Development and application of a prosthetist-specific rectification template based on artificial intelligence for the fabrication of transfemoral prosthetic sockets

  • Maria Grazia Santi ,

    Contributed equally to this work with: Maria Grazia Santi, Stefania Fatone, Andrea Giovanni Cutti

    Roles Data curation, Formal analysis, Investigation, Software, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliations INAIL, Vigorso di Budrio, Bologna, Italy, Department of Industrial Engineering, University of Padova, Padova, Province of Padua, Italy

    ⨯
  • Stefania Fatone ,

    Contributed equally to this work with: Maria Grazia Santi, Stefania Fatone, Andrea Giovanni Cutti

    Roles Funding acquisition, Investigation, Project administration, Resources, Writing – review & editing

    sfaton@uw.edu

    Affiliation Department of Rehabilitation Medicine, University of Washington, Seattle, Washington, United States of America

    ⨯
  • Andrew H. Hansen ,

    Roles Investigation, Project administration, Resources, Writing – review & editing

    ‡ AHH and SAG also contributed equally to this work.

    Affiliations Minneapolis VA Health Care System, Minneapolis, Minnesota, United States of America, Department of Family Medicine and Community Health, Department of Biomedical Engineering, University of Minnesota, Minneapolis, Minnesota, United States of America

    ⨯
  • Steven A. Gard ,

    Roles Funding acquisition, Project administration, Writing – review & editing

    ‡ AHH and SAG also contributed equally to this work.

    Affiliations Northwestern University Prosthetics-Orthotics Center, Department of Physical Medicine and Rehabilitation, Feinberg School of Medicine, Chicago, Illinois, United States of America, Jesse Brown VA Medical Center, Chicago, Illinois, United States of America

    ⨯
  • Andrea Giovanni Cutti ,

    Contributed equally to this work with: Maria Grazia Santi, Stefania Fatone, Andrea Giovanni Cutti

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Writing – original draft, Writing – review & editing

    Affiliation INAIL, Vigorso di Budrio, Bologna, Italy

    ⨯
  • Residual Limb Shape Capture Group

    Membership of the Residual Limb Shape Capture Group is provided in the Acknowledgments.

    ⨯

Abstract

The prosthetic socket is the most critical component of a lower limb prosthesis, requiring precise customization to the individual’s residual limb. This study presents proof-of-concept for a novel artificial intelligence (AI)-driven rectification template for transfemoral sockets tailored to a single prosthetist. Using a dataset of nine persons with transfemoral amputation, the study workflow required the manual casting and 3D scanning of unrectified and rectified plaster positives, anatomical landmark identification, and unsupervised training of an algorithm using Principal Component Analysis (PCA). The AI captured both explicit and implicit rectification strategies, generating an average rectification template and synergistic modes of variation. Validation was conducted via a leave-one-out approach, comparing AI-generated versus manually crafted rectified positives using clinically-relevant metrics: perimeter and volume differences. The first four PCA modes explained 78% of rectification variability, with key modifications observed in distal and medial regions. Volume differences between AI and manually rectified positives were within clinically acceptable limits for all participants, with 44% rated “good” and 56% “acceptable”. This proof-of-concept provides a new perspective on the application of AI to replicate prosthetist-specific rectification strategies for transfemoral sockets. It may potentially help streamline fabrication by capturing both explicit and implicit prosthetist knowledge. The approach may be useful in clinical training, documentation, and socket fabrication, particularly in resource-limited settings, and may contribute to more consistent and efficient prosthetics care.

Introduction

The socket is the most important component of a lower limb prosthesis. While all other components are mass-produced, the socket is a person-specific element, fabricated by prosthetists based on characteristics of the individual [1–5]. The socket is usually handcrafted from a negative mould of the individual’s residual limb, typically made by wrapping wet plaster or fiberglass bandages around the limb and allowing them to harden. A positive mould is then made by filling the negative mould with liquid plaster, allowing it to harden, and then stripping away the plaster or fiberglass bandages. This initial positive mould, also known as an “unrectified positive” (UP), is then shaped by the prosthetist using a process called “rectification”, in which material is added or removed in specific anatomical areas corresponding to the chosen socket design. The result is a “rectified positive” (RP) that has the shape of the socket interface with the residual limb [6–11]. From this shape, the socket is then manufactured using either composite lamination, vacuum forming of a thermoplastic material, 3D printing, or a combination thereof. The shape and volume of the socket are then fine-tuned with the prosthesis user by static evaluation and, after assembly with the other prosthetic components, by dynamic evaluation to obtain a comfortable and functional prosthesis.

The socket casting process just described is currently the most widely used in clinical practice. It reflects that over the last few decades, prosthetists have developed various techniques for both casting and positive shape rectification that produce well-fitting prosthetic sockets able to provide comfort, protect soft tissue, and allow effective control of the prosthesis [2,3]. Design approaches include the quadrilateral, ischial containment (IC), and subischial (SI) sockets [2,12]. There are also variants of the IC such as the Contour Adducted Trochanteric-Controlled Alignment Method (CAT-CAM) [13] and the Marlo Anatomical Socket (MAS) – an ischial-ramal containment (IRC) socket [14,15]. These approaches primarily differ in the management of the proximal socket region (i.e., brim), which is considered critical for stability, load transfer, and comfort. In the quadrilateral design, a narrow anterior-posterior dimension helps to keep the ischial tuberosity seated on a horizontal posterior shelf. A wider medio-lateral dimension allows proximal soft tissue to enter the socket but sometimes at the expense of coronal support [16]. Issues with medio-lateral stability led to the IC design and its variants wherein the ischial tuberosity and part of the ischial ramus are enclosed within the socket to enhance medio-lateral stability and control coronal plane femoral alignment. In the IRC-MAS design, the ischial ramus is enclosed more medially, allowing for the posterior and anterior sections of the proximal brim to be lowered, thus increasing hip motion and socket comfort [15]. These designs require precise shaping of the containment region and appropriate distribution of pressure over load-tolerant anatomical areas. Across these designs, various proximal dimensions (e.g., medial anterior-posterior, skeletal medio-lateral, diagonal medio-lateral, and ramus angle) are critical for appropriately designing the proximal region and are considered key determinants of socket fit and performance. These approaches also rely on global volume adjustments of the thigh tissue for some load transfer [17]. By comparison, subischial (SI) sockets have much lower proximal brims that aim to avoid contacting the pelvis with a similar goal of improving hip mobility and comfort [18,19]. In the SI design load transfer occurs exclusively through thigh tissue, but the proximal medio-lateral dimension of the socket and contouring posterior-lateral to the femur (similar to the CAT-CAM socket [13]) remain important criteria for socket stability [11]. The exact choice of socket, and whether it explicitly follows a particular design, is hybridized or further customized, is based on many factors including the patient’s residual limb, their daily activities, and the prosthetist’s expertise. Hence, transfemoral socket design is considered a complex process with the contribution of different parameters to the assessment of “socket fit” dependent on the specific socket design approach and patient presentation.

