe the relationship between the dependent and independent variables. The p-value for the lack of fit for all models were greater than 0.05, suggesting absence of any lack of fit of the models, and that also strengthened the reliability of the models. The values of the coefficients X1, X2 and X3 were related to the effect of these variables on the response. The result showed that the aqueous phase (X3) had more influence degree on the RT, whereas organic phases (X1 and X2) had stronger effect on TF for both analytes. The phenomena might be attribute to the reverse phase column was used for both hydrophobic analysts in this study. The increment of organic phase resulted in elution time and tailing factors of hydrophobic compounds decrease. From the response surface plots (Fig. 2), it was found that the appropriate analytical method with lower TF and RT for both analysts could be reach at specific mobile phase composition. Hence, the RSM was used to optimize the multiple purposes. An optimized mobile phase composition was predicted to yield RT and TF values of 12.81 min and 1.31 for adapalene, as well as 3.68 min and 1.15 for benzoyl peroxide, respectively, when level of X1, X2 and X3 were 25%, 50% and 25%, respectively. To validate the predictive ability of the hypothesized model for each response, a new chromatography was performed with these levels of the independent variables to yield retention time and tailing factor values of 12.09 min and 1.18 for adapalene, as well as 3.54 min and 1.15 for, benzoyl peroxide, respectively. The observed values were remarkably close to the predicted values (predicted error < 11.1%), demonstrating that the RSM could be successfully used to optimize the proportion of mobile phase.

thumbnail
Table 2. Optimal regression equation for each response variable.

https://doi.org/10.1371/journal.pone.0120171.t002

thumbnail
Fig 2. The three-dimensional contour diagrams illustrating the effect of tetrahydrofuran (X1), acetonitrile (X2) and water containing 0.1% acetic acid (X3) on the retention time (RT) and tailing factor (TF).

https://doi.org/10.1371/journal.pone.0120171.g002

3.2. Method validation

System suitability of the chromatographic analytical method was tested by estimating the repeatability, retention time, tailing factor and theoretical plates of six replicate injections of standard solution. The results listed in Table 3, the TF (1.27 and 1.12), TP (8307 and 8860), and the relative standard deviation (RSD, %) (0.69% and 0.71%) for adapalene and benzoyl peroxide, respectively, were in agreement with the recommendations of FDA [26], indicating the analytical method could provide reliable data.

In term of specificity, placebo interference was carried out to investigate the specificity of the analytes (adapalene and benzoyl peroxide) in pharmaceutical preparation. The chromatograms of the blank solution, standard solution (nominal concentration), blank sample solution, and real sample solution were used to justify the specificity of the analytical method. As shown in Fig. 3, there was no interference from the other compounds, which, therefore, confirms the specificity of the method. The running time of this analytical method was 15 min, which was shorter than USP reported [18,21].

thumbnail
Fig 3. Chromatograms obtained for (A) real sample, (B) standard solution (nominal concentration), (C) blank sample, and (D) blank solution.

https://doi.org/10.1371/journal.pone.0120171.g003

The linearity of the analytical method was determined by linear regression analysis obtaining a calibration curve by plotting the peak area versus a series of concentrations including 80%, 90%, 100%, 110% and 120% of nominal concentration of standard solution. The calibration curve obtained by analyzing adapalene was y = 32922+23913x with a correlation coefficient (r) of 0.9992, whereas the equation of a curve obtained by assaying benzoyl peroxide was y = 229017+3155x (r = 0.9996). The limits of detection of adapalene and benzoyl peroxide were 4 and 7 μg/ml, respectively. Results for accuracy, precision and recovery are presented in Table 4. All values of accuracy and precision were within acceptable limits with E% and RSD no greater than 15% [26]. The recovery ratios were greater than 97.97%. Table 5 shows the results obtained from study of robustness determined by performing small changes or variations in the method’s parameters such as detecting wavelength, flow rate and tetrahydrofuran concentration in mobile phase. None of the alterations caused a significant effect on the determination of both analytes, indicating the robustness of the developed method.

thumbnail
Table 4. Accuracy, precision and recovery of the analytes.

https://doi.org/10.1371/journal.pone.0120171.t004

3.3. Application of the analytical method to pharmaceutical gel

In order to test the method on commercial product, the absolute recoveries of adapalene and benzoyl peroxide from commercial gel (Epiduo) containing 0.1% adapalene and 2.5% benzoyl peroxide was determined. The recoveries of adapalene and benzoyl peroxide were 93.11±0.78% and 95.17±1.02%, respectively, that fall within limits of the USP monograph.

Conclusion

The RSM was successfully to optimize the composition of mobile phase. The selective, precise, accurate, linear and robust of this developed analytical method was met ICH guideline indicated its stability.

Author Contributions

Conceived and designed the experiments: PCW PJT. Performed the experiments: YCC. Analyzed the data: PCW PJT YBH. Contributed reagents/materials/analysis tools: PJT. Wrote the paper: YBH PCW.