Unfortunately, it is difficult to make the processes associated with socket fabrication repeatable and reproducible within and between prosthetists because they are based on implicit knowledge [9,10,20]. This is one reason why, since the 1980s, the prosthetic field has moved toward digitization of design and manufacturing processes [6,20,21]: first by introducing scanning techniques that digitally capture the residual limb/UP/RP/socket shapes, allowing for digital storage and visualization; then by introducing computer-aided design (CAD) software digital rectification templates [22–30], and computer-aided manufacturing (CAM), to digitally rectify the meshes [11,31]; and finally, by fabricating the socket with new materials and techniques (e.g., 3D printing) [32–37].

These innovations (3D scanners, CAD-CAM technologies, new materials and techniques) have benefitted prosthesis users, prosthetists, and clinics [20], by reducing fabrication time, physical storage space, and exposure of prosthetists to unhealthy biomechanical factors and irritants, and by allowing for documentation of fabrication steps [6,10,11,31,36,38–42]. Despite these benefits, current techniques rely on the prosthetist’s ability to make CAD rectifications or to adapt a predefined rectification template to an individual’s unique residual limb characteristics [20].

More recent advancements have explored Artificial Intelligence (AI) algorithms. For example, digital tools such as Instalimb (Sumida-ku, Tokyo, Japan) use AI to recognise anatomical landmarks on coloured scans of the residual limb speeding-up the rectification process [43]. In other examples, researchers trained AI (e.g., Principal Component Analysis, K-means, DiffusionNet) to capture the knowledge used by prosthetists when completing rectifications by analysing differences between the digital meshes captured before and after rectification [20,22,44–49], relying mostly on geometric parameters. These approaches capture both the typical rectifications adopted by prosthetists on whom the AI was trained, but also the variations used to adjust the typical rectification for individuals with amputation. This knowledge has two components: “explicit” and “implicit” [20], with prosthetists able to verbally describe the first but not the second. Importantly, the proposed approach to AI algorithms can be trained on a pool of prosthetists resulting in the acquisition of “shared knowledge” [22,44–46,48] or on a single prosthetist to capture their specific rectification method [20]. The result of this AI learning process is a new type of “AI rectification template” that assists a prosthetist in preparing a socket for a new individual with amputation, by applying the typical learned template and its variations, while also preserving their location, magnitude, and synergistic relationships.

All AI algorithms proposed to date have focused on the rectification of transtibial sockets [20–22,44–46,50–52], with only two exceptions for transradial sockets [47,49]. No AI-driven algorithms have been reported yet for people with transfemoral amputation, despite the provision of these sockets being challenging due to the mass of soft tissue surrounding the bone. To tackle this gap in knowledge, this proof-of-concept work aimed to demonstrate the feasibility of defining a prosthetist-specific AI rectification template to support the provision of transfemoral sockets, leveraging the workflow described in Cutti et al. [20], and to test its validity using a leave-one-out approach.

Materials & methods

Participant characteristics

This work constitutes retrospective analysis of existing data. Data from nine individuals with transfemoral amputation who participated in a larger, multisite clinical trial (NCT04141748) [20] were accessed for this study on November, 10, 2025. The study from which data were drawn was approved by the Institutional Review Boards of the primary institutions (STU00210416; VAM-19–00466; 714–2019-SPERAUSLBO-19134), as well as the Department of Defense Office of Human Research Oversight (E01091). The original clinical trial recruitment took place between September 1, 2020, and ended on September 1, 2023, and all participants gave written informed consent to participate in the original clinical trial. All authors were investigators in the original clinical trial and, as such, had access to information that could identify the individual participants during conduct of the original clinical trial. The subset of data that is the focus of this analysis came from a prosthetist with 20 years of experience in fitting sockets to people with transfemoral amputation. Table 1 reports the participant demographics (sex, age, mass, height, aetiology of amputation) and residual limb characteristics (side, length, and tissue type).

thumbnail
Table 1. Characteristics of the participants with transfemoral amputation included in the study. ID: identifier; M: male; F: female; BMI: body mass index; R: right; L: left.; SD: Standard deviation.

https://doi.org/10.1371/journal.pone.0356483.t001

Process for generating a prosthetist-specific rectification template

To build the prosthetist-specific AI rectification template we performed the following steps: (1) UP manual casting; (2) UP digital 3D scanning; (3) digital identification of landmarks on the UP (in short “UP virtual landmarking”); (4) RP manual fabrication; (5) RP digital 3D scanning; and (6) unsupervised training of the AI algorithm. These steps are summarized below, but detailed information can be found in Cutti et al. [20]. The only exception is Step 3, which is unique to application of this process for transfemoral sockets.

In Step 1, the prosthetist manually fabricates the UP by taking a plaster cast of each participant’s residual limb over a liner (Seal-In X, Ossur, Reykjavík, Iceland), with the subject standing. The key actions taken during this step are illustrated in Fig 1. During casting the prosthetist identified a set of landmarks (LMs) (Fig 2) on the participant’s residual limb, used stickers to identify those LMs on the liner worn over the residual limb, and marked the location of the stickers with a felt pen on the cling film wrapped around the liner (Figs 1a-1b). These ink LMs were transferred to the inner surface of the negative cast by the wet plaster bandages capturing the ink (Figs 1c-1d). Then, the negative cast was extended proximally by 15 cm using plaster bandage wrapped in a ring. A set of about 20 physical markers 7 mm in diameter and 2 mm in thickness were glued to the inside of the cast over the ink landmarks and randomly on the plaster extension (Fig 1e). By filling the negative cast with liquid plaster, the UP was obtained, and all marks were transferred to it as indentations (Fig 1f). Some of the LMs reported in Fig 2 represent anatomical locations identified by palpation (“type A”), while others are geometric locations (“type G”). “Type B” LMs are not used during this stage of the process but are required for, and will be described during, Step 3. The prosthetist used a soft ethylene-vinyl-acetate (EVA) strip to mark a line connecting the proximal and distal LMs, and a tape measure to identify the intermediate LMs (Fig 1a).

thumbnail
Fig 1. Unrectified positive manual casting.

Step 1 of manual casting. a-c: identify landmarks (LMs) on the participant’s residual limb and mark them on the liner; d: hand casting with wet plaster bandages during which the ink LMs transfer to the negative cast; e: internal view of the negative cast extended proximally using a plaster ring, with physical markers glued over the ink LMs; f: plaster positive exhibiting indentations from the physical markers.

https://doi.org/10.1371/journal.pone.0356483.g001

thumbnail
Fig 2. Identification of landmarks.

Landmarks identified and marked by the prosthetist on participants’ liner-clad residual limb or directly on the digital scan of the unrectified positive (UP). Abbreviations: A – anatomical location; G – geometrical location; B – locations along the proximal brim.

https://doi.org/10.1371/journal.pone.0356483.g002

In Step 2, the UP was digitally scanned using the EinScan Pro 2X Plus structured light scanner coupled with the high-definition accessory (Shining3D, Hangzhou, China). Accuracy and repeatability of this scanner for this specific application have been previously reported [53,54].

In Step 3, the type “B” LMs described in Fig 2 were virtually identified in the scan. These are geometric LMs that aim to describe the shape of the proximal brim of the socket. Virtual identification was completed by the prosthetist using SocketFactory, a custom Python software (version 3.8) that takes advantage of MeshLab (MeshLab_64bit_fp v2020.07) [55] and VTK libraries (version 9.0.0) [56] as computational geometry engines, with Qt (PyQt5 version 5.14.2) [57] to support the graphical user interface [53].

In Step 4, the prosthetist manually rectified the UP making sure to leave the marker indentations on the proximal extension of the positive mould untouched. Details about the rectification process are provided in the next section. Once completed, the prosthetist moved on to Step 5, digitally scanning the RP using the same process as described in Step 2.

Finally, in Step 6, each pair of UP-RP meshes were digitally processed using SocketFactory. As described in Cutti et al. [20] this included: digital labelling of the LMs described in Fig 2 (Fig 3a); alignment of the UPs in the virtual global coordinate frame (Fig 3b); spatial registration of the RPs with respect to the paired UPs using the LMs on the proximal untouched extension (Fig 3c); digital removal of the proximal extension and cleaning the meshes of any imperfections due to the scanning process (Fig 3d); mirroring of all right-sided UP-RP couples (to obtain a dataset of only left-sided meshes) (Fig 3e); projection of LMs from the UP to the corresponding RP (Fig 3f); morphing of UP-RP mesh couples with a Reference Mesh, to obtain a dataset of morphed UP* and RP* meshes with the same topology (Fig 3g) [58]; and evaluation of UP*-RP* displacement vectors between homologous vertices (Fig 3h). Once the dataset was processed, the Principal Component Analysis (PCA) was trained with the dataset of displacement vectors, using the approach described by Dickinson et al. [20,46]. The PCA extracted an average displacement matrix (ADM) and learned and clustered the “modes of variation”, i.e., the synergistic changes the prosthetist applied when adapting their typical rectifications to a specific individual with amputation.

thumbnail
Fig 3. Unsupervised training of the AI algorithm.

Processing steps for each Unrectified Positive-Rectified Positive (UP-RP) pair using SocketFactory software. a: digital labelling of the landmarks (LMs) described in Fig 2; b: alignment of the UPs in the virtual global coordinate frame; c: spatial registration of the RPs with respect to the paired UPs; d: digital removal of the proximal extension and cleaning of the meshes; e: mirroring of all right-sided UP-RP couples; f: projection of LMs from the UP to the corresponding RP; g: morphing of UP-RP mesh pairs with a Reference Mesh; h: evaluation of UP*-RP* displacement vectors between homologous vertices.

https://doi.org/10.1371/journal.pone.0356483.g003

Manual rectification procedure and check socket fabrication

As illustrated in Figs. 2, 3 and 4, the socket design executed by the prosthetist in this study was a variation of the IC socket with a lower posterior brim inspired by the “extremely low” posterior trimlines of the MAS design [14,59]. The lower posterior brim relieves the gluteus maximus muscle, allowing for freer contraction during hip flexion-extension, without sacrificing containment of the ischial tuberosity or necessitating that containment be moved medially as in the MAS socket. The remainder of the socket geometry follows primarily IC principles and not the proximal dimensions of the MAS socket. The rectification approach as described by the prosthetist is illustrated in Fig 4. First, the distal end of the residual limb is extended by approximately 10 mm for firm tissues and 15–20 mm for soft tissues to accommodate distal tissue deformation under load. An additional 10 mm localized increase is made to the antero-medial adductor region. A general volume reduction is then performed over the entire surface, excluding the distal end and the antero-medial adductor area. Finally, localized reductions are made in the proximal regions: 10 mm in the posterior-proximal area and 5 mm in the antero-lateral proximal area corresponding to the quadriceps insertion.

thumbnail
Fig 4. Rectification approach described by prosthetist.