References

  1. 1. Ghodsi SZ, Orawa H, Zouboulis CC. Prevalence, severity, and severity risk factors of acne in high school pupils: a community-based study. J Invest Dermatol. 2009;129: 2136–2141. pmid:19282841
  2. 2. Leyden JJ. New understandings of the pathogenesis of acne. J Am Acad Dermatol. 1995; 32: S15–25. pmid:7738222
  3. 3. Magin P, Adams J, Heading G, Pond D, Smith W, Psychological sequelae of acne vulgaris: results of a qualitative study. Can Fam Physician. 2006;52: 978–979. pmid:17273501
  4. 4. Leyden JJ. A review of the use of combination therapies for the treatment of acne vulgaris. J Am Acad Dermatol. 2003;49: S200–210. pmid:12963896
  5. 5. Strauss JS, Krowchuk DP, Leyden JJ, Lucky AW, Shalita AR, Siegfried EC, et al. Guidelines of care for acne vulgaris management. J Am Acad Dermatol. 2007;56: 651–663. pmid:17276540
  6. 6. Fu LW, Vender RB. Newer approaches in topical combination therapy for acne. Skin Therapy Lett. 2011;16: 3–6. pmid:22089505
  7. 7. Goreshi R, Samrao A, Ehst BD. A double-blind, randomized, bilateral comparison of skin irritancy following application of the combination acne products clindamycin/tretinoin and benzoyl peroxide/adapalene. J Drugs Dermatol. 2012;11: 1422–1426. pmid:23377511
  8. 8. Babaeinejad SH, Fouladi RF. The efficacy, safety and tolerability of adapalene versus benzoyl peroxide in the treatment of mild acne vulgaris; a randomized trial. J Drugs Dermatol. 2013;12: 1033–1038. pmid:24002152
  9. 9. Tan JK. Adapalene 0.1% and benzoyl peroxide 2.5%: a novel combination for treatment of acne vulgaris. Skin Therapy Lett. 2009;14: 4–5. pmid:20054505
  10. 10. Cunliffe WJ, Caputo R, Dreno B, Forstrom L, Heenen M, Orfanos CE, et al. Clinical efficacy and safety comparison of adapalene gel and tretinoin gel in the treatment of acne vulgaris: Europe and U.S. multicenter trials. J Am Acad Dermatol. 1997;36: S126–134. pmid:9204091
  11. 11. Shalita A, Weiss JS, Chalker DK, Ellis CN, Greenspan A, Katz HI, et al. A comparison of the efficacy and safety of adapalene gel 0.1% and tretinoin gel 0.025% in the treatment of acne vulgaris: a multicenter trial. J Am Acad Dermatol. 1996;34: 482–485. pmid:8609263
  12. 12. Berson D, Alexis A. Adapalene 0.3% for the treatment of acne in women. J Clin Aesthet Dermatol. 2013;6: 32–35. pmid:24307923
  13. 13. Babaeinejad SH, Fouladi RF. The efficacy, safety, and tolerability of adapalene versus benzoyl peroxide in the treatment of mild acne vulgaris: a randomized trial. J Drugs Dermatol. 2013;12: 790–794. pmid:23884492
  14. 14. Thakur NK, Bharti P, Mahant S, Rao R. Formulation and characterization of benzoyl peroxide gellified emulsions. Sci Pharm. 2012;80: 1045–1060. pmid:23264949
  15. 15. Martin B, Meunier C, Montels D, Watts O. Chemical stability of adapalene and tretinoin when combined with benzoyl peroxide in presence and in absence of visible light and ultraviolet radiation. Br J Dermatol. 1998;139 Suppl 52: 8–11. pmid:9990414
  16. 16. FDA Drug Approval Package: Epiduo (adapalene and benzoyl peroxide) NDA #022320. http://www.accessdata.fda.gov/drugsatfda_docs/nda/2008/022320_epiduo_toc.cfm. Accessed 2013 Oct 2.
  17. 17. Martins LA, Meneghini LZ, Junqueira CA, Ceni DC, Bergold AM. A simple HPLC-DAD method for determination of adapalene in topical gel formulation. J Chromatogr Sci. 2011;49: 796–800. pmid:22080808
  18. 18. Convention TUSP. The United States Pharmacopeia (USP 36) and the National Formulary (NF 31). 2013
  19. 19. Mu G, Liu H, Gao Y, Luan F. Determination of benzoyl peroxide, as benzoic acid, in wheat flour by capillary electrophoresis compared with HPLC. J Sci Food Agric. 2012;92: 960–964. pmid:21997699
  20. 20. Ali MS, Rafiuddin S, Munir MI, Ghori M, Khatri AR. An isocratic reversed-phase high-performance liquid chromatographic method for the simultaneous determination of benzoyl peroxide and the related compounds benzoic acid, benzaldehyde, ethyl benzoate, methylparaben, and propylparaben in dermal preparations. J AOAC Int. 2008;91:1025–1036. pmid:18980115
  21. 21. Convention TUSP. The United States Pharmacopeia (USP 34) and the National Formulary (NF 29). 2011
  22. 22. Lewis GA, Mathieu D, Phan-Tan-Luu R. Pharmaceutical Experimental Design. Dekker New York. 1999
  23. 23. Malzert-Freon A, Hennequin D, Rault S. Partial least squares analysis and mixture design for the study of the influence of composition variables on lipidic nanoparticle characteristics. J Pharm Sci. 2010;99: 4603–4615. pmid:20845459
  24. 24. ICH International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use: Q2(R1): Validation of Analytical Procedures: Text and Methodology, Geneva, 1994.
  25. 25. Tsai PJ, Huang CT, Lee CC, Li CL, Huang YB, Tsai YH, et al. Isotretinoin oil-based capsule formulation optimization. ScientificWorldJournal. 2013;2013:856967. pmid:24068886
  26. 26. FDA Food and Drug Administration. “FDA Guidance for Industry,Bioanalytical Method Validation.” ‹http://www.fda.gov/downloads/Drugs/…/Guidances/ucm070107.pdf›, cited May 1, 2001.