Verbal and graphical description of the prosthetist-specific rectification approach. Red shading is associated with a volume decrease, blue shading is associated with a volume increase, and grey shading represents unmodified areas (constant volume).

https://doi.org/10.1371/journal.pone.0356483.g004

Using the RP, the prosthetist fabricated a check (i.e., diagnostic) socket consisting of a flexible inner socket made from thermoplastic EVA, placed within a more rigid Polyethylene Terephthalate Glycol (PETG) socket with a U-shaped frame. The check socket was fit to the participant as part of the larger clinical trial [60]. Each participant assessed the socket during static standing and provided the Socket Comfort Score (SCS) [61] and verbal feedback reported in Table 2.

thumbnail
Table 2. Socket Comfort Scores (SCS) and verbal feedback provided by participants for the check sockets made by the prosthetist using manual techniques. Std: Standard Deviation.

https://doi.org/10.1371/journal.pone.0356483.t002

Digital rectification quality assessment

To qualitatively verify the typical rectification rules described by the prosthetist, the average UP* mesh was calculated based on the UP* meshes of the nine participants, and then the ADM was applied to the average UP* to obtain an average RP* mesh. To visually inspect the rectifications embedded in the AI prosthetist-specific rectification template, a distance colour map was generated between the average UP* mesh and the average RP* mesh. Moreover, to assess the synergistic modifications implemented by the prosthetist to account for the specific needs of each participant, the modes of variations learned by the AI algorithm were reported in rank order based on the explained variance. To visually capture the effect of each mode of variation, the average UP* mesh was digitally rectified twice by applying the ADM in combination with that mode, first with a weighting factor of +3 standard deviations and, second, with a weighting factor of −3 standard deviations. Therefore, to illustrate the effect of each mode of variation, two RP* meshes were generated.

Leave-one-out validation of the AI generated prosthetist-specific template

To validate the AI generated prosthetist-specific template, a leave-one-out approach was adopted. Steps 1–5 of the procedure previously presented were applied to the whole dataset and then Step 6 was applied to the processed dataset each time excluding data from one of the participants, overall generating nine slightly different templates, trained with slightly different datasets comprised of eight participants. Hence, nine RP* meshes were obtained.

To quantitatively analyse the results obtained, the RP* meshes were compared with the RP*s made manually by the prosthetist (referred to as “Hand RP”), by calculating ”clinically-relevant measures” [53] consisting of differences in the perimeters of the cross sections and volumes. Agreement between Hand and Digital RP measures is presented using “modified” Bland-Altman plots as introduced by Cutti et al. [53]. Specifically, to ease interpretation of perimeters (often referred to by prosthetists as circumferences) collected at different heights along the residual limb, the heights of the cross sections were reported in percentage on the Y axis of the plot, to resemble the vertical orientation of a residual limb model (instead of the X axis as is typical for Bland–Altman plots) [53]. Since there is no available literature describing the limits within which variation in perimeters or volume of a transfemoral socket will result in a clinically acceptable socket, the results are discussed with respect to the limits formulated by consensus of six prosthetists who participated in the original clinical trial and had clinical experience ranging from 7 to 20 years [60]. The limits for volumes are summarized in Table 3.

thumbnail
Table 3. Clinically acceptable limits of variation in socket volume based on the consensus of six prosthetists.

https://doi.org/10.1371/journal.pone.0356483.t003

The limits of variation for perimeters are reported in Fig 5. Changes in the perimeter of the proximal region of the residual limb are generally more tolerable than those occurring distally. This is because proximal circumferences are typically larger, so a given absolute variation has a proportionally smaller impact compared with the same change in the distal region. Hence, perimeter variations were divided according to whether proximal or distal regions were affected by a volume decrease or increase.

thumbnail
Fig 5. Limits of variation for perimeters.

Clinically acceptable limits of variation in perimeters based on the consensus of six prosthetists. V: volume; P: perimeter; R: radius.

https://doi.org/10.1371/journal.pone.0356483.g005

Results

Assessment of the rectification rule learned by the AI

Fig 6A depicts a colour map of the average rectifications typically performed by the prosthetist. Here, red is associated with material decrease, blue with material increase, and white with no changes. Fig 6B reports the distribution of the explained variance over the modes of variation: specifically, about 78% of variation was explained by modes 1–4. The variance explained by additional modes were less than 10% each. For modes 1–4, Fig 6B also depicts their geometric effect using the + 3 or −3 standard deviation weighting factor. Specifically, the first mode generates a distal anterior (+3 SD) or posterior (−3 SD) widening of the socket; the second mode generates a medial (+3) or lateral (−3) widening; the third mode shortens (+3) or elongates (−3) the distal portion of the socket; and the fourth mode seems related to the creation of reliefs for distal bony prominences or painful spots.

thumbnail
Fig 6. Colour map of the average rectifications.

(A) Distance colour maps in the Red–White–Blue scale describing the average rectifications made by the prosthetist during the rectification process. Red was associated with a decrease in material, blue with an increase in material, and white with no change. (B) Plot of the variance explained by each mode of variation and the corresponding qualitative description of the rectifications included in each of the first four modes of variation. Black arrows indicate the location, direction and relative magnitude of the rectifications captured in each mode. SD: Standard deviation.

https://doi.org/10.1371/journal.pone.0356483.g006

Digital rectification quality assessment

To assess the quality of the AI-based rectification through the leave-one-out approach, the clinically-relevant measures are reported in Fig 7. This is a graphical representation of the differences between RPs in terms of perimeters at six equally spaced cross sections along the longitudinal axis of the residual limb. The percentage of unacceptable measures are reported in Fig 7. This was calculated as the percentage of total participants outside the acceptable limit of variation.

thumbnail
Fig 7. Differences between rectified positives in terms of perimeters.

Modified Bland-Altman plot of variation in perimeter differences between RPs at 6 equally spaced cross sections along the longitudinal axis of the residual limb. Each participants’ measurement is reported using their ID inscribed in yellow circles. The green band represents the most stringent limit of variation identified by the prosthetists involved in the larger clinical study [60]. The percentage of unacceptable measures is reported in red for each cross section. Data used to create Fig 7 are provided as S1 File.

https://doi.org/10.1371/journal.pone.0356483.g007

Finally, Table 4 reports the volume differences between RPs in absolute and percentage values for each participant. To evaluate the results obtained in terms of percentage volume variations, prosthetists’ opinions are reported in Table 3.

thumbnail
Table 4. Volume differences between RPs are reported in absolute and percentage values for each participant. Average values are reported in the last row. Green shading indicates volume differences falling in the “good” fit range, while yellow shading indicates volume differences falling in the “acceptable” fit range.

https://doi.org/10.1371/journal.pone.0356483.t004

Discussions

This proof-of-concept study aimed to develop an AI-driven rectification template for transfemoral sockets, leveraging the procedure reported in our previous work on transtibial sockets [20]. The template was based on an unsupervised machine learning algorithm, PCA, which can automatically generate a suggestion for the RP shape based on the UP of a participant with a transfemoral amputation. The current study should be interpreted as a proof-of-concept focused on knowledge capture, not performance optimization of the socket.

To qualitatively evaluate the consistency of the average rectification template learned by the AI algorithm with the rectification technique described by the prosthetist, we compared Fig 4 with Fig 6. This comparison illustrated agreement between the two, i.e., the algorithm correctly identified the areas where the prosthetist typically added or removed material from the UP. Specifically, the UP underwent a general reduction in volume to improve coupling of the socket with the residual limb. Furthermore, material was removed locally from the proximal-lateral and posterior regions. Material was added to the distal end of the positive residual limb mould, creating an extension of approximately 15–20 mm and to the proximal antero-medial portion intended to make the proximal brim more comfortable at the level of the ischial support.

The synergistic modifications that the prosthetist applied to customize the average template to the specific participant (i.e., the implicit knowledge of the prosthetist), were captured by analysing the modes of variation identified by the PCA as described in Fig 6b. The first four modes (which explained 78.1% of the total variance), described rectifications applied at the distal end, which appear to be the areas subjected to participant-specific adaptation of the average template. In detail, the first, second and fourth modes of variation described build-ups either anteriorly/posteriorly (± 3 SD), medially/laterally (± 3 SD), or postero-medial/antero-lateral (± 3 SD), respectively. Finally, the third mode described elongation (−3 SD) or shortening (+3 SD) of the residual limb length. These modifications are likely aimed at creating relief areas for painful spots, or to off-load pressure from the distal end of the residual limb, which is commonly sensitive and painful in those with amputation that transects a long bone such as the femur.

Also, the study aimed to assess the validity of the AI predictions by comparing the Digital RP to the Hand RP, using a leave-one-out approach assessing perimeter and volume differences. Fig 7 shows that, for the proximal region, all participant perimeters fell within limits agreed to by the six prosthetists (see Table 3). However, it also shows that distally, the RP predicted by the AI for Participant 7 was wider than the Hand RP. The opposite occurred for Participant 4, whose Digital RP was less wide than the corresponding Hand RP at the most distal cross section. Interestingly, these outliers are consistent with those who reported low Socket Comfort Scores for the check sockets based on the Hand RP, i.e., 6.5 for Participant 4 and 5 for Participant 7. Additionally, during the fitting process, Participant 7 reported pressure at the anterior end of the distal femur, while Participant 4 reported that the socket was too wide (Table 2). The Digital RPs have a shape that would seem to address these complaints. While we may speculate that the AI-generated sockets may have resulted in better comfort, further studies are needed to confirm this hypothesis. While we believe that AI can assist prosthetists in their decision making and facilitate workflow, its superiority in autonomous rectification is not yet supported and was not the aim of this study. These data simply suggest that further efforts in AI training may consider Socket Comfort Score as an additional outcome measure.

The results reported in Fig 7 for perimeters were mostly acceptable. Similarly, the results reported in Table 4 suggest that for all participants the volume differences between Hand and Digital RPs fall within a “good” or “acceptable fit” range. However, evaluation of the acceptability of our results was based on perimeter and volume thresholds identified through consultation with six experienced prosthetists. This approach was necessary given that no literature was available on this topic for people with transfemoral amputation. As such, we believe this study is the first to propose acceptability thresholds for transfemoral socket fit based on perimeters and volumes. However, these thresholds need additional exploration and validation.

While our results are promising, this study has limitations. First, it should be considered exploratory given that the sample size cannot reflect the extent of variability in the clinical presentation of people with transfemoral amputation. For example, all participants had traumatic amputations. Other common amputation aetiologies, such as vascular disease, were not represented. Second, the study included a single prosthetist providing high intra-prosthetist consistency in terms of the outcome parameters but limiting generalizability of the approach to rectification. Future studies are needed to demonstrate that this same approach can be used effectively across many prosthetists and rectification patterns intended for different socket designs. Third, the workflow required identification of an extensive number of landmarks. We chose intentionally to have high redundancy in this first attempt. Future studies may perform sensitivity analyses to determine the minimum number of landmarks required to generate models with acceptable performance that may be more efficient to use in clinical practice. Currently, the location of Type B landmarks is specific to the brim shape of the socket used in this study and may need to be redefined depending on the socket design. Additionally, based on the clinical trial protocol from which these data were drawn [60], socket comfort was evaluated only under static conditions. Socket comfort under dynamic conditions would be required to obtain a more comprehensive understanding of overall socket performance.

Finally, results rely primarily on geometric agreement (perimeter and volume differences), which may not fully reflect clinical socket performance, particularly in transfemoral populations where pressure distribution and stability are critical. Other participant characteristics (e.g., residual limb tissue consistency, scars, position of neuromas, painful areas, volume fluctuation) and the inclusion of patient-centered outcomes regarding the quality of fit should be considered in future studies to design a multiparametric training dataset that, together with geometric parameters, could improve AI predictions. Further clinical parameters may also be included describing the quality of socket fit, but these might depend on the specific socket design. For instance, for an IC socket, prosthetists may need to evaluate if the adductor longus tendon is properly positioned in the medial-anterior aspect of the socket without excessive pressure, or whether the anterior-medial wall allows the pubic ramus to exit without discomfort, or whether there is any gapping or localized pressure at the trim lines. Additionally, it is important to verify that no excessive pressure is experienced with increased weight-bearing, that the socket maintains appropriate tightness under lateral forces, and that key anatomical landmarks, such as the greater trochanter, remain comfortable.

Conclusions

This proof-of-concept study illustrates application of an AI algorithm to transfemoral prosthetic sockets. We demonstrated that the algorithm could learn the rectifications typically made by a single prosthetist. A key factor in the success of this method was the introduction of a novel, step-by-step approach to data collection and processing, combined with a “morphing” stage that ensured consistent topology between UP and RP mould pairs. Once optimized, this technique could serve as a valuable time-saving resource for prosthetists by automating rectifications that result in a well-fitting initial socket for individuals with transfemoral amputation. This approach has a wide range of potential applications, from supporting digital rectification, to the creation of average rectification maps as a teaching aid. This may be particularly beneficial in resource-limited settings and isolated communities where experienced prosthetists are scarce.

Supporting information

Acknowledgments

Residual Limb Shape Capture Group (in alphabetical order): Dennis Anco, CP.2; Ilaria Annese, M.Eng.1; Liridona Ashiku, BS4; Kyle Barrons, CPO4; Michela Bisighini, M.Eng.1; Ryan Caldwell, CP4; Kristin Carnahan, CPO4; Juan Cave, CPO2; Kierra Falbo, CPO2,3; Anita Fazzini, M.Eng.1; Marco Folcio, M.Eng.1; Fabrizio Giacchi, CPO1; Amy Gravely, MA2; Gianni Gregori, CPO1; Syeda Hussain, BS4; John M. Looft, Ph.D.2,3; Claudia Mele, M.Eng.1; Gianluca Migliore, CPO1; Katherine Muschler2; Giovanni Osti, CPO1; Ioana Madalina Raileanu, M.Eng.1; Pericle Randi, PT1; Nicole Walker, CPO2,3; Jessica Yohay, BS4; Kelly Yun2 (1INAIL, Budrio, Italy; 2Minneapolis VA, MN, USA; 3University of Minnesota, MN, USA; 4Northwestern University, IL, USA). We would also like to acknowledge Marco Evangelos Biancolini (Universitá di Roma Tor Vergata) and RBF Morph Srl. (Rome, Italy) for developing the Python version of the software needed for application in this work.

References

  1. 1. Cutti AG, Morosato F, Gentile C, Gariboldi F, Hamoui G, Santi MG, et al. A Workflow for Studying the Stump-Socket Interface in Persons with Transtibial Amputation through 3D Thermographic Mapping. Sensors (Basel). 2023;23(11):5035. pmid:37299763
  2. 2. Paterno L, Ibrahimi M, Gruppioni E, Menciassi A, Ricotti L. Sockets for Limb Prostheses: A Review of Existing Technologies and Open Challenges. IEEE Trans Biomed Eng. 2018;65(9):1996–2010. pmid:29993506
  3. 3. Mak AF, Zhang M, Boone DA. State-of-the-art research in lower-limb prosthetic biomechanics-socket interface: a review. J Rehabil Res Dev. 2001;38(2):161–74. pmid:11392649
  4. 4. International Organization for Standardization. ISO 29782:2022 Prostheses and Orthoses — Factors to be considered when specifying a prosthesis for a person who has had a lower limb amputation. 2022.
  5. 5. International Organization for Standardization. ISO 21065:2017 Prosthetics and Orthotics — Terms Relating to the Treatment and Rehabilitation of Persons Having a Lower Limb Amputation. 2017.
  6. 6. Mayo AL, Gould S, Cimino SR, Glasford S, Harvey E, Ratto M, et al. A qualitative study on stakeholder perceptions of digital prosthetic socket fabrication for transtibial amputations. Prosthet Orthot Int. 2022;46(6):607–13. pmid:36515905
  7. 7. Boone DA, Burgess EM. Automated Fabrication of Mobility Aids: Clinical Demonstration of the UCL Computer Aided Socket Design System. JPO Journal of Prosthetics and Orthotics. 1989;1(3):187–90.
  8. 8. Cullen S, Mackay R, Mohagheghi A, Du X. The Use of Smartphone Photogrammetry to Digitise Transtibial Sockets: Optimisation of Method and Quantitative Evaluation of Suitability. Sensors (Basel). 2021;21(24):8405. pmid:34960503
  9. 9. Safari MR, Rowe P, McFadyen A, Buis A. Hands-off and hands-on casting consistency of amputee below knee sockets using magnetic resonance imaging. ScientificWorldJournal. 2013;2013:486146. pmid:24348164
  10. 10. Olsen J, Turner S, Chadwell A, Dickinson A, Ostler C, Armitage L, et al. The Impact of Limited Prosthetic Socket Documentation: A Researcher Perspective. Front Rehabil Sci. 2022;3:853414. pmid:36189046
  11. 11. Fatone S, Johnson WB, Tran L, Tucker K, Mowrer C, Caldwell R. Quantification of rectifications for the Northwestern University Flexible Sub-Ischial Vacuum Socket. Prosthet Orthot Int. 2017;41(3):251–7. pmid:28094689
  12. 12. Brodie M, Murray L, McGarry A. Transfemoral prosthetic socket designs: A review of the literature. J Prosthet Orthot. 2022;34:e73–92.
  13. 13. Sabolich JA. Contoured adducted trochanteric-controlled alignment method (CAT-CAM); Introduction and basic principles. Clinical Prosthetic and Orthotics. 1985;9:15–26.
  14. 14. Traballesi M, Delussu AS, Averna T, Pellegrini R, Paradisi F, Brunelli S. Energy cost of walking in transfemoral amputees: Comparison between Marlo Anatomical Socket and Ischial Containment Socket. Gait Posture. 2011;34(2):270–4. pmid:21684165
  15. 15. Klotz R, Colobert B, Botino M, Permentiers I. Influence of different types of sockets on the range of motion of the hip joint by the transfemoral amputee. Ann Phys Rehabil Med. 2011;54(7):399–410. pmid:21996548
  16. 16. Muller M. Transfemoral amputation: prosthetic management. In: Krajbich J, Pinzur M, Potter B, Stevens P, editors. Atlas of amputations and limb deficiencies: surgical, prosthetic and rehabilitation principles. 4 ed. Rosemont, IL: American Academy of Orthopaedic Surgeons. 2016. p. 537–54.
  17. 17. Zavaleta-Ruiz JL, Fatone S, Major MJ, Pankaj P. Mechanical Stimulation in the Residual Femur During Gait in Transfemoral Prosthesis Users Provides a Potential Reason for Bone Mineral Density Loss. Int J Numer Method Biomed Eng. 2025;41(10):e70103. pmid:41068549
  18. 18. Fatone S, Caldwell R. Northwestern University Flexible Subischial Vacuum Socket for Persons with Transfemoral Amputation: Part 1 Description of Technique. Prosthet Orthot Int. 2017;41(3):237–45.
  19. 19. Fatone S, Caldwell R, Angelico J, Stine R, Kim K-Y, Gard S, et al. Comparison of Ischial Containment and Subischial Sockets on Comfort, Function, Quality of Life, and Satisfaction With Device in Persons With Unilateral Transfemoral Amputation: A Randomized Crossover Trial. Arch Phys Med Rehabil. 2021;102(11):2063-2073.e2. pmid:34214499
  20. 20. Cutti AG, Santi MG, Hansen AH, Fatone S. Prosthetist-Specific Rectification Templates Based on Artificial Intelligence for the Digital Fabrication of Custom Transtibial Sockets. Prosthesis. 2024;6(5):1149–69.
  21. 21. Raschke SU. Limb Prostheses: Industry 1.0 to 4.0: Perspectives on Technological Advances in Prosthetic Care. Front Rehabil Sci. 2022;3:854404. pmid:36188935
  22. 22. Steer J, Stocks O, Parsons J, Worsley P, Dickinson A. ampscan: A lightweight Python package for shape analysis of prosthetics and orthotics. JOSS. 2020;5(48):2060.
  23. 23. Karakoç M, Batmaz İ, Sariyildiz MA, Yazmalar L, Aydin A, Em S. Sockets Manufactured by CAD/CAM Method Have Positive Effects on the Quality of Life of Patients With Transtibial Amputation. Am J Phys Med Rehabil. 2017;96(8):578–81. pmid:28085736
  24. 24. Tay FEH, Manna MA, Liu LX. A CASD/CASM method for prosthetic socket fabrication using the FDM technology. Rapid Prototyping Journal. 2002;8(4):258–62.
  25. 25. Travis RP, Dewar ME. Computer-aided socket design for trans-femoral amputees. Prosthet Orthot Int. 1993;17(3):172–9. pmid:8134277
  26. 26. Houston VL, Burgess EM, Childress DS, Lehneis HR, Mason CP, Garbarini MA, et al. Automated fabrication of mobility aids (AFMA): below-knee CASD/CAM testing and evaluation program results. J Rehabil Res Dev. 1992;29(4):78–124. pmid:1432729
  27. 27. Engsberg JR, Clynch GS, Lee AG, Allan JS, Harder JA. A CAD CAM method for custom below-knee sockets. Prosthet Orthot Int. 1992;16(3):183–8. pmid:1491952
  28. 28. Torres-Moreno R, Saunders CG, Foort J, Morrison JB. Computer-aided design and manufacture of an above-knee amputee socket. J Biomed Eng. 1991;13(1):3–9. pmid:2002669
  29. 29. Köhler P, Lindh L, Netz P. Comparison of CAD-CAM and hand made sockets for PTB prostheses. Prosthet Orthot Int. 1989;13(1):19–24. pmid:2717380
  30. 30. Krouskop TA, Muilenberg AL, Doughtery DR, Winningham DJ. Computer-aided design of a prosthetic socket for an above-knee amputee. J Rehabil Res Dev. 1987;24(2):31–8. pmid:3585783
  31. 31. Sidles JA, Boone DA, Harlan JS, Burgess EM. Rectification Maps: A New Method for Describing Residual Limb and Socket Shapes. JPO Journal of Prosthetics and Orthotics. 1989;1(3):149–53.
  32. 32. Daublys D, Domeika A, Sinkute B. Transfemoral prostheses: importance of proper fitting and materials for socket 3D-printing. Mechanika. 2023;29(5):419–24.
  33. 33. Kim S, Yalla SV, Shetty S, Rosenblatt NJ. Structural integrity of custom-designed additive manufactured prosthetic sockets compared to traditional sockets. Results in Materials. 2024;21:100549.
  34. 34. Marinopoulos T, Li S, Silberschmidt VV. Structural integrity of 3D-printed prosthetic sockets: Experimental study for paediatric applications. Journal of Materials Research and Technology. 2023;24:2734–42.
  35. 35. van der Stelt M, Verhamme L, Slump CH, Brouwers L, Maal TJ. Strength testing of low-cost 3D-printed transtibial prosthetic socket. Proc Inst Mech Eng H. 2022;236(3):367–75. pmid:34852701
  36. 36. van der Stelt M, Grobusch MP, Koroma AR, Papenburg M, Kebbie I, Masanga 3D prosthesis printing research group, et al. Pioneering low-cost 3D-printed transtibial prosthetics to serve a rural population in Sierra Leone - an observational cohort study. EClinicalMedicine. 2021;35:100874. pmid:34027333
  37. 37. Plesec V, Humar J, Dobnik-Dubrovski P, Harih G. Numerical Analysis of a Transtibial Prosthesis Socket Using 3D-Printed Bio-Based PLA. Materials (Basel). 2023;16(5):1985. pmid:36903100
  38. 38. Tsai TF, Maibach HI. How irritant is water? An overview. Contact Dermatitis. 1999;41(6):311–4. pmid:10617210
  39. 39. Kim S, Chen J, Cheng T, Gindulyte A, He J, He S, et al. PubChem 2023 update. Nucleic Acids Res. 2023;51(D1):D1373–80. pmid:36305812
  40. 40. Anderson S, Stuckey R, Oakman J. Work-related musculoskeletal injuries in prosthetists and orthotists in Australia. Int J Occup Saf Ergon. 2021;27(3):708–13. pmid:29893181
  41. 41. Fernie GR, Halsall AP, Ruder K. Shape sensing as an educational aid for student prosthetists. Prosthet Orthot Int. 1984;8(2):87–90. pmid:6483595
  42. 42. Klasson B. Computer aided design, computer aided manufacture and other computer aids in prosthetics and orthotics. Prosthet Orthot Int. 1985;9(1):3–11. pmid:4000909
  43. 43. Instalimb. Life-accelerating technology: The world’s first 3D printed prosthetic leg provider. https://india.instalimb.com/our-story/. 2024. Accessed 2025 August.
  44. 44. Steer JW, Grudniewski PA, Browne M, Worsley PR, Sobey AJ, Dickinson AS. Predictive prosthetic socket design: part 2-generating person-specific candidate designs using multi-objective genetic algorithms. Biomech Model Mechanobiol. 2020;19(4):1347–60. pmid:31741116
  45. 45. Steer JW, Worsley PR, Browne M, Dickinson AS. Predictive prosthetic socket design: part 1-population-based evaluation of transtibial prosthetic sockets by FEA-driven surrogate modelling. Biomech Model Mechanobiol. 2020;19(4):1331–46. pmid:31256276
  46. 46. Dickinson A, Diment L, Morris R, Pearson E, Hannett D, Steer J. Characterising Residual Limb Morphology and Prosthetic Socket Design Based on Expert Clinician Practice. Prosthesis. 2021;3(4):280–99.
  47. 47. Ngan CC, Pendse V, Sivasambu H, Ouellette E, Ready N, Andrysek J. Preliminary characterization of rectification for transradial prosthetic sockets. Sci Rep. 2024;14(1):5759. pmid:38459106
  48. 48. van der Stelt M, Berends B, Papenburg M, Langenhuyzen T, Maal T, Brouwers L, et al. Evaluating the Effectiveness of Transtibial Prosthetic Socket Shape Design Using Artificial Intelligence: A Clinical Comparison With Traditional Plaster Cast Socket Designs. Arch Phys Med Rehabil. 2025;106(2):239–46. pmid:39304077
  49. 49. Pendse V, Ngan CC, Ouellette E, Ready N, Andrysek J. PACT: A practice-driven predictive algorithm for customized transradial prosthetic socket design. PLoS One. 2026;21(1):e0340831. pmid:41505424
  50. 50. Choo YJ, Chang MC. Use of machine learning in the field of prosthetics and orthotics: A systematic narrative review. Prosthet Orthot Int. 2023;47(3):226–40. pmid:36811961
  51. 51. Safari MR, Meier MR. Systematic review of effects of current transtibial prosthetic socket designs-Part 1: Qualitative outcomes. J Rehabil Res Dev. 2015;52(5):491–508. pmid:26436666
  52. 52. Afiqah Hamzah N, Razak NAA, Sayuti Ab Karim M, Gholizadeh H. A review of history of CAD/CAM system application in the production of transtibial prosthetic socket in developing countries (from 1980 to 2019). Proc Inst Mech Eng H. 2021;235(12):1359–74. pmid:34304625
  53. 53. Cutti AG, Santi MG, Hansen AH, Fatone S, Residual Limb Shape Capture Group. Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics. Sensors (Basel). 2024;24(7):2350. pmid:38610559
  54. 54. Verein Deutscher Ingenieure eV. VDI/VDE 2634, Optical 3D measurement systems - Multiple view systems based on area scanning. Dusseldorf, Germany: VDI/VDE-Gesellschaft Mess- und Automatisierungstechnik. 2008. https://www.vdi.de/en/home/vdi-standards/details/vdivde-2634-blatt-3-optical-3d-measuring-systems-multiple-viewsystems-based-on-area-scanning
  55. 55. Cignoni P, Callieri M, Corsini M, Dellepiane M, Ganovelli F, Ranzuglia G. MeshLab: an open-source mesh processing tool. Computing. 2008;012008:129–36.
  56. 56. Schroeder W, Martin K, Lorensen B, Kitware Inc. The Visualization Toolkit: an object-oriented approach to 3D graphics. Kitware. 2006.
  57. 57. Qt Company Ltd. Qt for Python. Available from: https://doc.qt.io/qtforpython-6/index.html#documentation. Accessed on: June 21, 2023.
  58. 58. Biancolini M. Fast radial basis functions for engineering applications. Cham, Switzerland: Springer. 2017.
  59. 59. Gusev MG, Lein GA, Alzoba SV. Advantages of the Marlo anatomical socket (MAS) for transfemoral prosthesis. GO. 2020;26(3):359–63.
  60. 60. Fatone S, Gravely A, Cutti AG, Hansen AH, Gard SA. Randomized crossover trial of hand and hydrostatic casting for custom lower limb prosthetic sockets: Assessing socket comfort and fabrication time. PLOS One. 2025;20(11):e0337185.
  61. 61. Hanspal RS, Fisher K, Nieveen R. Prosthetic socket fit comfort score. Disabil Rehabil. 2003;25(22):1278–80. pmid:14617445